We describe the clinical and radiological results
of cementless primary total hip replacement (THR) in 25 patients
(18 women and seven men; 30 THRs) with severe developmental dysplasia
of the hip (DDH). Their mean age at surgery was 47 years (23 to
89). In all, 21 hips had Crowe type III dysplasia and nine had Crowe
type IV. Cementless acetabular components with standard polyethylene
liners were introduced as close to the level of the true acetabulum
as possible. The modular cementless S-ROM femoral component was
used with a low resection of the femoral neck. A total of 21 patients (25 THRs) were available for review at
a mean follow-up of 18.7 years (15.8 to 21.8). The mean modified
Harris hip score improved from 46 points pre-operatively to 90 at
final follow up (p <
0.001). A total of 15 patients (17 THRs; 57%) underwent revision of the
acetabular component at a mean of 14.6 years (7 to 20.8), all for
osteolysis. Two patients (two THRs) had symptomatic loosening. No
patient underwent femoral revision. Survival with revision of either
component for any indication was 81% at 15 years (95% CI 60.1 to
92.3), with 21 patients at risk. This technique may reduce the need for femoral osteotomy in severe
DDH, while providing a good long-term functional result. Cite this article:
Aims. Pelvic discontinuity is a challenging acetabular defect without a consensus on surgical management. Cup-cage reconstruction is an increasingly used treatment strategy. The present study evaluated implant survival, clinical and radiological outcomes, and complications associated with the cup-cage construct. Methods. We included 53 cup-cage construct (51 patients) implants used for hip revision procedures for pelvic discontinuity between January 2003 and January 2022 in this retrospective review. Mean age at surgery was 71.8 years (50.0 to 92.0; SD 10.3), 43/53 (81.1%) were female, and mean follow-up was 6.4 years (0.02 to 20.0; SD 4.6). Patients were implanted with a Trabecular Metal Revision Shell with either a ZCA cage (n = 12) or a TMARS cage (n = 40, all Zimmer Biomet). Pelvic discontinuity was diagnosed on preoperative radiographs and/or intraoperatively. Kaplan-Meier survival analysis was performed, with failure defined as revision of the cup-cage reconstruction. Results. The five-year all-cause survival for cup-cage reconstruction was 73.4% (95% confidence interval (CI) 61.4 to 85.4), while the ten- and 15-year survival was 63.7% (95% CI 46.8 to 80.6). Survival due to aseptic loosening was 93.4% (95% CI 86.2 to 100.0) at five, ten, and 15 years. The rate of revision for aseptic loosening, infection, and dislocation was 3/53 (5.7%), 7/53 (13.2%), and 6/53 (11.3%), respectively. The mean leg length discrepancy improved (p < 0.001) preoperatively from a mean of 18.2 mm (0 to 80; SD 15.8) to 7.0 mm (0 to 35; SD 9.8) at latest follow-up. The horizontal and vertical
Aims. Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the
Our aim was to determine if the height of the cup, lateralisation or the abduction angle correlated with functional outcome or survivorship in revision total hip replacement in patients with a previous diagnosis of developmental dysplasia of the hip. A retrospective investigation of 51 patients (63 hips) who had undergone revision total hip replacement was performed. The mean duration of follow-up was 119 months. Forty-one patients (52 hips) were available for both determination of functional outcome and survivorship analysis. Ten patients (11 hips) were only available for survivorship analysis. The height of the cup was found to have a statistically significant correlation with functional outcome and a high
Aims. Bone stock restoration of acetabular bone defects using impaction bone grafting (IBG) in total hip arthroplasty may facilitate future re-revision in the event of failure of the reconstruction. We hypothesized that the acetabular bone defect during re-revision surgery after IBG was smaller than during the previous revision surgery. The clinical and radiological results of re-revisions with repeated use of IBG were also analyzed. Methods. In a series of 382 acetabular revisions using IBG and a cemented component, 45 hips (45 patients) that had failed due to aseptic loosening were re-revised between 1992 and 2016. Acetabular bone defects graded according to Paprosky during the first and the re-revision surgery were compared. Clinical and radiological findings were analyzed over time. Survival analysis was performed using a competing risk analysis. Results. Intraoperative bone defect during the initial revision included 19 Paprosky type IIIA and 29 Paprosky type IIIB hips; at re-revision, seven hips were Paprosky type II, 27 type IIIA and 11 were type IIIB (p = 0.020). The mean preoperative Harris Hip Score was 45.4 (SD 6.4), becoming 80.7 (SD 12.7) at the final follow-up. In all, 12 hips showed radiological migration of the acetabular component, and three required further revision surgery. The nine-year cumulative failure incidence (nine patients at risk) of the acetabular component for further revision surgery was 9.6% (95% confidence interval (CI) 2.9 to 21.0) for any cause, and 7.5% (95% CI 1.9 to 18.5) for aseptic loosening. Hips with a greater hip height had a higher risk for radiological migration (odds ratio 1.09, 95% CI 1.02 to 1.17; p = 0.008). Conclusion. Bone stock restoration can be obtained using IBG in revision hip surgery. This technique is also useful in re-revision surgery; however, a better surgical technique including a closer distance to
The accurate reconstruction of hip anatomy and
biomechanics is thought to be important in achieveing good clinical
outcomes following total hip arthroplasty (THA). To this end some
newer hip designs have introduced further modularity into the design
of the femoral component such that neckshaft angle and anteversion,
which can be adjusted intra-operatively. The clinical effect of
this increased modularity is unknown. We have investigated the changes
in these anatomical parameters following conventional THA with a
prosthesis of predetermined neck–shaft angle and assessed the effect
of changes in the hip anatomy on clinical outcomes. . In total, 44 patients (mean age 65.3 years (standard deviation
(. sd. ) 7); 17 male/27 female; mean body mass index 26.9 (kg/m²)
(. sd. 3.1)) underwent a pre- and post-operative three-dimensional
CT scanning of the hip. The pre- and post-operative neck–shaft angle,
offset,
The ‘jumbo’ acetabular component is now commonly
used in acetabular revision surgery where there is extensive bone
loss. It offers high surface contact, permits weight bearing over
a large area of the pelvis, the need for bone grafting is reduced
and it is usually possible to restore centre of rotation of the
hip. Disadvantages of its use include a technique in which bone
structure may not be restored, a risk of excessive posterior bone
loss during reaming, an obligation to employ screw fixation, limited
bone ingrowth with late failure and high
Trabecular metal (TM) augments are a relatively
new option for reconstructing segmental bone loss during acetabular
revision. We studied 34 failed hip replacements in 34 patients that
were revised between October 2003 and March 2010 using a TM acetabular
shell and one or two augments. The mean age of the patients at the
time of surgery was 69.3 years (46 to 86) and the mean follow-up
was 64.5 months (27 to 107). In all, 18 patients had a minor column
defect, 14 had a major column defect, and two were associated with
pelvic discontinuity. The
Digital radiography is becoming widespread. Accurate pre-operative templating of digital images of the hip traditionally involves positioning a calibration object at its centre. This can be difficult and cause embarrassment. We have devised a method whereby a planar disc placed on the radiographic cassette accounts for the expected magnification. Initial examination of 50 pelvic CT scans showed a mean
The increasing need for total hip replacement
(THR) in an ageing population will inevitably generate a larger number
of revision procedures. The difficulties encountered in dealing
with the bone deficient acetabulum are amongst the greatest challenges
in hip surgery. The failed acetabular component requires reconstruction
to restore the
To investigate the extent of bone development around the scaffold of custom triflange acetabular components (CTACs) over time. We performed a single-centre historical prospective cohort study, including all patients with revision THA using the aMace CTAC between January 2017 and March 2021. A total of 18 patients (18 CTACs) were included. Models of the hemipelvis and the scaffold component of the CTACs were created by segmentation of CT scans. The CT scans were performed immediately postoperatively and at least one year after surgery. The amount of bone in contact with the scaffold was analyzed at both times, and the difference was calculated.Aims
Methods
Brace treatment is the cornerstone of managing developmental dysplasia of the hip (DDH), yet there is a lack of evidence-based treatment protocols, which results in wide variations in practice. To resolve this, we have developed a comprehensive nonoperative treatment protocol conforming to published consensus principles, with well-defined a priori criteria for inclusion and successful treatment. This was a single-centre, prospective, longitudinal cohort study of a consecutive series of infants with ultrasound-confirmed DDH who underwent a comprehensive nonoperative brace management protocol in a unified multidisciplinary clinic between January 2012 and December 2016 with five-year follow-up radiographs. The radiological outcomes were acetabular index-lateral edge (AI-L), acetabular index-sourcil (AI-S), centre-edge angle (CEA), acetabular depth ratio (ADR), International Hip Dysplasia Institute (IHDI) grade, and evidence of avascular necrosis (AVN). At five years, each hip was classified as normal (< 1 SD), borderline dysplastic (1 to 2 SDs), or dysplastic (> 2 SDs) based on validated radiological norm-referenced values.Aims
Methods
The use of a porous metal shell supported by two augments with the ‘footing’ technique is one solution to manage Paprosky IIIB acetabular defects in revision total hip arthroplasty. The aim of this study was to assess the medium-term implant survival and radiological and clinical outcomes of this technique. We undertook a retrospective, two-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the ‘footing’ technique for Paprosky IIIB acetabular defects between 2007 and 2020. The median age at the time of surgery was 64.4 years (interquartile range (IQR) 54.4 to 71.0). The median follow-up was 3.9 years (IQR 3.1 to 7.0).Aims
Methods
Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation. A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707).Aims
Methods
The aim of this study is to evaluate whether acetabular retroversion (AR) represents a structural anatomical abnormality of the pelvis or is a functional phenomenon of pelvic positioning in the sagittal plane, and to what extent the changes that result from patient-specific functional position affect the extent of AR. A comparative radiological study of 19 patients (38 hips) with AR were compared with a control group of 30 asymptomatic patients (60 hips). CT scans were corrected for rotation in the axial and coronal planes, and the sagittal plane was then aligned to the anterior pelvic plane. External rotation of the hemipelvis was assessed using the superior iliac wing and inferior iliac wing angles as well as quadrilateral plate angles, and correlated with cranial and central acetabular version. Sagittal anatomical parameters were also measured and correlated to version measurements. In 12 AR patients (24 hips), the axial measurements were repeated after matching sagittal pelvic rotation with standing and supine anteroposterior radiographs.Aims
Methods
The aim of this study was to examine the implant accuracy of custom-made partial pelvis replacements (PPRs) in revision total hip arthroplasty (rTHA). Custom-made implants offer an option to achieve a reconstruction in cases with severe acetabular bone loss. By analyzing implant deviation in CT and radiograph imaging and correlating early clinical complications, we aimed to optimize the usage of custom-made implants. A consecutive series of 45 (2014 to 2019) PPRs for Paprosky III defects at rTHA were analyzed comparing the preoperative planning CT scans used to manufacture the implants with postoperative CT scans and radiographs. The anteversion (AV), inclination (IC), deviation from the preoperatively planned implant position, and deviation of the centre of rotation (COR) were explored. Early postoperative complications were recorded, and factors for malpositioning were sought. The mean follow-up was 30 months (SD 19; 6 to 74), with four patients lost to follow-up.Aims
Methods
We investigated factors that were thought to be associated with an increased incidence of squeaking of ceramic-on-ceramic total hip replacements. Between June 1997 and December 2008 the three senior authors implanted 2406 primary total hip replacements with a ceramic-on-ceramic bearing surface. The mean follow-up was 10.6 years. The diagnosis was primary osteoarthritis in each case, and no patient had undergone previous surgery to the hip. We identified 74 squeaking hips (73 patients) giving an incidence of 3.1% at a mean follow-up of 9.5 years (4.1 to 13.3). Taller, heavier and younger patients were significantly more likely to have hips that squeaked. Squeaking hips had a significantly higher range of post-operative internal (p = 0.001) and external rotation (p = 0.003) compared with silent hips. Patients with squeaking hips had significantly higher activity levels (p = 0.009). A squeaking hip was not associated with a significant difference in patient satisfaction (p = 0.24) or Harris hip score (p = 0.34). Four implant position factors enabled good prediction of squeaking. These were high acetabular component inclination, high femoral offset, lateralisation of the
A national screening programme has existed in the UK for the diagnosis of developmental dysplasia of the hip (DDH) since 1969. However, every aspect of screening and treatment remains controversial. Screening programmes throughout the world vary enormously, and in the UK there is significant variation in screening practice and treatment pathways. We report the results of an attempt by the British Society for Children’s Orthopaedic Surgery (BSCOS) to identify a nationwide consensus for the management of DDH in order to unify treatment and suggest an approach for screening. A Delphi consensus study was performed among the membership of BSCOS. Statements were generated by a steering group regarding aspects of the management of DDH in children aged under three months, namely screening and surveillance (15 questions), the technique of ultrasound scanning (eight questions), the initiation of treatment (19 questions), care during treatment with a splint (ten questions), and on quality, governance, and research (eight questions). A two-round Delphi process was used and a consensus document was produced at the final meeting of the steering group.Aims
Methods
A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.Aims
Methods
Radiological residual acetabular dysplasia (RAD) has been reported in up to 30% of children who had successful brace treatment of infant developmental dysplasia of the hip (DDH). Predicting those who will resolve and those who may need corrective surgery is important to optimize follow-up protocols. In this study we have aimed to identify the prevalence and predictors of RAD at two years and five years post-bracing. This was a single-centre, prospective longitudinal cohort study of infants with DDH managed using a published, standardized Pavlik harness protocol between January 2012 and December 2016. RAD was measured at two years’ mean follow-up using acetabular index-lateral edge (AI-L) and acetabular index-sourcil (AI-S), and at five years using AI-L, AI-S, centre-edge angle (CEA), and acetabular depth ratio (ADR). Each hip was classified based on published normative values for normal, borderline (1 to 2 standard deviations (SDs)), or dysplastic (> 2 SDs) based on sex, age, and laterality.Aims
Methods