Nine cases of acetabular erosion following insertion of the Monk "hard top"
We report the clinical and operative details of seven cases of fracture of the femoral stem of the Ring TiMESH cementless
A 70-year-old man with an uncemented metal-on-polyethylene
total
The Kent hip is a distally-locked femoral stem which was developed to address severe proximal bone loss, severe bony deformity and peri-prosthetic fracture. We reviewed the results of 145 consecutive Kent hips implanted into 141 patients between 1987 and 2000. The indications for implantation were aseptic loosening (75 hips), septic loosening (two), peri-prosthetic and prosthetic fracture (37), severe bony deformity (24), and fracture through a proximal femoral metastasis (seven). The median time to full weight-bearing after surgery was two days and the mean length of follow-up was 5.1 years (2 to 15). Further revisions were required for 13 femoral stems. With removal of the stem for any reason as an end-point, the cumulative survival at five, ten and 15 years was 93%, 89% and 77%, respectively. In patients aged ≥ 70 years, the cumulative survival at 15 years was 92%, compared with 68% in those aged <
70 years. Because of these findings, we recommend the use of interlocking stems in patients aged ≥ 70 years, particularly in those with a peri-prosthetic fracture, for whom alternative methods are limited. Outcome scores and survival data, compared with other systems, indicate that the Kent hip should be used with caution in younger patients.
The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components. A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry.Aims
Patients and Methods
A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis.
We reviewed 82 primary arthroplasties (in 71 patients) in which cementless porous-coated hip prostheses were used. The mean age of the patients at operation was 52 years (24 to 86); they were followed up for an average of 62.1 months (60 to 66). The diagnosis was avascular necrosis of the femoral head in 35%, fracture of the femoral neck in 24%, primary osteoarthritis in 16% and miscellaneous in 25%. The average preoperative Harris hip score was 56.7 points and the average postoperative score was 83.3 points. Eight hips (10%) had component loosening; four had been revised and four were awaiting revision. In 27 hips (33%) there was a radiolucent line wider than 2 mm in zones 1 and 7. In 55 hips (67%) there was calcar resorption of more than 10 mm. Twenty patients (28%) complained of thigh pain although they had no radiographic evidence of loosening of a component. Factors that may have contributed to the poor clinical and radiographic results were: 1) inadequate surface area for bone ingrowth, particularly on the lateral aspect of the upper part of the prosthesis, 2) poor initial fit of the stem in the metaphysis, which resulted in cantilever motion of the proximal part of the stem about the well-fixed distal stem, and 3) the collar of the prosthesis, which prevented it from subsiding to a naturally stable position and caused damage to the calcar.
Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20). The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for recurrent dislocation. On the basis of these long-term results, the porous-coated anatomic femoral component survived for a minimum of 18 years after operation while the acetabular component was less durable. The findings identify the principles of uncemented joint replacement which can be applied to current practice.
In a frail elderly patient a pathological fracture of the femur at the lower end of a loose femoral stem was treated by the retrograde insertion of an intramedullary nail over the tip of the prosthesis.
A prospective, randomised and independently assessed trial of the Ring UPM total hip replacement showed that the quality of the early result was better if the femoral prosthesis was cemented than if it was not. More patients with cemented prostheses were painfree at four months (58% cemented:42% uncemented) and at one year (63% cemented:50% uncemented), but at two years pain relief was equal in both groups. At two years significantly more patients with cemented prostheses could walk without support (96% cemented:62% uncemented, p = 0.01 to 0.05). There is a need for more similar trials to compare the results of contemporary designs of cemented and uncemented total hip prostheses.
We report the clinical outcome and survival of 100 total hip arthroplasties with Link RS cementless hip prostheses after a mean of 63 months (51 to 93). The average preoperative Merle d'Aubigne hip score was 9.4 points and the average score at the latest follow-up was 15.3 points. Thirteen prostheses have been revised and seven await revision. Survivorship analysis based on intention to revise showed only 68% survival at seven years (95% CI 53% to 81%). The survivals of femoral and acetabular components verified to be loose at surgery were 78% and 89% respectively at seven years. A number of other patients have pain and it seems likely that there will be substantial reduction in these survival figures in the future.
Hip implant retrieval analysis is the most important
source of insight into the performance of new materials and designs
of hip arthroplasties. Even the most rigorous
Seven hips with protrusio acetabuli which showed complete or partial resolution of the protrusion after replacement with Ring's prosthesis are reported. It is suggested that this prosthesis offers a simple and effective method of treating painful protrusion into the pelvis; it allows healing of the medial wall of the acetabulum while avoiding many of the hazards of other methods of treatment.
Sixty-five total hip arthroplastics were reinserted after sepsis around the hip, positive cultures being obtained from fifty-six. Although 65 per cent of patients still have their implant in position, only sixteen of sixty-five (25 per cent) show an excellent or good result on a Harris rating. Twenty-three of sixty-five (35 per cent) subsequently required an excision arthroplasty. The indications and contraindications for this procedure are given.
We have followed up for a period of seven to nine years 100 consecutive arthroplasties of the hip in which an entirely HA-coated implant had been used. The clinical results were excellent and bony incorporation was extensive in all components. No stem became loose or subsided but five cups were revised because of loosening after 3.8 to 5.5 years, having functioned painlessly and shown radiological ingrowth. Revision procedures because of excessive polyethylene wear have been performed on 18 hips and are planned for six more. Two eroded metal backings with worn-through polyethylene were exchanged; six hips showed metallosis without polyethylene wear-through. There were two cases of granulomatous cysts in the groin and 66 hips had osteolysis located periarticularly, in the greater trochanter or in the acetabulum.
We enrolled 98 patients (107 hips) with a mean age of 47 years (SD 8.6) into a prospective study of the Madreporic Lord THR; 34 hips had primary and 73 secondary osteoarthritis. After ten years, the survival rate using revision as the endpoint for failure was 70% (±9) for the cup and 98% (±0.3) for the stem. The combined clinical and radiological survival rates were 46% (±11) and 81% (±10), respectively. Osteoporosis due to stress-shielding was observed in the proximal femur. Hips with radiologically dense bone postoperatively showed the most pronounced bone loss. We recommend continued radiological follow-up of patients with this type of implant to allow revision to be performed before there is severe bony destruction of the pelvis.