There is little evidence on the cost effectiveness
of different
The aim of this study was to compare the design of the generic
OptiStem XTR femoral stem with the established Exeter femoral stem. We obtained five boxed, as manufactured, implants of both designs
at random (ten in total). Two examiners were blinded to the implant
design and independently measured the mass, volume, trunnion surface
topography, trunnion roughness, trunnion cone angle, Caput-Collum-Diaphyseal
(CCD) angle, femoral offset, stem length, neck length, and the width
and roughness of the polished stem shaft using peer-reviewed methods.
We then compared the stems using these parameters.Aims
Materials and Methods
Aims. Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. Patients and Methods. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of England, Wales and Northern Ireland to determine
if a cemented femoral stem
Aims. We evaluated the short-term functional outcome and prevalence
of bearing-specific generation of audible noise in 301 patients
(336 hips) operated on with fourth generation (Delta) medium diameter
head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs). Patients and Methods. There were 191 female (63%) and 110 male patients (37%) with
a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years
(1.3 to 3.4). Patients completed three questionnaires: Oxford Hip
Score (OHS), Research and Development 36-item health survey (RAND-36)
and a noise-specific symptom questionnaire. Plain radiographs were also
analysed. A total of three hips (0.9%) were revised. . Results. There were 52 patients (54 hips, 17%) who reported noise, and
in 25 (48%) of them the noise was frequently heard. In the multiple
regression analysis, the only independent risk factor for noise
was a specific THA
Aims. Natural Language Processing (NLP) offers an automated method to extract data from unstructured free text fields for arthroplasty registry participation. Our objective was to investigate how accurately NLP can be used to extract structured clinical data from unstructured clinical notes when compared with manual data extraction. Methods. A group of 1,000 randomly selected clinical and hospital notes from eight different surgeons were collected for patients undergoing primary arthroplasty between 2012 and 2018. In all, 19 preoperative, 17 operative, and two postoperative variables of interest were manually extracted from these notes. A NLP algorithm was created to automatically extract these variables from a training sample of these notes, and the algorithm was tested on a random test sample of notes. Performance of the NLP algorithm was measured in Statistical Analysis System (SAS) by calculating the accuracy of the variables collected, the ability of the algorithm to collect the correct information when it was indeed in the note (sensitivity), and the ability of the algorithm to not collect a certain data element when it was not in the note (specificity). Results. The NLP algorithm performed well at extracting variables from unstructured data in our random test dataset (accuracy = 96.3%, sensitivity = 95.2%, and specificity = 97.4%). It performed better at extracting data that were in a structured, templated format such as range of movement (ROM) (accuracy = 98%) and implant
Modern metal-on-metal hip resurfacing has been
widely performed in the United Kingdom for over a decade. However,
the literature reports conflicting views of the benefits: excellent
medium- to long-term results with some
New
Despite excellent results, the use of cemented
total hip replacement (THR) is declining. This retrospective cohort study
records survival time to revision following primary cemented THR
using the most common combination of components that accounted for
almost a quarter of all cemented THRs, exploring risk factors independently associated
with failure. All patients with osteoarthritis who had an Exeter
V40/Contemporary THR (Stryker) implanted before 31 December 2010
and recorded in the National Joint Registry for England and Wales
were included in the analysis. Cox’s proportional hazard models
were used to analyse the extent to which risk of revision was related
to patient, surgeon and implant covariates, with a significance
threshold of p <
0.01. A total of 34 721 THRs were included in
the study. The overall seven-year rate of revision for any reason
was 1.70% (99% confidence interval (CI) 1.28 to 2.12). In the final
adjusted model the risk of revision was significantly higher in
THRs with the Contemporary hooded component (hazard ratio (HR) 1.88,
p <
0.001) than with the flanged version, and in smaller head
sizes (<
28 mm) compared with 28 mm diameter heads (HR 1.50,
p = 0.005). The seven-year revision rate was 1.16% (99% CI 0.69
to 1.63) with a 28 mm diameter head and flanged component. The overall
risk of revision was independent of age, gender, American Society
of Anesthesiologists grade, body mass index, surgeon volume, surgical
approach,
The Exeter V40 femoral stem is the most implanted stem in the National Joint Registry (NJR) for primary total hip arthroplasty (THA). In 2004, the 44/00/125 stem was released for use in ‘cement-in-cement’ revision cases. It has, however, been used ‘off-label’ as a primary stem when patient anatomy requires a smaller stem with a 44 mm offset. We aimed to investigate survival of this implant in comparison to others in the range when used in primary THAs recorded in the NJR. We analyzed 328,737 primary THAs using the Exeter V40 stem, comprising 34.3% of the 958,869 from the start of the NJR to December 2018. Our exposure was the stem, and the outcome was all-cause construct revision. We stratified analyses into four groups: constructs using the 44/00/125 stem, those using the 44/0/150 stem, those including a 35.5/125 stem, and constructs using any other Exeter V40 stem.Aims
Methods
Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.Aims
Methods
The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
Aims. We wished to quantify the extent of soft-tissue damage sustained
during minimally invasive total hip arthroplasty through the direct
anterior (DA) and direct superior (DS) approaches. Materials and Methods. In eight cadavers, the DA approach was performed on one side,
and the DS approach on the other, a single
The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component. We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems.Aims
Methods
The aim of this study was to evaluate the performance of first-generation annealed highly cross-linked polyethylene (HXLPE) in cementless total hip arthroplasty (THA). We retrospectively evaluated 29 patients (35 hips) who underwent THA between December 2000 and February 2002. The survival rate was estimated using the Kaplan-Meier method. Hip joint function was evaluated using the Japanese Orthopaedic Association (JOA) score. Two-dimensional polyethylene wear was estimated using Martell’s Hip Analysis Suite. We calculated the wear rates between years 1 and 5, 5 and 10, 10 and 15, and 15 and final follow-up.Aims
Methods
To determine if primary cemented acetabular component geometry (long posterior wall (LPW), hooded, or offset reorientating) influences the risk of revision total hip arthroplasty (THA) for instability or loosening. The National Joint Registry (NJR) dataset was analyzed for primary THAs performed between 2003 and 2017. A cohort of 224,874 cemented acetabular components were included. The effect of acetabular component geometry on the risk of revision for instability or for loosening was investigated using log-binomial regression adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, operating surgeon grade, surgical approach, polyethylene crosslinking, and prosthetic head size. A competing risk survival analysis was performed with the competing risks being revision for other indications or death.Aims
Methods
We have assessed the effect of different regimes of antibiotic prophylaxis on the survival of total hip implants, comparing antibiotics administered both systemically and in the bone cement, systemically only, in the bone cement only and with no antibiotics given. We studied 10 905 primary cemented total hip replacements, performed for osteoarthritis of the hip and reported to the Norwegian arthroplasty register between 1987 and 1995. Cox-estimated failure-rate ratios (FRR) are presented with adjustment for gender, age, the
The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk regression analyses adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, surgeon grade, surgical approach, head size, and polyethylene crosslinking. Stratified analyses by surgical approach were performed, including pairwise comparisons of liner geometries.Aims
Methods
This aim of this study was to assess the reliability and validity of the Unified Classification System (UCS) for postoperative periprosthetic femoral fractures (PFFs) around cemented polished taper-slip (PTS) stems. Radiographs of 71 patients with a PFF admitted consecutively at two centres between 25 February 2012 and 19 May 2020 were collated by an independent investigator. Six observers (three hip consultants and three trainees) were familiarized with the UCS. Each PFF was classified on two separate occasions, with a mean time between assessments of 22.7 days (16 to 29). Interobserver reliability for more than two observers was assessed using percentage agreement and Fleiss’ kappa statistic. Intraobserver reliability between two observers was calculated with Cohen kappa statistic. Validity was tested on surgically managed UCS type B PFFs where stem stability was documented in operation notes (n = 50). Validity was assessed using percentage agreement and Cohen kappa statistic between radiological assessment and intraoperative findings. Kappa statistics were interpreted using Landis and Koch criteria. All six observers were blinded to operation notes and postoperative radiographs.Aims
Methods
Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Aims
Methods
Aims
Patients and Methods