Advertisement for orthosearch.org.uk
Results 1 - 7 of 7
Results per page:
The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 263 - 269
1 Feb 2014
Batta V Coathup MJ Parratt MT Pollock RC Aston WJ Cannon SR Skinner JA Briggs TW Blunn GW

We reviewed the outcome of 69 uncemented, custom-made, distal femoral endoprosthetic replacements performed in 69 patients between 1994 and 2006. There were 31 women and 38 men with a mean age at implantation of 16.5 years (5 to 37). All procedures were performed for primary malignant bone tumours of the distal femur. At a mean follow-up of 124.2 months (4 to 212), 53 patients were alive, with one patient lost to follow-up. All nine implants (13.0%) were revised due to aseptic loosening at a mean of 52 months (8 to 91); three implants (4.3%) were revised due to fracture of the shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic fracture. Bone remodelling associated with periosteal cortical thinning adjacent to the uncemented intramedullary stem was seen in 24 patients but this did not predispose to failure. All aseptically loose implants in this series were diagnosed to be loose within the first five years.

The results from this study suggest that custom-made uncemented distal femur replacements have a higher rate of aseptic loosening compared to published results for this design when used with cemented fixation. Loosening of uncemented replacements occurs early indicating that initial fixation of the implant is crucial.

Cite this article: Bone Joint J 2014;96-B:263–9.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 522 - 529
1 Mar 2021
Nichol T Callaghan J Townsend R Stockley I Hatton PV Le Maitre C Smith TJ Akid R

Aims. The aim of this study was to develop a single-layer hybrid organic-inorganic sol-gel coating that is capable of a controlled antibiotic release for cementless hydroxyapatite (HA)-coated titanium orthopaedic prostheses. Methods. Coatings containing gentamicin at a concentration of 1.25% weight/volume (wt/vol), similar to that found in commercially available antibiotic-loaded bone cement, were prepared and tested in the laboratory for: kinetics of antibiotic release; activity against planktonic and biofilm bacterial cultures; biocompatibility with cultured mammalian cells; and physical bonding to the material (n = 3 in all tests). The sol-gel coatings and controls were then tested in vivo in a small animal healing model (four materials tested; n = 6 per material), and applied to the surface of commercially pure HA-coated titanium rods. Results. The coating released gentamicin at > 10 × minimum inhibitory concentration (MIC) for sensitive staphylococcal strains within one hour thereby potentially giving effective prophylaxis for arthroplasty surgery, and showed > 99% elution of the antibiotic within the coating after 48 hours. There was total eradication of both planktonic bacteria and established bacterial biofilms of a panel of clinically relevant staphylococci. Mesenchymal stem cells adhered to the coated surfaces and differentiated towards osteoblasts, depositing calcium and expressing the bone marker protein, osteopontin. In the in vivo small animal bone healing model, the antibiotic sol-gel coated titanium (Ti)/HA rod led to osseointegration equivalent to that of the conventional HA-coated surface. Conclusion. In this study we report a new sol-gel technology that can release gentamicin from a bioceramic-coated cementless arthroplasty material. In vitro, local gentamicin levels are in excess of what can be achieved by antibiotic-loaded bone cement. In vivo, bone healing in an animal model is not impaired. This, thus, represents a biomaterial modification that may have the potential to protect at-risk patients from implant-related deep infection. Cite this article: Bone Joint J 2021;103-B(3):522–529


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1248 - 1255
1 Oct 2019
Pineda A Pabbruwe MB Kop AM Vlaskovsky P Hurworth M

Aims. The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Materials and Methods. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed. Results. In 104 implants, a decrease in the odds of loosening and an increase in metallosis and tissue staining in the Duofix group relative to Porocoat group was detected (p = 0.028). There was an increased presence of embedded metallic debris in the Duofix group (p < 0.001). Decreased tissue ingrowth was associated with the Duofix surface (p < 0.001). The attached beads had reduced microhardness, indicative of adverse thermal processing, which resulted in bead shedding, particulate debris, and metallosis. Conclusion. Hydroxyapatite coating of the LCS femoral component produced unexpected results and led to its recall. The root cause was likely a combination of retained alumina grit and a reduction in bead microhardness (mechanical strength) resulting in increased particle debris, metallosis, and early revision. The Duofix LCS femoral component was not equivalent to the Porocoat version despite its approval through the Food and Drug Administration (FDA) 510(k) equivalance approval process. Regulation of the introduction of modified existing devices needs to be improved and the Duofix LCS should have been considered to be a new device for which equivalence had not been demonstrated at the point of introduction. Cite this article: Bone Joint J 2019;101-B:1248–1255


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 85 - 91
1 Jan 1996
Önsten I Carlsson ÅS Sanzén L Besjakov J

A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 3 | Pages 351 - 360
1 May 1997
Donnelly WJ Kobayashi A Freeman MAR Chin TW Yeo H West M Scott G

We compared the radiological appearances and survival of four methods of fixation of a femoral stem in 538 hips after follow-up for five or ten years. The fixation groups were: 1) press-fit shot-blasted smooth Ti-Al-V stem; 2) press-fit shot-blasted proximally ridged stem; 3) proximal hydroxyapatite (HA) coating; and 4) cementing. Survival analysis at five to ten years showed better results in the HA-coated (100% at five to six years) and cemented stems (100% at 5 to 6 years) than in the two press-fit groups. There was a higher mean rate of migration in the smooth and ridged Ti-Al-V shot-blasted press-fit groups (0.8 mm/year and 0.6 mm/year, respectively) when compared with the HA-coated and cemented prostheses (both 0.3 mm/year). More radiolucent lines and osteolytic lesions were seen in the press-fit groups than in either the HA-coated or cemented implants, with a trend for a lower incidence of both in the HA compared with the cemented group. Proximal osteopenia increased in the press-fit and cemented prostheses with time, but did not do so in the HA group. There was a higher incidence of resorption of the femoral neck with time in the cemented group than in the other three. We conclude that the HA and the cemented interfaces both provide secure fixation with a trend in favour of HA. The cemented prosthesis meets the suggested National Institutes of Health definition of ‘efficacious’ at ten years


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 455 - 460
1 Apr 2018
Mumith A Thomas M Shah Z Coathup M Blunn G

Increasing innovation in rapid prototyping (RP) and additive manufacturing (AM), also known as 3D printing, is bringing about major changes in translational surgical research.

This review describes the current position in the use of additive manufacturing in orthopaedic surgery.

Cite this article: Bone Joint J 2018;100-B:455-60.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1360 - 1365
1 Oct 2009
Sewell MD Spiegelberg BGI Hanna SA Aston WJS Meswania JM Blunn GW Henry C Cannon SR Briggs TWR

We describe the application of a non-invasive extendible endoprosthetic replacement in skeletally-mature patients undergoing revision for failed joint replacement with resultant limb-length inequality after malignant or non-malignant disease. This prosthesis was developed for tumour surgery in skeletally-immature patients but has now been adapted for use in revision procedures to reconstruct the joint or facilitate an arthrodesis, replace bony defects and allow limb length to be restored gradually in the post-operative period.

We record the short-term results in nine patients who have had this procedure after multiple previous reconstructive operations. In six, the initial reconstruction had been performed with either allograft or endoprosthetic replacement for neoplastic disease and in three for non-neoplastic disease. The essential components of the prosthesis are a magnetic disc, a gearbox and a drive screw which allows painless lengthening of the prosthesis using the principle of electromagnetic induction. The mean age of the patients was 37 years (18 to 68) with a mean follow-up of 34 months (12 to 62). They had previously undergone a mean of six (2 to 14) open procedures on the affected limb before revision with the non-invasive extendible endoprosthesis.

The mean length gained was 56 mm (19 to 107) requiring a mean of nine (3 to 20) lengthening episodes performed in the outpatient department. There was one case of recurrent infection after revision of a previously infected implant and one fracture of the prosthesis after a fall. No amputations were performed. Planned exchange of the prosthesis was required in three patients after attainment of the maximum lengthening capacity of the implant. There was no failure of the lengthening mechanism. The Mean Musculoskeletal Tumour Society rating score was 22 of 30 available points (18 to 28).

The use of a non-invasive extendible endoprosthesis in this manner provided patients with good functional results and restoration of leg-length equality, without the need for multiple open lengthening procedures.