header advert
Results 1 - 1 of 1
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 5 | Pages 726 - 728
1 Sep 1995
McCallum J Scott R

Osteoarthritis of the medial compartment of the knee often shows a specific pattern of anterior wear. Review of our revisions from a series of medial metal-backed Brigham unicondylar knee replacements performed between 1983 and 1989 showed that this wear pattern was common on the tibial polyethylene surface. We reviewed these cases retrospectively to compare the pattern of preoperative erosion with the wear of the prosthesis. In all 14 knees with severe anterior wear in a unicompartmental replacement, the prearthroplasty radiographs showed similar patterns, suggesting that the implanted tibial component may continue to be subjected to the same localised stresses that precipitated the failure of the original articular cartilage. Many tibial components implanted during the 1980s had an unacceptably thin anterior rim of polyethylene and it seems that greater thickness is essential at the anterior and peripheral margins of the tibial plateau.