Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 833 - 843
1 Jul 2022
Kayani B Baawa-Ameyaw J Fontalis A Tahmassebi J Wardle N Middleton R Stephen A Hutchinson J Haddad FS

Aims

This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA).

Methods

Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1615 - 1622
1 Dec 2015
Müller M Abdel MP Wassilew GI Duda G Perka C

The accurate reconstruction of hip anatomy and biomechanics is thought to be important in achieveing good clinical outcomes following total hip arthroplasty (THA). To this end some newer hip designs have introduced further modularity into the design of the femoral component such that neck­shaft angle and anteversion, which can be adjusted intra-operatively. The clinical effect of this increased modularity is unknown. We have investigated the changes in these anatomical parameters following conventional THA with a prosthesis of predetermined neck–shaft angle and assessed the effect of changes in the hip anatomy on clinical outcomes.

In total, 44 patients (mean age 65.3 years (standard deviation (sd) 7); 17 male/27 female; mean body mass index 26.9 (kg/m²) (sd 3.1)) underwent a pre- and post-operative three-dimensional CT scanning of the hip. The pre- and post-operative neck–shaft angle, offset, hip centre of rotation, femoral anteversion, and stem alignment were measured. Additionally, a functional assessment and pain score were evaluated before surgery and at one year post-operatively and related to the post-operative anatomical changes.

The mean pre-operative neck–shaft angle was significantly increased by 2.8° from 128° (sd 6.2; 119° to 147°) to 131° (sd 2.1; 127° to 136°) (p = 0.009). The mean pre-operative anteversion was 24.9° (sd 8; 7.9 to 39.1) and reduced to 7.4° (sd 7.3; -11.6° to 25.9°) post-operatively (p < 0.001). The post-operative changes had no influence on function and pain. Using a standard uncemented femoral component, high pre- and post-operative variability of femoral anteversion and neck–shaft angles was found with a significant decrease of the post-operative anteversion and slight increase of the neck–shaft angles, but without any impact on clinical outcome.

Cite this article: Bone Joint J 2015;97-B:1615–22.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1603 - 1609
1 Dec 2013
Chana R Facek M Tilley S Walter WK Zicat B Walter WL

We report the clinical and radiological outcomes of a series of contemporary cementless ceramic-on-ceramic total hip replacements (THRs) at ten years in patients aged ≤ 55 years of age. Pre- and post-operative activity levels are described. A total of 120 consecutive ceramic cementless THRs were performed at a single centre in 110 patients from 1997 to 1999. The mean age of the patients at operation was 45 years (20 to 55). At ten years, four patients had died and six were lost to follow-up, comprising ten hips. The mean post-operative Harris hip score was 94.7 (55 to 100). Radiological analysis was undertaken in 90 available THRs of the surviving 106 hips at final review: all had evidence of stable bony ingrowth, with no cases of osteolysis. Wear was undetectable. There were four revisions. The survival for both components with revision for any cause as an endpoint was 96.5% (95% confidence interval 94.5 to 98.7). The mean modified University of California, Los Angeles activity level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6 to 10) at the ten-year post-operative period.

Alumina ceramic-on-ceramic bearings in cementless primary THR in this series have resulted in good clinical and radiological outcomes with undetectable rates of wear and excellent function in the demanding younger patient group at ten years.

Cite this article: Bone Joint J 2013;95-B:1603–9.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 217 - 223
1 Feb 2014
Namba RS Inacio MCS Cafri G

The outcome of total knee replacement (TKR) using components designed to increase the range of flexion is not fully understood. The short- to mid-term risk of aseptic revision in high flexion TKR was evaluated. The endpoint of the study was aseptic revision and the following variables were investigated: implant design (high flexion vs non-high flexion), the thickness of the tibial insert (≤ 14 mm vs > 14 mm), cruciate ligament (posterior stabilised (PS) vs cruciate retaining), mobility (fixed vs rotating), and the manufacturer (Zimmer, Smith & Nephew and DePuy). Covariates included patient, implant, surgeon and hospital factors. Marginal Cox proportional hazard models were used.

In a cohort of 64 000 TKRs, high flexion components were used in 8035 (12.5%). The high flexion knees with tibial liners of thickness > 14 mm had a density of revision of 1.45/100 years of observation, compared with 0.37/100 in non-high flexion TKR with liners ≤ 14 mm thick. Relative to a standard fixed PS TKR, the NexGen (Zimmer, Warsaw, Indiana) Gender Specific Female high flexion fixed PS TKR had an increased risk of revision (hazard ratio (HR) 2.27 (95% confidence interval (CI) 1.48 to 3.50)), an effect that was magnified when a thicker tibial insert was used (HR 8.10 (95% CI 4.41 to 14.89)).

Surgeons should be cautious when choosing high flexion TKRs, particularly when thicker tibial liners might be required.

Cite this article: Bone Joint J 2014;96-B:217–23.