Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.Aims
Methods
Aims. The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. Methods. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time. Results. The estimated survivorship free of all-cause revision was 98% (95% confidence interval (CI) 97% to 98%), 95% (95% CI 95% to 96%), and 86% (95% CI 83% to 88%) at two, five, and ten years, respectively. In 209/369 revisions there was a consistent constellation of findings with varying severity that included polyethylene wear and associated synovitis, osteolysis, and component loosening. This
Aims. This study aimed to determine outcomes of isolated tibial insert exchange (ITIE) during revision total knee arthroplasty (TKA). Methods. From 1985 to 2016, 270 ITIEs were performed at one institution for instability (55%, n = 148), polyethylene wear (39%, n = 105), insert fracture/dissociation (5%, n = 14), or stiffness (1%, n = 3). Patients with component loosening, implant malposition, infection, and extensor mechanism problems were excluded. Results. Survivorship free of any re-revision was 68% at ten years. For the indication of insert wear, survivorship free of any re-revision at ten years was 74%. Re-revisions were more frequent for index diagnoses other than wear (hazard ratio (HR) 1.9; p = 0.013), with ten-year survivorships of 69% for instability and 37% for insert fracture/dissociation. Following ITIE for wear, the most common reason for re-revision was aseptic loosening (33%, n = 7). For other indications, the most common reason for re-revision was recurrence of the original diagnosis. Mean Knee Society Scores improved from 54 (0 to 94) preoperatively to 77 (38 to 94) at ten years. Conclusion. After ITIE, the risk and reasons for re-revision correlated with preoperative indications. The best results were for polyethylene wear. For other diagnoses, the re-revision rate was higher and the
Aims. We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. Methods. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson
Aims. Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. Methods. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation. Results. Multivariate Fine-Gray regression analyses revealed different hazards for each Henderson
Our aim was to compare the biomechanical properties of suturing methods to determine a better method for the repair of lacerated skeletal muscle. We tested Kessler stitches and the combination of Mason-Allen and perimeter stitches. Individual stitches were placed in the muscle belly of quadriceps femoris from a pig cadaver and were tensioned mechanically. The maximum loads and strains were measured and
The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
Nonagenarians (aged 90 to 99 years) have experienced the fastest percent decile population growth in the USA recently, with a consequent increase in the prevalence of nonagenarians living with joint arthroplasties. As such, the number of revision total hip arthroplasties (THAs) and total knee arthroplasties (TKAs) in nonagenarians is expected to increase. We aimed to determine the mortality rate, implant survivorship, and complications of nonagenarians undergoing aseptic revision THAs and revision TKAs. Our institutional total joint registry was used to identify 96 nonagenarians who underwent 97 aseptic revisions (78 hips and 19 knees) between 1997 and 2018. The most common indications were aseptic loosening and periprosthetic fracture for both revision THAs and revision TKAs. Mean age at revision was 92 years (90 to 98), mean BMI was 27 kg/m2 (16 to 47), and 67% (n = 65) were female. Mean time between primary and revision was 18 years (SD 9). Kaplan-Meier survival was used for patient mortality, and compared to age- and sex-matched control populations. Reoperation risk was assessed using cumulative incidence with death as a competing risk. Mean follow-up was five years.Aims
Methods
Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components. We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.Aims
Methods
Total femoral arthroplasty (TFA) is a rare procedure used in cases of significant femoral bone loss, commonly from cancer, infection, and trauma. Low patient numbers have resulted in limited published work on long-term outcomes, and even less regarding TFA undertaken for non-oncological indications. The aim of this study was to evaluate the long-term clinical outcomes of all TFAs in our unit. Data were collected retrospectively from a large tertiary referral revision arthroplasty unit’s database. Inclusion criteria included all patients who underwent TFA in our unit. Preoperative demographics, operative factors, and short- and long-term outcomes were collected for analysis. Outcome was defined using the Musculoskeletal Infection Society (MSIS) outcome reporting tool.Aims
Methods
Aims. Unicompartmental knee arthroplasty (UKA) has higher revision rates than total knee arthroplasty (TKA). As revision of UKA may be less technically demanding than revision TKA, UKA patients with poor functional outcomes may be more likely to be offered revision than TKA patients with similar outcomes. The aim of this study was to compare clinical thresholds for revisions between TKA and UKA using revision incidence and patient-reported outcomes, in a large, matched cohort at early, mid-, and late-term follow-up. Methods. Analyses were performed on propensity score-matched patient cohorts of TKAs and UKAs (2:1) registered in the New Zealand Joint Registry between 1 January 1999 and 31 December 2019 with an Oxford Knee Score (OKS) response at six months (n, TKA: 16,774; UKA: 8,387), five years (TKA: 6,718; UKA: 3,359), or ten years (TKA: 3,486; UKA: 1,743). Associations between OKS and revision within two years following the score were examined. Thresholds were compared using receiver operating characteristic analysis. Reasons for aseptic revision were compared using cumulative incidence with competing risk. Results. Fewer TKA patients with ‘poor’ outcomes (≤ 25) subsequently underwent revision compared with UKA at six months (5.1% vs 19.6%; p < 0.001), five years (4.3% vs 12.5%; p < 0.001), and ten years (6.4% vs 15.0%; p = 0.024). Compared with TKA, the relative risk for UKA was 2.5-times higher for ‘unknown’ reasons, bearing dislocations, and disease progression. Conclusion. Compared with TKA, more UKA patients with poor outcomes underwent revision from early to long-term follow-up, and were more likely to undergo revision for ‘unknown’ reasons, which suggest a lower clinical threshold for UKA. For UKA, revision risk was higher for bearing dislocations and disease progression. There is supporting evidence that the higher revision UKA rates are associated with lower clinical thresholds for revision and additional
Adverse spinal motion or balance (spine mobility) and adverse pelvic mobility, in combination, are often referred to as adverse spinopelvic mobility (SPM). A stiff lumbar spine, large posterior standing pelvic tilt, and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Adverse SPM can create functional malposition of the acetabular components and hence is an instability risk. Adverse pelvic mobility is often, but not always, associated with abnormal spinal motion parameters. Dislocation rates for dual-mobility articulations (DMAs) have been reported to be between 0% and 1.1%. The aim of this study was to determine the early survivorship from the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a DMA. A multicentre study was performed using data from 227 patients undergoing primary total hip arthroplasty (THA), enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameter had a DMA inserted at the time of their surgery. The mean age was 76 years (22 to 93) and 63% were female (n = 145). At a mean of 14 months (5 to 31) postoperatively, the AOANJRR was analyzed for follow-up information. Reasons for revision and types of revision were identified.Aims
Methods
Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model.Aims
Methods
The aim of this study was to measure the effect of hospital case volume on the survival of revision total knee arthroplasty (RTKA). This is a retrospective analysis of Scottish Arthroplasty Project data, a nationwide audit which prospectively collects data on all arthroplasty procedures performed in Scotland. The primary outcome was RTKA survival at ten years. The primary explanatory variable was the effect of hospital case volume per year on RTKA survival. Kaplan-Meier survival curves were plotted with 95% confidence intervals (CIs) to determine the lifespan of RTKA. Multivariate Cox proportional hazards were used to estimate relative revision risks over time. Hazard ratios (HRs) were reported with 95% CI, and p-value < 0.05 was considered statistically significant.Aims
Methods
We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.Aims
Methods
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to
We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of
Aims. The aims of this study were to validate the outcome of total elbow arthroplasty (TEA) in patients with rheumatoid arthritis (RA), and to identify factors that affect the outcome. Methods. We searched PubMed, MEDLINE, Cochrane Reviews, and Embase from between January 2003 and March 2019. The primary aim was to determine the implant failure rate, the
The aim of this study was to describe implant and patient-reported outcome in patients with a unilateral transfemoral amputation (TFA) treated with a bone-anchored, transcutaneous prosthesis. In this cohort study, all patients with a unilateral TFA treated with the Osseointegrated Prostheses for the Rehabilitation of Amputees (OPRA) implant system in Sahlgrenska University Hospital, Gothenburg, Sweden, between January 1999 and December 2017 were included. The cohort comprised 111 patients (78 male (70%)), with a mean age 45 years (17 to 70). The main reason for amputation was trauma in 75 (68%) and tumours in 23 (21%). Patients answered the Questionnaire for Persons with Transfemoral Amputation (Q-TFA) before treatment and at two, five, seven, ten, and 15 years’ follow-up. A prosthetic activity grade was assigned to each patient at each timepoint. All mechanical complications, defined as fracture, bending, or wear to any part of the implant system resulting in removal or change, were recorded.Aims
Methods
The aim of this meta-analysis was to compare the outcome of total elbow arthroplasty (TEA) undertaken for rheumatoid arthritis (RA) with TEA performed for post-traumatic conditions with regard to implant failure, functional outcome, and perioperative complications. We completed a comprehensive literature search on PubMed, Web of Science, Embase, and the Cochrane Library and conducted a systematic review and meta-analysis. Nine cohort studies investigated the outcome of TEA between RA and post-traumatic conditions. The preferred reporting items for systematic reviews and meta-analysis (Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA)) guidelines and Newcastle-Ottawa scale were applied to assess the quality of the included studies. We assessed three major outcome domains: implant failures (including aseptic loosening, septic loosening, bushing wear, axle failure, component disassembly, or component fracture); functional outcomes (including arc of range of movement, Mayo Elbow Performance Score (MEPS), and the Disabilities of the Arm, Shoulder and Hand (DASH) questionnaire), and perioperative complications (including deep infection, intraoperative fracture, postoperative fracture, and ulnar neuropathy).Aims
Materials and Methods
The aim of this study was to report the results of custom-made endoprostheses with extracortical plates plus or minus a short, intramedullary stem aimed at preserving the physis after resection of bone sarcomas in children. Between 2007 and 2017, 18 children aged less than 16 years old who underwent resection of bone sarcomas, leaving ≤ 5 cm of bone from the physis, and reconstruction with a custom-made endoprosthesis were reviewed. Median follow-up was 67 months (interquartile range 45 to 91). The tumours were located in the femur in 11 patients, proximal humerus in six, and proximal tibia in one.Aims
Patients and Methods
Aims. Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and
Aims. Joint-preserving surgery of the hip (JPSH) has evolved considerably
and now includes a number of procedures, including arthroscopy,
surgical dislocation, and redirectional osteotomies of the femur
and acetabulum. There are a number of different factors which lead
to failure of JPSH. Consequently, it is of interest to assess the
various
Unicompartmental knee arthroplasty (UKA) is a potential treatment
for isolated bone on bone osteoarthritis when limited to a single
compartment. The risk for revision of UKA is three times higher
than for total knee arthroplasty (TKA). The aim of this review was
to discuss the different revision options after UKA failure. A search was performed for English language articles published
between 2006 and 2016. After reviewing titles and abstracts, 105
papers were selected for further analysis. Of these, 39 papers were
deemed to contain clinically relevant data to be included in this review.Objectives
Materials and Methods
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
The aim of this study was to investigate the effect of a posterior
malleolar fragment (PMF), with <
25% ankle joint surface, on
pressure distribution and joint-stability. There is still little
scientific evidence available to advise on the size of PMF, which
is essential to provide treatment. To date, studies show inconsistent
results and recommendations for surgical treatment date from 1940. A total of 12 cadaveric ankles were assigned to two study groups.
A trimalleolar fracture was created, followed by open reduction
and internal fixation. PMF was fixed in Group I, but not in Group
II. Intra-articular pressure was measured and cyclic loading was
performed.Aims
Materials and Methods
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,
and therefore, improve patient satisfaction post-operatively. Cite this article:
The aim of this to study was to compare the previously unreported
long-term survival outcome of the Oxford medial unicompartmental
knee arthroplasty (UKA) performed by trainee surgeons and consultants. We therefore identified a previously unreported cohort of 1084
knees in 947 patients who had a UKA inserted for anteromedial knee
arthritis by consultants and surgeons in training, at a tertiary
arthroplasty centre and performed survival analysis on the group
with revision as the endpoint.Aims
Patients and Methods
Since redesign of the Oxford phase III mobile-bearing unicompartmental
knee arthroplasty (UKA) femoral component to a twin-peg design,
there has not been a direct comparison to total knee arthroplasty
(TKA). Thus, we explored differences between the two cohorts. A total of 168 patients (201 knees) underwent medial UKA with
the Oxford Partial Knee Twin-Peg. These patients were compared with
a randomly selected group of 177 patients (189 knees) with primary
Vanguard TKA. Patient demographics, Knee Society (KS) scores and
range of movement (ROM) were compared between the two cohorts. Additionally,
revision, re-operation and manipulation under anaesthesia rates
were analysed.Aims
Patients and Methods
Hip resurfacing has been proposed as an alternative
to traditional total hip arthroplasty in young, active patients.
Much has been learned following the introduction of metal-on-metal resurfacing
devices in the 1990s. The triad of a well-designed device, implanted
accurately, in the correct patient has never been more critical
than with these implants. Following Food and Drug Administration approval in 2006, we studied
the safety and effectiveness of one hip resurfacing device (Birmingham
Hip Resurfacing) at our hospital in a large, single-surgeon series.
We report our early to mid–term results in 1333 cases followed for
a mean of 4.3 years (2 to 5.7) using a prospective, observational
registry. The mean patient age was 53.1 years (12 to 84); 70% were
male and 91% had osteoarthritis. Complications were few, including
no dislocations, no femoral component loosening, two femoral neck
fractures (0.15%), one socket loosening (0.08%), three deep infections
(0.23%), and three cases of metallosis (0.23%). There were no destructive
pseudotumours. Overall survivorship at up to 5.7 years was 99.2%. Aseptic survivorship
in males under the age of 50 was 100%. We believe this is the largest
United States series of a single surgeon using a single resurfacing
system. Cite this article:
A retrospective study was conducted to investigate
the changes in metal ion levels in a consecutive series of Birmingham
Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed
250 BHRs implanted in 232 patients between 1998 and 2001. Implant
survival, clinical outcome (Harris hip score), radiographs and serum chromium
(Cr) and cobalt (Co) ion levels were assessed. Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost
to follow-up and ten had been revised. The remaining 202 BHRs in
190 patients (136 men and 54 women; mean age at surgery 50.5 years
(17 to 76)) were evaluated at a minimum follow-up of ten years (mean
10.8 years (10 to 13.6)). The overall implant survival at 13.2 years
was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris
hip score was 97.7 (median 100; 65 to 100). Median and mean ion
levels were low for unilateral resurfacings (Cr: median 1.3 µg/l,
mean
1.95 µg/l (<
0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l
(<
0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l,
mean 3.46 µg/l (<
0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66
µg/l (<
0.5 to 9.5)). In 80 unilateral BHRs with sequential ion
measurements, Cr and Co levels were found to decrease significantly
(p <
0.001) from the initial assessment at a median of six years
(4 to 8) to the last assessment at a median of 11 years (9 to 13),
with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co.
Three female patients had a >
2.5 µg/l increase of Co ions, associated with
head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there
was no significant difference in change of ion levels between genders
(Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505;
Co, p = 0.370). Higher acetabular component inclination angles correlated
with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002).
Patients with increased ion levels had lower Harris hip scores (p
= 0.038). In conclusion, in well-functioning BHRs the metal ion levels
decreased significantly at ten years. An increase >
2.5 µg/l was
associated with poor function. Cite this article:
We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). Our study shows that the performance of the BHR continues to
be good at 12- to 15-year follow-up. Men have better implant survival
(98.0%; 95% CI 97.4 to 98.6) at 15 years than women (91.5%; 95%
CI 89.8 to 93.2), and women <
60 years (90.5%; 95% CI 88.3 to
92.7) fare worse than others. Hip dysplasia and osteonecrosis are
risk factors for failure. Patients under 50 years with osteoarthritis
fare best (99.4%; 95% CI 98.8 to 100 survival at 15 years), with
no failures in men in this group. Cite this article:
Metal-on-metal hip arthroplasty gained significant
favor in the first decade of the millennium. However, the past several
years have seen increasing reports of failure, pseudotumor and other
adverse reactions. This study presents the results of a single center’s
15-year experience with metal-on-metal total hip arthroplasty as
strong evidence that metal-on-metal is going, going, gone.
The Gamma nail is frequently used in unstable
peri-trochanteric hip fractures. We hypothesised that mechanical failure
of the Gamma nail was associated with inadequate proximal three-point
fixation. We identified a consecutive series of 299 Gamma nails
implanted in 299 patients over a five-year period, 223 of whom fulfilled
our inclusion criteria for investigation. The series included 61
men and 162 women with a mean age of 81 years (20 to 101). Their fractures
were classified according to the Modified Evans’ classification
and the quality of fracture reduction was graded. The technical
adequacy of three points of proximal fixation was recorded from
intra-operative fluoroscopic images, and technical inadequacy for
each point was defined. All patients were followed to final follow-up
and mechanical failures were identified. A multivariate statistical
analysis was performed, adjusting for confounders. A total of 16
failures (7.2%) were identified. The position of the lag screw relative
to the lateral cortex was the most important point of proximal fixation,
and when inadequate the failure rate was 25.8% (eight of 31: odds
ratio 7.5 (95% confidence interval 2.5 to 22.7), p <
0.001). Mechanical failure of the Gamma nail in peri-trochanteric femoral
fractures is rare (<
1%) when three-point proximal fixation is
achieved. However, when proximal fixation is inadequate, failure
rates increase. The strongest predictor of failure is positioning
the lateral end of the lag screw short of the lateral cortex. Adherence
to simple technical points minimises the risk of fixation failure
in this vulnerable patient group. Cite this article:
Radiological evidence of post-traumatic osteoarthritis
(PTOA) after fracture of the tibial plateau is common but end-stage arthritis
which requires total knee arthroplasty is much rarer. The aim of this study was to examine the indications for, and
outcomes of, total knee arthroplasty after fracture of the tibial
plateau and to compare this with an age and gender-matched cohort
of TKAs carried out for primary osteoarthritis. Between 1997 and 2011, 31 consecutive patients (23 women, eight
men) with a mean age of 65 years (40 to 89) underwent TKA at a mean
of 24 months (2 to 124) after a fracture of the tibial plateau.
Of these, 24 had undergone ORIF and seven had been treated non-operatively.
Patients were assessed pre-operatively and at 6, 12 and >
60 months
using the Short Form-12, Oxford Knee Score and a patient satisfaction
score. Patients with instability or nonunion needed total knee arthroplasty
earlier (14 and 13.3 months post-injury) than those with intra-articular
malunion (50 months, p <
0.001). Primary cruciate-retaining implants
were used in 27 (87%) patients. Complication rates were higher in
the PTOA cohort and included wound complications (13% Total knee arthroplasty undertaken after fracture of the tibial
plateau has a higher rate of complications than that undertaken
for primary osteoarthritis, but patient-reported outcomes and satisfaction
are comparable. Cite this article:
We evaluated the biomechanical properties of two different methods of fixation for unstable fractures of the proximal humerus. Biomechanical testing of the two groups, locking plate alone (LP), and locking plate with a fibular strut graft (LPSG), was performed using seven pairs of human cadaveric humeri. Cyclical loads between 10 N and 80 N at 5 Hz were applied for 1 000 000 cycles. Immediately after cycling, an increasing axial load was applied at a rate of displacement of 5 mm/min. The displacement of the construct, maximum failure load, stiffness and
Despite the increasing interest and subsequent published literature on hip resurfacing arthroplasty, little is known about the prevalence of its complications and in particular the less common
We compared the bulking and tensile strength of the Pennington modified Kessler, Cruciate and the Savage repairs in an ex vivo model. A total of 60 porcine tendons were randomised to three groups, half repaired using a core suture alone and the remainder employing a core and peripheral technique. The tendons were distracted to failure. The force required to produce a 3 mm gap, the ultimate strength, the
We report the effects of local administration of osteogenic protein-1 on the biomechanical properties of the overstretched anterior cruciate ligament in an animal model. An injury in the anterior cruciate ligament was created in 45 rabbits. They were divided into three equal groups. In group 1, no treatment was applied, in group II, phosphate-buffered saline was applied around the injured ligament, and in group III, 12.5 μg of osteogenic protein-1 mixed with phosphate-buffered saline was applied around the injured ligament. A control group of 15 rabbits was assembled from randomly-selected injured knees from among the first three groups. Each rabbit was killed at 12 weeks. The maximum load and stiffness of the anterior cruciate ligament was found to be significantly greater in group III than either group 1 (p = 0.002, p = 0.014) or group II (p = 0.032, p = 0.025). The tensile strength and the tangent modulus of fascicles from the ligament were also significantly greater in group III than either group I (p = 0.002, p = 0.0174) or II (p = 0.005, p = 0.022). The application of osteogenic protein-1 enhanced the healing in the injured anterior cruciate ligament, but compared with the control group the treated ligament remained lengthened. The administration of osteogenic protein-1 may have a therapeutic role in treating the overstretched anterior cruciate ligament.
Filling the empty holes in peri-articular locking
plates may improve the fatigue strength of the fixation. The purpose of
this A locking/compression plate was applied to 33 synthetic femurs
and then a 6 cm metaphyseal defect was created (AO Type 33-A3).
The specimens were then divided into three groups: unplugged, plugged
with locking screw only and fully plugged holes. They were then
tested using a stepwise or run-out fatigue protocol, each applying
cyclic physiological multiaxial loads. All specimens in the stepwise group failed at the 770 N load
level. The mean number of cycles to failure for the stepwise specimen
was 25 500 cycles ( In conclusion, filling the empty combination locking/compression
holes in peri-articular distal femur locking plates at the level
of supracondylar comminution does not increase the fatigue life
of the fixation in a comminuted supracondylar femoral fracture model
(AO 33-A3) with a 6 cm gap.
We investigated the effect of stimulation with a pulsed electromagnetic field on the osseointegration of hydroxyapatite in cortical bone in rabbits. Implants were inserted into femoral cortical bone and were stimulated for six hours per day for three weeks. Electromagnetic stimulation improved osseointegration of hydroxyapatite compared with animals which did not receive this treatment in terms of direct contact with the bone, the maturity of the bone and mechanical fixation. The highest values of maximum push-out force (Fmax) and ultimate shear strength (σu) were observed in the treated group and differed significantly from those of the control group at three weeks (Fmax; p <
0.0001; σu, p <
0.0005).