Cil A Veillette CJH Sanchez-Sotelo J Sperling JW Schleck C Cofield RH

Between 1976 and 2004, 38 revision arthroplasties (35 patients) were performed for aseptic loosening of the humeral component. The mean interval from primary arthroplasty to revision was 7.1 years (0.4 to 16.6). A total of 35 shoulders (32 patients) were available for review at a mean follow-up of seven years (2 to 19.3). Pre-operatively, 34 patients (97%) had moderate or severe pain; at final follow-up, 29 (83%) had no or only mild pain (p < 0.0001). The mean active abduction improved from 88° to 107° (p < 0.01); and the mean external rotation from 37° to 46° (p = 0.27). Excellent or satisfactory results were achieved in 25 patients (71%) according to the modified Neer rating system. Humeral components were cemented in 29, with ingrowth implants used in nine cases. There were 19 of standard length and 17 were longer (two were custom replacements and are not included). Bone grafting was required for defects in 11 humeri. Only two glenoid components were left unrevised. Intra-operative complications included cement extrusion in eight cases, fracture of the shaft of the humerus is two and of the tuberosity in four. There were four re-operations, one for recurrent humeral loosening, with 89% survival free of re-operations at ten years. Revision surgery for aseptic loosening of the humeral component provides reliable pain relief and modest improvement of movement, although there is a substantial risk of intra-operative complications. Revision to a total shoulder replacement gives better results than to a hemiarthroplasty


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1198 - 1204
1 Sep 2008
Peden JP Morrey BF

This study reports our experience with total elbow replacement for fused elbows.

Between 1982 and 2004, 13 patients with spontaneously ankylosed elbows were treated with a linked semi-constrained non-custom total elbow implant. The mean age at operation was 54 years (24 to 80). The stiffness was a result of trauma in ten elbows, juvenile rheumatoid arthritis in one, and rheumatoid arthritis in two. The patients were followed for a mean of 12 years (2 to 26) and were evaluated clinically using the Mayo Elbow Performance Score, as well as radiologically.

A mean arc from 37° of extension to 118° of flexion was achieved. Outcomes were good or excellent for seven elbows at final review. Ten patients felt better or much better after total elbow replacement. However, there was a high complication rate and re-operation was required in over half of patients. Two developed peri-operative soft-tissue breakdown requiring debridement. A muscle flap with skin grafting was used for soft-tissue cover in one. Revision was undertaken in one elbow following fracture of the ulnar component. Three patients developed a deep infection. Three elbows were manipulated under anaesthesia for post-operative stiffness. Prophylactic measures for heterotopic ossification were unsuccessful.

Total elbow replacement for the ankylosed elbow should be performed with caution. However, the outcome can be reliable in the long term and have a markedly positive impact on patient function and satisfaction. The high potential for complications must be considered. We consider total elbow replacement to be an acceptable procedure in selected patients with reasonable expectations.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 62 - 65
1 Jan 2007
Ito H Matsumoto T Yoshitomi H Kakinoki R Nakamura T

We compared the outcome of peri-operative humeral condylar fractures in patients undergoing a Coonrad-Morrey semiconstrained total elbow replacement with that of patients with rheumatoid arthritis undergoing the same procedure without fractures. In a consecutive series of 40 elbows in 33 patients, 13 elbows had a fracture in either condyle peri-operatively, and 27 elbows were intact. The fractured condyle was either fixed internally or excised. We found no statistical difference in the patients’ background, such as age, length of follow-up, immobilisation period, Larsen’s radiological grade, or Steinbrocker’s stage and functional class. There was also no statistical difference between the groups in relation to the Mayo Elbow Performance Score, muscle strength, range of movement, or radiolucency around the implants at a mean of 4.8 years (1.1 to 8.0) follow-up.

We conclude that fractured condyles can be successfully treated with either internal fixation or excision, and cause no harmful effect.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 78 - 83
1 Jan 2006
Sneftrup SB Jensen SL Johannsen HV Søjbjerg JO

We studied retrospectively the results of revision arthroplasty of the elbow using a linked Coonrad-Morrey implant in 23 patients (24 elbows) after a mean follow-up period of 55 months. According to the Mayo Elbow Performance Score, 19 elbows were satisfactory, nine were excellent and ten good. The median total score had improved from 35 points (20 to 75) before the primary arthroplasty to 85 points (40 to 100) at the latest follow-up. There was a marked relief of pain, but the range of movement showed no overall improvement. Two patients had a second revision because of infection and two for aseptic loosening. The estimated five-year survival rate of the prosthesis was 83.1% (95% confidence interval 61.1 to 93.3). Revision elbow arthroplasty using the Coonrad-Morrey implant provided satisfactory results but with complications occurring in 13 cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 387 - 392
1 Mar 2010
Kasten P Pape G Raiss P Bruckner T Rickert M Zeifang F Loew M

We have investigated the mid-term outcome of total shoulder replacement using a keeled cemented glenoid component and a modern cementing technique with regard to the causes of failure and loosening of the components.

Between 1997 and 2003 we performed 96 total shoulder replacements on 88 patients, 24 men and 64 women with a mean age of 69.7 years (31 to 82). The minimum follow-up was five years and at the time of review 87 shoulders (77 patients) were examined at a mean follow-up of 89.1 months (60 to 127). Cumulative survival curves were generated with re-operations (accomplished and planned), survivorship of the proshesis, loosening of the glenoid (defined as tilt > 5° or subsidence > 5 mm), the presence of radiolucent lines and a Constant score of < 30 as the endpoints.

There were two re-operations not involving revision of the implants and the survival rate of the prosthesis was 100.0% for the follow-up period, with an absolute Constant score of > 30 as the endpoint the survival rate was 98%. Radiological glenoid loosening was 9% after five years, and 33% after nine years. There was an incidence of 8% of radiolucent lines in more than three of six zones in the immediate post-operative period, of 37.0% after the first year which increased to 87.0% after nine years. There was no correlation between the score of Boileau and the total Constant score at the latest follow-up, but there was correlation between glenoid loosening and pain (p = 0.001).

We found that total shoulder replacement had an excellent mid-term survivorship and clinical outcome. The surgical and cementing techniques were related to the decrease in radiolucent lines around the glenoid compared with earlier studies. One concern, however, was the fact that radiolucent lines increased over time and there was a rate of glenoid loosening of 9% after five years and 33% after nine years. This suggests that the design of the glenoid component, and the implantation and cementing techniques may need further improvement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 189 - 193
1 Feb 2008
Söderlund T Mattila VM Visuri TI Pihlajamäki HK

We reviewed the outcome of arthroscopic stabilisation of anterior glenohumeral instability in young adults using the transglenoid suture technique. A questionnaire was sent to 455 consecutive patients who had undergone this procedure between 1992 and 2000. Of these, 312 patients (68.5%) with 313 affected shoulders and a mean age of 20 years (18 to 28) responded.

Outcome was determined by the number of re-dislocations or, in patients who had not re-dislocated, by the disease-specific quality of life as measured by the Western Ontario Shoulder Instability index. During a mean follow-up of 6.4 years (1 to 14), 177 patients (56%) sustained a re-dislocation, including 70 who required a further operation. In 136 patients (44%) who reported neither re-dislocation nor re-operation, the index scores were good (median 90.4%; 28.9% to 100%). No significant peri- or pre-operative predictors of re-dislocation or re-operation were found.

We found a high rate of re-dislocation after transglenoid suture repair in young, physically active patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 330 - 334
1 Mar 2007
Cesar M Roussanne Y Bonnel F Canovas F

Between 1993 and 2002, 58 GSB III total elbow replacements were implanted in 45 patients with rheumatoid arthritis by the same surgeon. At the most recent follow-up, five patients had died (five elbows) and six (nine elbows) had been lost to follow-up, leaving 44 total elbow replacements in 34 patients available for clinical and radiological review at a mean follow-up of 74 months (25 to 143). There were 26 women and eight men with a mean age at operation of 55.7 years (24 to 77).

At the latest follow-up, 31 excellent (70%), six good (14%), three fair (7%) and four poor (9%) results were noted according to the Mayo elbow performance score. Five humeral (11%) and one ulnar (2%) component were loose according to radiological criteria (type III or type IV). Of the 44 prostheses, two (5%) had been revised, one for type-IV humeral loosening after follow-up for ten years and one for fracture of the ulnar component. Seven elbows had post-operative dysfunction of the ulnar nerve, which was transient in five and permanent in two.