Advertisement for orthosearch.org.uk
Results 1 - 19 of 19
Results per page:
The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 54 - 59
1 Jan 2016
González Della Valle A Sharrock N Barlow M Caceres L Go G Salvati EA

We describe our technique and rationale using hybrid fixation for primary total hip arthroplasty (THA) at the Hospital for Special Surgery. Modern uncemented acetabular components have few screw holes, or no holes, polished inner surfaces, improved locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented sockets can be combined with highly cross-linked polyethylene liners, which have demonstrated very low wear and osteolysis rates after ten to 15 years of implantation. The results of cement fixation with a smooth or polished surface finished stem have been excellent, virtually eliminating complications seen with cementless fixation like peri-operative femoral fractures and thigh pain. Although mid-term results of modern cementless stems are encouraging, the long-term data do not show reduced revision rates for cementless stems compared with cemented smooth stems. In this paper we review the conduct of a hybrid THA, with emphasis on pre-operative planning, surgical technique, hypotensive epidural anaesthesia, and intra-operative physiology.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):54–9.


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 768 - 774
1 Jul 2023
Wooster BM Kennedy NI Dugdale EM Sierra RJ Perry KI Berry DJ Abdel MP

Aims

Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis.

Methods

We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m2 (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18).


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 206 - 211
1 Feb 2022
Bloch BV White JJE Matar HE Berber R Manktelow ARJ

Aims. Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost. Methods. In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA. Results. Cemented fixation had the lowest implant survival in all age groups, with a total ten-year survivorship of 97.0% (95% confidence interval (CI) 95.8 to 97.8) in the cemented group, 97.6% (95% CI 96.9 to 98.1) in the hybrid group, and 97.9% (95% CI 96.9 to 98.6) in the cementless group. This was not statistically significant (p = 0.092). There was no age group where cemented fixation outperformed hybrid or cementless fixation. Conclusion. While all fixation techniques performed well at long-term follow-up, cemented fixation was associated with the lowest implant survival in all age groups, including in more elderly patients. We recommend that surgeons should carefully monitor their own outcomes and use fixation techniques that they are familiar with, and deliver the best outcomes in their own hands. Cite this article: Bone Joint J 2022;104-B(2):206–211


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 129 - 134
1 Jul 2021
Ayekoloye CI Abu Qa'oud M Radi M Leon SA Kuzyk P Safir O Gross AE

Aims

Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants.

Methods

A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 414 - 422
1 Apr 2020
Gwynne-Jones DP Gray AR

Aims

To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem.

Methods

Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or uncemented acetabular component (454) were collected from local records and the New Zealand Joint Registry. A competing risks regression survival analysis was performed with death as the competing risk with adjustments made for age, sex, approach, and bearing.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 997 - 1002
1 Aug 2020
Leong JW Cook MJ O’Neill TW Board TN

Aims

The aim of this study was to investigate whether the use of antibiotic-loaded bone cement influenced the risk of revision surgery after primary total hip arthroplasty (THA) for osteoarthritis.

Methods

The study involved data collected by the National Joint Registry (NJR) for England and Wales, Northern Ireland and the Isle of Man between 1 September 2005 and 31 August 2017. Cox proportional hazards were used to investigate the association between use of antibiotic-loaded bone cement and the risk of revision due to prosthetic joint infection (PJI), with adjustments made for the year of the initial procedure, age at the time of surgery, sex, American Society of Anesthesiologists (ASA) grade, head size, and body mass index (BMI). We looked also at the association between use of antibiotic-loaded bone cement and the risk of revision due to aseptic loosening or osteolysis.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1050 - 1057
1 Sep 2019
Lampropoulou-Adamidou K Hartofilakidis G

Aims

To our knowledge, no study has compared the long-term results of cemented and hybrid total hip arthroplasty (THA) in patients with osteoarthritis (OA) secondary to congenital hip disease (CHD). This is a demanding procedure that may require special techniques and implants. Our aim was to compare the long-term outcome of cemented low-friction arthroplasty (LFA) and hybrid THA performed by one surgeon.

Patients and Methods

Between January 1989 and December 1997, 58 hips (44 patients; one man, 43 woman; mean age 56.6 years (25 to 77)) with OA secondary to CHD were treated with a cemented Charnley LFA (group A), and 55 hips (39 patients; two men, 37 women; mean age 49.1 years (27 to 70)) were treated with a hybrid THA (group B), by the senior author (GH). The clinical outcome and survivorship were compared.


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 37 - 43
1 Jan 2017
Garland A Gordon M Garellick G Kärrholm J Sköldenberg O Hailer NP

Aims

It has been suggested that cemented fixation of total hip arthroplasty (THA) is associated with an increased peri-operative mortality compared with cementless THA. Our aim was to investigate this through a nationwide matched cohort study adjusting for age, comorbidity, and socioeconomic background.

Patients and Methods

A total of 178 784 patients with osteoarthritis who underwent either cemented or cementless THA from the Swedish Hip Arthroplasty Register were matched with 862 294 controls from the general population. Information about the causes of death, comorbidities, and socioeconomic background was obtained. Mortality within the first 90 days after the operation was the primary outcome measure.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1010 - 1017
1 Aug 2018
Jain S Magra M Dube B Veysi VT Whitwell GS Aderinto JB Emerton ME Stone MH Pandit HG

Aims

This study aimed to evaluate implant survival of reverse hybrid total hip arthroplasty (THA) at medium-term follow-up.

Patients and Methods

A consecutive series of 1082 THAs in 982 patients with mean follow-up of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21 to 94). Of these, 194 (17.9%) were in patients under 60 years, 663 (61.3%) in female patients and 348 (32.2%) performed by a trainee. Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%). Survival analysis was performed and subgroups compared using log rank tests.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1347 - 1354
1 Oct 2016
Palan J Smith MC Gregg P Mellon S Kulkarni A Tucker K Blom AW Murray DW Pandit H

Aims

Periprosthetic fracture (PF) after primary total hip arthroplasty (THA) is an uncommon but potentially devastating complication. This study aims to investigate the influence of cemented stem designs on the risk of needing a revision for a PF.

Patients and Methods

We analysed data on 257 202 primary THAs with cemented stems and 390 linked first revisions for PF recorded in the National Joint Registry (NJR) of England, Wales and Northern Ireland to determine if a cemented femoral stem brand was associated with the risk of having revision for a PF after primary THA. All cemented femoral stem brands with more than 10 000 primary operations recorded in the NJR were identified. The four most commonly used cemented femoral stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem (n = 15 113) and Charnley (n = 20 182).

We compared the revision risk ratios due to PF amongst the stems using a Poisson regression model adjusting for patient factors. Compared with the Exeter V40, the age, gender and ASA grade adjusted revision rate ratio was 3.89 for the cemented CPT stem (95% confidence interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to 1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 451 - 458
1 Apr 2009
Hooper GJ Rothwell AG Stringer M Frampton C

We have reviewed the rate of revision of fully cemented, hybrid and uncemented primary total hip replacements (THRs) registered in the New Zealand Joint Registry between 1999 and December 2006 to determine whether there was any statistically significant difference in the early survival and reason for revision in these different types of fixation. The percentage rate of revision was calculated per 100 component years and compared with the reason for revision, the type of fixation and the age of the patients.

Of the 42 665 primary THRs registered, 920 (2.16%) underwent revision requiring change of at least one component. Fully-cemented THRs had a lower rate of revision when considering all causes for failure (p < 0.001), but below the age of 65 years uncemented THRs had a lower rate (p < 0.01). The rate of revision of the acetabular component for aseptic loosening was less in the uncemented and hybrid groups compared with that in the fully cemented group (p < 0.001), and the rate of revision of cemented and uncemented femoral components was similar, except in patients over 75 years of age in whom revision of cemented femoral components was significantly less frequent (p < 0.02). Revision for infection was more common in patients aged below 65 years and in cemented and hybrid THRs compared with cementless THRs (p < 0.001). Dislocation was the most common cause of revision for all types of fixation and was more frequent in both uncemented acetabular groups (p < 0.001). The experience of the surgeon did not affect the findings.

Although cemented THR had the lowest rate of revision for all causes in the short term (90 days), uncemented THR had the lowest rate of aseptic loosening in patients under 65 years of age and had rates comparable with international rates of aseptic loosening in those over 65 years.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 167 - 170
1 Feb 2005
Treacy RBC McBryde CW Pynsent PB

We report the survival at five years of 144 consecutive metal-on-metal resurfacings of the hip implanted between August 1997 and May 1998. Failure was defined as revision of either the acetabular or femoral component for any reason during the study period. The survival at the end of five years was 98% overall and 99% for aseptic revisions only. The mean age of the patients at implantation was 52.1 years.

Three femoral components failed during the first two years, two were infected and one fractured. A single stage revision was carried out in each case. No other revisions were performed or are impending. No patients were lost to follow-up. Four died from unrelated causes during the study period.

This study confirms that hip resurfacing using a metal-on-metal bearing of known provenance can provide a solution in the medium term for the younger more active adult who requires surgical intervention for hip disease.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 587 - 592
1 May 2011
Kim Y Kim J Park J Joo J

We reviewed the results of 84 total hip replacements performed with a short metaphyseal-fitting anatomical cementless femoral component in 84 unselected consecutive patients with a mean age of 78.9 years (70 to 88). The mean follow-up was 4.6 years (4 to 5). The mean pre-operative Harris hip score was 26 points (0 to 56), which improved to 89 (61 to 100) at the final follow-up. No patient had thigh pain. The mean pre-operative Western Ontario and McMaster Universities osteoarthritis index score was 61 points (48 to 75), which improved to 21 (6 to 46). The mean University of California, Los Angeles activity score was 5.5 points (3 to 7) at the final follow-up. Osseointegration was seen in all femoral and acetabular components. All hips had grade 1 stress shielding of the proximal femur. No acetabular or femoral osteolysis was identified.

These results demonstrate that a short metaphyseal-fitting femoral component achieves optimal fixation without diaphyseal anchorage in elderly patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 158 - 163
1 Feb 2011
Baker RP Pollard TCB Eastaugh-Waring SJ Bannister GC

We compared the medium-term clinical and radiological results of hybrid total hip replacement (THR) with metal-on-metal Birmingham hip resurfacing (BHR) in two groups of 54 young patients matched for age, gender, body mass index and pre-operative levels of activity.

The clinical outcome was assessed by the University of California, Los Angeles (UCLA) activity score, the Oxford Hip Score (OHS) and the EuroQol scores. Radiologically, all hips were assessed for migration and osteolysis, the hybrid THRs for polyethylene wear and the BHRs for a pedestal sign. The mean follow-up of the patients with a hybrid THR was ten years and for those with a BHR, nine years. Four patients with a hybrid THR and one with a BHR had died. In each group five were lost to follow-up. The revision rate of the hybrid THRs was 16.7% (9 of 54) and of the BHRs 9.3% (5 of 54) (p = 0.195). Radiographs of a further eight hybrid THRs demonstrated wear and osteolysis, and they await revision (p = 0.008). Of the unrevised BHRs 90% had radiological changes, of which approximately 50% had progressed over the previous four years. All hybrid THRs demonstrated linear polyethylene wear with a mean of 1.24 mm (0.06 to 3.03). The BHRs recorded superior OHS (p = 0.013), UCLA (p = 0.008), and EuroQol visual analogue scores (p = 0.009).

After nine years, patients with BHRs remained more active and had a lower rate of revision than those with hybrid THRs. Both groups demonstrated progressive radiological changes at medium-term follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 12 - 19
1 Jan 2010
Vendittoli P Roy A Mottard S Girard J Lusignan D Lavigne M

We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed coating for fixation of the acetabular component.

Although significantly higher blood ion levels of Cr and Co were observed at three months in the resurfaced group than in total hip replacement, no significant difference was found at two years post-operatively for Cr, 1.58 μg/L and 1.62 μg/L respectively (p = 0.819) and for Co, 0.67 μg/L and 0.94 μg/L respectively (p = 0.207). A steady state was reached at one year in the resurfaced group and after three months in the total hip replacement group. Interestingly, Ti, which is not part of the bearing surfaces with its release resulting from metal corrosion, had significantly elevated ion levels after implantation in both groups. The hip resurfacing group had significantly higher Ti levels than the total hip replacement group for all periods of follow-up. At two years the mean blood levels of Ti ions were 1.87 μg/L in hip resurfacing and and 1.30 μg/L in total hip replacement (p = 0.001).

The study confirms even with different bearing diameters and clearances, hip replacement and 28 mm metal-on-metal total hip replacement produced similar Cr and Co metal ion levels in this randomised controlled trial study design, but apart from wear on bearing surfaces, passive corrosion of exposed metallic surfaces is a factor which influences ion concentrations. Ti plasma spray coating the acetabular components for hip resurfacing produces significantly higher release of Ti than Ti grit-blasted surfaces in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 16 - 20
1 Jan 2007
Gill HS Campbell PA Murray DW De Smet KA

Resurfacing arthroplasty of the hip is being used increasingly as an alternative to total hip replacement, especially for young active patients. There is concern about necrosis of the femoral head after resurfacing which can result in fracture and loosening. Most systems use a cemented femoral component, with the potential for thermal necrosis of the cancellous bone of the reamed femoral head. We used thermal probes to record temperatures close to the cement-bone interface during resurfacing arthroplasty.

The maximum temperature recorded at the cement-bone interface in four cases was approximately 68°C which was higher than that reported to kill osteocytes. A modified surgical technique using insertion of a suction cannula into the lesser trochanter, generous pulsed lavage and early reduction of the joint significantly reduced the maximum recorded cancellous bone temperature to approximately 36°C in five cases (p = 0.014).

We recommend the modified technique since it significantly reduces temperatures at the cement-bone interface.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1149 - 1154
1 Sep 2007
Lian Y Yoo M Pei F Cho Y Cheng J Chun S

We performed 52 total hip replacements in 52 patients using a cementless acetabular component combined with a circumferential osteotomy of the medial acetabular wall for the late sequelae of childhood septic arthritis of the hip. The mean age of the patients at operation was 44.5 years (22 to 66) and the mean follow-up was 7.8 years (5 to 11.8). The mean improvement in the Harris Hip Score was 29.6 points (19 to 51) at final follow-up. The mean cover of the acetabular component was 98.5% (87.8% to 100%). The medial acetabular wall was preserved with a mean thickness of 8.3 mm (1.7 to 17.4) and the mean length of abductor lever arm increased from 43.4 mm (19.1 to 62) to 54.2 mm (36.5 to 68.6). One acetabular component was revised for loosening and osteolysis 4.5 years postoperatively, and one had radiolucent lines in all acetabular zones at final review. Kaplan-Meier survival was 94.2% (95% confidence interval 85.8% to 100%) at 7.3 years, with revision or radiological loosening as an end-point when two hips were at risk.

A cementless acetabular component combined with circumferential medial acetabular wall osteotomy provides favourable results for acetabular reconstruction in patients who present with late sequelae of childhood septic hip arthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1143 - 1151
1 Sep 2008
Langton DJ Jameson SS Joyce TJ Webb J Nargol AVF

Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p < 0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 441 - 448
1 Apr 2007
Vendittoli P Mottard S Roy AG Dupont C Lavigne M

We evaluated the concentrations of chromium and cobalt ions in blood after metal-on-metal surface replacement arthroplasty using a wrought-forged, high carbon content chromium-cobalt alloy implant in 64 patients. At one year, mean whole blood ion levels were 1.61 μg/L (0.4 to 5.5) for chromium and 0.67 μg/L (0.23 to 2.09) for cobalt. The pre-operative ion levels, component size, female gender and the inclination of the acetabular component were inversely proportional to the values of chromium and/or cobalt ions at one year postoperatively. Other factors, such as age and level of activity, did not correlate with the levels of metal ions. We found that the levels of the ions in the serum were 1.39 and 1.37 times higher for chromium and cobalt respectively than those in the whole blood.

The levels of metal ions obtained may be specific to the hip resurfacing implant and reflect its manufacturing process.