In 100 patients the fulcrum axis which is the line connecting the anterior tip of the coracoid and the posterolateral angle of the acromion, was used to position
The glenopolar angle assesses the rotational
alignment of the glenoid and may provide prognostic information
and aid the management of scapula fractures. We have analysed the
effect of the anteroposterior (AP) shoulder radiograph rotational
offset on the glenopolar angle in a laboratory setting and used
this to assess the accuracy of shoulder imaging employed in routine
clinical practice. Fluoroscopic imaging was performed on 25 non-paired scapulae
tagged with 2 mm steel spheres to determine the orientation of true
AP views. The glenopolar angle was measured on all the bony specimens
rotated at 10° increments. The mean glenopolar angle measured on
the bone specimens in rotations between 0° and 20° and thereafter
was found to be significantly different (p <
0.001). We also obtained
the AP radiographs of the uninjured shoulder of 30 patients treated
for fractures at our centre and found that none fitted the criteria
of a true AP shoulder radiograph. The mean angular offset from the
true AP view was 38° (10° to 65°) for this cohort. Radiological
AP shoulder views may not fully project the normal anatomy of the scapular
body and the measured glenopolar angle. The absence of a true AP
view may compromise the clinical management of a scapular fracture. Cite this article:
Radiological changes and differences between cemented and uncemented components of Grammont reverse shoulder arthroplasties (DePuy) were analysed at a mean follow-up of 9.6 years (8 to 12). Of 122 reverse shoulder arthroplasties implanted in five shoulder centres between 1993 and 2000, a total of 68 (65 patients) were available for study. The indications for reversed shoulder arthroplasty were cuff tear arthropathy in 48 shoulders, revision of shoulder prostheses of various types in 11 and massive cuff tear in nine. The development of scapular notching, bony scapular spur formation, heterotopic ossification, glenoid and humeral radiolucencies, stem subsidence, radiological signs of stress shielding and resorption of the tuberosities were assessed on standardised
Eleven articulated scoliotic spines were examined radiographically and morphometrically. Measurement of the curve on anteroposterior radiographs of the specimens gave a mean Cobb angle of 70 degrees, though
The incidence of loosening of a cemented glenoid componentin total shoulder arthroplasty, detected by means of radiolucent lines or positional shift of the component on
This study aimed to assess the impact of using the metal-augmented glenoid baseplate (AGB) on improving clinical and radiological outcomes, as well as reducing complications, in patients with superior glenoid wear undergoing reverse shoulder arthroplasty (RSA). From January 2016 to June 2021, out of 235 patients who underwent primary RSA, 24 received a superior-AGB after off-axis reaming (Group A). Subsequently, we conducted propensity score matching in a 1:3 ratio, considering sex, age, follow-up duration, and glenoid wear (superior-inclination and retroversion), and selected 72 well-balanced matched patients who received a standard glenoid baseplate (STB) after eccentric reaming (Group B). Superior-inclination, retroversion, and lateral humeral offset (LHO) were measured to assess preoperative glenoid wear and postoperative correction, as well as to identify any complications. Clinical outcomes were measured at each outpatient visit before and after surgery.Aims
Methods
Favourable short-term outcomes have been reported following latissimus dorsi tendon transfer for patients with an irreparable subscapularis (SSC) tendon tear. The aim of this study was to investigate the long-term outcomes of this transfer in these patients. This was a retrospective study involving 30 patients with an irreparable SSC tear and those with a SSC tear combined with a reparable supraspinatus tear, who underwent a latissimus dorsi tendon transfer. Clinical scores and active range of motion (aROM), SSC-specific physical examination and the rate of return to work were assessed. Radiological assessment included recording the acromiohumeral distance (AHD), the Hamada grade of cuff tear arthropathy and the integrity of the transferred tendon. Statistical analysis compared preoperative, short-term (two years), and final follow-up at a mean of 8.7 years (7 to 10).Aims
Methods
Conservative treatment of moderately displaced proximal humeral head fractures yields good clinical results, but secondary fragment displacement may occur. Identification of those fractures at risk of displacement may influence initial decision-making. A total of 163 shoulders in 162 patients with conservatively treated isolated proximal humeral fractures were included. The fractures occurred between January 2015 and May 2018. The mean age of the patients was 69 years (26 to 100) and the mean follow-up was 144 days (42 to 779). The fractures were classified according to Neer. Scores for osteoporosis (Tingart, Deltoid Tuberosity Index (DTI)) and osteoarthritis (OA) of the glenohumeral joint were assessed. Translation of the head on follow-up radiographs of more than 10 mm was defined as displacement. Eccentric head index (EHI) describes the offset of the humeral head centre in relation to the diaphyseal axis. The ratio was estimated on anteroposterior (AP) and Neer views. Medial hinge was considered intact if the medial cortex proximal and distal to the fracture was in line on AP view.Aims
Methods
This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber. For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current.Aims
Methods
The aim of this study was to report the outcomes of different treatment options for glenoid loosening following reverse shoulder arthroplasty (RSA) at a minimum follow-up of two years. We retrospectively studied the records of 79 patients (19 men, 60 women; 84 shoulders) aged 70.4 years (21 to 87) treated for aseptic loosening of the glenosphere following RSA. Clinical evaluation included pre- and post-treatment active anterior elevation (AAE), external rotation, and Constant score.Aims
Patients and Methods
Aims
Materials and Methods
The aim of this study was to evaluate the clinical and radiological
outcome of using an anatomical short-stem shoulder prosthesis to
treat primary osteoarthritis of the glenohumeral joint. A total of 66 patients (67 shoulders) with a mean age of 76 years
(63 to 92) were available for clinical and radiological follow-up
at two different timepoints (T1, mean 2.6 years, Aims
Patients and Methods
Promising medium-term results from total shoulder arthroplasty
(TSA) have been reported for the treatment of primary osteoarthritis
in young and middle-aged patients. The aim of this study was to
evaluate the long-term functional and radiological outcome of TSA
in the middle-aged patient. The data of all patients from the previous medium-term study
were available. At a mean follow-up of 13 years (8 to 17), we reviewed
21 patients (12 men, nine women, 21 shoulders) with a mean age of
55 years (37 to 60). The Constant-Murley score (CS) with its subgroups
and subjective satisfaction were measured. Radiological signs of implant
loosening were analysed.Aims
Patients and Methods
Recently, there has been considerable interest in quantifying
the associations between bony abnormalities around and in the hip
joint and osteoarthritis (OA). Our aim was to investigate the relationships
between acetabular undercoverage, acetabular overcoverage, and femoroacetabular
impingement (FAI) with OA of the hip, which currently remain controversial. A total of 545 cadaveric skeletons (1090 hips) from the Hamann-Todd
osteological collection were obtained. Femoral head volume (FHV),
acetabular volume (AV), the FHV/AV ratio, acetabular version, alpha
angle and anterior femoral neck offset (AFNO) were measured. A validated
grading system was used to quantify OA of the hip as minimal, moderate,
or severe. Multiple linear and multinomial logistic regression were
used to determine the factors that correlated independently with
the FHV, AV, and the FHV/AV ratio. Aims
Materials and Methods
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
In this study we evaluated whether the Instability
Severity Index Score (ISIS) and the Western Ontario Shoulder Instability
Index (WOSI) could detect those patients at risk of failure following
arthroscopic Bankart repair. Between April 2008 and June 2010, the
ISIS and WOSI were recorded pre-operatively in 110 patients (87
male, 79%) with a mean age of 25.1 years (16 to 61) who underwent
this procedure for recurrent anterior glenohumeral instability. A telephone interview was performed two-years post-operatively
to determine whether patients had experienced a recurrent dislocation
and whether they had returned to pre-injury activity levels. In
all, six (5%) patients had an ISIS >
6 points (0 to 9). Of 100 (91%)
patients available two years post-operatively, six (6%) had a recurrent dislocation,
and 28 (28%) did not return to pre-injury activity. No patient who
dislocated had an ISIS >
6 (p = 1.0). There was no difference in
the mean pre-operative WOSI in those who had a re-dislocation and
those who did not (p = 0.99). The pre-operative WOSI was significantly
lower (p = 0.02) in those who did not return to pre-injury activity, whereas
the ISIS was not associated with return to pre-injury activity (p
= 0.13). In conclusion, neither the pre-operative ISIS nor WOSI predicted
recurrent dislocation within two years of arthroscopic Bankart repair.
Patients with a lower pre-operative WOSI were less likely to return
to pre-injury activity. Cite this article:
The indications for reverse shoulder arthroplasty
(RSA) continue to be expanded. Associated impairment of the deltoid
muscle has been considered a contraindication to its use, as function
of the RSA depends on the deltoid and impairment of the deltoid
may increase the risk of dislocation. The aim of this retrospective
study was to determine the functional outcome and risk of dislocation
following the use of an RSA in patients with impaired deltoid function.
Between 1999 and 2010, 49 patients (49 shoulders) with impairment
of the deltoid underwent RSA and were reviewed at a mean of 38 months
(12 to 142) post-operatively. There were nine post-operative complications (18%),
including two dislocations. The mean forward elevation improved
from 50° ( These results suggest that pre-operative deltoid impairment,
in certain circumstances, is not an absolute contraindication to
RSA. This form of treatment can yield reliable improvement in function
without excessive risk of post-operative dislocation. Cite this article:
We compared extrusion of the allograft after
medial and lateral meniscal allograft transplantation and examined
the correlation between the extent of extrusion and the clinical
outcome. A total of 73 lateral and 26 medial meniscus allografts
were evaluated by MRI at a mean of 32 months (24 to 59) in 99 patients
(67 men, 32 women) with a mean age of 35 years (21 to 52). The absolute
values and the proportional widths of extruded menisci as a percentage were
measured in coronal images that showed maximum extrusion. Functional
assessments were performed using Lysholm scores. The mean extrusion
was 4.7 mm (1.8 to 7.7) for lateral menisci and 2.9 mm (1.2 to 6.5)
for medial menisci (p <
0.001), and the mean percentage extrusions
were 52.0% (23.8% to 81.8%) and 31.2% (11.6% to 63.4%), respectively
(p <
0.001). Mean Lysholm scores increased significantly from
49.0 (10 to 83) pre-operatively to 86.6 (33 to 99) at final follow-up
for lateral menisci (p = 0.001) and from 50.9 (15 to 88) to 88.3
(32 to 100) for medial menisci (p <
0.001). The final mean Lysholm
scores were similar in the two groups (p = 0.312). Furthermore,
Lysholm scores were not found to be correlated with degree of extrusion
(p = 0.242). Thus, transplanted lateral menisci extrude more significantly
than transplanted medial menisci. However, the clinical outcome
after meniscal transplantation was not found to be adversely affected
by extrusion of the allograft.
We carried out a prospective investigation into
the radiological outcomes of uncemented Oxford medial compartment
unicondylar replacement in 220 consecutive patients (231 knees)
performed in a single centre with a minimum two-year follow-up.
The functional outcomes using the mean Oxford knee score and the
mean high-activity arthroplasty score were significantly improved
over the pre-operative scores (p <
0.001). There were 196 patients
with a two-year radiological examination performed under fluoroscopic
guidance, aiming to provide images acceptable for analysis of the
bone–implant interface. Of the six tibial zones examined on each
knee on the anteroposterior radiograph, only three had a partial
radiolucent line. All were in the medial aspect of the tibial base plate
(zone 1) and all measured <
1 mm. All of these patients were
asymptomatic. There were no radiolucent lines seen around the femoral
component or on the lateral view. There was one revision for loosening
at one year due to initial inadequate seating of the tibial component.
These results confirm that the early uncemented Oxford medial unicompartmental
compartmental knee replacements were reliable and the incidence
of radiolucent lines was significantly decreased compared with the
reported results of cemented versions of this implant. These independent
results confirm those of the designing centre.
Labral tears are commonly associated with femoroacetabular impingement. We reviewed 151 patients (156 hips) with femoroacetabular impingement and labral tears who had been treated arthroscopically. These were subdivided into those who had undergone a labral repair (group 1) and those who had undergone resection of the labrum (group 2). In order to ensure the groups were suitably matched for comparison of treatment effects, patients with advanced degenerative changes (Tönnis grade >
2, lateral sourcil height <
2 mm and Outerbridge grade 4 changes in the weight-bearing area of the femoral head) were excluded, leaving 96 patients (101 hips) in the study. At a mean follow-up of 2.44 years (2 to 4), the mean modified Harris hip score in the labral repair group (group 1, 69 hips) improved from 60.2 (24 to 85) pre-operatively to 93.6 (55 to 100), and in the labral resection group (group 2, 32 hips) from 62.8 (29 to 96) pre-operatively to 88.8 (35 to 100). The mean modified Harris hip score in the labral repair group was 7.3 points greater than in the resection group (p = 0.036, 95% confidence interval 0.51 to 14.09). Labral detachments were found more frequently in the labral repair group and labral flap tears in the resection group. No patient in our study group required a subsequent hip replacement during the period of follow-up. This study shows that patients without advanced degenerative changes in the hip can achieve significant improvement in their symptoms after arthroscopic treatment of femoroacetabular impingement. Where appropriate, labral repair provides a superior result to labral resection.
The combination of an irreparable tear of the rotator cuff and destructive arthritis of the shoulder joint may cause severe pain, disability and loss of independence in the aged. Standard anatomical shoulder replacements depend on a functioning rotator cuff, and hence may fail in the presence of tears in the cuff. Many designs of non-anatomical constrained or semi-constrained prostheses have been developed for cuff tear arthropathy, but have proved unsatisfactory and were abandoned. The DePuy Delta III reverse prosthesis, designed by Grammont, medialises and stabilises the centre of rotation of the shoulder joint and has shown early promise. This study evaluated the mid-term clinical and radiological results of this arthroplasty in a consecutive series of 50 shoulders in 43 patients with a painful pseudoparalysis due to an irreparable cuff tear and destructive arthritis, performed over a period of seven years by a single surgeon. A follow-up of 98% was achieved, with a mean duration of 39 months (8 to 81). The mean age of the patients at the time of surgery was 81 years (59 to 95). The female to male ratio was 5:1. During the seven years, six patients died of natural causes. The clinical outcome was assessed using the American Shoulder and Elbow score, the Oxford Shoulder Score and the Short-form 36 score. A radiological review was performed using the Sirveaux score for scapular notching. The mean American Shoulder and Elbow score was 19 (95% confidence interval (CI) 14 to 23) pre-operatively, and 65 (95% CI 48 to 82) (paired The mean maximum elevation improved from 55° pre-operatively to 105° at final follow-up. There were seven complications during the whole series, although only four patients required further surgery.
We prospectively evaluated the long-term outcome of 158 consecutive patients who underwent revision total hip replacement using uncemented computer-assisted design-computer-assisted manufacture femoral components. There were 97 men and 61 women. Their mean age was 63.1 years (34.6 to 85.9). The mean follow-up was 10.8 years (10 to 12). The mean Oxford, Harris and Western Ontario and McMaster hip scores improved from 41.1, 44.2 and 52.4 pre-operatively to 18.2, 89.3 and 12.3, respectively (p <
0.0001, for each). Six patients required further surgery. The overall survival of the femoral component was 97% (95% confidence interval 94.5 to 99.7). These results are comparable to those of previously published reports for revision total hip replacement using either cemented or uncemented components.
Total shoulder replacement is a successful procedure for degenerative or some inflammatory diseases of the shoulder. However, fixation of the glenoid seems to be the main weakness with a high rate of loosening. The results using all-polyethylene components have been better than those using metal-backed components. We describe our experience with 35 consecutive total shoulder replacements using a new metal-backed glenoid component with a mean follow-up of 75.4 months (48 to 154). Our implant differs from others because of its mechanism of fixation. It has a convex metal-backed bone interface and the main stabilising factor is a large hollow central peg. The patients were evaulated with standard radiographs and with the Constant Score, the Simple Shoulder Test and a visual analogue scale. All the scores improved and there was no loosening, no polyethylene-glenoid disassembly and no other implant-related complications. We conclude that a metal-backed glenoid component is a good option in total shoulder replacement with no worse results than of those using a cemented all-polyethylene prosthesis.
We present the 10- to 17-year results of 112 computer-assisted design computer-assisted manufacture femoral components. The total hip replacements were performed between 1992 and 1998 in 111 patients, comprising 53 men and 58 women. Their mean age was 46.2 years (24.6 to 62.2) with a mean follow-up of 13 years (10 to 17). The mean Harris Hip Score improved from 42.4 (7 to 99) to 90.3 (38 to 100), the mean Oxford Hip Score from 43.1 (12 to 59) to 18.2 (12 to 51) and the mean Western Ontario MacMasters University Osteoarthritis Index score from 57.0 (7 to 96) to 11.9 (0 to 85). There was one revision due to failure of the acetabular component but no failures of the femoral component. There were no revisions for aseptic loosening. The worst-case survival in this cohort of custom femoral components at 13.2 years follow-up was 98.2% (95% confidence interval 95 to 99). Overall survival of this series of total hip replacements was 97.3% (95% confidence interval 95 to 99). These results are comparable with the best medium- to long-term results for femoral components used in primary total hip replacement with any means of fixation.
Fractures of the proximal humerus can lead to malalignment of the humeral head, necrosis and post-traumatic osteoarthritis. In such cases surface replacement might be a promising option. A total of 28 shoulders with glenohumeral arthritis subsequent to a fracture underwent surface replacement arthroplasty of the humeral head in patients with a mean age of 60 years (35 to 83). On the basis of the inclination of the impacted head, post-traumatic arthritis was divided into three types: type 1, an impacted fracture of the head in an anatomical position (seven cases); type 2, a valgus impacted fracture (13 cases); type 3, a varus impacted fracture (eight cases). The outcome was measured by means of the Constant score. According to the Boileau classification of the sequelae of fractures of the proximal humerus, all 28 patients had a final result of intra-capsular category 1. The mean Constant score for the 28 shoulders increased from 23.2 points (2 to 45) pre-operatively to 55.1 points (20 to 89) at a mean of 31 months (24 to 66) post-operatively. Valgus impacted fractures had significantly better results (p <
0.039). Surface replacement arthroplasty can provide good results for patients with post-traumatic osteoarthritis of the shoulder. Their use avoids post-operative complications of the humeral shaft, such as peri-prosthetic fractures. Further surgery can be undertaken more easily as the bone stock is preserved.
We report the clinical and radiological outcome of total ankle replacement performed in conjunction with hindfoot fusion or in isolation. Between May 2003 and June 2008, 60 ankles were treated with total ankle replacement with either subtalar or triple fusion, and the results were compared with a control group of 288 ankles treated with total ankle replacement alone. After the mean follow-up of 39.5 months (12 to 73), the ankles with hindfoot fusion showed significant improvement in the mean visual analogue score for pain (p <
0.001), the mean American Orthopaedic Foot and Ankle Society score (p <
0.001), and the mean of a modified version of this score (p <
0.001). The mean visual analogue pain score (p = 0.304) and mean modified American Orthopaedic Foot and Ankle Society score (p = 0.119) were not significantly different between the hindfoot fusion and the control groups. However, the hindfoot fusion group had a significantly lower mean range of movement (p = 0.009) and a higher rate of posterior focal osteolysis (p = 0.04). Both groups showed various complications (p = 0.131) and failure occurring at a similar rate (p = 0.685). Subtalar or triple fusion is feasible and has minimal adverse effects on ankles treated with total ankle replacement up to midterm follow-up. The clinical outcome of total ankle replacement when combined with hindfoot fusion is comparable to that of ankle replacement alone. Thus, hindfoot fusion should be performed in conjunction with total ankle replacement when indicated.
We describe a case of rotational dislocation of the proximal interphalangeal joint of the ring finger. This injury was not initially appreciated and therefore closed reduction failed. The clinical findings included puckering of the skin on the dorsum of the joint and rotational incongruity on radiographs. At open reduction there was interposition of the lateral band. Identification of the pathology allows early successful treatment.
We describe the outcome at a mean follow-up of 8.75 years (7.6 to 9.8) of seven patients who had undergone osteochondral autologous transplantation for full-thickness cartilage defects of the shoulder between 1998 and 2000. These patients have been described previously at a mean of 32.6 months when eight were included. One patient has been lost to follow-up. The outcome was assessed by the Constant shoulder score and the Lysholm knee score to assess any donor-site morbidity. Standard radiographs and MR scores were obtained and compared with the pre-operative findings and the results from the previous review. No patient required any further surgery on the shoulder. The mean Constant score improved significantly until the final follow-up (p = 0.018). The Lysholm score remained excellent throughout. There was a significant progression of osteoarthritic changes from the initial surgery to the first and final follow-up but this did not appear to be related to the size of the defect, the number of cylinders required or the Constant score (p = 0.016). MRI showed that all except one patient had a congruent joint surface at the defect with full bony integration of all osteochondral cylinders. The results have remained satisfactory over a longer period with very good objective and subjective findings.
We reviewed the outcome of 28 patients who had been treated using the Aequalis fracture prosthesis for an acute fracture of the proximal humerus at a mean follow-up of 39.3 months (24 to 63). The mean age of the patients at the time of the fracture was 66.3 years (38 to 80). The mean Constant score was 68.2 (37 to 84) for the operated shoulder, which represented 89.5% of the mean score for the uninjured side (p <
0.001). The quality of the reconstruction as shown on the immediate post-operative radiographs was categorised into three types, anatomical, acceptable, and unacceptable, depending on the position of the tuberosities relative to the prosthetic head and the humeral shaft. Anatomical reconstruction was associated with a higher mean Constant score as well as higher mean values of anterior forward elevation, abduction and external rotation than the other types, but the differences were not statistically significant (p >
0.231). A total of 18 patients had active anterior elevation ≥150°. Their mean active abduction and external rotation were 163.6° and 31.3°, respectively. In seven of the 28 patients, the mean active anterior elevation, abduction and external rotation were 130.7°, 129.2° and 22.8°, respectively. In all, 12 patients were very satisfied with the results, 12 were satisfied, two were dissatisfied and two were disappointed; 26 reported no or only mild pain while only two had moderate pain. In five patients proximal migration of the humeral head was shown on the anteroposterior radiographs of the shoulder. No evidence of loosening was found in any prosthesis.
A cadaver study using six pairs of lower limbs was conducted to investigate the accuracy of computer navigation and standard instrumentation for the placement of the Birmingham Hip Resurfacing femoral component. The aim was to place all the femoral components with a stem-shaft angle of 135°. The mean stem-shaft angle obtained in the standard instrumentation group was 127.7° (120° to 132°), compared with 133.3° (131° to 139°) in the computer navigation group (p = 0.03). The scatter obtained with computer-assisted navigation was approximately half that found using the conventional jig. Computer navigation was more accurate and more consistent in its placement of the femoral component than standard instrumentation. We suggest that image-free computer-assisted navigation may have an application in aligning the femoral component during hip resurfacing.
Our aim in this prospective study was to evaluate the outcome of total shoulder replacement in the treatment of young and middle-aged active patients with primary glenohumeral osteoarthritis. We reviewed 21 patients (21 shoulders) with a mean age of 55 years (37 to 60). The mean follow-up was seven years (5 to 9). The same anatomical, third-generation, cemented implant had been used in all patients. All the patients were evaluated radiologically and clinically using the Constant and Murley score. No patients required revision. In one a tear of the supraspinatus tendon occurred. Overall, 20 patients (95%) were either very satisfied (n = 18) or satisfied (n = 2) with the outcome. Significant differences (p <
0.0001) were found for all categories of the Constant and Murley score pre- and post-operatively. The mean Constant and Murley score increased from 24.1 points (10 to 45) to 64.5 points (39 to 93), and the relative score from 30.4% (11% to 50%) to 83% (54% to 116%). No clinical or radiological signs of loosening of the implant were seen. For young and middle-aged patients with osteoarthritis, third-generation total shoulder replacement is a viable method of treatment with a low rate of complications and excellent results in the mid-term.
Between 1998 and 2007, 22 patients with fractures of the scapula had operative treatment more than three weeks after injury. The indications for operation included displaced intra-articular fractures, medialisation of the glenohumeral joint, angular deformity, or displaced double lesions of the superior shoulder suspensory complex. Radiological and functional outcomes were obtained for 16 of 22 patients. Disabilities of the Arm, Shoulder, Hand (DASH) and Short form-36 scores were collected for 14 patients who were operated on after March 2002. The mean delay from injury to surgery was 30 days (21 to 57). The mean follow-up was for 27 months (12 to 72). At the last review the mean DASH score was 14 (0 to 41). Of the 16 patients with follow-up, 13 returned to their previous employment and recreational activities without restrictions. No wound complications, infection or nonunion occurred. Malunion of the scapula can be prevented by surgical treatment of fractures in patients with delayed presentation. Surgery is safe, effective, and gives acceptable functional results.
The floating shoulder is defined as ipsilateral fractures of the midshaft of the clavicle and the neck of the glenoid. This rare injury can be difficult to manage without a thorough understanding of the complex anatomy of the shoulder girdle. Surgical intervention needs to be considered for all of these injuries. While acceptable results can be expected with non-operative management of minimally-displaced fractures, displacement at one or both sites is best managed with surgical reduction and fixation.
We studied retrospectively the results of revision arthroplasty of the elbow using a linked Coonrad-Morrey implant in 23 patients (24 elbows) after a mean follow-up period of 55 months. According to the Mayo Elbow Performance Score, 19 elbows were satisfactory, nine were excellent and ten good. The median total score had improved from 35 points (20 to 75) before the primary arthroplasty to 85 points (40 to 100) at the latest follow-up. There was a marked relief of pain, but the range of movement showed no overall improvement. Two patients had a second revision because of infection and two for aseptic loosening. The estimated five-year survival rate of the prosthesis was 83.1% (95% confidence interval 61.1 to 93.3). Revision elbow arthroplasty using the Coonrad-Morrey implant provided satisfactory results but with complications occurring in 13 cases.