Advertisement for orthosearch.org.uk
Results 1 - 9 of 9
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1731 - 1735
1 Nov 2021
Iobst CA Frost MW Rölfing JD Rahbek O Bafor A Duncan M Kold S

Aims. Limb-lengthening nails have largely replaced external fixation in limb-lengthening and reconstructive surgery. However, the adverse events and high prevalence of radiological changes recently noted with the STRYDE lengthening nail have raised concerns about the use of internal lengthening nails. The aim of this study was to compare the prevalence of radiological bone abnormalities between STRYDE, PRECICE, and FITBONE nails prior to nail removal. Methods. This was a retrospective case series from three centres. Patients were included if they had either of the three limb-lengthening nails (STYDE, PRECICE, or FITBONE) removed. Standard orthogonal radiographs immediately prior to nail removal were examined for bone abnormalities at the junction of the telescoping nail parts. Results. In total, 306 patients (168 male, 138 female) had 366 limb-lengthening nails removed. The mean time from nail insertion to radiological evaluation was 434 days (36 to 3,015). Overall, 77% of STRYDE nails (20/26) had bone abnormalities at the interface compared with only 2% of FITBONE (4/242) and 1% of PRECICE nails (1/98; p < 0.001). Focal osteolysis in conjunction with periosteal reaction at the telescoping interface was only observed in STRYDE nails. Conclusion. Bone abnormalities at the interface of telescoping nail parts were seen in the majority of STRYDE nails, but only very rarely with FITBONE or PRECICE nails. We conclude that the low prevalence of radiological changes at the junctional interface of 242 FITBONE and 98 PRECICE nails at the time of nail removal does not warrant clinical concerns. Cite this article: Bone Joint J 2021;103-B(11):1731–1735


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 178 - 188
1 Feb 2019
Chaudhary MM Lakhani PH

Aims

Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator.

Patients and Methods

A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 938 - 942
1 Jul 2006
Singh S Lahiri A Iqbal M

Limb lengthening by callus distraction and external fixation has a high rate of complications. We describe our experience using an intramedullary nail (Fitbone) which contains a motorised and programmable sliding mechanism for limb lengthening and bone transport. Between 2001 and 2004 we lengthened 13 femora and 11 tibiae in ten patients (seven men and three women) with a mean age of 32 years (21 to 47) using this nail. The indications for operation were short stature in six patients and developmental or acquired disorders in the rest.

The mean lengthening achieved was 40 mm (27 to 60). The mean length of stay in hospital was seven days (5 to 9). The mean healing index was 35 days/cm (18.8 to 70.9). There were no cases of implant-related infection or malunion.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 1000 - 1004
1 Jul 2005
Baumgart R Bürklein D Hinterwimmer S Thaller P Mutschler W

Ollier’s disease is characterised by severe deformity of the extremities and retarded growth because of multiple enchondromas. For correction of deformity, the Ilizarov method has been used although it has many complications. A 17-year-old boy with Ollier’s disease had a limb-length discrepancy of 17.4 cm, with a valgus deformity of the right knee and recurvatum of the femur of 23°. He had undergone three unsuccessful attempts to correct the deformities by using external fixators. We used a fully implantable, motorised, lengthening and correction nail (Fitbone) to achieve full correction of all the deformities without complications. We decided to carry out the procedure in three stages. First, we lengthened the femur by 3.6 cm and the tibia by 4 cm. We then exchanged the femoral nail for a longer implant and achieved a further 6 cm of length. This reduced the shortening to 3.8 cm. When the boy has finished secondary school we will adjust the remaining discrepancy.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 639 - 643
1 May 2011
Burghardt RD Herzenberg JE Specht SC Paley D

Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device.

The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1382 - 1388
1 Oct 2016
Laubscher M Mitchell C Timms A Goodier D Calder P

Aims

Patients undergoing femoral lengthening by external fixation tolerate treatment less well when compared to tibial lengthening. Lengthening of the femur with an intramedullary device may have advantages.

Patients and Methods

We reviewed all cases of simple femoral lengthening performed at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities, congenital limb deficiencies and lengthening with an unstable hip were excluded, leaving 33 cases (in 22 patients; 11 patients had bilateral procedures) for review. Healing index, implant tolerance and complications were compared.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1241 - 1245
1 Sep 2012
Burghardt RD Paley D Specht SC Herzenberg JE

Internal lengthening devices in the femur lengthen along the anatomical axis, potentially creating lateral shift of the mechanical axis. We aimed to determine whether femoral lengthening along the anatomical axis has an inadvertent effect on lower limb alignment. Isolated femoral lengthening using the Intramedullary Skeletal Kinetic Distractor was performed in 27 femora in 24 patients (mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment procedures or concurrent tibial lengthening, or who developed mal- or nonunion, were excluded. Pre-operative and six-month post-operative radiographs were used to measure lower limb alignment. The mean lengthening achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0 to 3.5). In one femur that was initially in varus, a 3 mm medial shift occurred during a lengthening of 2.2 cm.

In a normally aligned limb, intramedullary lengthening along the anatomical axis of the femur results in a lateral shift of the mechanical axis by approximately 1 mm for each 1 cm of lengthening.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 788 - 792
1 Jun 2011
Schiedel FM Pip S Wacker S Pöpping J Tretow H Leidinger B Rödl R

We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties.

Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful tibial lengthening was only achieved in five of 11 patients (45%).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 955 - 961
1 Jul 2009
Simpson AHWR Shalaby H Keenan G

Femoral lengthening using the Intramedullary Skeletal Kinetic Distractor is a new technique. However, with intramedullary distraction the surgeon has less control over the lengthening process. Therefore, 33 femora lengthened with this device were assessed to evaluate the effect of operative variables under the surgeon’s control on the course of lengthening. The desired lengthening was achieved in 32 of 33 limbs. Problems encountered included difficulty in achieving length in eight femora (24%) and uncontrolled lengthening in seven (21%). Uncontrolled lengthening was more likely if the osteotomy was placed with less than 80 mm of the thick portion of the nail in the distal fragment (p = 0.052), and a failure to lengthen was more likely if there was over 125 mm in the distal fragment (p = 0.008). The latter problem was reduced with over-reaming by 2.5 mm to 3 mm. Previous intramedullary nailing also predisposed to uncontrolled lengthening (p = 0.042), and these patients required less reaming.

Using the Intramedullary Skeletal Kinetic Distractor, good outcomes were obtained; problems were minimised by optimising the position of the osteotomy and the amount of over-reaming performed.