Aims. We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following
Aims. To report our experience with trunnion corrosion following metal-on-polyethylene
total hip arthroplasty, in particular to report the spectrum of
presentation and determine the mean time to presentation. Patients and Methods. We report the presenting symptoms and signs, intraoperative findings,
and early results and complications of operative treatment in nine
patients with a mean age of 74 years (60 to 86). The onset of symptoms
was at a mean of seven years (3 to 18) after index surgery. Results. Patients presented with a variety of symptoms including pain,
limp and rash. The preoperative mean serum cobalt level was 7.1 ppb
(2.2 to 12.8) and mean serum chromium level was 2.2 ppb (0.5 to
5.2). Metal artifact reduction sequence (MARS) MRI showed fluid
collection and possible pseudotumour formation in five hips, fluid
collection in two hips, and synovitis/debris in one hip, with no
MRI in one patient. Acetabular revision was performed in three patients,
six patients underwent liner and head exchange only. The postoperative
metal levels decreased in all patients: mean cobalt 0.5 ppb (0 to
1.8) and mean chromium 0.9 ppb (0 to 2.6) at a mean of five months
(3 to 8) postoperatively. Seven patients had good pain relief and
no complications at one year. There were two major complications
requiring reoperation: acute infection at six weeks, for which the
patient required two-stage reimplantation; and recurrent dislocation,
for which the patient was revised to a dual mobility component. Conclusion. Trunnion corrosion in
Aims. In
The long term biological effects of wear products
following total hip arthroplasty (THA) are unclear. However, the indications
for THA are expanding, with increasingly younger patients undergoing
the procedure. This prospective, randomised study compared two groups of patients
undergoing THA after being randomised to receive one of two different
bearing surfaces:
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional
A 70-year-old man with an uncemented metal-on-polyethylene
total hip prosthesis underwent revision arthroplasty 33 months later
because of pain, swelling and recurrent dislocation. There appeared
to be corrosion and metal release from the prosthetic head, resulting
in pseudotumour formation and severe local soft-tissue destruction.
The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron
taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.
Aims
Patients and Methods
Aims. To analyze the short-term outcome of two types of total wrist arthroplasty (TWA) in terms of wrist function, migration, and periprosthetic bone behaviour. Methods. A total of 40 patients suffering from non-rheumatoid wrist arthritis were enrolled in a randomized controlled trial comparing the ReMotion and Motec TWAs. Patient-rated and functional outcomes, radiological changes, blood metal ion levels, migration measured by model-based radiostereometric analysis (RSA), bone mineral density (BMD) measured by dual-energy X-ray absorptiometry (DXA), complications, loosening, and revision rates at two years were compared. Results. Patient-Rated Wrist and Hand Evaluation (PRWHE) scores, abbreviated version of the Disabilities of the Arm, Shoulder and Hand questionnaire (QuickDASH) scores, and pain improved similarly and significantly in both groups. Wrist motion improved significantly in the Motec group only, and forearm rotation in the ReMotion group only. Cobalt (Co) and chromium (Cr) blood ion levels were significantly higher in the metal-on-metal (MoM) Motec group than in the
The biological significance of cobalt-chromium wear particles from metal-on-metal hip replacements may be different to the effects of the constituent metal ions in solution. Bacteria may be able to discriminate between particulate and ionic forms of these metals because of a transmembrane nickel/cobalt-permease. It is not known whether wear particles are bacteriocidal. We compared the doubling time of coagulase negative staphylococcus, Staphylococcus aureus and methicillin resistant S. aureus when cultured in either wear particles from a metal-on-metal hip simulator, wear particles from a
We have studied the relationship between metal ion levels and lymphocyte counts in patients with metal-on-metal hip resurfacings. Peripheral blood samples were analysed for lymphocyte subtypes and whole blood cobalt and chromium ion levels in 68 patients (34 with metal-on-metal hip resurfacings and 34 with standard
Dislocation is a common reason for revision following total hip replacement. This study investigated the relationship between the bearing surface and the risk of revision due to dislocation. It was based on 110 239 primary total hip replacements with a diagnosis of osteoarthritis collected by the Australian Orthopaedic Association National Joint Replacement Registry between September 1999 and December 2007. A total of 862 (0.78%) were revised because of dislocation. Ceramic-on-ceramic bearing surfaces had a lower risk of requiring revision due to dislocation than did
Previous research has shown an increase in chromosomal aberrations in patients with worn implants. The type of aberration depended on the type of metal alloy in the prosthesis. We have investigated the metal-specific difference in the level of DNA damage (DNA stand breaks and alkali labile sites) induced by culturing human fibroblasts in synovial fluid retrieved at revision arthroplasty. All six samples from revision cobalt-chromium metal-on-metal and four of six samples from cobalt-chromium
Aims. There are limited published data detailing the volumetric material loss from tapers of conventional
Aims. The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. Patients and Methods. A descriptive observational study was undertaken using data from the National Joint Registry (NJR) for England, Wales, Northern Ireland and the Isle of Man between April 2003 and March 2017. All patients aged ≤ 20 years at the time of THA were included and the primary outcome was revision surgery. Descriptive statistics were used to summarize the data and Kaplan–Meier estimates calculated for the cumulative implant survival. Results. A total of 769 THAs were performed in 703 patients. The median follow-up was 5.1 years (interquartile range (IQR) 2.6 to 7.8). Eight patients died and 35 THAs were revised. The use of metal-on-metal (MoM) bearings and resurfacing procedures declined after 2008. The most frequently recorded indications for revision were loosening (20%) and infection (20%), although the absolute risk of these events occurring was low (0.9%). Factors associated with lower implant survival were MoM and
Aims. Periprosthetic joint infection (PJI) is a serious complication
of total hip arthroplasty (THA). Different bearing surface materials
have different surface properties and it has been suggested that
the choice of bearing surface may influence the risk of PJI after
THA. The objective of this meta-analysis was to compare the rate
of PJI between
Aims. Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of
Symptomatic cobalt toxicity from a failed total
hip replacement is a rare but devastating complication. It has been reported
following revision of fractured ceramic components, as well as in
patients with failed metal-on-metal articulations. Potential clinical
findings include fatigue, weakness, hypothyroidism, cardiomyopathy,
polycythaemia, visual and hearing impairment, cognitive dysfunction,
and neuropathy. We report a case of an otherwise healthy 46-year-old
patient, who developed progressively worsening symptoms of cobalt
toxicity beginning approximately six months following synovectomy
and revision of a fractured ceramic-on-ceramic total hip replacement
to a
Aims. We present a case series of ten
We carried out a cross-sectional study with analysis of the demographic, clinical and laboratory characteristics of patients with metal-on-metal hip resurfacing, ceramic-on-ceramic and
Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey. NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.Aims
Methods