Advertisement for orthosearch.org.uk
Results 1 - 20 of 74
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1502 - 1506
1 Nov 2005
Arora J Ogden AC

We analysed at a mean follow-up of 7.25 years the clinical and radiological outcome of 117 patients (125 knees) who had undergone a primary, cemented, modular Freeman-Samuelson total knee replacement. While the tibial and femoral components were cemented, the patellar component was uncemented. A surface-cementing technique was used to secure the tibial components. A total of 82 knees was available for radiological assessment. Radiolucent lines were seen in 41 knees (50%) and osteolytic lesions were seen in 13 knees (16%). Asymptomatic, rotational loosening of the patellar implant was seen in four patients and osteolysis was more common in patients with a patellar resurfacing. Functional outcome scores were available for 41 patients (41 knees, 35%) and the mean Western Ontario McMasters Universities score was 77.5 (. sd. 19.5) and the cumulative survival was 93.4% at ten years with revision for aseptic loosening as an endpoint. Increased polyethylene wear from modular components, a rotationally-loose patella, and the surface-cementing technique may have contributed to the high rate of osteolysis seen in our study


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 609 - 614
1 May 2007
Himanen A Belt EA Lehto MUK Hämäläinen MMJ

We evaluated the survival of moulded monoblock and modular tibial components of the AGC total knee replacement in patients with rheumatoid arthritis. Between 1985 and 1995, 751 knees with this diagnosis were replaced at our institution. A total of 256 tibial components were of the moulded design and 495 of the modular design. The mean follow-up of the moulded subgroup was 9.6 years (0.5 to 14.7), and that of the modular group 7.0 years (0.1 to 14.7). The groups differed significantly from each other in Larsen grade, cementing of components and patellar resurfacing, but no statistically significant difference between the survival of the components was found (Log rank test, p = 0.91). The cumulative success rate of the moulded group was 96.8% (95% confidence interval 93.6% to 98.4%) at five years and 94.4% (95% confidence interval 90.4% to 96.7%) at ten years, and of the modular group 96.2% (95% confidence interval 94% to 97.6%) and 93.6% (95% confidence interval 89.7% to 96%), respectively. Revision was required in 37 total knee replacements, the main causes were infection, pain, loosening of the tibial component and patellar problems. Survival rates for both components were satisfactory


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 376 - 385
1 Mar 2022
Gramlich Y Hofmann L Kress S Ruckes C Kemmerer M Klug A Hoffmann R Kremer M

Aims

This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant.

Methods

A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 137 - 144
1 Jun 2021
Lachiewicz PF Steele JR Wellman SS

Aims

To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period.

Methods

This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1594 - 1596
1 Dec 2008
Dalury DF Barrett WP Mason JB Goldstein WM Murphy JA Roche MW

This retrospective study evaluated the midterm clinical and radiographic outcomes of a second-generation total knee replacement system. In a multicentre consecutive series of 1512 patients, 1970 knees were treated with the PFC Sigma knee system (Depuy, Warsaw, Indiana). The patients were reviewed for functional outcome, and underwent independent radiographic evaluation at a mean follow-up of 7.3 years (5 to 10). A total of 40 knees (2%) required revision, 17 (0.9%) for infection. The incidence of osteolysis was 2.2%. The ten-year survival with revision for any cause other than infection as the endpoint was 97.2% (95% CI 95.4 to 99.1).

The PFC Sigma knee system appears to provide excellent results in the medium term.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims. Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. Methods. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate. Results. Five patient-specific variables and seven implant-specific variables were evaluated for significant association with lower insert wear rate. Six were significant when controlling for other factors: greater patient age, female sex, shorter duration in vivo, polished tray, highly cross-linked PE (HXLPE), and constrained knee design. Conclusion. This study confirmed that knee wear rate increased with duration in vivo. Older patients and females had significantly lower wear rates. Polished modular tibial tray surfaces, HXLPE, and constrained TKA designs were device design factors associated with significantly reduced wear rate. Cite this article: Bone Joint J 2021;103-B(11):1695–1701


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 33 - 39
1 Jul 2019
Lachiewicz PF O’Dell JA

Aims. There is insufficient evidence to recommend the use of alternative polyethylene bearings in modular, fixed-bearing total knee arthroplasty (TKA). The purpose of this study was to compare standard polyethylene (SP) and highly crosslinked polyethylene (XLP) tibial liners in posterior-stabilized TKA, with osteolysis as the primary outcome and clinical results and the rate of re-operation as the secondary outcomes. Patients and Methods. This is a single-surgeon, prospective randomized study involving one design of modular posterior-stabilized TKA. An analysis of 122 TKAs with an SP compression moulded liner and 123 with an XLP liner was performed, with a mean follow-up of six years (2 to 11). Patients were evaluated clinically using the Knee Society score, Lower Extremity Activity Score (LEAS), and the presence of an effusion, and standard radiographs were assessed for radiolucent lines and osteolytic lesions. Results. Osteolysis was present in four TKAs (3.3%) in the SP group, and no knees in the XLP group (p = 0.06). There were no significant differences between the Knee Society total score, change in total score, knee function score, change in function score, LEAS, and change in LEAS in the two groups. There was a significant difference in the presence of an effusion (10/122 with SP liners, 1/123 with XLP liners; p = 0.02). There was no significant difference in the rate of re-operation between the two groups (p = 0.36). There were no complications related to the XLP liner. Conclusion. At this length of follow-up, there were no advantages and no complications related to the use of this XLP tibial liner. The presence of effusion and small osteolytic lesions was more frequent with SP than XLP liners, but of unknown clinical significance. Cite this article: Bone Joint J 2019;101-B(7 Supple C):33–39


The Bone & Joint Journal
Vol. 106-B, Issue 7 | Pages 669 - 679
1 Jul 2024
Schnetz M Maluki R Ewald L Klug A Hoffmann R Gramlich Y

Aims

In cases of severe periprosthetic joint infection (PJI) of the knee, salvage procedures such as knee arthrodesis (KA) or above-knee amputation (AKA) must be considered. As both treatments result in limitations in quality of life (QoL), we aimed to compare outcomes and factors influencing complication rates, mortality, and mobility.

Methods

Patients with PJI of the knee and subsequent KA or AKA between June 2011 and May 2021 were included. Demographic data, comorbidities, and patient history were analyzed. Functional outcomes and QoL were prospectively assessed in both groups with additional treatment-specific scores after AKA. Outcomes, complications, and mortality were evaluated.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1317 - 1323
1 Oct 2007
Kim Y Yoon S Kim J

We compared the results of 146 patients who received an anatomic modular knee fixed-bearing total knee replacement (TKR) in one knee and a low contact stress rotating platform mobile-bearing TKR in the other. There were 138 women and eight men with a mean age of 69.8 years (42 to 80). The mean follow-up was 13.2 years (11.0 to 14.5). The patients were assessed clinically and radiologically using the rating systems of the Hospital for Special Surgery and the Knee Society at three months, six months, one year, and annually thereafter. The assessment scores of both rating systems pre-operatively and at the final review did not show any statistically significant differences between the two designs of implant. In the anatomic modular knee group, one knee was revised because of aseptic loosening of the tibial component and one because of infection. In addition, three knees were revised because of wear of the polyethylene tibial bearing. In the low contact stress group, two knees were revised because of instability requiring exchange of the polyethylene insert and one because of infection. The radiological analysis found no statistical difference in the incidence of radiolucent lines at the final review (Student’s t-test, p = 0.08), most of which occurred at tibial zone 1. The Kaplan-Meier survivorship for aseptic loosening of the anatomic modular knee and the low contact stress implants at 14.5 years was 99% and 100%, respectively, with a 95% confidence interval of 94% to 100% for both designs. We found no evidence of the superiority of one design over the other at long-term follow-up


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1025 - 1032
1 Aug 2020
Hampton M Mansoor J Getty J Sutton PM

Aims. Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Methods. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m. 2. (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up. Results. At final review all patients were between 11 and 15 years following surgery. In total, 41 of the knees were cemented and 36 uncemented. There were no revisions in the cemented group and one revision in the uncemented group for fracture. The uncemented group reported better outcomes with both statistically and clinically significant (p = 0.001) improvements in knee-specific Oxford and Knee Society scores compared with the cemented group. The global SF-12 scores demonstrated no statistical difference (p = 0.812). Uncemented knees had better radiological analysis compared with the cemented group (p < 0.001). Conclusion. Use of an uncemented trabecular metal tibial implant can afford better long-term clinical outcomes when compared to cemented tibial components of a matched design. However, both have excellent survivorship up to 15 years after implantation. Cite this article: Bone Joint J 2020;102-B(8):1025–1032


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 47 - 55
1 Jan 2023
Clement ND Avery P Mason J Baker PN Deehan DJ

Aims

The aim of this study was to identify variables associated with time to revision, demographic details associated with revision indication, and type of prosthesis employed, and to describe the survival of hinge knee arthroplasty (HKA) when used for first-time knee revision surgery and factors that were associated with re-revision.

Methods

Patient demographic details, BMI, American Society of Anesthesiologists (ASA) grade, indication for revision, surgical approach, surgeon grade, implant type (fixed and rotating), time of revision from primary implantation, and re-revision if undertaken were obtained from the National Joint Registry data for England, Wales, Northern Ireland, and the Isle of Man over an 18-year period (2003 to 2021).


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 647 - 652
1 May 2017
Abdel MP Ledford CK Kobic A Taunton MJ Hanssen AD

Aims. The number of revision total knee arthroplasties (TKA) that are performed is expected to increase. However, previous reports of the causes of failure after TKA are limited in that they report the causes at specific institutions, which are often dependent on referral patterns. Our aim was to report the most common indications for re-operations and revisions in a large series of posterior-stabilised TKAs undertaken at a single institution, excluding referrals from elsewhere, which may bias the causes of failure. Patients and Methods. A total of 5098 TKAs which were undertaken between 2000 and 2012 were included in the study. Re-operations, revisions with modular component exchange, and revisions with non-modular component replacement or removal were identified from the medical records. The mean follow-up was five years (two to 12). Results. The Kaplan-Meier ten-year survival without a re-operation, modular component revision and non-modular component revision was 95.7%, 99.3% and 95.3%, respectively. The most common indications for a re-operation were: post-operative stiffness (58%), delayed wound healing (21%), and patellar clunk (11%). The indications for isolated modular component revision were acute periprosthetic joint infection (PJI) (64%) and instability (36%). The most common indications for non-modular component revision were chronic PJI (52%), aseptic loosening (17%), periprosthetic fracture (10%), and instability (10%). Conclusion. Post-operative stiffness remains the most common indication for re-operation after TKA. Infection is the most common indication for modular and non-modular component revision. Aseptic loosening was not an uncommon cause of failure, however, it was much less common than in national registry and non-registry data. Focusing on posterior-stabilised TKAs initially performed at our institution allowed for an accurate assessment of the causes of failure in a contemporary specialty practice. Cite this article: Bone Joint J 2017;99-B:647–52


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 875 - 883
1 Jul 2022
Mills K Wymenga AB van Hellemondt GG Heesterbeek PJC

Aims

Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA).

Methods

This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction).


Aims

The aim of this study was to compare any differences in the primary outcome (biphasic flexion knee moment during gait) of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) at one year post-surgery.

Methods

A total of 76 patients (34 bi-UKA and 42 TKA patients) were analyzed in a prospective, single-centre, randomized controlled trial. Flat ground shod gait analysis was performed preoperatively and one year postoperatively. Knee flexion moment was calculated from motion capture markers and force plates. The same setup determined proprioception outcomes during a joint position sense test and one-leg standing. Surgery allocation, surgeon, and secondary outcomes were analyzed for prediction of the primary outcome from a binary regression model.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 786 - 792
1 Jun 2015
Hutt JRB Farhadnia P Massé V Lavigne M Vendittoli P

This randomised trial evaluated the outcome of a single design of unicompartmental arthroplasty of the knee (UKA) with either a cemented all-polyethylene or a metal-backed modular tibial component. A total of 63 knees in 45 patients (17 male, 28 female) were included, 27 in the all-polyethylene group and 36 in the metal-backed group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and two metal-backed components were revised (at one and five years). One revision in both groups was for unexplained pain, one in the metal-backed group was for progression of osteoarthritis. The others in the all-polyethylene group were for aseptic loosening. The survivorship at seven years calculated by the Kaplan–Meier method for the all-polyethylene group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at risk 16) This difference was statistically significant (p <  0.001). At the most recent follow-up, significantly better mean Western Ontario and McMaster Universities Arthritis Index Scores were found in the all-polyethylene group (13.4 vs 23.0, p = 0.03) but there was no difference in the mean Knee injury and Osteoarthritis Outcome scores (68.8; 41.4 to 99.0 vs 62.6; 24.0 to 100.0), p = 0.36). There were no significant differences for range of movement (p = 0.36) or satisfaction (p = 0.23). This randomised study demonstrates that all-polyethylene components in this design of fixed bearing UKA had unsatisfactory results with significantly higher rates of failure before ten years compared with the metal-back components. Cite this article: Bone Joint J 2015;97-B:786–92


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 444 - 451
1 Apr 2022
Laende EK Mills Flemming J Astephen Wilson JL Cantoni E Dunbar MJ

Aims

Thresholds of acceptable early migration of the components in total knee arthroplasty (TKA) have traditionally ignored the effects of patient and implant factors that may influence migration. The aim of this study was to determine which of these factors are associated with overall longitudinal migration of well-fixed tibial components following TKA.

Methods

Radiostereometric analysis (RSA) data over a two-year period were available for 419 successful primary TKAs (267 cemented and 152 uncemented in 257 female and 162 male patients). Longitudinal analysis of data using marginal models was performed to examine the associations of patient factors (age, sex, BMI, smoking status) and implant factors (cemented or uncemented, the size of the implant) with maximum total point motion (MTPM) migration. Analyses were also performed on subgroups based on sex and fixation.


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1555 - 1560
4 Oct 2021
Phillips JRA Tucker K

Aims

Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations.

Methods

Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated.


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 840 - 845
1 May 2021
Rossi SMP Perticarini L Clocchiatti S Ghiara M Benazzo F

Aims

In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up.

Methods

A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 185 - 190
1 Jun 2021
Kildow BJ Patel SP Otero JE Fehring KA Curtin BM Springer BD Fehring TK

Aims

Debridement, antibiotics, and implant retention (DAIR) remains one option for the treatment of acute periprosthetic joint infection (PJI) despite imperfect success rates. Intraosseous (IO) administration of vancomycin results in significantly increased local bone and tissue concentrations compared to systemic antibiotics alone. The purpose of this study was to evaluate if the addition of a single dose of IO regional antibiotics to our protocol at the time of DAIR would improve outcomes.

Methods

A retrospective case series of 35 PJI TKA patients, with a median age of 67 years (interquartile range (IQR) 61 to 75), who underwent DAIR combined with IO vancomycin (500 mg), was performed with minimum 12 months' follow-up. A total of 26 patients with primary implants were treated for acute perioperative or acute haematogenous infections. Additionally, nine patients were treated for chronic infections with components that were considered unresectable. Primary outcome was defined by no reoperations for infection, nor clinical signs or symptoms of PJI.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 484 - 489
1 Apr 2011
Charalambous CP Abiddin Z Mills SP Rogers S Sutton P Parkinson R

The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed mobile polyethylene bearing. We present the early results of the use of this prosthesis for established isolated patellofemoral arthritis in 51 consecutive patellofemoral replacements in 35 patients. The mean follow-up was 25 months (5 to 60). The estimated survival rate at three years was 63% (95% confidence interval 47 to 80) with revision as the endpoint and 46% (95% confidence interval 30 to 63) with revision and ongoing moderate or severe pain as the endpoint. The early results of the use of the low contact stress patellofemoral replacement are disappointing with a high rate of revision. We cannot therefore recommend its use