The aims of this study were to determine the
functional impact and financial burden of isolated and recurrent dislocation
after total hip arthroplasty (THA). Our secondary goal was to determine
whether there was a difference between patients who were treated
non-operatively and those who were treated operatively. . We retrospectively reviewed 71 patients who had suffered dislocation
of a primary THA. Their mean age was 67 years (41 to 92) and the
mean follow-up was 3.8 years (2.1 to 8.2). . Because patients with
We treated 34 patients with
Between April 1992 and July 2005, 310 posterior lip augmentation devices were used for the treatment of
We present our experience with a double-mobility
acetabular component in 155 consecutive revision total hip replacements
in 149 patients undertaken between 2005 and 2009, with particular
emphasis on the incidence of further dislocation. The mean age of
the patients was 77 years (42 to 89) with 59 males and 90 females.
In all, five patients died and seven were lost to follow-up. Indications
for revision were aseptic loosening in 113 hips, recurrent instability
in 29, peri-prosthetic fracture in 11 and sepsis in two. The mean
follow-up was 42 months (18 to 68). Three hips (2%) in three patients
dislocated within six weeks of surgery; one of these dislocated
again after one year. All three were managed successfully with closed
reduction. Two of the three dislocations occurred in patients who
had undergone revision for
Aims. Dislocation remains a significant complication after total hip arthroplasty (THA), being the third leading indication for revision. We present a series of acetabular revision using a dual mobility cup (DMC) and compare this with our previous series using the posterior lip augmentation device (PLAD). Methods. A retrospective review of patients treated with either a DMC or PLAD for dislocation in patients with a Charnley THA was performed. They were identified using electronic patient records (EPR). EPR data and radiographs were evaluated to determine operating time, length of stay, and the incidence of complications and
Aims. We investigated the long-term performance of the Tripolar Trident acetabular component used for
Aims. We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. . Methods. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid
We have studied the natural history of a first episode of dislocation after primary total hip replacement (THR) to clarify the incidence of
Aims. To report our experience with trunnion corrosion following metal-on-polyethylene
total hip arthroplasty, in particular to report the spectrum of
presentation and determine the mean time to presentation. Patients and Methods. We report the presenting symptoms and signs, intraoperative findings,
and early results and complications of operative treatment in nine
patients with a mean age of 74 years (60 to 86). The onset of symptoms
was at a mean of seven years (3 to 18) after index surgery. Results. Patients presented with a variety of symptoms including pain,
limp and rash. The preoperative mean serum cobalt level was 7.1 ppb
(2.2 to 12.8) and mean serum chromium level was 2.2 ppb (0.5 to
5.2). Metal artifact reduction sequence (MARS) MRI showed fluid
collection and possible pseudotumour formation in five hips, fluid
collection in two hips, and synovitis/debris in one hip, with no
MRI in one patient. Acetabular revision was performed in three patients,
six patients underwent liner and head exchange only. The postoperative
metal levels decreased in all patients: mean cobalt 0.5 ppb (0 to
1.8) and mean chromium 0.9 ppb (0 to 2.6) at a mean of five months
(3 to 8) postoperatively. Seven patients had good pain relief and
no complications at one year. There were two major complications
requiring reoperation: acute infection at six weeks, for which the
patient required two-stage reimplantation; and
Periprosthetic femoral fracture (PFF) is a potentially
devastating complication after total hip arthroplasty, with historically
high rates of complication and failure because of the technical
challenges of surgery, as well as the prevalence of advanced age
and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty
using a modular, titanium, tapered, conical stem for PFF in a series
of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of
27 patients had an American Society of Anesthesiologists grade of
at least 3. At a mean follow-up of 35 months (4 to 66) the mean
Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly
associated with a poorer OHS. All fractures united and no stem needed
to be revised. Three hips in three patients required further surgery
for infection, recurrent PFF and
Aims. We report on the outcome of the Exeter Contemporary flanged cemented
all-polyethylene acetabular component with a mean follow-up of 12
years (10 to 13.9). This study reviewed 203 hips in 194 patients.
129 hips in 122 patients are still in situ; 66
hips in 64 patients were in patients who died before ten years,
and eight hips (eight patients) were revised. Clinical outcome scores
were available for 108 hips (104 patients) and radiographs for 103
hips (100 patients). Patients and Methods. A retrospective review was undertaken of a consecutive series
of 203 routine primary cemented total hip arthroplasties (THA) in
194 patients. Results. There were no acetabular component revisions for aseptic loosening.
Acetabular revision was undertaken in eight hips. In four hips revision
was necessitated by periprosthetic femoral fractures, in two hips
by
We report the use of porous metal acetabular
revision shells in the treatment of contained bone loss. The outcomes of
53 patients with ≤ 50% acetabular bleeding host bone
contact were compared with a control group of 49 patients with >
50% to 85% bleeding host bone contact. All patients were treated
with the same type of trabecular metal acetabular revision shell.
The mean age at revision was 62.4 years (42 to 80) and the mean
follow-up of both groups was 72.4 months (60 to 102). Clinical,
radiological and functional outcomes were assessed. There were four
(7.5%) mechanical failures in the ≤ 50% host bone contact
group and no failures in the >
50% host bone contact group (p = 0.068).
Out of both groups combined there were four infections (3.9%) and
five
In this retrospective study, we investigated
the results of revision total hip replacement (THR) using a cemented long-stemmed
Exeter femoral component, with a minimum length of 205 mm in patients
with extensive femoral bone defects. The study included 37 consecutive
patients with a mean age of 76 years (39 to 93) and a mean follow-up
of nine years (5 to 16). A total of 26 patients (70%) had a pre-operative
Endo-Klinik score of 3 or 4. Impaction bone grafting was used in
24 patients (65%). At the time of evaluation, 22 patients (59%)
were still alive and were evaluated clinically and radiologically.
A total of 14 patients died during follow-up and their data were
included until the time of their death. One reconstruction failed
after five years and five months owing to
We present the medium-term results of hybrid total hip arthroplasties using pre-coated stems with a second-generation cementing technique. The 128 hips in 111 patients (18 men and 93 women) were followed up at a mean of 11 years after surgery. The mean age at the time of surgery was 61 years. Both components of one hip were removed at ten months after surgery for infection. None of the other 127 femoral components showed possible, probable, or definite loosening at the most recent follow-up. Five acetabular components were revised for aseptic loosening,
The prevalence of obesity is increasing substantially around the world. Elevated BMI increases the risk of complications following total hip arthroplasty (THA). We sought to evaluate trends in BMI and complication rates of obese patients undergoing primary THA over the last 30 years. Through our institutional total joint registry, we identified 15,455 primary THAs performed for osteoarthritis from 1990 to 2019. Patients were categorized according to the World Health Organization (WHO) obesity classification and groups were trended over time. Cox proportional hazards regression analysis controlling for confounders was used to investigate the association between year of surgery and two-year risk of any reoperation, any revision, dislocation, periprosthetic joint infection (PJI), venous thromboembolism (VTE), and periprosthetic fracture. Regression was stratified by three separate groups: non-obese; WHO Class I and Class II (BMI 30 to 39 kg/m2); and WHO Class III patients (BMI ≥ 40 kg/m2).Aims
Methods
A reduced femoral offset in total hip replacement has been thought to be disadvantageous. We reviewed the results of 54 consecutive primary total hip replacements in 49 patients (mean age of 68 years) performed between August 1990 and December 1994, with a mean follow-up of 8.8 years (. sd. 2.1). The mean pre-operative femoral offset for these hips was 41 mm (. sd. 7.4). All patients received a low-offset Charnley stem and a polyethylene cup inserted by a single surgeon. At their latest follow-up, surviving patients had a significant improvement in the performance of their hip. Three had undergone revision, one each for deep infection,
Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20). The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for
Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases,
Between 1980 and 2000, 63 support rings were used in the management of acetabular deficiency in a series of 60 patients, with a mean follow-up of 8.75 years (2 months to 23.8 years). There was a minimum five-year follow-up for successful reconstructions. The indication for revision surgery was aseptic loosening in 30 cases and infection in 33. All cases were Paprosky III defects; IIIA in 33 patients (52.4%) and IIIB in 30 (47.6%), including four with pelvic dissociation. A total of 26 patients (43.3%) have died since surgery, and 34 (56.7%) remain under clinical review. With acetabular revision for infection or aseptic loosening as the definition of failure, we report success in 53 (84%) of the reconstructions. A total of 12 failures (19%) required further surgery, four (6.3%) for aseptic loosening of the acetabular construct, six (9.5%) for recurrent infection and two (3.2%) for
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) vs 5.8%
(6 of 104), p = 0.569) or re-operation rates without revision (5.1%
(5 of 99) vs 2.9% (3 of 104), p = 0.428). In the
THR group one