The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1049 - 1053
1 Aug 2011
Putzer D Mayr E Haid C Reinthaler A Nogler M

In revision total hip replacement, bone loss can be managed by impacting porous bone chips. In order to guarantee sufficient mechanical strength, the bone chips have to be compacted. The aim of this study was to determine in an in vitro simulation whether the use of a pneumatic hammer leads to higher primary stability than manual impaction. Bone mass characteristics were measured by force and distance variation of a penetrating punch, which was lowered into a plastic cup filled with bone chips. From these measurements bulk density, contact stiffness, impaction hardness and penetration resistance were calculated for different durations of impaction.

We found that the pneumatic method reached higher values of impaction hardness, contact stiffness and bulk density suggesting an increase in stability of the implant. No significant differences were found between the two different methods concerning the penetration resistance. The pneumatic method might reduce the risk of fracture in vivo, as force peaks are smaller and applied for a shorter period. Results from manual impaction showed higher variability and depend much on the experience of the surgeon. The pneumatic hammer is a suitable tool to standardise the impaction process.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 874 - 878
1 Jul 2007
Atilla B Ali H Aksoy MC Caglar O Tokgozoglu AM Alpaslan M

We have reviewed 54 patients who had undergone 61 total hip replacements using bulk femoral autografts to augment a congenitally dysplastic acetabulum. There were 52 women and two men with a mean age of 42.4 years (29 to 76) at the time of the index operation. A variety of different prostheses was used: 28 (45.9%) were cemented and 33 (54.1%) uncemented. The graft technique remained unchanged throughout the series.

Follow-up was at a mean of 8.3 years (3 to 20). The Hospital for Special Surgery hip score improved from a mean of 10.7 (4 to 18) pre-operatively to a mean of 35 (28 to 38) at follow-up.

The position of the acetabular component was anatomical in 37 hips (60.7%), displaced less than 1 cm in 20 (32.7%) and displaced more than 1 cm in four (6.6%). Its cover was between 50% and 75% in 34 hips (55.7%) and less than 50% in 25 (41%). In two cases (3.3%), it was more than 75%.

There was no graft resorption in 36 hips (59%), mild resorption in 21 (34%) and severe resorption in four (6%).

Six hips (9.8%) were revised for aseptic loosening. The overall rate of loosening and revision was 14.8%. Overall survival at 8.3 years was 93.4%.

The only significant factor which predicted failure was the implantation of the acetabular component more than 1 cm from the anatomical centre of rotation of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 173 - 178
1 Feb 2006
de Jong PT Haverkamp D van der Vis HM Marti RK

We evaluated the long-term results of 116 total hip replacements with a superolateral shelfplasty in 102 patients with osteoarthritis secondary to developmental dysplasia of the hip. After a mean follow-up of 19.5 years (11.5 to 26.0), 14 acetabular components (12%) had been revised. The cumulative survival at 20 years was 78%, with revision for loosening of the acetabular component as the end-point. All grafts were well integrated and showed remodelling. In six grafts some resorption had occurred under the heads of the screws where the graft was not supporting the socket. Apart from these 14 revisions, seven acetabular components had possible radiological signs of loosening at a mean follow-up of 14.5 years, one had signs of probable loosening, and five had signs of definite loosening. These results indicate that this technique of bone grafting for acetabular reconstruction in hip dysplasia is a durable solution for cemented acetabular components.