Aims. Revision total hip arthroplasty in patients with Vancouver type B3 fractures with Paprosky type IIIA, IIIB, and IV femoral defects are difficult to treat. One option for Paprovsky type IIIB and IV defects involves modular cementless, tapered, revision femoral components in conjunction with distal interlocking screws. The aim of this study was to analyze the rate of reoperations and complications and union of the fracture, subsidence of the stem, mortality, and the clinical outcomes in these patients. Methods. A total of 46 femoral components in patients with Vancouver B3 fractures (23 with Paprosky type IIIA, 19 with type IIIB, and four with type IV defects) in 46 patients were revised with a transfemoral approach using a modular, tapered, cementless revision Revitan curved femoral component with distal cone-in-cone fixation and prospectively followed for a mean of 48.8 months (SD 23.9; 24 to 112). The mean age of the patients was 80.4 years (66 to 100). Additional distal interlocking was also used in 23 fractures in which distal cone-in-cone fixation in the isthmus was < 3 cm. Results. One patient (2.2%) died during the first postoperative year. After six months, 43 patients (93.5%) had osseous, and three had fibrous consolidation of the fracture and the bony flap, 42 (91.3%) had bony ingrowth and four had stable fibrous fixation of the stem. No patient had radiolucency around the interlocking screws and no screw broke. One patient had non-progressive subsidence and two had a dislocation. The mean Harris Hip Score increased from of 57.8 points (SD 7.9) three months postoperatively to 76.1 points (SD 10.7) 24 months postoperatively. Conclusion. The 2° tapered, fluted revision femoral component with distal cone-in-cone-fixation, combined with additional distal interlocking in patients with bony deficiency at the isthmus, led to reproducibly good results in patients with a Vancouver B3 fracture and Paprosky type IIIA, IIIB, and IV defects with regard to union of the fracture, subsidence or loosening of the stem, and clinical outcomes. Cite this article: Bone Joint J 2024;106-B(4):344–351


The Bone & Joint Journal
Vol. 102-B, Issue 5 | Pages 573 - 579
1 May 2020
Krueger DR Guenther K Deml MC Perka C

Aims. We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed. . Methods. All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation. Results. There were 113 recorded cases with fracture at the modular junction, resulting in a calculated fracture rate of 0.30% (113/37,600). The fracture rate of the implant without signs of improper use was 0.11% (41/37,600). In 79% (89/113) of cases with a failed implant, either a lateralized (high offset) neck segment, an extralong head, or the combination of both were used. Logistic regression analysis revealed male sex, high body mass index (BMI), straight component design, and small neck segments were significant risk factors for failure. Investigation of the implants (76/113) showed at least one sign of improper use in 72 cases. Conclusion. Implant failure at the modular junction is associated with patient- and implant-specific risk factors as well as technical errors during implantation. Whenever possible, the use of short and lateralized neck segments should be avoided with this revision system. Implantation instructions and contraindications need to be adhered to and respected. Cite this article: Bone Joint J 2020;102-B(5):573–579


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 709 - 715
1 Jun 2020
Abdelsamie KR Elhawary I Ali H Ali M EL-Shafie M Dominic Meek RM

Aims. Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. Methods. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared. Results. In total, 65 TM and 35 PCM cases were included. At mean follow-up of seven years (1 to 13), subsidence was noted in both cohorts during the initial three months postoperatively (p < 0.001) then implants stabilized. Subsidence noted in 58.7% (38/65 cases) of the TM cohort (mean 2.3 mm, SD 3.5 mm) compared to 48.8% (17/35) of PCM cohort (mean 1.9 mm, SD 2.6 mm; p = 0.344). Subsidence of PCM cohort were significantly associated with extended trochanteric osteotomy (ETO) (p < 0.041). Although the ETO was used less frequently in PCM stem cohort (7/35), subsidence was noted in 85% (6/7) of them. Significant improvement of the final mean Oxford Hip Score (OHS) was reported in both treatment groups (p < 0.001). Conclusion. Both modular TM and PCM revision femoral components subsided within the femur. TM implants subsided more frequently than PCM components if the femur was intact but with no difference in clinical outcomes. However, if an ETO is performed then a PCM component will subside significantly more and suggests the use of a TM implant may be advisable. Cite this article: Bone Joint J 2020;102-B(6):709–715


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 124 - 134
1 Feb 2023
Jain S Farook MZ Aslam-Pervez N Amer M Martin DH Unnithan A Middleton R Dunlop DG Scott CEH West R Pandit H

Aims

The aim of this study was to compare open reduction and internal fixation (ORIF) with revision surgery for the surgical management of Unified Classification System (UCS) type B periprosthetic femoral fractures around cemented polished taper-slip femoral components following primary total hip arthroplasty (THA).

Methods

Data were collected for patients admitted to five UK centres. The primary outcome measure was the two-year reoperation rate. Secondary outcomes were time to surgery, transfusion requirements, critical care requirements, length of stay, two-year local complication rates, six-month systemic complication rates, and mortality rates. Comparisons were made by the form of treatment (ORIF vs revision) and UCS type (B1 vs B2/B3). Kaplan-Meier survival analysis was performed with two-year reoperation for any reason as the endpoint.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 125 - 132
1 May 2024
Carender CN Perry KI Sierra RJ Trousdale RT Berry DJ Abdel MP

Aims

Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components.

Methods

We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1215 - 1221
1 Jul 2021
Kennedy JW Ng NYB Young D Kane N Marsh AG Meek RMD

Aims

Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components.

Methods

From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1678 - 1685
1 Nov 2021
Abdelaziz H Schröder M Shum Tien C Ibrahim K Gehrke T Salber J Citak M

Aims

One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions.

Methods

In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1222 - 1230
1 Jul 2021
Slullitel PA Garcia-Barreiro GG Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims

We aimed to compare the implant survival, complications, readmissions, and mortality of Vancouver B2 periprosthetic femoral fractures (PFFs) treated with internal fixation with that of B1 PFFs treated with internal fixation and B2 fractures treated with revision arthroplasty.

Methods

We retrospectively reviewed the data of 112 PFFs, of which 47 (42%) B1 and 27 (24%) B2 PFFs were treated with internal fixation, whereas 38 (34%) B2 fractures underwent revision arthroplasty. Decision to perform internal fixation for B2 PFFs was based on specific radiological (polished femoral components, intact bone-cement interface) and clinical criteria (low-demand patient). Median follow-up was 36.4 months (24 to 60). Implant survival and mortality over time were estimated with the Kaplan-Meier method. Adverse events (measured with a modified Dindo-Clavien classification) and 90-day readmissions were additionally compared between groups.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 150 - 157
1 Jun 2021
Anderson LA Christie M Blackburn BE Mahan C Earl C Pelt CE Peters CL Gililland J

Aims

Porous metaphyseal cones can be used for fixation in revision total knee arthroplasty (rTKA) and complex TKAs. This metaphyseal fixation has led to some surgeons using shorter cemented stems instead of diaphyseal engaging cementless stems with a potential benefit of ease of obtaining proper alignment without being beholden to the diaphysis. The purpose of this study was to evaluate short term clinical and radiographic outcomes of a series of TKA cases performed using 3D-printed metaphyseal cones.

Methods

A retrospective review of 86 rTKAs and nine complex primary TKAs, with an average age of 63.2 years (SD 8.2) and BMI of 34.0 kg/m2 (SD 8.7), in which metaphyseal cones were used for both femoral and tibial fixation were compared for their knee alignment based on the type of stem used. Overall, 22 knees had cementless stems on both sides, 52 had cemented stems on both sides, and 15 had mixed stems. Postoperative long-standing radiographs were evaluated for coronal and sagittal plane alignment. Adjusted logistic regression models were run to assess malalignment hip-knee-ankle (HKA) alignment beyond ± 3° and sagittal alignment of the tibial and femoral components ± 3° by stem type.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1447 - 1458
1 Nov 2019
Chatziagorou G Lindahl H Kärrholm J

Aims. We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis. Patients and Methods. The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR. Results. Out of 1381 Vancouver type B fractures that fulfilled the inclusion criteria, 257 underwent further reoperation by the end of 2013. Interprosthetic and Type B1 fractures had a higher risk for reoperation. For B1 fractures, the rate of reoperation did not differ (p = 0.322) after use of conventional (26%) or locking plate osteosynthesis (19%). No significant differences were observed between cemented, cementless monoblock, and cementless modular revision components for the treatment of type B2 and B3 fractures. Conclusion. In this country-specific study, the choice of locking or conventional plates for the treatment of type B1, and cemented or cementless femoral components fixation for B2 and B3 fractures, had no significant influence on risk for reoperation. Interprosthetic fractures adversely affected the outcome of treatment of type B fractures. Differences in the patient characteristics of the compared groups were observed. Cite this article: Bone Joint J 2019;101-B:1447–1458


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1319 - 1323
3 Oct 2020
Khatun F Gill DF Atrey A Porteous M

Aims. We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital. Methods. All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with National Joint Registry (NJR) data and general practitioner (GP) records to assess whether they had undergone revision for any reason. Results. During this period of study 144 patients (157 hips) died (69.78%). Two patients were lost to follow-up, leaving 61 patients (66 hips) available for review (29.33%). Of the 225 hips, three underwent revision for femoral failure with osteolysis. One underwent femoral component revision for treatment of a periprosthetic fracture. Eight underwent revision of the acetabular component only for loosening. Two hips had both components revised, when components were found to be loose at time of revision for acetabular loosening, though no radiological femoral osteolysis. Two patients underwent revision for infection. Using femoral loosening as an endpoint, the survival of the Exeter Universal femoral component was 98.7% (n = 220, 95% confidence interval (CI) 96.1% to 100%) at 20 to 22 years. Survival with an endpoint of revision for any reason was 92.6% (n = 209, 95% CI 89.4 to 95.55), with a ‘worst-case scenario’ (considering two patients lost to follow-up to have failed), the overall survival rate was 91.7% (n = 207, 95% CI 87.8 to 95.9) at 20 to 22 years. Conclusion. Our results confirm excellent long-term results for the cemented Exeter Universal femoral componentimplanted outside of the originating centre. Cite this article: Bone Joint J 2020;102-B(10):1319–1323


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 615 - 618
1 May 2012
ten Have BLEF Brouwer MD RW van Biezen FC Verhaar JAN

The purpose of this prospective study was to evaluate the long-term clinical and radiological outcomes of revision of the femoral component of a total hip replacement using impaction bone grafting. Femoral revision with an impacted allograft was performed on 29 patients (31 hips). In all, 21 hips (68%) had grade III or IV femoral defects according to the Endo-Klinik classification. A total of 11 patients (12 hips) died before the ten-year follow-up period. Of the remaining patients, 18 patients (19 hips) were followed for 10 to 15 years; three further patients died during this time. None of the 31 stems underwent further revision of their stem. However, four stems showed extensive subsidence (> 15 mm). One of these patients had a femoral fracture that required fixation. Three other patients had a femoral fracture, two of which required fixation and the other was treated conservatively. Patients with a femoral fracture and/or severe subsidence had significantly more grade IV defects (six of seven hips; p = 0.004). One patient needed a closed reduction for dislocation. Impaction allografting in revision hip surgery gives good long-term results for femora with grades I, II and III Endo-Klinik-classified defects. Extensive subsidence and femoral fractures were seen mainly in patients with grade IV damaged femora


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 3 | Pages 378 - 382
1 Apr 2000
McNally SA Shepperd JAN Mann CV Walczak JP

We describe the clinical and radiological outcome of 100 consecutive total hip replacements in 86 patients using the JRI Furlong hydroxyapatite-coated femoral component. The follow-up was 100% at a mean of ten years (9.0 to 12.0). All 54 living patients (62 hips) were examined annually. Their mean Merle d’Aubigné score was 5.8/5.6/4.6. A total of 32 patients had died. Their notes at the last annual assessment recorded a similar outcome to that of the living patients. Radiographs showed no radiolucent lines around any femoral component and there were no signs of impending failure. One patient required excision arthroplasty, but the femoral component had been bonded satisfactorily. There have been no other revisions of femoral components. Our findings suggest that the hydroxyapatite coating gives a satisfactory prosthesis-bone interface which is preferable to any other system


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 57 - 61
1 Jun 2019
Chalmers BP Mangold DG Hanssen AD Pagnano MW Trousdale RT Abdel MP

Aims

Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR.

Patients and Methods

Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 50 - 53
1 Jan 2016
Konan S Garbuz DS Masri BA Duncan CP

Tapered fluted titanium stems are increasingly used for femoral revision arthroplasty. They are available in modular and non-modular forms. Modularity has advantages when the bone loss is severe, the proximal femur is mis shapen or the surgeon is unfamiliar with the implant, but it introduces the risk of fracture of the stem at the junction between it and the proximal body segment. For that reason, and while awaiting intermediate-term results of more recently introduced designs of this junction, non-modularity has attracted attention, at least for straightforward revision cases.

We review the risks and causes of fracture of tapered titanium modular revision stems and present an argument in favour of the more selective use of modular designs.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):50–3.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1197 - 1203
1 Sep 2015
Kim Y Park J Kim J

A number of studies have reported satisfactory results from the isolated revision of an acetabular component. However, many of these studies reported only the short- to intermediate-term results of heterogeneous bearing surfaces in a mixed age group.

We present our experience of using a ceramic-on-ceramic (CoC) bearing for isolated revision of an uncemented acetabular component in 166 patients (187 hips) who were under the age of 50 years at the time of revision. There were 78 men and 88 women with a mean age of 47.4 years (28 to 49). The most common reason for revision was polyethylene wear and acetabular osteolysis in 123 hips (66%), followed by aseptic loosening in 49 hips (26%).

We report the clinical and radiological outcome, complication rate, and survivorship of this group. The mean duration of follow-up was 15.6 years (11 to 19).

The mean pre-operative Harris hip score was 33 points (1 to 58), and improved to a mean of 88 points (51 to 100) at follow-up. The mean pre-operative total Western Ontario and McMaster Universities Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8 points (9 to 61) post-operatively. Overall, 153 of 166 patients (92%) were satisfied with their outcome. Kaplan–Meier survivorship analysis, with revision or radiological evidence of implant failure (13 patients, 8%) as end-points, was 92% at 15 years (95% confidence interval 0.89 to 0.97).

Isolated revision of a cementless acetabular component using a CoC bearing gives good results in patients under 50 years of age.

Cite this article: Bone Joint J 2015;97-B:1197–1203.


Aims

To investigate the longevity of uncemented fixation of a femoral component in total hip arthroplasty (THA) in patients with Dorr type C proximal femoral morphology.

Patients and Methods

A total of 350 consecutive uncemented THA in 320 patients were performed between 1983 and 1987, by a single surgeon using the Taperloc femoral component. The 63 patients (68 hips) with Dorr type C proximal femoral morphology were the focus of this review. The mean age of the patients was 69 years (24 to 88) and mean follow-up was 16.6 years (ten to 29). Survival analysis included eight patients (eight hips) who died without undergoing revision surgery prior to obtaining ten years follow-up. All 55 surviving patients (60 hips) were available for clinical assessment and radiographic review. As a comparator group, the survival and implant fixation in the remaining 282 THAs (257 patients) with Dorr type A and B morphology were evaluated. The mean age of these patients was 52 years (20 to 82).


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 622 - 627
1 May 2016
Donaldson JR Tudor F Gollish J

Aims

The aim of this study was to examine the results of revision total knee arthroplasty (TKA) undertaken for stiffness in the absence of sepsis or loosening.

Patients and Methods

We present the results of revision surgery for stiff TKA in 48 cases (35 (72.9%) women and 13 (27.1%) men). The mean age at revision surgery was 65.5 years (42 to 83). All surgeries were performed by a single surgeon. Stiffness was defined as an arc of flexion of < 70° or a flexion contracture of > 15°. The changes in the range of movement (ROM) and the Western Ontario and McMasters Osteoarthritis index scores (WOMAC) were recorded.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 217 - 223
1 Feb 2014
Namba RS Inacio MCS Cafri G

The outcome of total knee replacement (TKR) using components designed to increase the range of flexion is not fully understood. The short- to mid-term risk of aseptic revision in high flexion TKR was evaluated. The endpoint of the study was aseptic revision and the following variables were investigated: implant design (high flexion vs non-high flexion), the thickness of the tibial insert (≤ 14 mm vs > 14 mm), cruciate ligament (posterior stabilised (PS) vs cruciate retaining), mobility (fixed vs rotating), and the manufacturer (Zimmer, Smith & Nephew and DePuy). Covariates included patient, implant, surgeon and hospital factors. Marginal Cox proportional hazard models were used.

In a cohort of 64 000 TKRs, high flexion components were used in 8035 (12.5%). The high flexion knees with tibial liners of thickness > 14 mm had a density of revision of 1.45/100 years of observation, compared with 0.37/100 in non-high flexion TKR with liners ≤ 14 mm thick. Relative to a standard fixed PS TKR, the NexGen (Zimmer, Warsaw, Indiana) Gender Specific Female high flexion fixed PS TKR had an increased risk of revision (hazard ratio (HR) 2.27 (95% confidence interval (CI) 1.48 to 3.50)), an effect that was magnified when a thicker tibial insert was used (HR 8.10 (95% CI 4.41 to 14.89)).

Surgeons should be cautious when choosing high flexion TKRs, particularly when thicker tibial liners might be required.

Cite this article: Bone Joint J 2014;96-B:217–23.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 118 - 121
1 Nov 2014
Lachiewicz PF Watters TS

Metaphyseal bone loss is common with revision total knee replacement (RTKR). Using the Anderson Orthopaedic Research Institute (AORI) classification, type 2-B and type 3 defects usually require large metal blocks, bulk structural allograft or highly porous metal cones. Tibial and femoral trabecular metal metaphyseal cones are a unique solution for large bone defects. These cones substitute for bone loss, improve metaphyseal fixation, help correct malalignment, restore the joint line and may permit use of a shorter stem. The technique for insertion involves sculpturing of the remaining bone with a high speed burr and rasp, followed by press-fit of the cone into the metaphysis. The fixation and osteoconductive properties of the porous cone outer surface allow ingrowth and encourage long-term biological fixation. The revision knee component is then cemented into the porous cone inner surface, which provides superior fixation compared with cementing into native but deficient metaphyseal bone. The advantages of the cone compared with allograft include: technical ease, biological fixation, no resorption, and possibly a lower risk of infection. The disadvantages include: difficult extraction and relatively short-term follow-up. Several studies using cones report promising short-term results for the reconstruction of large bone defects in RTKR.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):118–21.