We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article:
The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
We used the stainless steel cable grip system described by Dall and Miles in 1983 to fix trochanters in 40 hips after total arthroplasty with trochanteric osteotomy. The cable broke in 32.5% of the hips; the trochanter failed to unite in 37.5%. Significantly more cables broke when placed inside the femoral canal than when the cable was placed round the femoral shaft (58% as against 9.5%, difference p less than 0.01). The high incidence of breakage may have resulted from contact between the stainless steel cable and the titanium prosthesis, from the acute angulation, or because of the lower
There are three basic concepts that are important to the biomechanics of pedicle screw-based instrumentation. First, the outer diameter of the screw determines pullout strength, while the inner diameter determines
Filling the empty holes in peri-articular locking
plates may improve the
Sterilisation by gamma irradiation in the presence of air causes free radicals generated in polyethylene (PE) to react with oxygen, which could lead to loss of physical properties and reduction in
We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.Aims
Methods
Temporary epiphysiodesis (ED) is commonly applied in children and adolescents to treat leg length discrepancies (LLDs) and tall stature. Traditional Blount staples or modern two-hole plates are used in clinical practice. However, they require accurate planning, precise surgical techniques, and attentive follow-up to achieve the desired outcome without complications. This study reports the results of ED using a novel rigid staple (RigidTack) incorporating safety, as well as technical and procedural success according to the idea, development, evaluation, assessment, long-term (IDEAL) study framework. A cohort of 56 patients, including 45 unilateral EDs for LLD and 11 bilateral EDs for tall stature, were prospectively analyzed. ED was performed with 222 rigid staples with a mean follow-up of 24.4 months (8 to 49). Patients with a predicted LLD of ≥ 2 cm at skeletal maturity were included. Mean age at surgery was 12.1 years (8 to 14). Correction and complication rates including implant-associated problems, and secondary deformities as well as perioperative parameters, were recorded (IDEAL stage 2a). These results were compared to historical cohorts treated for correction of LLD with two-hole plates or Blount staples.Aims
Methods
We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed. All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation.Aims
Methods
The objective of this five-year prospective, blinded, randomised
controlled trial (RCT) was to compare femoral head penetration into
a vitamin E diffused highly cross-linked polyethylene (HXLPE) liner
with penetration into a medium cross-linked polyethylene control
liner using radiostereometric analysis. Patients scheduled for total hip arthroplasty (THA) were randomised
to receive either the study E1 (32 patients) or the control ArComXL
polyethylene (35 patients). The median age (range) of the overall
cohort was 66 years (40 to 76).Aims
Patients and Methods
A flexed knee gait is common in patients with bilateral spastic
cerebral palsy and occurs with increased age. There is a risk for
the recurrence of a flexed knee gait when treated in childhood,
and the aim of this study was to investigate whether multilevel
procedures might also be undertaken in adulthood. At a mean of 22.9 months (standard deviation 12.9), after single
event multi level surgery, 3D gait analysis was undertaken pre-
and post-operatively for 37 adult patients with bilateral cerebral
palsy and a fixed knee gait.Aims
Patients and Methods
Coloured bone cements have been introduced to
make the removal of cement debris easier at the time of primary and
revision joint replacement. We evaluated the physical, mechanical
and pharmacological effects of adding methylene blue to bone cement
with or without antibiotics (gentamicin, vancomycin or both). The
addition of methylene blue to plain cement significantly decreased
its mean setting time (570 seconds (
We investigated the effect of pre-heating a femoral component on the porosity and strength of bone cement, with or without vacuum mixing used for total hip replacement. Cement mantles were moulded in a manner simulating clinical practice for cemented hip replacement. During polymerisation, the temperature was monitored. Specimens of cement extracted from the mantles underwent bending or fatigue tests, and were examined for porosity. Pre-heating the stem alone significantly increased the mean temperature values measured within the mantle (+14.2°C) (p <
0.001) and reduced the mean curing time (−1.5 min) (p <
0.001). The addition of vacuum mixing modulated the mean rise in the temperature of polymerisation to 11°C and reduced the mean duration of the process by one minute and 50 seconds (p = 0.01 and p <
0.001, respectively). In all cases, the maximum temperature values measured in the mould simulating the femur were <
50°C. The mixing technique and pre-heating the stem slightly increased the static mechanical strength of bone cement. However, the fatigue life of the cement was improved by both vacuum mixing and pre-heating the stem, but was most marked (+ 280°C) when these methods were combined. Pre-heating the stem appears to be an effective way of improving the quality of the cement mantle, which might enhance the long-term performance of bone cement, especially when combined with vacuum mixing.
Management of bisphosphonate-associated subtrochanteric
fractures remains opinion- or consensus-based. There are limited
data regarding the outcomes of this fracture. We retrospectively reviewed 33 consecutive female patients with
a mean age of 67.5 years (47 to 91) who were treated surgically
between May 2004 and October 2009. The mean follow-up was 21.7 months
(0 to 53). Medical records and radiographs were reviewed to determine
the post-operative ambulatory status, time to clinical and radiological
union and post-fixation complications such as implant failure and
need for second surgery. The predominant fixation method was with an extramedullary device
in 23 patients. 25 (75%) patients were placed on wheelchair mobilisation
or no weight-bearing initially. The mean time to full weight-bearing
was 7.1 months (2.2 to 29.7). The mean time for fracture site pain
to cease was 6.2 months (1.2 to 17.1). The mean time to radiological
union was 10.0 months (2.2 to 27.5). Implant failure was seen in
seven patients (23%, 95 confidence interval (CI) 11.8 to 40.9).
Revision surgery was required in ten patients (33%, 95 CI 19.2 to
51.2). A large proportion of the patients required revision surgery
and suffered implant failure. This fracture is associated with slow
healing and prolonged post-operative immobility. Cite this article:
Focal femoral inlay resurfacing has been developed
for the treatment of full-thickness chondral defects of the knee. This
technique involves implanting a defect-sized metallic or ceramic
cap that is anchored to the subchondral bone through a screw or
pin. The use of these experimental caps has been advocated in middle-aged
patients who have failed non-operative methods or biological repair
techniques and are deemed unsuitable for conventional arthroplasty
because of their age. This paper outlines the implant design, surgical
technique and biomechanical principles underlying their use. Outcomes
following implantation in both animal and human studies are also reviewed. Cite this article:
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article:
The menisci of the knee have an important role in load-bearing and shock absorption within the joint. They may also function as secondary stabilisers, have a proprioceptive role, and aid the lubrication and nutrition of the articular cartilage. Complete or partial loss of a meniscus can have damaging effects on a knee, leading to serious long-term sequelae. This paper reviews the consequences of meniscectomy and summarises the body of evidence in the literature regarding those factors most relevant to long-term outcome.
Bone allografts can store and release high levels of vancomycin. We present our results of a two-stage treatment for infected hip arthroplasty with acetabular and femoral impaction grafting using vancomycin-loaded allografts. We treated 29 patients (30 hips) by removal of the implants, meticulous debridement, parenteral antibiotic therapy and second-stage reconstruction using vancomycin-supplemented impacted bone allografts and a standard cemented Charnley femoral component. The mean follow-up was 32.4 months (24 to 60). Infection control was obtained in 29 cases (re-infection rate of 3.3%; 95% confidence interval 0.08 to 17) without evidence of progressive radiolucent lines, demarcation or graft resorption. One patient had a further infection ten months after revision caused by a different pathogen. Associated post-operative complications were one traumatic periprosthetic fracture at 14 months, a single dislocation in two hips and four displacements of the greater trochanter. Vancomycin-supplemented allografts restored bone stock and provided sound fixation with a low incidence of further infection.