Aims. Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using
Aims. This study aimed to quantify the shoulder kinematics during an apprehension-relocation test in patients with anterior shoulder instability (ASI) and glenoid bone loss using the
Aims. The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using
Aims. Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using
Aims. Thresholds of acceptable early migration of the components in total knee arthroplasty (TKA) have traditionally ignored the effects of patient and implant factors that may influence migration. The aim of this study was to determine which of these factors are associated with overall longitudinal migration of well-fixed tibial components following TKA. Methods.
Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using
Aims. Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. Methods. In this single-blind, randomized controlled, clinical
Aims. Stemless humeral implants have been developed to overcome stem-related complications in total shoulder arthroplasty (TSA). However, stemless implant designs may hypothetically result in less stable initial fixation, potentially affecting long-term survival. The aim of this study is to investigate early fixation and migration patterns of the stemless humeral component of the Simpliciti Shoulder System and to evaluate clinical outcomes. Methods. In this prospective cohort study,
Refobacin Bone Cement R and Palacos Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.Aims
Methods
Aims. The aim of this study was to compare the migration of the femoral component, five years postoperatively, between patients with a highly cross-linked polyethylene (HXLPE) insert and those with a conventional polyethylene (PE) insert in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary aims included clinical outcomes and patient-reported outcome measures (PROMs). We have previously reported the migration and outcome of the tibial components in these patients. Methods. A double-blinded randomized controlled trial was conducted including 96 TKAs. The migration of the femoral component was measured with
Aims. The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). Methods. A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with
Aims. The aim of this study was to compare a bicruciate-retaining (BCR) total knee arthroplasty (TKA) with a posterior cruciate-retaining (CR) TKA design in terms of kinematics, measured using fluoroscopy and stability as micromotion using
Aims. The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Methods. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using
Aims. We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. Methods. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with
Aims. The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS). Methods. A randomized controlled
Aims. The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using
Aims.
Aims. Vitamin E-diffused, highly crosslinked polyethylene (VEPE) and porous titanium-coated (PTC) shells were introduced in total hip arthroplasty (THA) to reduce the risk of aseptic loosening. The purpose of this study was: 1) to compare the wear properties of VEPE to moderately crosslinked polyethylene; 2) to assess the stability of PTC shells; and 3) to report their clinical outcomes at seven years. Patients and Methods. A total of 89 patients were enrolled into a prospective study. All patients received a PTC shell and were randomized to receive a VEPE liner (n = 44) or a moderately crosslinked polyethylene (ModXLPE) liner (n = 45).
Aims. For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using
Aims. Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. Methods. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with
Aims. Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by
Aims. The direct anterior (DA) approach has been associated with rapid patient recovery after total hip arthroplasty (THA) but may be associated with more frequent femoral complications including implant loosening. The objective of this study was to determine whether the addition of a collar to the femoral stem affects implant migration, patient activity, and patient function following primary THA using the DA approach. Methods. Patients were randomized to either a collared (n = 23) or collarless (n = 26) cementless femoral stem implanted using the DA approach. Canal fill ratio (CFR) was measured on the first postoperative radiographs. Patients underwent a supine
Aims. Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using
Aims. The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs. . Patients and Methods. The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (. sd. 8.4) in the PSI group and 69.4 years (. sd. 8.4) in the CSI group. All patients received the same TKA implant. Intraoperative surgical resources and any surgical waste generated were recorded. Patients underwent
In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by
Aims. The introduction of a novel design of total knee arthroplasty (TKA) must achieve outcomes at least as good as existing designs. A novel design of TKA with a reducing radius of the femoral component and a modified cam-post articulation has been released and requires assessment of the fixation to bone.
Aims. Early implant migration measured with
Aims. Post-operative migration of cemented acetabular components as
measured by
Aims. The aim of this study was to evaluate the long-term inducible
displacement of cemented tibial components ten years after total
knee arthroplasty (TKA). Patients and Methods. A total of 15 patients from a previously reported prospective
trial of fixation using
Aims.
Aims. The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure. Patients and Methods. A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving
Aims. The increased in vivo resistance to wear of
highly crosslinked polyethylene (HXLPE) in total hip arthroplasty
(THA) has led to an increased use of larger articulations which
have been shown to reduce the incidence of early dislocation. To
date, there are few reports of the wear of larger articulations
using second generation HXLPE liners. Our prospective cohort study
measured the bedding-in and early wear of large (36 mm and 40 mm
diameter) articulations involving a second generation X3 HXLPE liner
and compared our findings with previous clinical and in vitro studies
of the same material. Patients and Methods. The proximal penetration of the femoral head five years post-operatively
was measured for 15 patients using
The Exeter femoral stem is a double-tapered highly
polished collarless cemented implant with good long-term clinical
results. In order to determine why the stem functions well we have
undertaken a long-term
We performed a three-year
We report the five-year outcome of a randomised
controlled trial which used
This was a randomised controlled trial studying
the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised
that it would have similar early clinical results and micromovement
to those of a standard-length tapered Synergy metaphyseal fixation
stem. Using
We analysed the axis of movement in the normal elbow during flexion in vivo using
In a single-blind, randomised series of knee replacements in 116 patients, we used
We studied the quality of fixation of the tibial component using
We obtained simultaneous measurements of sagittal knee laxity in 12 consecutive patients after reconstruction of the anterior cruciate ligament (ACL), using the Stryker laxity tester and
We studied the stability of cemented all-polyethylene keeled glenoid components by
BoneMaster is a thin electrochemically applied hydroxyapatite (HA) coating for orthopaedic implants that is quickly resorbed during osseointegration. Early stabilization is a surrogacy marker of good survival of femoral stems. The hypothesis of this study was that a BoneMaster coating yields a fast early and lasting fixation of stems. A total of 53 patients were randomized to be treated using Bi-Metric cementless femoral stems with either only a porous titanium plasma-sprayed coating (P group) or a porous titanium plasma-sprayed coating with an additional BoneMaster coating (PBM group). The patients were examined with radiostereometry until five years after surgery.Aims
Methods
Oxidized zirconium (OxZi) and highly cross-linked polyethylene (HXLPE) were developed to minimize wear and risk of osteolysis in total hip arthroplasty (THA). However, retrieval studies have shown that scratched femoral heads may lead to runaway wear, and few reports of long-term results have been published. The purpose of this investigation is to report minimum ten-year wear rates and clinical outcomes of THA with OxZi femoral heads on HXLPE, and to compare them with a retrospective control group of cobalt chrome (CoCr) or ceramic heads on HXLPE. From 2003 to 2006, 108 THAs were performed on 96 patients using an OxZi head with a HXLPE liner with minimum ten-year follow-up. Harris Hip Scores (HHS) were collected preoperatively and at the most recent follow-up (mean 13.3 years). Linear and volumetric liner wear was measured on radiographs of 85 hips with a minimum ten-year follow-up (mean 14.5 years). This was compared to a retrospective control group of 45 THAs using ceramic or CoCr heads from October 1999 to February 2005, with a minimum of ten years’ follow-up.Aims
Methods
The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components. Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).Aims
Methods
The aim of this study was to compare the biomechanical models of two frequently used techniques for reconstructing severe acetabular defects with pelvic discontinuity in revision total hip arthroplasty (THA) – the Trabecular Metal Acetabular Revision System (TMARS) and custom triflange acetabular components (CTACs) – using virtual modelling. Pre- and postoperative CT scans from ten patients who underwent revision with the TMARS for a Paprosky IIIB acetabular defect with pelvic discontinuity were retrospectively collated. Computer models of a CTAC implant were designed from the preoperative CT scans of these patients. Computer models of the TMARS reconstruction were segmented from postoperative CT scans using a semi-automated method. The amount of bone removed, the implant-bone apposition that was achieved, and the restoration of the centre of rotation of the hip were compared between all the actual TMARS and the virtual CTAC implants.Aims
Methods
In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased taper corrosion, with the release of metal ions and adverse reactions. The aim of this study was to investigate the relationship between the size of the femoral head and the levels of metal ions in the blood in these patients. A total of 96 patients were enrolled at two centres and randomized to undergo MoP THA using either a 32 mm metal head or a femoral head of between 36 mm and 44 mm in size, being the largest possible to fit the thinnest available polyethylene insert. The levels of metal ions and patient-reported outcome measures (Oxford Hip Score, University of California, Los Angeles Activity Scale) were recorded at two and five years postoperatively.Aims
Methods
The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA). A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.Aims
Methods
To analyze the short-term outcome of two types of total wrist arthroplasty (TWA) in terms of wrist function, migration, and periprosthetic bone behaviour. A total of 40 patients suffering from non-rheumatoid wrist arthritis were enrolled in a randomized controlled trial comparing the ReMotion and Motec TWAs. Patient-rated and functional outcomes, radiological changes, blood metal ion levels, migration measured by model-based radiostereometric analysis (RSA), bone mineral density (BMD) measured by dual-energy X-ray absorptiometry (DXA), complications, loosening, and revision rates at two years were compared.Aims
Methods
Many surgeons choose to perform total knee arthroplasty (TKA) surgery with the aid of a tourniquet. A tourniquet is a device that fits around the leg and restricts blood flow to the limb. There is a need to understand whether tourniquets are safe, and if they benefit, or harm, patients. The aim of this study was to determine the benefits and harms of tourniquet use in TKA surgery. We searched MEDLINE, EMBASE, Cochrane Central Register of Controlled trials, and trial registries up to 26 March 2020. We included randomized controlled trials (RCTs), comparing TKA with a tourniquet versus without a tourniquet. Outcomes included: pain, function, serious adverse events (SAEs), blood loss, implant stability, duration of surgery, and length of hospital stay.Aims
Methods
Higher osteoblastic bone activity is expected in aseptic loosening and painful unicompartmental knee arthroplasty (UKA). However, insights into normal bone activity patterns after medial UKAs are lacking. The aim of this study was to identify the evolution in bone activity pattern in well-functioning medial mobile-bearing UKAs. In total, 34 patients (13 female, 21 male; mean age 62 years (41 to 79); BMI 29.7 kg/m2 (23.6 to 42.1)) with 38 medial Oxford partial UKAs (20 left, 18 right; 19 cementless, 14 cemented, and five hybrid) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively, and at one and two years postoperatively. Changes in mean osteoblastic activity were investigated using a tracer localization scheme with volumes of interest (VOIs), reported by normalized mean tracer values. A SPECT/CT registration platform additionally explored cortical tracer evolution in zones of interest identified by previous experimental research.Aims
Methods
To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period. This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.Aims
Methods
The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.Aims
Methods
Our intention was to investigate if the highly porous biological fixation surfaces of a new 3D-printed total knee arthroplasty (TKA) achieved adequate fixation of the tibial and patellar components to the underlying bone. A total of 29 patients undergoing primary TKA consented to participate in this prospective cohort study. All patients received a highly porous tibial baseplate and metal-backed patella. Patient-reported outcomes measures were recorded and implant migration was assessed using radiostereometric analysis.Aims
Patients and Methods
The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad).Aims
Methods
A retrospective study was conducted to measure short-term Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear.Aims
Patients and Methods
Patient-specific instrumentation of total knee arthroplasty (TKA) is a technique permitting the targeting of individual kinematic alignment, but deviation from a neutral mechanical axis may have implications on implant fixation and therefore survivorship. The primary objective of this randomized controlled study was to compare the fixation of tibial components implanted with patient-specific instrumentation targeting kinematic alignment (KA+PSI) A total of 47 patients due to undergo TKA were randomized to KA+PSI (n = 24) or MA+CAS (n = 23). In the KA+PSI group, there were 16 female and eight male patients with a mean age of 64 years (Aims
Patients and Methods
The aim of this study was to determine whether there is a difference
in the rate of wear between acetabular components positioned within
and outside the ‘safe zones’ of anteversion and inclination angle. We reviewed 100 hips in 94 patients who had undergone primary
total hip arthroplasty (THA) at least ten years previously. Patients
all had the same type of acetabular component with a bearing couple
which consisted of a 28 mm cobalt-chromium head on a highly crosslinked
polyethylene (HXLPE) liner. A supine radiostereometric analysis
(RSA) examination was carried out which acquired anteroposterior
(AP) and lateral paired images. Acetabular component anteversion
and inclination angles were measured as well as total femoral head
penetration, which was divided by the length of implantation to
determine the rate of polyethylene wear.Aims
Patients and Methods
The aim of this study was to determine the stability of a new
short femoral stem compared with a conventional femoral stem in
patients undergoing cementless total hip arthroplasty (THA), in
a prospective randomized controlled trial using radiostereometric
analysis (RSA). A total of 53 patients were randomized to receive cementless
THA with either a short femoral stem (MiniHip, 26 patients, mean
age: 52 years, nine male) or a conventional length femoral stem
(MetaFix, 23 patients, mean age: 53 years, 11 male). All patients
received the same cementless acetabular component. Two-year follow-up
was available on 38 patients. Stability was assessed through migration
and dynamically inducible micromotion. Radiographs for RSA were
taken postoperatively and at three, six, 12, 18, and 24 months.Aims
Patients and Methods
Our aim in this study was to describe a continuing review of
11 total hip arthroplasties using 22.225 mm Alumina ceramic femoral
heads on a Charnley flanged femoral component, articulating against
a silane crosslinked polyethylene. Nine patients (11 THAs) were reviewed at a mean of 27.5 years
(26 to 28) post-operatively. Outcome was assessed using the d’Aubigne
and Postel, and Charnley scores and penetration was recorded on
radiographs. In addition, the oxidation of a 29-year-old shelf-aged
acetabular component was analysed.Aims
Patients and Methods
Hydroxyapatite coatings for uncemented fixation in total knee
arthroplasty can theoretically provide a long-lasting biological
interface with the host bone. The objective of this study was to
test this hypothesis with propriety hydroxyapatite, peri-apatite,
coated tibial components using component migration measured with radiostereometric
analysis over two years as an indicator of long-term fixation. A total of 29 patients at two centres received uncemented PA-coated
tibial components and were followed for two years with radiostereometric
analysis exams to quantify the migration of the component.Aims
Patients and Methods
To compare the structural durability of hydroxyapatite-tricalcium
phosphate (HATCP) to autologous iliac crest bone graft in calcaneal
lengthening osteotomy (CLO) for pes planovalgus in childhood. We present the interim results of ten patients (HATCP, n = 6
and autograft, n = 5) with a mean age of 11.5 years (8.2 to 14.2)
from a randomised controlled non-inferiority trial with six months
follow-up. The primary outcome was the stability of the osteotomy
as measured by radiostereometric analysis. A non-inferiority margin
of ≤ 2 mm osteotomy compression was set.Aims
Patients and Methods
The aim of this study was to quantify the stability
of fracture-implant complex in fractures after fixation. A total
of 15 patients with an undisplaced fracture of the femoral neck,
treated with either a dynamic hip screw or three cannulated hip
screws, and 16 patients with an AO31-A2 trochanteric fracture treated
with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric
analysis was used at six weeks, four months and 12 months post-operatively
to evaluate shortening and rotation. Migration could be assessed in ten patients with a fracture of
the femoral neck and seven with a trochanteric fracture. By four
months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1)
had occurred in the fracture of the femoral neck group and 5.0 mm
(-0.13 to 12.9) in the trochanteric fracture group. A wide range
of rotation occurred in both types of fracture. Right-sided trochanteric
fractures seem more rotationally stable than left-sided fractures. This prospective study shows that migration at the fracture site
occurs continuously during the first four post-operative months,
after which stabilisation occurs. This information may allow the
early recognition of patients at risk of failure of fixation. Cite this article:
The optimal method of tibial component fixation remains uncertain
in total knee arthroplasty (TKA). Hydroxyapatite coatings have been
applied to improve bone ingrowth in uncemented designs, but may
only coat the directly accessible surface. As peri-apatite (PA)
is solution deposited, this may increase the coverage of the implant
surface and thereby fixation. We assessed the tibial component fixation
of uncemented PA-coated TKAs Patients were randomised to PA-coated or cemented TKAs. In 60
patients (30 in each group), radiostereometric analysis of tibial
component migration was evaluated as the primary outcome at baseline,
three months post-operatively and at one, two and five years. A
linear mixed-effects model was used to analyse the repeated measurements.Aims
Patients and Methods
The aim of this study was to compare the incidence of aseptic
loosening after the use of a cemented acetabular component and a
Trabecular Metal (TM) acetabular component (Zimmer Inc., Warsaw,
Indiana) at acetabular revision with bone impaction grafting. A total of 42 patients were included in the study. Patients were
randomised to receive an all- polyethylene cemented acetabular component
(n = 19) or a TM component (n = 23). Radiostereometric analysis
and conventional radiographic examinations were performed regularly
up to two years post-operatively or until further revision.Aims
Patients and Methods
This study aimed to determine the diagnostic performance of radiographic
criteria to detect aseptic acetabular loosening after revision total
hip arthroplasty (THA). Secondary aims were to determine the predictive
values of different thresholds of migration and to determine the
predictive values of radiolucency criteria. Acetabular component migration to re-revision was measured retrospectively
using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements
(Sutherland method) in two groups: Group A, 52 components (48 patients) found
not loose at re-revision and Group B, 42 components (36 patients)
found loose at re-revision between 1980 and 2015. The presence and
extent of radiolucent lines was also assessed.Aims
Patients and Methods
Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck. We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years. Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D. Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years. The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.
We performed a randomised, radiostereometric study comparing two different bone cements, one of which has been sparsely clinically documented. Randomisation of 60 total hip replacements (57 patients) into two groups of 30 was undertaken. All the patients were operated on using a cemented Charnley total hip replacement, the only difference between groups being the bone cement used to secure the femoral component. The two cements used were Palamed G and Palacos R with gentamicin. The patients were followed up with repeated clinical and radiostereometric examinations for two years to assess the micromovement of the femoral component and the clinical outcome. The mean subsidence was 0.18 mm and 0.21 mm, and the mean internal rotation was 1.7° and 2.0° at two years for the Palamed G and Palacos R with gentamicin bone cements, respectively. We found no statistically significant differences between the groups. Micromovement occurred between the femoral component and the cement, while the cement mantle was stable inside the bone. The Harris hip score improved from a mean of 38 points (14 to 54) and 36 (10 to 57) pre-operatively to a mean of 92 (77 to 100) and 91 (63 to 100) at two years in the Palamed G and Palacos R groups, respectively. No differences were found between the groups. Both bone cements provided good initial fixation of the femoral component and good clinical results at two years.
Femoral impaction bone grafting was first developed in 1987 using
morselised cancellous bone graft impacted into the femoral canal
in combination with a cemented, tapered, polished stem. We describe
the evolution of this technique and instrumentation since that time. Between 1987 and 2005, 705 revision total hip arthroplasties
(56 bilateral) were performed with femoral impaction grafting using
a cemented femoral stem. All surviving patients were prospectively
followed for a mean of 14.7 years (9.8 to 28.3) with no loss to
follow-up. By the time of the final review, 404 patients had died.Aims
Patients and Methods
The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol. Migration of the tibial component in
all planes was compared between initial and ten-year follow-up.
Outcome scores including the Knee Society Score, Western Ontario
and McMaster Universities Arthritis Index, 12-item Short Form Health
Survey, Forgotten Joint Score, and University of California, Los
Angeles Activity Score were recorded.Aims
Patients and Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
We investigated the stability of seven Schatzker type II fractures of the lateral tibial plateau treated by subchondral screws and a buttress plate followed by immediate partial weight-bearing. In order to assess the stability of the fracture, weight-bearing inducible displacements of the fracture fragments and their migration over a one-year period were measured by differentially loaded radiostereometric analysis and standard radiostereometric analysis, respectively. The mean inducible craniocaudal fracture fragment displacements measured −0.30 mm (−0.73 to 0.02) at two weeks and 0.00 mm (−0.12 to 0.15) at 52 weeks. All inducible displacements were elastic in nature under all loads at each examination during follow-up. At one year, the mean craniocaudal migration of the fracture fragments was −0.34 mm (−1.64 to 1.51). Using radiostereometric methods, this case series has shown that in the Schatzker type II fractures investigated, internal fixation with subchondral screws and a buttress plate provided adequate stability to allow immediate post-operative partial weight-bearing, without harmful consequences.
Metaphyseal bone loss is common with revision
total knee replacement (RTKR). Using the Anderson Orthopaedic Research
Institute (AORI) classification, type 2-B and type 3 defects usually
require large metal blocks, bulk structural allograft or highly
porous metal cones. Tibial and femoral trabecular metal metaphyseal
cones are a unique solution for large bone defects. These cones
substitute for bone loss, improve metaphyseal fixation, help correct
malalignment, restore the joint line and may permit use of a shorter
stem. The technique for insertion involves sculpturing of the remaining
bone with a high speed burr and rasp, followed by press-fit of the
cone into the metaphysis. The fixation and osteoconductive properties
of the porous cone outer surface allow ingrowth and encourage long-term
biological fixation. The revision knee component is then cemented
into the porous cone inner surface, which provides superior fixation
compared with cementing into native but deficient metaphyseal bone.
The advantages of the cone compared with allograft include: technical
ease, biological fixation, no resorption, and possibly a lower risk
of infection. The disadvantages include: difficult extraction and
relatively short-term follow-up. Several studies using cones report
promising short-term results for the reconstruction of large bone
defects in RTKR. Cite this article:
Percutaneous physiodesis is an established technique for treating mild leg-length discrepancy and problems of expected extreme height. Angular deformities resulting from incomplete physeal arrest have been reported, and little is known about the time interval from percutaneous physiodesis to actual physeal arrest. This procedure was carried out in ten children, six with leg-length discrepancy and four with expected extreme height. Radiostereometric analysis was used to determine the three-dimensional dynamics of growth retardation. Errors of measurement of translation were less than 0.05 mm and of rotation less than 0.06°. Physeal arrest was obtained in all but one child within 12 weeks after physiodesis and no clinically-relevant angular deformities occurred. This is a suitable method for following up patients after percutaneous physiodesis. Incomplete physeal arrest can be detected at an early stage and the procedure repeated before corrective osteotomy is required.
Resurfacing of the humeral head is commonly used
within the UK to treat osteoarthritis (OA) of the shoulder. We present
the results of a small prospective randomised study of this procedure
using the Global CAP prosthesis with two different coatings, Porocoat
and DuoFix hydroxyapatite (HA). We followed two groups of ten patients
with OA of the shoulder for two years after insertion of the prosthesis
with tantalum marker beads, recording pain, Constant–Murley and
American Shoulder and Elbow Surgeons (ASES) outcome scores, and
using radiostereometric analysis to assess migration. The outcomes
were similar to those of other series, with significant reductions
in pain (p = 0.003) and an improvement in the Constant (p = 0.001)
and ASES scores (p = 0.006). The mean migration of the prosthesis
three months post-operatively was 0.78 mm (0.51 to 1.69) and 0.72
mm (0.33 to 1.45) for the Porocoat and DuoFix groups, respectively.
Analysis of variance indicated that the rate of migration reached
a plateau after three months post-operatively in both groups. At
follow-up of two years the mean migration was 1 mm ( The addition of a coating of HA to the sintered surface does
not improve fixation of this prosthesis. Cite this article:
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented acetabular component fixation in
total hip arthroplasty (THA) offers advantages over retention by
improving the cement-bone interface, without jeopardising implant
stability. We have previously published two-year follow-up data
of a randomised controlled trial (RCT), in which 50 patients with
primary osteoarthritis were randomised to either retention or removal
of the SCBP. The mean age of the retention group (n = 25, 13 males)
was 70.0 years ( Cite this article:
The most common reasons for revision of unicompartmental
knee arthroplasty (UKA) are loosening and pain. Cementless components
may reduce the revision rate. The aim of this study was to compare
the fixation and clinical outcome of cementless and cemented Oxford
UKAs. A total of 43 patients were randomised to receive either a cemented
or a cementless Oxford UKA and were followed for two years with
radiostereometric analysis (RSA), radiographs aligned with the bone–implant
interfaces and clinical scores. The femoral components migrated significantly during the first
year (mean 0.2 mm) but not during the second. There was no significant
difference in the extent of migration between cemented and cementless
femoral components in either the first or the second year. In the
first year the cementless tibial components subsided significantly
more than the cemented components (mean 0.28 mm ( As second-year migration is predictive of subsequent loosening,
and as radiolucency is suggestive of reduced implant–bone contact,
these data suggest that fixation of the cementless components is
at least as good as, if not better than, that of cemented devices. Cite this article:
Total knee arthroplasty (TKA) is known to lead
to a reduction in periprosthetic bone mineral density (BMD). In theory,
this may lead to migration, instability and aseptic loosening of
the prosthetic components. Bisphosphonates inhibit bone resorption
and may reduce this loss in BMD. We hypothesised that treatment
with bisphosphonates and calcium would lead to improved BMD and
clinical outcomes compared with treatment with calcium supplementation
alone following TKA. A total of 26 patients, (nine male and 17 female,
mean age 67 years) were prospectively randomised into two study
groups: alendronate and calcium (bisphosphonate group, n = 14) or calcium
only (control group, n = 12). Dual energy X-ray absorptiometry (DEXA)
measurements were performed post-operatively, and at three months,
six months, one, two, four, and seven years post-operatively. Mean femoral metaphyseal BMD was significantly higher in the
bisphosphonate group compared with controls, up to four years following
surgery in some areas of the femur (p = 0.045). BMD was observed
to increase in the lateral tibial metaphysis in the bisphosphonate
group until seven years (p = 0.002), and was significantly higher than
that observed in the control group throughout (p = 0.024). There
were no significant differences between the groups in the central
femoral metaphyseal, tibial medial metaphyseal or diaphyseal regions
of interest (ROI) of either the femur or tibia. Bisphosphonate treatment after TKA may be of benefit for patients
with poor bone quality. However, further studies with a larger number
of patients are necessary to assess whether this is clinically beneficial. Cite this article:
We report the results of revision total knee
replacement (TKR) in 26 patients with major metaphyseal osteolytic defects
using 29 trabecular metal cones in conjunction with a rotating hinged
total knee prosthesis. The osteolytic defects were types II and
III (A or B) according to the Anderson Orthopaedic Research Institute
(AORI) classification. The mean age of the patients was 72 years
(62 to 84) and there were 15 men and 11 women. In this series patients had
undergone a mean of 2.34 previous total knee arthroplasties. The
main objective was to restore anatomy along with stability and function
of the knee joint to allow immediate full weight-bearing and active
knee movement. Outcomes were measured using Knee Society scores,
Oxford knee scores, range of movement of the knee and serial radiographs.
Patients were followed for a mean of 36 months (24 to 49). The mean
Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20
(32 to 38) (p <
0.001) and mean American Knee Society scores
improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p <
0.001).
No radiolucent lines suggestive of loosening were seen around the trabecular
metal cones, and by one year all the radiographs showed good osteo-integration.
There was no evidence of any collapse or implant migration. Our
early results confirm the findings of others that trabecular metal
cones offer a useful way of managing severe bone loss in revision
TKR. Cite this article:
The optimum cementing technique for the tibial
component in cemented primary total knee replacement (TKR) remains
controversial. The technique of cementing, the volume of cement
and the penetration are largely dependent on the operator, and hence
large variations can occur. Clinical, experimental and computational
studies have been performed, with conflicting results. Early implant
migration is an indication of loosening. Aseptic loosening is the
most common cause of failure in primary TKR and is the product of
several factors. Sufficient penetration of cement has been shown
to increase implant stability. This review discusses the relevant literature regarding all aspects
of the cementing of the tibial component at primary TKR. Cite this article:
An international faculty of orthopaedic surgeons
presented their work on the current challenges in hip surgery at
the London Hip Meeting which was attended by over
400 delegates. The topics covered included femoroacetabular impingement, thromboembolic
phenomena associated with hip surgery, bearing surfaces (including metal-on-metal
articulations), outcomes of hip replacement surgery and revision
hip replacement. We present a concise report of the current opinions
on hip surgery from this meeting with appropriate references to
the current literature.
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the mobile bearing was also evaluated. A series of 42 knees was randomised
to MB or FB tibial components with appropriate polyethylene inserts
and followed for between ten and 12 years, or until the death of
the patient. The polyethylene in the MB group was superior in that
it was gamma-irradiated in inert gas and was calcium-stearate free;
the polyethylene in the FB group was gamma-irradiated in air and
contained calcium stearate. In theory this should be advantageous
to the wear rate of the MB group. At final follow-up the overall
mean migration was 0.75 mm ( For the MB group, the mean linear wear rate on the under-surface
was 0.026 mm/year (
In this prospective study we studied the effect
of the inclination angle of the acetabular component on polyethylene wear
and component migration in cemented acetabular sockets using radiostereometric
analysis. A total of 120 patients received either a cemented Reflection
All-Poly ultra-high-molecular-weight polyethylene or a cemented
Reflection All-Poly highly cross-linked polyethylene acetabular
component, combined with either cobalt–chrome or Oxinium femoral
heads. Femoral head penetration and migration of the acetabular
component were assessed with repeated radiostereometric analysis
for two years. The inclination angle was measured on a standard
post-operative anteroposterior pelvic radiograph. Linear regression
analysis was used to determine the relationship between the inclination
angle and femoral head penetration and migration of the acetabular component. We found no relationship between the inclination angle and penetration
of the femoral head at two years’ follow-up (p = 0.9). Similarly,
our data failed to reveal any statistically significant correlation
between inclination angle and migration of these cemented acetabular
components (p = 0.07 to p = 0.9).
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
Previously, radiostereometric analysis following hip revision performed using impacted morsellised allograft bone and a cemented Exeter stem has shown continuous subsidence of the stem for up to five years. It is not known whether the subsidence continues thereafter. In our study, 17 of 25 consecutive osteo-arthritic patients with aseptically loose stems who underwent first-time revision using impacted morsellised allograft bone and a cemented Exeter stem were followed by yearly radiostereometric examinations for nine years. The mean subsidence at six weeks was 1.1 mm (0.1 to 2.3), from six weeks to one year 1.3 mm (0 to 2.6), from one to five years 0.7 mm (0 to 2.0), and from five to nine years 0.7 mm (0.1 to 3.1). That from six weeks to nine years was 2.7 mm (0 to 6.4) (95% confidence interval 2.0 to 3.5). The Charnley pain score significantly improved after revision, and was maintained at nine years, but walking ability deteriorated slightly as follow-up extended. Of the eight patients who were not followed for nine years, two had early subsidence exceeding 11 mm. Our findings show that in osteo-arthritic patients who undergo revision for aseptic loosening of the stem using impacted morsellised allograft bone and a cemented Exeter stem, migration of the stem continues over nine years at a slower rate after the first year, but without clinical deterioration or radiological loosening.
The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively. The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year). The linear wear rate of the Oxford UKR remains very low into the third decade. We believe that phase-2 bearings had lower wear rates than phase-1 implants because of the improved bearing design and surgical technique which decreased the incidence of impingement. We conclude that the design of the Oxford UKR gives low rates of wear in the long term.
This was a safety study where the hypothesis was that the newer-design CPCS femoral stem would demonstrate similar early clinical results and micromovement to the well-established Exeter stem. Both are collarless, tapered, polished cemented stems, the only difference being a slight lateral to medial taper with the CPCS stem. A total of 34 patients were enrolled in a single-blinded randomised controlled trial in which 17 patients received a dedicated radiostereometric CPCS stem and 17 a radiostereometric Exeter stem. No difference was found in any of the outcome measures pre-operatively or post-operatively between groups. At two years, the mean subsidence for the CPCS stem was nearly half that seen for the Exeter stem (0.77 mm (−0.943 to 1.77) and 1.25 mm (0.719 to 1.625), respectively; p = 0.032). In contrast, the mean internal rotation of the CPCS stem was approximately twice that of the Exeter (1.61° (−1.07° to 4.33°) and 0.59° (0.97° to 1.64°), respectively; p = 0.048). Other migration patterns were not significantly different between the stems. The subtle differences in designs may explain the different patterns of migration. Comparable migration with the Exeter stem suggests that the CPCS design will perform well in the long term.
In a prospective study between 2000 and 2005, 22 patients with primary osteoarthritis of the shoulder had a total shoulder arthroplasty with a standard five-pegged glenoid component, 12 with non-offset humeral head and ten with offset humeral head components. Over a period of 24 months the relative movement of the glenoid component with respect to the scapula was measured using radiostereometric analysis. Nine glenoids needed reaming for erosion. There was a significant increase in rotation about all three axes with time (p <
0.001), the largest occurring about the longitudinal axis (anteversion-retroversion), with mean values of 3.8° and 1.9° for the non-offset and offset humeral head eroded subgroups, respectively. There was also a significant difference in rotation about the anteversion-retroversion axis (p = 0.01) and the varus-valgus (p <
0.001) z-axis between the two groups. The offset humeral head group reached a plateau at early follow-up with rotation about the z-axis, whereas the mean of the non-offset humeral head group at 24 months was three times greater than that of the offset group accounting for the highly significant difference between them.
Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse - femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers. The accuracy of the method was found to be ± 1.6 mm for the x-direction and ± 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.
The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis. We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years. After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial. We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.
Using meta-analysis we compared the survival and clinical outcomes of cemented and uncemented techniques in primary total knee replacement. We reviewed randomised controlled trials and observational studies comparing cemented and uncemented fixation. Our primary outcome was survival of the implant free of aseptic loosening. Our secondary outcome was joint function as measured by the Knee Society score. We identified 15 studies that met our final eligibility criteria. The combined odds ratio for failure of the implant due to aseptic loosening for the uncemented group was 4.2 (95% confidence interval (CI) 2.7 to 6.5) (p <
0.0001). Subgroup analysis of data only from randomised controlled trials showed no differences between the groups for odds of aseptic loosening (odds ratio 1.9, 95% CI 0.55 to 6.40, p = 0.314). The weighted mean difference for the Knee Society score was 0.005 (95% CI −0.26 to 0.26) (p = 0.972). There was improved survival of the cemented compared to uncemented implants, with no statistically significant difference in the mean Knee Society score between groups for all pooled data.
We carried out a prospective randomised study designed to compare the penetration rate of acetabular polyethylene inserts of identical design but different levels of cross-linking at a minimum of four years follow-up. A total of 102 patients (102 hips) were randomised to receive either highly cross-linked Durasul, or contemporary Sulene polyethylene inserts at total hip replacement. A single blinded observer used the Martell system to assess penetration of the femoral head. At a mean follow-up of 4.9 years (4.2 to 6.1) the mean femoral head penetration rate was 0.025 mm/year (
We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from It is clear that both design principles are capable of producing successful long-term results, providing that their specific requirements of stem metallurgy, shape and surface finish, preparation of the bone and handling of the cement are observed.