Impingement by prominence at the femoral
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods
There is increasing global awareness of adverse
reactions to metal debris and elevated serum metal ion concentrations
following the use of second generation metal-on-metal total hip
arthroplasties. The high incidence of these complications can be
largely attributed to corrosion at the head-neck interface. Severe
corrosion of the taper is identified most commonly in association
with larger diameter femoral heads. However, there is emerging evidence
of varying levels of corrosion observed in retrieved components
with smaller diameter femoral heads. This same mechanism of galvanic
and mechanically-assisted crevice corrosion has been observed in
metal-on-polyethylene and ceramic components, suggesting an inherent
biomechanical problem with current designs of the head-neck interface. We provide a review of the fundamental questions and answers
clinicians and researchers must understand regarding corrosion of
the taper, and its relevance to current orthopaedic practice. Cite this article:
In 12 patients, we measured the oxygen concentration in the femoral
Aims. Asphericity of the femoral
We examined the morphology of mammalian hips asking whether evolution can explain the morphology of impingement in human hips. We describe two stereotypical mammalian hips, coxa recta and coxa rotunda. Coxa recta is characterised by a straight or aspherical section on the femoral head or
A 30-year-old man presented with pain and limitation of movement of the right hip. The symptoms had failed to respond to conservative treatment. Radiographs and CT scans revealed evidence of impingement between the femoral
The β-angle is a radiological tool for measuring the distance between the pathological
Femoroacetabular impingement is recognised as being a cause of labral tears and chondral damage. We report a series of five patients who presented with persistent pain in the hip after arthroscopy for isolated labral debridement. All five had a bony abnormality consistent with cam-type femoroacetabular impingement. They had a further operation to correct the abnormality by chondro-osteoplasty of the femoral
Open reduction of the prominence at the femoral
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. . Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. . Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The
Nine uncemented hip prostheses of modular design were revised because of late infection (2 cases), femoral stem loosening or fracture (2), loosening of threaded cups (3) and protrusion of bipolar cups (2). At surgery we found tissue discoloration and macroscopic corrosion in four of the nine prostheses, located at the
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
To clarify the mid-term results of transposition osteotomy of the acetabulum (TOA), a type of spherical periacetabular osteotomy, combined with structural allograft bone grafting for severe hip dysplasia. We reviewed patients with severe hip dysplasia, defined as Severin IVb or V (lateral centre-edge angle (LCEA) < 0°), who underwent TOA with a structural bone allograft between 1998 and 2019. A medical chart review was conducted to extract demographic data, complications related to the osteotomy, and modified Harris Hip Score (mHHS). Radiological parameters of hip dysplasia were measured on pre- and postoperative radiographs. The cumulative probability of TOA failure (progression to Tönnis grade 3 or conversion to total hip arthroplasty) was estimated using the Kaplan–Meier product-limited method, and a multivariate Cox proportional hazard model was used to identify predictors for failure.Aims
Methods
In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased taper corrosion, with the release of metal ions and adverse reactions. The aim of this study was to investigate the relationship between the size of the femoral head and the levels of metal ions in the blood in these patients. A total of 96 patients were enrolled at two centres and randomized to undergo MoP THA using either a 32 mm metal head or a femoral head of between 36 mm and 44 mm in size, being the largest possible to fit the thinnest available polyethylene insert. The levels of metal ions and patient-reported outcome measures (Oxford Hip Score, University of California, Los Angeles Activity Scale) were recorded at two and five years postoperatively.Aims
Methods
Hip arthroscopy has gained prominence as a primary surgical intervention for symptomatic femoroacetabular impingement (FAI). This study aimed to identify radiological features, and their combinations, that predict the outcome of hip arthroscopy for FAI. A prognostic cross-sectional cohort study was conducted involving patients from a single centre who underwent hip arthroscopy between January 2013 and April 2021. Radiological metrics measured on conventional radiographs and magnetic resonance arthrography were systematically assessed. The study analyzed the relationship between these metrics and complication rates, revision rates, and patient-reported outcomes.Aims
Methods
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
There has been a marked increase in the number of hip arthroscopies performed over the past 16 years, primarily in the management of femoroacetabular impingement (FAI). Insights into the pathoanatomy of FAI, and high-level evidence supporting the clinical effectiveness of arthroscopy in the management of FAI, have fuelled this trend. Arthroscopic management of labral tears with repair may have superior results compared with debridement, and there is now emerging evidence to support reconstructive options where repair is not possible. In situations where an interportal capsulotomy is performed to facilitate access, data now support closure of the capsule in selective cases where there is an increased risk of postoperative instability. Preoperative planning is an integral component of bony corrective surgery in FAI, and this has evolved to include computer-planned resection. However, the benefit of this remains controversial. Hip instability is now widely accepted, and diagnostic criteria and treatment are becoming increasingly refined. Instability can also be present with FAI or develop as a result of FAI treatment. In this annotation, we outline major current controversies relating to decision-making in hip arthroscopy for FAI. Cite this article:
The direct anterior approach (DAA) for total hip arthroplasty (THA) has potential advantages over other approaches and is most commonly performed with the patient in the supine position. We describe a technique for DAA THA with the patient in the lateral decubitus position and report the early clinical and radiological outcomes, the characteristics of the learning curve, and perioperative complications. All primary DAA THAs performed in the lateral position by a single surgeon over a four-year period from the surgeon’s first case using the technique were identified from a prospectively collected database. Modified Harris Hip Scores (mHHS) were collected to assess clinical outcome, and routine radiological analysis was performed. Retrospective review of the medical records identified perioperative complications, the characteristics of the learning curve, and revisions.Aims
Methods