1 Oct 2008
Kim T Lee D Bin S

We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired t-test, p < 0.05). At the last follow-up, the mean maximum flexion was 135° (110° to 150°). Two knees showed radiolucent lines, but revision was not required because the patients were asymptomatic. Revision was required in one case because of infection, but there were no prosthesis-related revisions. There were no other complications. The estimated survival rate at four years with revision for any reason and prosthesis-related problems was 99.6% and 100%, respectively. This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 37 - 42
1 Jan 2008
Nutton RW van der Linden ML Rowe PJ Gaston P Wade FA

Modifications in the design of knee replacements have been proposed in order to maximise flexion. We performed a prospective double-blind randomised controlled trial to compare the functional outcome, including maximum knee flexion, in patients receiving either a standard or a high flexion version of the NexGen legacy posterior stabilised total knee replacement. A total of 56 patients, half of whom received each design, were assessed pre-operatively and at one year after operation using knee scores and analysis of range of movement using electrogoniometry. For both implant designs there was a significant improvement in the function component of the knee scores (p < 0.001) and the maximum range of flexion when walking on the level, ascending and descending a slope or stairs (all p < 0.001), squatting (p = 0.020) and stepping into a bath (p = 0.024). There was no significant difference in outcome, including the maximum knee flexion, between patients receiving the standard and high flexion designs of this implant


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 221 - 226
1 Mar 2023
Wilton T Skinner JA Haddad FS

Recent publications have drawn attention to the fact that some brands of joint replacement may contain variants which perform significantly worse (or better) than their ‘siblings’. As a result, the National Joint Registry has performed much more detailed analysis on the larger families of knee arthroplasties in order to identify exactly where these differences may be present and may hitherto have remained hidden. The analysis of the Nexgen knee arthroplasty brand identified that some posterior-stabilized combinations have particularly high revision rates for aseptic loosening of the tibia, and consequently a medical device recall has been issued for the Nexgen ‘option’ tibial component which was implicated. More elaborate signal detection is required in order to identify such variation in results in a routine fashion if patients are to be protected from such variation in outcomes between closely related implant types. Cite this article: Bone Joint J 2023;105-B(3):221–226


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims. The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). Methods. At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays. Results. A total of 114 explanted fixed-bearing TKAs were examined. This included 76 used with contemporary PE inserts which were compared with 15 used with older generation PEs. The Attune and NexGen (central locking) trays were found to have significantly less cement cover than Triathlon and PFC trays (peripheral locking group) (p = 0.001). The median planicity values of the PE inserts used with central locking trays were significantly greater than of those with peripheral locking inserts (205 vs 85 microns; p < 0.001). Attune and NexGen inserts had a characteristic pattern of backside deformation, with the outer edges of the PE deviating inferiorly, leaving the PE margins as the primary areas of articulation. Conclusion. Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the frictional torque generated at this interface. Cite this article: Bone Joint J 2021;103-B(12):1791–1801


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 48 - 54
1 Jul 2019
Kahlenberg CA Lyman S Joseph AD Chiu Y Padgett DE

Aims. The outcomes of total knee arthroplasty (TKA) depend on many factors. The impact of implant design on patient-reported outcomes is unknown. Our goal was to evaluate the patient-reported outcomes and satisfaction after primary TKA in patients with osteoarthritis undergoing primary TKA using five different brands of posterior-stabilized implant. Patients and Methods. Using our institutional registry, we identified 4135 patients who underwent TKA using one of the five most common brands of implant. These included Biomet Vanguard (Zimmer Biomet, Warsaw, Indiana) in 211 patients, DePuy/Johnson & Johnson Sigma (DePuy Synthes, Raynham, Massachusetts) in 222, Exactech Optetrak Logic (Exactech, Gainesville, Florida) in 1508, Smith & Nephew Genesis II (Smith & Nephew, London, United Kingdom) in 1415, and Zimmer NexGen (Zimmer Biomet) in 779 patients. Patients were evaluated preoperatively using the Knee Injury and Osteoarthritis Outcome Score (KOOS), Lower Extremity Activity Scale (LEAS), and 12-Item Short-Form Health Survey questionnaire (SF-12). Demographics including age, body mass index, Charlson Comorbidity Index, American Society of Anethesiologists status, sex, and smoking status were collected. Postoperatively, two-year KOOS, LEAS, SF-12, and satisfaction scores were compared between groups. Results. Outcomes were available for 4069 patients (98%) at two years postoperatively. In multiple regression analysis, which separately compared each implant group with the aggregate of all others, there were no clinically significant differences in the change of KOOS score from baseline to two-year follow-up between any of the groups. More than 80% of patients in each group were satisfied at this time in all domains. In a multivariate regression model, patients in the NexGen group were the most likely to be satisfied (odds ratio (OR) 1.63; p = 0.006) and Optetrak Logic patients were the least likely to be satisfied (OR 0.60; p < 0.001). Conclusion. TKA provides improvement in function and satisfaction regardless of the type of implant. We could not demonstrate superiority of one design above others across these groups of implants, and any price premium for one above the other systems may not be justified. Healthcare administrators may find these similarities in outcomes helpful when negotiating purchasing contracts. Cite this article: Bone Joint J 2019;101-B(7 Supple C):48–54


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 483 - 488
1 Apr 2005
Nakayama K Matsuda S Miura H Higaki H Otsuka K Iwamoto Y

We measured the contact areas and contact stresses at the post-cam mechanism of a posterior-stabilised total knee arthroplasty when a posterior force of 500 N was applied to the Kirschner Performance, Scorpio Superflex, NexGen LPS Flex Fixed, and NexGen LPS Flex Mobile knee systems. Measurements were made at 90°, 120°, and 150° of flexion both in neutral rotation and 10° of internal rotation of the tibial component. Peak contact stresses at 90°, 120°, and 150° were 24.0, 33.9, and 28.8 MPa, respectively, for the Kirschner; 26.0, 32.4, and 22.1 MPa, respectively, for the Scorpio; and 34.1, 31.5, and 32.5 MPa, respectively, for the NexGen LPS Flex Fixed. With an internally rotated tibia, the contact stress increased significantly with all the fixed-bearing arthroplasties but not with the NexGen LPS Flex Mobile arthroplasty. The post-cam design should be modified in order to provide a larger contact area whilst avoiding any impingement and edge loading


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1585 - 1593
1 Dec 2008
Henricson A Linder L Nilsson KG

We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 639 - 645
1 May 2010
Kim Y Choi Y Kim J

We undertook a study in which 138 female patients with a mean age of 71.2 years (51 to 82) received a standard NexGen CR-flex prosthesis in one knee and a gender-specific NexGen CR-flex prosthesis in the other. The mean follow-up period was 3.25 years (3.1 to 3.5). The aspect ratios of the standard and gender-specific prostheses were compared with that of the distal femur. The mean post-operative Knee Society knee scores were 94 (70 to 100) and 93 (70 to 100) points and the function scores were 83 (60 to 100) and 84 (60 to 100) points for the standard implants and the gender-specific designs, respectively. The mean post-operative Western Ontario and McMaster Universities score was 26.4 points (0 to 76). Patient satisfaction, the radiological results and the complication rates were similar in the two groups. In those with a standard prosthesis, the femoral component was closely matched in 80 knees (58.0%), overhung in 14 (10.1%) and undercovered the bone in 44 (31.9%). In those with a gender-specific prosthesis, it was closely matched in 15 knees (10.9%) and undercovered the bone in 123 (89.1%). Since we found no significant differences between the two groups with regard to the clinical and radiological results, patient satisfaction or complication rate, the goal of the design of the gender-specific CR-flex prosthesis to improve the outcome was not achieved in our patients


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1180 - 1185
1 Sep 2008
Hamai S Miura H Higaki H Shimoto T Matsuda S Iwamoto Y

Mechanical failure because of wear or fracture of the polyethylene tibial post in posteriorly-stabilised total knee replacements has been extensively described. In this study of 12 patients with a clinically and radiologically successful NexGen LPS posteriorly-stabilised prosthesis impingement of the anterior tibial post was evaluated in vivo in three dimensions during gait using radiologically-based image-matching techniques. Impingement was observed in all images of the patients during the stance phase, although the NexGen LPS was designed to accommodate 14° of hyperextension of the component before impingement occurred. Impingement arises as a result of posterior translation of the femur during the stance phase. Further attention must therefore be given to the configuration of the anterior portion of the femoral component and the polyethylene post when designing posteriorly-stabilised total knee replacements


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 408 - 412
1 Mar 2007
Ma H Lu Y Kwok T Ho F Huang C Huang C

One of the most controversial issues in total knee replacement is whether or not to resurface the patella. In order to determine the effects of different designs of femoral component on the conformity of the patellofemoral joint, five different knee prostheses were investigated. These were Low Contact Stress, the Miller-Galante II, the NexGen, the Porous-Coated Anatomic, and the Total Condylar prostheses. Three-dimensional models of the prostheses and a native patella were developed and assessed by computer. The conformity of the curvature of the five different prosthetic femoral components to their corresponding patellar implants and to the native patella at different angles of flexion was assessed by measuring the angles of intersection of tangential lines. The Total Condylar prosthesis had the lowest conformity with the native patella (mean 8.58°; 0.14° to 29.9°) and with its own patellar component (mean 11.36°; 0.55° to 39.19°). In the other four prostheses, the conformity was better (mean 2.25°; 0.02° to 10.52°) when articulated with the corresponding patellar component. The Porous-Coated Anatomic femoral component showed better conformity (mean 6.51°; 0.07° to 9.89°) than the Miller-Galante II prosthesis (mean 11.20°; 5.80° to 16.72°) when tested with the native patella. Although the Nexgen prosthesis had less conformity with the native patella at a low angle of flexion, this improved at mid (mean 3.57°; 1.40° to 4.56°) or high angles of flexion (mean 4.54°; 0.91° to 9.39°), respectively. The Low Contact Stress femoral component had the best conformity with the native patella (mean 2.39°; 0.04° to 4.56°). There was no significant difference (p > 0.208) between the conformity when tested with the native patella or its own patellar component at any angle of flexion. The geometry of the anterior flange of a femoral component affects the conformity of the patellofemoral joint when articulating with the native patella. A more anatomical design of femoral component is preferable if the surgeon decides not to resurface the patella at the time of operation


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 217 - 223
1 Feb 2014
Namba RS Inacio MCS Cafri G

The outcome of total knee replacement (TKR) using components designed to increase the range of flexion is not fully understood. The short- to mid-term risk of aseptic revision in high flexion TKR was evaluated. The endpoint of the study was aseptic revision and the following variables were investigated: implant design (high flexion vs non-high flexion), the thickness of the tibial insert (≤ 14 mm vs > 14 mm), cruciate ligament (posterior stabilised (PS) vs cruciate retaining), mobility (fixed vs rotating), and the manufacturer (Zimmer, Smith & Nephew and DePuy). Covariates included patient, implant, surgeon and hospital factors. Marginal Cox proportional hazard models were used. In a cohort of 64 000 TKRs, high flexion components were used in 8035 (12.5%). The high flexion knees with tibial liners of thickness > 14 mm had a density of revision of 1.45/100 years of observation, compared with 0.37/100 in non-high flexion TKR with liners ≤ 14 mm thick. Relative to a standard fixed PS TKR, the NexGen (Zimmer, Warsaw, Indiana) Gender Specific Female high flexion fixed PS TKR had an increased risk of revision (hazard ratio (HR) 2.27 (95% confidence interval (CI) 1.48 to 3.50)), an effect that was magnified when a thicker tibial insert was used (HR 8.10 (95% CI 4.41 to 14.89)). Surgeons should be cautious when choosing high flexion TKRs, particularly when thicker tibial liners might be required. Cite this article: Bone Joint J 2014;96-B:217–23


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1479 - 1486
1 Nov 2011
Park J Kim Y

The purpose of this prospective, randomised study was to evaluate the clinical and radiological results comparing the identical cemented or cementless NexGen total knee prostheses implanted bilaterally in the same patient. Sequential simultaneous bilateral total knee replacements were performed in 50 patients (100 knees). There were 39 women and 11 men with a mean age of 58.4 years (51 to 67) who received a cemented prosthesis in one knee and a cementless prosthesis in the other. The mean follow-up was 13.6 years (13 to 14). At final review, the mean Knee Society scores (96.2 (82 to 100) versus 97.7 (90 to 100)), the mean Western Ontario and McMaster Universities osteoarthritis index (34.5 (4 to 59) versus 35.6 (5 to 51)), the mean ranges of knee movement (124° (100° to 140°) versus 128° (110° to 140°)), mean patient satisfaction (8.1 (. sd. 1.9) versus 8.3 (. sd. 1.7)), and radiological results were similar in both groups. The rate of survival of the femoral components was 100% in both groups at 14 years. The rate of survival of the cemented tibial component was 100% and 98% in the cementless tibial component. No osteolysis was identified in either group. Our data have shown no advantage of cementless over cemented components in total knee replacement


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 367 - 370
1 Mar 2013
Bini SA Chen Y Khatod M Paxton EW

We evaluated the impact of pre-coating the tibial component with polymethylmethacrylate (PMMA) on implant survival in a cohort of 16 548 primary NexGen total knee replacements (TKRs) in 14 113 patients. In 13 835 TKRs a pre-coated tray was used while in 2713 TKRs the non-pre-coated version of the same tray was used. All the TKRs were performed between 2001 and 2009 and were cemented. TKRs implanted with a pre-coated tibial component had a lower cumulative survival than those with a non-pre-coated tibial component (p = 0.01). After adjusting for diagnosis, age, gender, body mass index, American Society of Anesthesiologists grade, femoral coupling design, surgeon volume and hospital volume, pre-coating was an independent risk factor for all-cause aseptic revision (hazard ratio 2.75, p = 0.006). Revision for aseptic loosening was uncommon for both pre-coated and non-pre-coated trays (rates of 0.12% and 0%, respectively). Pre-coating with PMMA does not appear to be protective of revision for this tibial tray design at short-term follow-up. Cite this article: Bone Joint J 2013;95-B:367–70


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1457 - 1461
1 Nov 2007
Han HS Kang S Yoon KS

We have examined the results obtained with 72 NexGen legacy posterior stabilised-flex fixed total knee replacements in 47 patients implanted by a single surgeon between March 2003 and September 2004. Aseptic loosening of the femoral component was found in 27 (38%) of the replacements at a mean follow-up of 32 months (30 to 48) and 15 knees (21%) required revision at a mean of 23 months (11 to 45). We compared the radiologically-loose and revised knees with those which had remained well-fixed to identify the factors which had contributed to this high rate of aseptic loosening. Post-operatively, the mean maximum flexion was 136° (110° to 140°) in the loosened group and 125° (95° to 140°) in the well-fixed group (independent t-test, p = 0.022). Squatting, kneeling, or sitting cross-legged could be achieved by 23 (85%) of the loosened knees, but only 22 (49%) of the well-fixed knees (chi-squared test, p = 0.001). The loosened femoral components were found to migrate into a more flexed position, but no migration was detected in the well-fixed group. These implants allowed a high degree of flexion, but showed a marked rate of early loosening of the femoral component, which was associated with weight-bearing in maximum flexion


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1168 - 1176
1 Nov 2023
Yüksel Y Koster LA Kaptein BL Nelissen RGHH den Hollander P

Aims

Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes.

Methods

A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 622 - 634
1 Jun 2023
Simpson CJRW Wright E Ng N Yap NJ Ndou S Scott CEH Clement ND

Aims

This systematic review and meta-analysis aimed to compare the influence of patellar resurfacing following cruciate-retaining (CR) and posterior-stabilized (PS) total knee arthroplasty (TKA) on the incidence of anterior knee pain, knee-specific patient-reported outcome measures, complication rates, and reoperation rates.

Methods

A systematic review of MEDLINE, PubMed, and Google Scholar was performed to identify randomized controlled trials (RCTs) according to search criteria. Search terms used included: arthroplasty, replacement, knee (Mesh), TKA, prosthesis, patella, patellar resurfacing, and patellar retaining. RCTs that compared patellar resurfacing versus unresurfaced in primary TKA were included for further analysis. Studies were evaluated using the Scottish Intercollegiate Guidelines Network assessment tool for quality and minimization of bias. Data were synthesized and meta-analysis performed.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1238 - 1244
1 Sep 2010
Nicoll D Rowley DI

This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003. The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p < 0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain. Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors


Aims

The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral.

Methods

The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1240 - 1248
1 Nov 2024
Smolle MA Keintzel M Staats K Böhler C Windhager R Koutp A Leithner A Donner S Reiner T Renkawitz T Sava M Hirschmann MT Sadoghi P

Aims

This multicentre retrospective observational study’s aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision.

Methods

A total of 780 patients (39% male; median age 70.7 years (IQR 62.0 to 77.2)) underwent cemented TKA using the Attune Knee System) at five centres, and with the latest radiograph available for the evaluation of RLL at between six and 36 months from surgery. Univariate and multivariate logistic regression models were performed to assess associations between patient and implant-associated factors on the presence of tibial and femoral RLLs. Differences in revision risk depending on RLLs and tibial baseplate design were investigated with the log-rank test.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 148 - 157
1 Feb 2023
Koster LA Rassir R Kaptein BL Sierevelt IN Schager M Nelissen RGHH Nolte PA

Aims

The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs.

Methods

A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1126 - 1131
1 Oct 2022
Hannon CP Kruckeberg BM Pagnano MW Berry DJ Hanssen AD Abdel MP

Aims

We have previously reported the mid-term outcomes of revision total knee arthroplasty (TKA) for flexion instability. At a mean of four years, there were no re-revisions for instability. The aim of this study was to report the implant survivorship and clinical and radiological outcomes of the same cohort of of patients at a mean follow-up of ten years.

Methods

The original publication included 60 revision TKAs in 60 patients which were undertaken between 2000 and 2010. The mean age of the patients at the time of revision TKA was 65 years, and 33 (55%) were female. Since that time, 21 patients died, leaving 39 patients (65%) available for analysis. The cumulative incidence of any re-revision with death as a competing risk was calculated. Knee Society Scores (KSSs) were also recorded, and updated radiographs were reviewed.


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 855 - 863
1 May 2021
Koster LA Meinardi JE Kaptein BL Van der Linden - Van der Zwaag E Nelissen RGHH

Aims

The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS).

Methods

A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims

Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance.

Methods

An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 444 - 451
1 Apr 2022
Laende EK Mills Flemming J Astephen Wilson JL Cantoni E Dunbar MJ

Aims

Thresholds of acceptable early migration of the components in total knee arthroplasty (TKA) have traditionally ignored the effects of patient and implant factors that may influence migration. The aim of this study was to determine which of these factors are associated with overall longitudinal migration of well-fixed tibial components following TKA.

Methods

Radiostereometric analysis (RSA) data over a two-year period were available for 419 successful primary TKAs (267 cemented and 152 uncemented in 257 female and 162 male patients). Longitudinal analysis of data using marginal models was performed to examine the associations of patient factors (age, sex, BMI, smoking status) and implant factors (cemented or uncemented, the size of the implant) with maximum total point motion (MTPM) migration. Analyses were also performed on subgroups based on sex and fixation.


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1561 - 1570
1 Oct 2021
Blyth MJG Banger MS Doonan J Jones BG MacLean AD Rowe PJ

Aims

The aim of this study was to compare the clinical outcomes of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) during the first six weeks and at one year postoperatively.

Methods

A per protocol analysis of 76 patients, 43 of whom underwent TKA and 34 of whom underwent bi-UKA, was performed from a prospective, single-centre, randomized controlled trial. Diaries kept by the patients recorded pain, function, and the use of analgesics daily throughout the first week and weekly between the second and sixth weeks. Patient-reported outcome measures (PROMs) were compared preoperatively, and at three months and one year postoperatively. Data were also compared longitudinally and a subgroup analysis was conducted, stratified by preoperative PROM status.


Aims

The aim of this study was to compare any differences in the primary outcome (biphasic flexion knee moment during gait) of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) at one year post-surgery.

Methods

A total of 76 patients (34 bi-UKA and 42 TKA patients) were analyzed in a prospective, single-centre, randomized controlled trial. Flat ground shod gait analysis was performed preoperatively and one year postoperatively. Knee flexion moment was calculated from motion capture markers and force plates. The same setup determined proprioception outcomes during a joint position sense test and one-leg standing. Surgery allocation, surgeon, and secondary outcomes were analyzed for prediction of the primary outcome from a binary regression model.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 51 - 58
1 Jun 2021
Yang J Heckmann ND Nahhas CR Salzano MB Ruzich GP Jacobs JJ Paprosky WG Rosenberg AG Nam D

Aims

Recent total knee arthroplasty (TKA) designs have featured more anatomical morphologies and shorter tibial keels. However, several reports have raised concerns about the impact of these modifications on implant longevity. The aim of this study was to report the early performance of a modern, cemented TKA design.

Methods

All patients who received a primary, cemented TKA between 2012 and 2017 with a minimum two-year follow-up were included. The implant investigated features an asymmetrical tibial baseplate and shortened keel. Patient demographic details, Knee Society Scores (KSS), component alignment, and the presence of radiolucent lines at final follow-up were recorded. Kaplan-Meier analyses were performed to estimate survivorship.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 137 - 144
1 Jun 2021
Lachiewicz PF Steele JR Wellman SS

Aims

To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period.

Methods

This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1025 - 1032
1 Aug 2020
Hampton M Mansoor J Getty J Sutton PM

Aims

Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design.

Methods

Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 98 - 104
1 Jan 2021
van Ooij B Sierevelt IN van der Vis HM Hoornenborg D Haverkamp D

Aims

For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using radiostereometric analysis (RSA) techniques have shown to be highly predictive in detecting late occurring aseptic loosening at an early stage. To investigate the difference in predicted long-term survival between cemented, uncemented, and hybrid fixation of TKA, we performed a randomized controlled trial using RSA.

Methods

A total of 105 patients were randomized into three groups (cemented, uncemented, and hybrid fixation of the ACS Mobile Bearing (ACS MB) knee system, implantcast). RSA examinations were performed on the first day after surgery and at scheduled follow-up visits at three months, six months, one year, and two years postoperatively. Patient-reported outcome measures (PROMs) were obtained preoperatively and after two years follow-up. Patients and follow-up investigators were blinded for the result of randomization.


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 960 - 969
1 Aug 2019
Odgaard A Laursen MB Gromov K Troelsen A Kristensen PW Schrøder H Madsen F Overgaard S

Aims

The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.

Materials and Methods

Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1511 - 1518
1 Nov 2020
Banger MS Johnston WD Razii N Doonan J Rowe PJ Jones BG MacLean AD Blyth MJG

Aims

The aim of this study was to compare robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) in order to determine the changes in the anatomy of the knee and alignment of the lower limb following surgery.

Methods

An analysis of 38 patients who underwent TKA and 32 who underwent bi-UKA was performed as a secondary study from a prospective, single-centre, randomized controlled trial. CT imaging was used to measure coronal, sagittal, and axial alignment of the knee preoperatively and at three months postoperatively to determine changes in anatomy that had occurred as a result of the surgery. The hip-knee-ankle angle (HKAA) was also measured to identify any differences between the two groups.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 99 - 104
1 Jul 2020
Shah RF Bini S Vail T

Aims

Natural Language Processing (NLP) offers an automated method to extract data from unstructured free text fields for arthroplasty registry participation. Our objective was to investigate how accurately NLP can be used to extract structured clinical data from unstructured clinical notes when compared with manual data extraction.

Methods

A group of 1,000 randomly selected clinical and hospital notes from eight different surgeons were collected for patients undergoing primary arthroplasty between 2012 and 2018. In all, 19 preoperative, 17 operative, and two postoperative variables of interest were manually extracted from these notes. A NLP algorithm was created to automatically extract these variables from a training sample of these notes, and the algorithm was tested on a random test sample of notes. Performance of the NLP algorithm was measured in Statistical Analysis System (SAS) by calculating the accuracy of the variables collected, the ability of the algorithm to collect the correct information when it was indeed in the note (sensitivity), and the ability of the algorithm to not collect a certain data element when it was not in the note (specificity).


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 85 - 90
1 Jun 2020
Blevins JL Rao V Chiu Y Lyman S Westrich GH

Aims

The purpose of this investigation was to determine the relationship between height, weight, and sex with implant size in total knee arthroplasty (TKA) using a multivariate linear regression model and a Bayesian model.

Methods

A retrospective review of an institutional registry was performed of primary TKAs performed between January 2005 and December 2016. Patient demographics including patient age, sex, height, weight, and body mass index (BMI) were obtained from registry and medical record review. In total, 8,100 primary TKAs were included. The mean age was 67.3 years (SD 9.5) with a mean BMI of 30.4 kg/m2 (SD 6.3). The TKAs were randomly split into a training cohort (n = 4,022) and a testing cohort (n = 4,078). A multivariate linear regression model was created on the training cohort and then applied to the testing cohort . A Bayesian model was created based on the frequencies of implant sizes in the training cohort. The model was then applied to the testing cohort to determine the accuracy of the model at 1%, 5%, and 10% tolerance of inaccuracy.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 426 - 433
1 Apr 2020
Boettner F Sculco P Faschingbauer M Rueckl K Windhager R Kasparek MF

Aims

To compare patients undergoing total knee arthroplasty (TKA) with ≤ 80° range of movement (ROM) operated with a 2 mm increase in the flexion gap with matched non-stiff patients with at least 100° of preoperative ROM and balanced flexion and extension gaps.

Methods

In a retrospective cohort study, 98 TKAs (91 patients) with a preoperative ROM of ≤ 80° were examined. Mean follow-up time was 53 months (24 to 112). All TKAs in stiff knees were performed with a 2 mm increased flexion gap. Data were compared to a matched control group of 98 TKAs (86 patients) with a mean follow-up of 43 months (24 to 89). Knees in the control group had a preoperative ROM of at least 100° and balanced flexion and extension gaps. In all stiff and non-stiff knees posterior stabilized (PS) TKAs with patellar resurfacing in combination with adequate soft tissue balancing were used.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1331 - 1340
3 Oct 2020
Attard V Li CY Self A Mann DA Borthwick LA O’Connor P Deehan DJ Kalson NS

Aims

Stiffness is a common complication after total knee arthroplasty (TKA). Pathogenesis is not understood, treatment options are limited, and diagnosis is challenging. The aim of this study was to investigate if MRI can be used to visualize intra-articular scarring in patients with stiff, painful knee arthroplasties.

Methods

Well-functioning primary TKAs (n = 11), failed non-fibrotic TKAs (n = 5), and patients with a clinical diagnosis of fibrosis1 (n = 8) underwent an MRI scan with advanced metal suppression (Slice Encoding for Metal Artefact Correction, SEMAC) with gadolinium contrast. Fibrotic tissue (low intensity on T1 and T2, low-moderate post-contrast enhancement) was quantified (presence and tissue thickness) in six compartments: supra/infrapatella, medial/lateral gutters, and posterior medial/lateral.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1087 - 1092
1 Sep 2019
Garceau S Warschawski Y Dahduli O Alshaygy I Wolfstadt J Backstein D

Aims

The aim of this study was to assess the effects of transferring patients to a specialized arthroplasty centre between the first and second stages (interstage) of prosthetic joint infection (PJI) of the knee.

Patients and Methods

A search of our institutional database was performed to identify patients having undergone two-stage revision total knee arthroplasty (TKA) for PJI. Two cohorts were created: continuous care (CC) and transferred care (TC). Baseline characteristics and outcomes were collected and compared between cohorts.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 77 - 83
1 Jul 2019
James EW Blevins JL Gausden EB Turcan S Denova TA Satalich JR Ranawat AS Warren RF Ranawat AS

Aims

Anterior cruciate ligament (ACL) and multiligament knee (MLK) injuries increase the risk of development of knee osteoarthritis and eventual need for total knee arthroplasty (TKA). There are limited data regarding implant use and outcomes in these patients. The aim of this study was to compare the use of constrained implants and outcomes among patients undergoing TKA with a history of prior knee ligament reconstruction (PKLR) versus a matched cohort of patients undergoing TKA with no history of PKLR.

Patients and Methods

Patients with a history of ACL or MLK reconstruction who underwent TKA between 2007 and 2017 were identified in a single-institution registry. There were 223 patients who met inclusion criteria (188 ACL reconstruction patients, 35 MLK reconstruction patients). A matched cohort, also of 223 patients, was identified based on patient age, body mass index (BMI), sex, and year of surgery. There were 144 male patients and 79 female patients in both cohorts. Mean age at the time of TKA was 57.2 years (31 to 88). Mean BMI was 29.7 kg/m2 (19.5 to 55.7).


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 310 - 318
1 Mar 2020
Joseph MN Achten J Parsons NR Costa ML

Aims

A pragmatic, single-centre, double-blind randomized clinical trial was conducted in a NHS teaching hospital to evaluate whether there is a difference in functional knee scores, quality-of-life outcome assessments, and complications at one-year after intervention between total knee arthroplasty (TKA) and patellofemoral arthroplasty (PFA) in patients with severe isolated patellofemoral arthritis.

Methods

This parallel, two-arm, superiority trial was powered at 80%, and involved 64 patients with severe isolated patellofemoral arthritis. The primary outcome measure was the functional section of the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score at 12 months. Secondary outcomes were the full 24-item WOMAC, Oxford Knee Score (OKS), American Knee Society Score (AKSS), EuroQol five dimension (EQ-5D) quality-of-life score, the University of California, Los Angeles (UCLA) Physical Activity Rating Scale, and complication rates collected at three, six, and 12 months. For longer-term follow-up, OKS, EQ-5D, and self-reported satisfaction score were collected at 24 and 60 months.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1356 - 1361
1 Nov 2019
Chalmers BP Mehrotra KG Sierra RJ Pagnano MW Taunton MJ Abdel MP

Aims

Knee osteonecrosis in advanced stages may lead to joint degeneration. Total knee arthroplasty (TKA) for osteonecrosis has traditionally been associated with suboptimal results. We analyzed outcomes of contemporary TKAs for osteonecrosis, with particular emphasis on: survivorship free from aseptic loosening, any revision, and any reoperation plus the clinical outcomes, complications, and radiological results.

Patients and Methods

In total, 156 patients undergoing 167 primary TKAs performed for osteonecrosis between 2004 and 2014 at a single institution were reviewed. The mean age at index TKA was 61 years (14 to 93) and the mean body mass index (BMI) was 30 kg/m2 (18 to 51) The mean follow-up was six years (2 to 12). A total of 110 TKAs (66%) were performed for primary osteonecrosis and 57 TKAs (34%) for secondary osteonecrosis. Overall, 15 TKAs (9%) had tibial stems, while 12 TKAs (7%) had femoral stems. Posterior-stabilized designs were used in 147 TKAs (88%) of TKAs. Bivariate Cox regression analysis was conducted to identify risk factors for revision and reoperation.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 55 - 60
1 Jul 2019
Laende EK Richardson CG Dunbar MJ

Aims

Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for cementless fixation, which may offer long-term advantages once osseointegration has occurred. The objective of this study was to compare the long-term migration with one- and two-year migration to evaluate the predictive ability of short-term migration data and to compare migration and inducible displacement between cemented and cementless (porous metal monoblock) components at least ten years postoperatively.

Patients and Methods

Patients who had participated in RSA migration studies with two-year follow-up were recruited to return for a long-term follow-up, at least ten years from surgery. Two cemented tibial designs from two manufacturers and one porous metal monoblock cementless tibial design were studied. At the long-term follow-up, patients had supine RSA examinations to determine migration and loaded examinations (single leg stance) to determine inducible displacement. In total, 79 patients (54 female) returned, with mean time since surgery of 12 years (10 to 14). There were 58 cemented and 21 cementless tibial components.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 925 - 929
1 Jul 2018
Abdel MP Tibbo ME Stuart MJ Trousdale RT Hanssen AD Pagnano MW

Aims

It has been suggested that mobile-bearing total knee arthroplasty (TKA) might lead to better outcomes by accommodating some femorotibial rotational mismatch, thereby reducing contact stresses and polyethylene wear. The aim of this study was to determine whether there is a difference between fixed- and mobile-bearing versions of a contemporary TKA with respect to durability, range of movement (ROM) and function, ten years postoperatively.

Patients and Methods

A total of 240 patients who were enrolled in this randomized controlled trial (RCT) underwent a primary cemented TKA with one of three tibial components (all-polyethylene fixed-bearing, modular metal-backed fixed-bearing and mobile-bearing). Patients were reviewed at a median follow-up of ten years (IQR 9.2 to 10.4).


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 996 - 1002
1 Aug 2017
Brown TS Van Citters DW Berry DJ Abdel MP

Advances in polyethylene (PE) in total hip arthroplasty have led to interest and increased use of highly crosslinked PE (HXLPE) in total knee arthroplasty (TKA). Biomechanical data suggest improved wear characteristics for HXLPE inserts over conventional PE in TKA. Short-term results from registry data and few clinical trials are promising. Our aim is to present a review of the history of HXLPEs, the use of HXLPE inserts in TKA, concerns regarding potential mechanical complications, and a thorough review of the available biomechanical and clinical data.

Cite this article: Bone Joint J 2017;99-B:996–1002.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 33 - 39
1 Jul 2019
Lachiewicz PF O’Dell JA

Aims

There is insufficient evidence to recommend the use of alternative polyethylene bearings in modular, fixed-bearing total knee arthroplasty (TKA). The purpose of this study was to compare standard polyethylene (SP) and highly crosslinked polyethylene (XLP) tibial liners in posterior-stabilized TKA, with osteolysis as the primary outcome and clinical results and the rate of re-operation as the secondary outcomes.

Patients and Methods

This is a single-surgeon, prospective randomized study involving one design of modular posterior-stabilized TKA. An analysis of 122 TKAs with an SP compression moulded liner and 123 with an XLP liner was performed, with a mean follow-up of six years (2 to 11). Patients were evaluated clinically using the Knee Society score, Lower Extremity Activity Score (LEAS), and the presence of an effusion, and standard radiographs were assessed for radiolucent lines and osteolytic lesions.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1330 - 1335
1 Oct 2018
Ponzio DY Weitzler L deMeireles A Esposito CI Wright TM Padgett DE

Aims

The aim of this study was to evaluate the surface damage, the density of crosslinking, and oxidation in retrieved antioxidant-stabilized highly crosslinked polyethylene (A-XLPE) tibial inserts from total knee arthroplasty (TKA), and to compare the results with a matched cohort of standard remelted highly crosslinked polyethylene (XLPE) inserts.

Materials and Methods

A total of 19 A-XLPE tibial inserts were retrieved during revision TKA and matched to 18 retrieved XLPE inserts according to the demographics of the patients, with a mean length of implantation of 15 months (1 to 42). The percentage areas of PE damage on the articular surfaces and the modes of damage were measured. The density of crosslinking of the PE and oxidation were measured at loaded and unloaded regions on these surfaces.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1209 - 1216
1 Sep 2013
van der Voort P Pijls BG Nouta KA Valstar ER Jacobs WCH Nelissen RGHH

Mobile-bearing (MB) total knee replacement (TKR) was introduced to reduce the risk of aseptic loosening and wear of polyethylene inserts. However, no consistent clinical advantages of mobile- over fixed-bearing (FB) TKR have been found. In this study we evaluated whether mobile bearings have an advantage over fixed bearings with regard to revision rates and clinical outcome scores. Furthermore, we determined which modifying variables affected the outcome.

A systematic search of the literature was conducted to collect clinical trials comparing MB and FB in primary TKR. The primary outcomes were revision rates for any reason, aseptic loosening and wear. Secondary outcomes included range of movement, Knee Society score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score and radiological parameters. Meta-regression techniques were used to explore factors modifying the observed effect.

Our search yielded 1827 publications, of which 41 studies met our inclusion criteria, comprising over 6000 TKRs. Meta-analyses showed no clinically relevant differences in terms of revision rates, clinical outcome scores or patient-reported outcome measures between MB and FB TKRs. It appears that theoretical assumptions of superiority of MB over FB TKR are not borne out in clinical practice.

Cite this article: Bone Joint J 2013;95-B:1209–16.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1057 - 1063
1 Aug 2013
Zeng Y Shen B Yang J Zhou ZK Kang PD Pei FX

The purpose of this study was to undertake a meta-analysis to determine whether there is lower polyethylene wear and longer survival when using mobile-bearing implants in total knee replacement when compared with fixed-bearing implants. Of 975 papers identified, 34 trials were eligible for data extraction and meta-analysis comprising 4754 patients (6861 knees). We found no statistically significant differences between the two designs in terms of the incidence of radiolucent lines, osteolysis, aseptic loosening or survival. There is thus currently no evidence to suggest that the use of mobile-bearing designs reduce polyethylene wear and prolong survival after total knee replacement.

Cite this article: Bone Joint J 2013;95-B:1057–63.


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 170 - 175
1 Feb 2018
Lam Tin Cheung K Lanting BA McCalden RW Yuan X MacDonald SJ Naudie DD Teeter MG

Aims

The aim of this study was to evaluate the long-term inducible displacement of cemented tibial components ten years after total knee arthroplasty (TKA).

Patients and Methods

A total of 15 patients from a previously reported prospective trial of fixation using radiostereometric analysis (RSA) were examined at a mean of 11 years (10 to 11) postoperatively. Longitudinal supine RSA examinations were acquired at one week, one year, and two years postoperatively and at final follow-up. Weight-bearing RSA examinations were also undertaken with the operated lower limb in neutral and in maximum internal rotation positions. Maximum total point motion (MTPM) was calculated for the longitudinal and inducible displacement examinations (supine versus standing, standing versus internal rotation, and supine versus standing with internal rotation).


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 636 - 642
1 May 2013
Gøthesen Ø Espehaug B Havelin L Petursson G Lygre S Ellison P Hallan G Furnes O

We evaluated the rates of survival and cause of revision of seven different brands of cemented primary total knee replacement (TKR) in the Norwegian Arthroplasty Register during the years 1994 to 2009. Revision for any cause, including resurfacing of the patella, was the primary endpoint. Specific causes of revision were secondary outcomes.

Three posterior cruciate-retaining (PCR) fixed modular-bearing TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing posterior cruciate-sacrificing TKRs were investigated in a total of 17 782 primary TKRs. The median follow-up for the implants ranged from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from 89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative to the fixed modular-bearing Profix knee (the most frequently used TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision for aseptic tibial loosening was higher in the mobile-bearing LCS Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared with the Profix. These implants (except AGC Universal) also had an increased risk of revision for femoral loosening (RR 2.3 (95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95% CI 1.1 to 11.0), respectively). These results suggest that aseptic loosening is related to design in TKR.

Cite this article: Bone Joint J 2013;95-B:636–42.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1355 - 1361
1 Oct 2011
Bollars P Luyckx J Innocenti B Labey L Victor J Bellemans J

High-flexion total knee replacement (TKR) designs have been introduced to improve flexion after TKR. Although the early results of such designs were promising, recent literature has raised concerns about the incidence of early loosening of the femoral component. We compared the minimum force required to cause femoral component loosening for six high-flexion and six conventional TKR designs in a laboratory experiment.

Each TKR design was implanted in a femoral bone model and placed in a loading frame in 135° of flexion. Loosening of the femoral component was induced by moving the tibial component at a constant rate of displacement while maintaining the same angle of flexion. A stereophotogrammetric system registered the relative movement between the femoral component and the underlying bone until loosening occurred.

Compared with high-flexion designs, conventional TKR designs required a significantly higher force before loosening occurred (p < 0.001). High-flexion designs with closed box geometry required significantly higher loosening forces than high-flexion designs with open box geometry (p = 0.0478). The presence of pegs further contributed to the fixation strength of components.

We conclude that high-flexion designs have a greater risk for femoral component loosening than conventional TKR designs. We believe this is attributable to the absence of femoral load sharing between the prosthetic component and the condylar bone during flexion.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 105 - 112
1 Jan 2016
Khan M Osman K Green G Haddad FS

Total knee arthroplasty (TKA) is a cost effective and extremely successful operation. As longevity increases, the demand for primary TKA will continue to rise. The success and survivorship of TKAs are dependent on the demographics of the patient, surgical technique and implant-related factors.

Currently the risk of failure of a TKA requiring revision surgery ten years post-operatively is 5%.

The most common indications for revision include aseptic loosening (29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses considerable clinical burdens on patients and financial burdens on healthcare systems.

We present a current concepts review on the epidemiology of failed TKAs using data from worldwide National Joint Registries.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):105–12.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1058 - 1066
1 Aug 2012
Baker PN Deehan DJ Lees D Jameson S Avery PJ Gregg PJ Reed MR

Patient-reported outcome measures (PROMs) are increasingly being used to assess functional outcome and patient satisfaction. They provide a framework for comparisons between surgical units, and individual surgeons for benchmarking and financial remuneration. Better performance may bring the reward of more customers as patients and commissioners seek out high performers for their elective procedures. Using National Joint Registry (NJR) data linked to PROMs we identified 22 691 primary total knee replacements (TKRs) undertaken for osteoarthritis in England and Wales between August 2008 and February 2011, and identified the surgical factors that influenced the improvements in the Oxford knee score (OKS) and EuroQol-5D (EQ-5D) assessment using multiple regression analysis. After correction for patient factors the only surgical factors that influenced PROMs were implant brand and hospital type (both p < 0.001). However, the effects of surgical factors upon the PROMs were modest compared with patient factors. For both the OKS and the EQ-5D the most important factors influencing the improvement in PROMs were the corresponding pre-operative score and the patient’s general health status. Despite having only a small effect on PROMs, this study has shown that both implant brand and hospital type do influence reported subjective functional scores following TKR. In the current climate of financial austerity, proposed performance-based remuneration and wider patient choice, it would seem unwise to ignore these effects and the influence of a range of additional patient factors.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 787 - 792
1 Jun 2012
Thomsen MG Husted H Bencke J Curtis D Holm G Troelsen A

The purpose of this study was to investigate whether a gender-specific high-flexion posterior-stabilised (PS) total knee replacement (TKR) would offer advantages over a high-flex PS TKR regarding range of movement (ROM), ‘feel’ of the knee, pain and satisfaction, as well as during activity. A total of 24 female patients with bilateral osteoarthritis entered this prospective, blind randomised trial in which they received a high-flex PS TKR in one knee and a gender-specific high-flexion PS TKR in the other knee. At follow-up, patients were assessed clinically measuring ROM, and questioned about pain, satisfaction and daily ‘feel’ of each knee. Patients underwent gait analysis pre-operatively and at one year, which yielded kinematic, kinetic and temporospatial parameters indicative of knee function during gait. At final follow-up we found no statistically significant differences in ROM (p = 0.82). The median pain score was 0 (0 to 8) in both groups (p = 0.95). The median satisfaction score was 9 (4 to 10) in the high-flex group and 8 (0 to 10) in the gender-specific group (p = 0.98). The median ‘feel’ score was 9 (3 to 10) in the high-flex group and 8 (0 to 10) in the gender-specific group (p = 0.66). Gait analysis showed no statistically significant differences between the two prosthetic designs in any kinematic, kinetic or temporospatial parameters.

Both designs produced good clinical results with significant improvements in several gait parameters without evidence of any advantage in the gender-specific design.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 337 - 345
1 Mar 2015
Jaroma AVJ Soininvaara TA Kröger H

Total knee arthroplasty (TKA) is known to lead to a reduction in periprosthetic bone mineral density (BMD). In theory, this may lead to migration, instability and aseptic loosening of the prosthetic components. Bisphosphonates inhibit bone resorption and may reduce this loss in BMD. We hypothesised that treatment with bisphosphonates and calcium would lead to improved BMD and clinical outcomes compared with treatment with calcium supplementation alone following TKA. A total of 26 patients, (nine male and 17 female, mean age 67 years) were prospectively randomised into two study groups: alendronate and calcium (bisphosphonate group, n = 14) or calcium only (control group, n = 12). Dual energy X-ray absorptiometry (DEXA) measurements were performed post-operatively, and at three months, six months, one, two, four, and seven years post-operatively.

Mean femoral metaphyseal BMD was significantly higher in the bisphosphonate group compared with controls, up to four years following surgery in some areas of the femur (p = 0.045). BMD was observed to increase in the lateral tibial metaphysis in the bisphosphonate group until seven years (p = 0.002), and was significantly higher than that observed in the control group throughout (p = 0.024). There were no significant differences between the groups in the central femoral metaphyseal, tibial medial metaphyseal or diaphyseal regions of interest (ROI) of either the femur or tibia.

Bisphosphonate treatment after TKA may be of benefit for patients with poor bone quality. However, further studies with a larger number of patients are necessary to assess whether this is clinically beneficial.

Cite this article: Bone Joint J 2015;97-B:337–45.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1549 - 1552
1 Nov 2005
Malhotra R Bhan S Kiran EK

We present seven patients with recurrent haemarthroses after total knee arthroplasty, caused by an inherent platelet function defect. These patients developed painful knee swelling, persistent bleeding and/or wound breakdown, a platelet factor 3 availability defect being identified in all cases. Surgical exploration, with joint debridement, lavage and synovectomy, was performed in four patients who did not improve with conservative therapy. Histopathological examination of synovium revealed a focal synovial reaction with histiocytic infiltration, and occasional foreign-body giant cells. One patient required an early revision because of aseptic loosening of their tibial component. The condition was treated by single-donor platelet transfusions with good results. The diagnosis, management, and relevance of this disorder are discussed.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 867 - 873
1 Jul 2016
Dalury DF

As the number of younger and more active patients treated with total knee arthroplasty (TKA) continues to increase, consideration of better fixation as a means of improving implant longevity is required. Cemented TKA remains the reference standard with the largest body of evidence and the longest follow-up to support its use. However, cementless TKA, may offer the opportunity of a more bone-sparing procedure with long lasting biological fixation to the bone. We undertook a review of the literature examining advances of cementless TKA and the reported results.

Cite this article: Bone Joint J 2016;98-B:867–73.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 622 - 627
1 May 2016
Donaldson JR Tudor F Gollish J

Aims

The aim of this study was to examine the results of revision total knee arthroplasty (TKA) undertaken for stiffness in the absence of sepsis or loosening.

Patients and Methods

We present the results of revision surgery for stiff TKA in 48 cases (35 (72.9%) women and 13 (27.1%) men). The mean age at revision surgery was 65.5 years (42 to 83). All surgeries were performed by a single surgeon. Stiffness was defined as an arc of flexion of < 70° or a flexion contracture of > 15°. The changes in the range of movement (ROM) and the Western Ontario and McMasters Osteoarthritis index scores (WOMAC) were recorded.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1054 - 1059
1 Aug 2011
van Jonbergen HPW Scholtes VAB van Kampen A Poolman RW

The efficacy of circumpatellar electrocautery in reducing the incidence of post-operative anterior knee pain is unknown. We conducted a single-centre, outcome-assessor and patient-blinded, parallel-group, randomised, controlled trial to compare circumpatellar electrocautery with no electrocautery in total knee replacement in the absence of patellar resurfacing. Patients requiring knee replacement for primary osteoarthritis were randomly assigned circumpatellar electrocautery (intervention group) or no electrocautery (control group). The primary outcome measure was the incidence of anterior knee pain. A secondary measure was the standardised clinical and patient-reported outcomes determined by the American Knee Society scores and the Western Ontario and McMaster Universities (WOMAC) osteoarthritis index. A total of 131 knees received circumpatellar electrocautery and 131 had no electrocautery.

The overall incidence of anterior knee pain at follow-up at one year was 26% (20% to 31%), with 19% (12% to 26%) in the intervention group and 32% (24% to 40%) in the control group (p = 0.02). The relative risk reduction from electrocautery was 40% (9% to 61%) and the number needed to treat was 7.7 (4.3 to 41.4). The intervention group had a better mean total WOMAC score at follow-up at one year compared with the control group (16.3 (0 to 77.7) versus 21.6 (0 to 76.7), p = 0.04). The mean post-operative American Knee Society knee scores and function scores were similar in the intervention and control groups (knee score: 92.4 (55 to 100) versus 90.4 (51 to 100), respectively (p = 0.14); function score: 86.5 (15 to 100) versus 84.5 (30 to 100), respectively (p = 0.49)).

Our study suggests that in the absence of patellar resurfacing electrocautery around the margin of the patella improves the outcome of total knee replacement.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 493 - 497
1 Apr 2013
Hino K Ishimaru M Iseki Y Watanabe S Onishi Y Miura H

There are several methods for evaluating stability of the joint during total knee replacement (TKR). Activities of daily living demand mechanical loading to the knee joint, not only in full extension, but also in mid-flexion. The purpose of this study was to compare the varus-valgus stability throughout flexion in knees treated with either cruciate-retaining or posterior-stabilised TKR, using an intra-operative navigation technique. A total of 34 knees underwent TKR with computer navigation, during which the investigator applied a maximum varus-valgus stress to the knee while steadily moving the leg from full extension to flexion both before and after prosthetic implantation. The femorotibial angle was measured simultaneously by the navigation system at every 10° throughout the range of movement. It was found that posterior-stabilised knees had more varus-valgus laxity than cruciate-retaining knees at all angles examined, and the differences were statistically significant at 10° (p = 0.0093), 20° (p = 0.0098) and 30° of flexion (p = 0.0252).

Cite this article: Bone Joint J 2013;95-B:493–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1311 - 1316
1 Oct 2008
Kim Y Kim J

The purpose of this study was to determine objectively the outcome of total knee replacement in patients with ankylosed knees.

There were 82 patients (99 knees) with ankylosed knees who underwent total knee replacement with a condylar constrained or a posterior stabilised prosthesis. Their mean age was 41.9 years (23 to 60) and the mean follow-up was for 8.9 years (6.6 to 14). Pre- and post-operative data included the Hospital for Special Surgery (HSS), the Knee Society (KS) and the Western Ontario and McMaster University Osteoarthritis index (WOMAC) scores.

The mean HSS, KS and WOMAC scores improved from 60, 53, and 79 pre-operatively to 81, 85, and 37 at follow-up. These improvements were statistically significant (p = 0.018, 0.001 and 0.014 respectively). The mean physical, social and emotional WOMAC scores also improved significantly (p = 0.032, p = 0.023 and p < 0.001 respectively). The mean satisfaction score was 8.5 (sd 1.5).

Total knee replacement gives good mid-term results in patients with ankylosed knees.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 357 - 363
1 Mar 2011
Gillespie RJ Levine A Fitzgerald SJ Kolaczko J DeMaio M Marcus RE Cooperman DR

Recently, gender-specific designs of total knee replacement have been developed to accommodate anatomical differences between males and females. We examined a group of male and female distal femora matched for age and height, to determine if there was a difference in the aspect ratio (mediolateral distance versus anteroposterior distance) and the height of the anterior flange between the genders. The Hamann-Todd Collection provided 1207 skeletally mature cadaver femora. The femoral length, the anteroposterior height, height of the lateral and medial flanges and the mediolateral width were measured in all the specimens. The mechanical axis of the femur, the cut articular width and the aspect ratio were assessed. Statistical analysis of the effect of gender upon the aspect ratio and the lateral and medial flanges was undertaken, controlling for age, height and race.

The mean aspect ratio of male femora was 1.21 (sd 0.07) and of female femora it was 1.16 (sd 0.06) (p < 0.001). There was no significant difference between male and female specimens in the mean size of the lateral flange (6.57 mm (sd 2.57) and 7.02 mm (sd 2.36), respectively; p = 0.099) or of the medial flange (3.03 mm (sd 2.47) and 3.56 mm (sd 2.32), respectively; p = 0.67). Future work in the design of knee prostheses should take into account the overall variability of the anatomy of the distal femur.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1644 - 1648
1 Dec 2014
Abdel MP Pulido L Severson EP Hanssen AD

Instability in flexion after total knee replacement (TKR) typically occurs as a result of mismatched flexion and extension gaps. The goals of this study were to identify factors leading to instability in flexion, the degree of correction, determined radiologically, required at revision surgery, and the subsequent clinical outcomes. Between 2000 and 2010, 60 TKRs in 60 patients underwent revision for instability in flexion associated with well-fixed components. There were 33 women (55%) and 27 men (45%); their mean age was 65 years (43 to 82). Radiological measurements and the Knee Society score (KSS) were used to assess outcome after revision surgery. The mean follow-up was 3.6 years (2 to 9.8). Decreased condylar offset (p < 0.001), distalisation of the joint line (p < 0.001) and increased posterior tibial slope (p < 0.001) contributed to instability in flexion and required correction at revision to regain stability. The combined mean correction of posterior condylar offset and joint line resection was 9.5 mm, and a mean of 5° of posterior tibial slope was removed. At the most recent follow-up, there was a significant improvement in the mean KSS for the knee and function (both p < 0.001), no patient reported instability and no patient underwent further surgery for instability.

The following step-wise approach is recommended: reduction of tibial slope, correction of malalignment, and improvement of condylar offset. Additional joint line elevation is needed if the above steps do not equalise the flexion and extension gaps.

Cite this article: Bone Joint J 2014;96-B:1644–8.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 902 - 906
1 Jul 2014
Chareancholvanich K Pornrattanamaneewong C

We have compared the time to recovery of isokinetic quadriceps strength after total knee replacement (TKR) using three different lengths of incision in the quadriceps. We prospectively randomised 60 patients into one of the three groups according to the length of incision in the quadriceps above the upper border of the patella (2 cm, 4 cm or 6 cm). The strength of the knees was measured pre-operatively and every month post-operatively until the peak quadriceps torque returned to its pre-operative level.

There was no significant difference in the mean operating time, blood loss, hospital stay, alignment or pre-operative isokinetic quadriceps strength between the three groups. Using the Kaplan–Meier method, group A had a similar mean recovery time to group B (2.0 ± 0.2 vs 2.5 ± 0.2 months, p = 0.176). Group C required a significantly longer recovery time (3.4 ± 0.3 months) than the other groups (p < 0.03). However, there were no significant differences in the mean Oxford knee scores one year post-operatively between the groups.

We conclude that an incision of up to 4 cm in the quadriceps does not delay the recovery of its isokinetic strength after TKR.

Cite this article: Bone Joint J 2014;96-B:902–6.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 473 - 478
1 Apr 2014
van Jonbergen HPW Scholtes VAB Poolman RW

In the absence of patellar resurfacing, we have previously shown that the use of electrocautery around the margin of the patella improved the one-year clinical outcome of total knee replacement (TKR). In this prospective randomised study we compared the mean 3.7 year (1.1 to 4.2) clinical outcomes of 300 TKRs performed with and without electrocautery of the patellar rim: this is an update of a previous report. The overall prevalence of anterior knee pain was 32% (95% confidence intervals [CI] 26 to 39), and 26% (95% CI 18 to 35) in the intervention group compared with 38% (95% CI 29 to 48) in the control group (chi-squared test; p = 0.06). The overall prevalence of anterior knee pain remained unchanged between the one-year and 3.7 year follow-up (chi-squared test; p = 0.12). The mean total Western Ontario McMasters Universities Osteoarthritis Indices and the American Knee Society knee and function scores at 3.7 years’ follow-up were similar in the intervention and control groups (repeated measures analysis of variance p = 0.43, p = 0.09 and p = 0.59, respectively). There were no complications. A total of ten patients (intervention group three, control group seven) required secondary patellar resurfacing after the first year.

Our study suggests that the improved clinical outcome with electrocautery denervation compared with no electrocautery is not maintained at a mean of 3.7 years’ follow-up.

Cite this article: Bone Joint J 2014;96-B:473–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 463 - 468
1 Apr 2009
Fukunaga K Kobayashi A Minoda Y Iwaki H Hashimoto Y Takaoka K

The patellar clunk syndrome describes painful catching, grinding or jumping of the patella when the knee moves from a flexed to an extended position after total knee replacement (TKR). The posterior stabilised TKR had been noted to have a higher incidence of this problem. Mobile-bearing posteriorly stabilised TKRs have been introduced to improve patellar tracking and related problems by a mechanism of self-alignment. We evaluated the patellar clunk syndrome in 113 knees in 93 patients with such a TKR at a mean follow-up of 2.3 years (2.0 to 3.2). The syndrome was identified in 15 knees (13.3%).

Logistic regression analysis showed that the absolute value of the post-operative angle of patellar tilt was significantly associated with the occurrence of patellar clunk (p = 0.025). Patellar tracking should be carefully checked during surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 646 - 655
1 May 2005
Victor J Banks S Bellemans J

We performed a prospective, randomised trial of 44 patients to compare the functional outcomes of a posterior-cruciate-ligament-retaining and posterior-cruciate-ligament-substituting total knee arthroplasty, and to gain a better understanding of the in vivo kinematic behaviour of both devices.

At follow-up at five years, no statistically significant differences were found in the clinical outcome measurements for either design. The prevalence of radiolucent lines and the survivorship were the same. In a subgroup of 15 knees, additional image-intensifier analysis in the horizontal and sagittal planes was performed during step-up and lunge activity. Our analysis revealed striking differences. Lunge activity showed a mean posterior displacement of both medial and lateral tibiofemoral contact areas (roll-back) which was greater and more consistent in the cruciate-substituting than in the cruciate-retaining group (medial p < 0.0001, lateral p = 0.011). The amount of posterior displacement could predict the maximum flexion which could be achieved (p = 0.018). Forward displacement of the tibiofemoral contact area in flexion during stair activity was seen more in the cruciate-retaining than in the cruciate-substituting group. This was attributed mainly to insufficiency of the posterior cruciate ligament and partially to that of the anterior cruciate ligament. We concluded that, despite similar clinical outcomes, there are significant kinematic differences between cruciate-retaining and cruciate-substituting arthroplasties.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 64 - 68
1 Jan 2009
Kim Y Choi Y Kim J

We wished to determine whether simultaneous bilateral sequential total knee replacement (TKR) carried increased rates of mortality and complications compared with unilateral TKR in low- and high-risk patients.

Our study included 2385 patients who had undergone bilateral sequential TKR under one anaesthetic and 719 who had unilateral TKR. There were no significant pre-operative differences between the groups in terms of age, gender, height, weight, body mass index, diagnosis, comorbidity and duration of follow-up, which was a mean of 10.2 years (5 to 14) in the bilateral and 10.4 years (5 to 14) in the unilateral group.

The peri-operative mortality rate (eight patients, 0.3%) of patients who had bilateral sequential TKR was similar to that (five patients, 0.7%) of those undergoing unilateral TKR. In bilateral cases the peri-operative mortality rate (three patients, 0.4%) of patients at high risk was similar to that (five patients, 0.3%) of patients at low risk as it was also in unilateral cases (two patients, 1.0% vs three patients, 0.6%). There was no significant difference (p = 0.735) in either the overall number of major complications between bilateral and unilateral cases or between low- (p = 0.57) and high-risk (p = 0.61) patients. Also, the overall number of minor complications was not significantly different between the bilateral and unilateral group (p = 0.143).

Simultaneous bilateral sequential TKR can be offered to patients at low and high risk and has an expected rate of complications similar to that of unilateral TKR.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1354 - 1358
1 Oct 2013
Singh G Tan JH Sng BY Awiszus F Lohmann CH Nathan SS

The optimal management of the tibial slope in achieving a high flexion angle in posterior-stabilised (PS) total knee replacement (TKR) is not well understood, and most studies evaluating the posterior tibial slope have been conducted on cruciate-retaining TKRs. We analysed pre- and post-operative tibial slope differences, pre- and post-operative coronal knee alignment and post-operative maximum flexion angle in 167 patients undergoing 209 TKRs. The mean pre-operative posterior tibial slope was 8.6° (1.3° to 17°) and post-operatively it was 8.0° (0.1° to 16.7°). Multiple linear regression analysis showed that the absolute difference between pre- and post-operative tibial slope (p < 0.001), post-operative coronal alignment (p = 0.02) and pre-operative range of movement (p < 0.001) predicted post-operative flexion. The variance of change in tibial slope became larger as the post-operative maximum flexion angle decreased. The odds ratio of having a post-operative flexion angle < 100° was 17.6 if the slope change was > 2°. Our data suggest that recreation of the anatomical tibial slope appears to improve maximum flexion after posterior-stabilised TKR, provided coronal alignment has been restored.

Cite this article: Bone Joint J 2013;95-B:1354–8.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1348 - 1353
1 Oct 2013
Valenzuela GA Jacobson NA Buzas D Korecki TD Valenzuela RG Teitge RA

The outcome of high tibial osteotomy (HTO) deteriorates with time, and additional procedures may be required. The aim of this study was to compare the clinical and radiological outcomes between unicompartmental knee replacement (UKR) and total knee replacement (TKR) after HTO as well as after primary UKR. A total of 63 patients (63 knees) were studied retrospectively and divided into three groups: UKR after HTO (group A; n = 22), TKR after HTO (group B; n = 18) and primary UKR (group C; n = 22). The Oxford knee score (OKS), Knee Society score (KSS), hip–knee–ankle angles, mechanical axis and patellar height were evaluated pre- and post-operatively. At a mean of 64 months (19 to 180) post-operatively the mean OKS was 43.8 (33 to 49), 43.3 (30 to 48) and 42.5 (29 to 48) for groups A, B and C, respectively (p = 0.73). The mean KSS knee score was 88.8 (54 to 100), 88.11 (51 to 100) and 85.3 (45 to 100) for groups A, B and C, respectively (p = 0.65), and the mean KSS function score was 85.0 (50 to 100) in group A, 85.8 (20 to 100) in group B and 79.3 (50 to 100) in group C (p = 0.48). Radiologically the results were comparable for all groups except for patellar height, with a higher incidence of patella infra following a previous HTO (p = 0.02).

Cite this article: Bone Joint J 2013;95-B:1348–53.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1025 - 1031
1 Aug 2008
Mizu-uchi H Matsuda S Miura H Okazaki K Akasaki Y Iwamoto Y

We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans.

The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group.

The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group.

The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group.

The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1317 - 1323
1 Oct 2007
Kim Y Yoon S Kim J

We compared the results of 146 patients who received an anatomic modular knee fixed-bearing total knee replacement (TKR) in one knee and a low contact stress rotating platform mobile-bearing TKR in the other. There were 138 women and eight men with a mean age of 69.8 years (42 to 80). The mean follow-up was 13.2 years (11.0 to 14.5). The patients were assessed clinically and radiologically using the rating systems of the Hospital for Special Surgery and the Knee Society at three months, six months, one year, and annually thereafter.

The assessment scores of both rating systems pre-operatively and at the final review did not show any statistically significant differences between the two designs of implant. In the anatomic modular knee group, one knee was revised because of aseptic loosening of the tibial component and one because of infection. In addition, three knees were revised because of wear of the polyethylene tibial bearing. In the low contact stress group, two knees were revised because of instability requiring exchange of the polyethylene insert and one because of infection.

The radiological analysis found no statistical difference in the incidence of radiolucent lines at the final review (Student’s t-test, p = 0.08), most of which occurred at tibial zone 1. The Kaplan-Meier survivorship for aseptic loosening of the anatomic modular knee and the low contact stress implants at 14.5 years was 99% and 100%, respectively, with a 95% confidence interval of 94% to 100% for both designs.

We found no evidence of the superiority of one design over the other at long-term follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 904 - 910
1 Jul 2007
Kim Y Kim D Kim J

We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1).

The total knee score, pain score, functional score and range of movement were not statistically different (p > 0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a mobile bearing.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 354 - 359
1 Mar 2013
Chareancholvanich K Narkbunnam R Pornrattanamaneewong C

Patient-specific cutting guides (PSCGs) are designed to improve the accuracy of alignment of total knee replacement (TKR). We compared the accuracy of limb alignment and component positioning after TKR performed using PSCGs or conventional instrumentation. A total of 80 patients were randomised to undergo TKR with either of the different forms of instrumentation, and radiological outcomes and peri-operative factors such as operating time were assessed. No significant difference was observed between the groups in terms of tibiofemoral angle or femoral component alignment. Although the tibial component in the PSCGs group was measurably closer to neutral alignment than in the conventional group, the size of the difference was very small (89.8° (sd 1.2) vs 90.5° (sd 1.6); p = 0.030). This new technology slightly shortened the bone-cutting time by a mean of 3.6 minutes (p < 0.001) and the operating time by a mean 5.1 minutes (p = 0.019), without tangible differences in post-operative blood loss (p = 0.528) or need for blood transfusion (p = 0.789). This study demonstrated that both PSCGs and conventional instrumentation restore limb alignment and place the components with the similar accuracy. The minimal advantages of PSCGs in terms of consistency of alignment or operative time are unlikely to be clinically relevant.

Cite this article: Bone Joint J 2013;95-B:354–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1457 - 1461
1 Nov 2012
Krishnan SP Dawood A Richards R Henckel J Hart AJ

Improvements in the surgical technique of total knee replacement (TKR) are continually being sought. There has recently been interest in three-dimensional (3D) pre-operative planning using magnetic resonance imaging (MRI) and CT. The 3D images are increasingly used for the production of patient-specific models, surgical guides and custom-made implants for TKR.

The users of patient-specific instrumentation (PSI) claim that they allow the optimum balance of technology and conventional surgery by reducing the complexity of conventional alignment and sizing tools. In this way the advantages of accuracy and precision claimed by computer navigation techniques are achieved without the disadvantages of additional intra-operative inventory, new skills or surgical time.

This review describes the terminology used in this area and debates the advantages and disadvantages of PSI.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 638 - 641
1 May 2012
Ha C Na S

We aimed to obtain anthropometric data on Korean knees and to compare these with data on commonly available total knee arthroplasties (TKAs). The dimensions of the femora and tibiae of 1168 knees were measured intra-operatively. The femoral components were found to show a tendency toward mediolateral (ML) under-coverage in small femurs and ML overhang in the large femurs. The ML under-coverage was most prominent for the small prostheses. The ML/anteroposterior (ML/AP) ratio of Korean tibiae was greater than that of tibial components.

This study shows that, for different reasons, current TKAs do not provide a reasonable fit for small or large Korean knees, and that the ‘gender-specific’ and ‘stature-specific’ components help for large Korean femurs but offer less satisfactory fits for small femurs. Specific modifications of prostheses are needed for Asian knees.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1009 - 1015
1 Aug 2012
Scott CEH Biant LC

Stems improve the mechanical stability of tibial components in total knee replacement (TKR), but come at a cost of stress shielding along their length. Their advantages include resistance to shear, reduced tibial lift-off and increased stability by reducing micromotion. Longer stems may have disadvantages including stress shielding along the length of the stem with associated reduction in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic fracture and end-of-stem pain. These features make long stems unattractive in the primary TKR setting, but often desirable in revision surgery with bone loss and instability. In the revision scenario, stems are beneficial in order to convey structural stability to the construct and protect the reconstruction of bony defects. Cemented and uncemented long stemmed implants have different roles depending on the nature of the bone loss involved.

This review discusses the biomechanics of the design of tibial components and stems to inform the selection of the component and the technique of implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1493 - 1496
1 Nov 2011
Lee JK Choi CH

Bone defects are occasionally encountered during primary total knee replacement (TKR) and cause difficulty in establishing a stable well-aligned bone-implant interface. Between March 1999 and November 2005, 59 knees in 43 patients underwent primary TKR with a metal block augmentation for tibial bone deficiency. In all, six patients (eight knees) died less than four years post-operatively, and four patients (five knees) were lost to follow-up leaving 46 knees in 33 patients available for review at a mean of 78.6 months (62 to 129). The clinical results obtained, including range of movement, American Knee Society and Oxford knee scores, and the Western Ontario and McMaster Universities osteoarthritis index, were good to excellent, with no failures. Radiolucent lines at the block-cement-bone interface were noted in five knees (11%) during the first post-operative year, but these did not progress.

Modular rectangular metal augmentation for tibial bone deficiency is a useful option. No deterioration of the block-prosthesis or block-cement-bone interface was seen at minimum of five years follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 908 - 913
1 Jul 2012
Seo JG Moon YW Park SH Lee JH Kang HM Kim SM

Peri-prosthetic patellar fracture following resurfacing as part of total knee replacement (TKR) is an infrequent yet challenging complication. This case-control study was performed to identify clinical, radiological and surgical factors that increase the risk of developing a spontaneous patellar fracture after TKR. Patellar fractures were identified in 74 patients (88 knees) from a series of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding those with a previous history of extensor mechanism realignment or a clear traumatic event, a metal-backed patella, any uncemented component or subsequent infection, the remaining 64 fractures were compared with a matched group of TKRs with an excellent outcome defined by the Knee Society score. The mean age of patients with a fracture was 70 years (51 to 81) at the time of TKR. Patellar fractures were detected at a mean of 13.4 months (2 to 84) after surgery. The incidence of patellar fracture was found to be strongly associated with the number of previous knee operations, greater pre-operative mechanical malalignment, smaller post-operative patellar tendon length, thinner post-resection patellar thickness, and a lower post-operative Insall-Salvati ratio.

An understanding of the risk factors associated with spontaneous patellar fracture following TKR provides a valuable insight into prevention of this challenging complication.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 56 - 61
1 Jan 2012
Kawahara S Matsuda S Fukagawa S Mitsuyasu H Nakahara H Higaki H Shimoto T Iwamoto Y

In posterior stabilised total knee replacement (TKR) a larger femoral component is sometimes selected to manage the increased flexion gap caused by resection of the posterior cruciate ligament. However, concerns remain regarding the adverse effect of the increased anteroposterior dimensions of the femoral component on the patellofemoral (PF) joint. Meanwhile, the gender-specific femoral component has a narrower and thinner anterior flange and is expected to reduce the PF contact force. PF contact forces were measured at 90°, 120°, 130° and 140° of flexion using the NexGen Legacy Posterior Stabilized (LPS)-Flex Fixed Bearing Knee system using Standard, Upsized and Gender femoral components during TKR. Increasing the size of the femoral component significantly increased mean PF forces at 120°, 130° and 140° of flexion (p = 0.005, p < 0.001 and p < 0.001, respectively). No difference was found in contact force between the Gender and the Standard components. Among the patients who had overhang of the Standard component, mean contact forces with the Gender component were slightly lower than those of the Standard component, but no statistical difference was found at 90°, 120°, 130° or 140° of flexion (p = 0.689, 0.615, 0.253 and 0.248, respectively).

Upsized femoral components would increase PF forces in deep knee flexion. Gender-specific implants would not reduce PF forces.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 497 - 503
1 Apr 2012
Bercovy M Beldame J Lefebvre B Duron A

This prospective study compares the outcome of 157 hydroxyapatite (HA)-coated tibial components with 164 cemented components in the ROCC Rotating Platform total knee replacement in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There were two revisions for loosening: one for an HA-coated and one for a cemented tibial component. Radiological evaluation demonstrated no radiolucent lines with the HA-coated femoral components. A total of three HA-coated tibial components exhibited radiolucent lines at three months post-operatively and these disappeared after three further months of protected weight-bearing. With HA-coated components the operating time was shorter (p <  0.006) and the radiological assessment of the tibial interface was more stable (p < 0.01). Using revision for aseptic loosening of the tibial component as the end point, the survival rates at nine years was identical for both groups at 99.1%.

Our results suggest that HA-coated components perform at least as well as the same design with cemented components and compare favourably with those of series describing cemented or porous-coated knee replacements, suggesting that fixation of both components with hydroxyapatite is a reliable option in primary total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 510 - 516
1 Apr 2012
Hickey BA Kempshall PJ Metcalfe AJ Forster MC

As part of the national initiative to reduce waiting times for joint replacement surgery in Wales, the Cardiff and Vale NHS Trust referred 224 patients to the NHS Treatment Centre in Weston-Super-Mare for total knee replacement (TKR). A total of 258 Kinemax TKRs were performed between November 2004 and August 2006. Of these, a total of 199 patients (232 TKRs, 90%) have been followed up for five years. This cohort was compared with 258 consecutive TKRs in 250 patients, performed at Cardiff and Vale Orthopaedic Centre (CAVOC) over a similar time period. The five year cumulative survival rate was 80.6% (95% confidence interval (CI) 74.0 to 86.0) in the Weston-Super-Mare cohort and 95.0% (95% CI 90.2 to 98.2) in the CAVOC cohort with revision for any reason as the endpoint. The relative risk for revision at Weston-Super-Mare compared with CAVOC was 3.88 (p < 0.001). For implants surviving five years, the mean Oxford knee scores (OKS) and mean EuroQol (EQ-5D) scores were similar (OKS: Weston-Super-Mare 29 (2 to 47) vs CAVOC 29.8 (3 to 48), p = 0.61; EQ-5D: Weston-Super-Mare 0.53 (-0.38 to 1.00) vs CAVOC 0.55 (-0.32 to 1.00), p = 0.79). Patients with revised TKRs had significantly lower Oxford knee and EQ-5D scores (p < 0.001).

The results show a higher revision rate for patients operated at Weston-Super-Mare Treatment Centre, with a reduction in functional outcome and quality of life after revision. This further confirms that patients moved from one area to another for joint replacement surgery fare poorly.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1487 - 1492
1 Nov 2011
Macheras GA Kateros K Galanakos SP Koutsostathis SD Kontou E Papadakis SA

We report the long-term results of the management of neglected chronically infected total knee replacements with a two-stage re-implantation protocol. In 18 of 34 patients (53%) a resistant organism was isolated. All cases were treated by the same surgical team in a specialist centre and had a mean follow-up of 12.1 years (10 to 14). They were evaluated clinically and radiologically using the Knee Society Score (KSS) and the American Knee Society Roentgenographic scoring system, respectively. One patient died after eight years from an unrelated cause and two were lost to follow-up. Three patients (8.8%) developed a recurrent infection for which further surgery was required. The infection was eradicated successfully in 31 patients (91.1%). There was one case of aseptic loosening after 13 years. We found a significant improvement in the KSS at final follow-up (p < 0.001).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 648 - 655
1 May 2012
Jolles BM Grzesiak A Eudier A Dejnabadi H Voracek C Pichonnaz C Aminian K Martin E

This study compared the outcome of total knee replacement (TKR) in adult patients with fixed- and mobile-bearing prostheses during the first post-operative year and at five years’ follow-up, using gait parameters as a new objective measure. This double-blind randomised controlled clinical trial included 55 patients with mobile-bearing (n = 26) and fixed-bearing (n = 29) prostheses of the same design, evaluated pre-operatively and post-operatively at six weeks, three months, six months, one year and five years. Each participant undertook two walking trials of 30 m and completed the EuroQol questionnaire, Western Ontario and McMaster Universities osteoarthritis index, Knee Society score, and visual analogue scales for pain and stiffness. Gait analysis was performed using five miniature angular rate sensors mounted on the trunk (sacrum), each thigh and calf. The study population was divided into two groups according to age (≤ 70 years versus > 70 years).

Improvements in most gait parameters at five years’ follow-up were greater for fixed-bearing TKRs in older patients (> 70 years), and greater for mobile-bearing TKRs in younger patients (≤ 70 years). These findings should be confirmed by an extended age controlled study, as the ideal choice of prosthesis might depend on the age of the patient at the time of surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1210 - 1216
1 Sep 2011
Mitsuyasu H Matsuda S Fukagawa S Okazaki K Tashiro Y Kawahara S Nakahara H Iwamoto Y

We investigated whether the extension gap in total knee replacement (TKR) would be changed when the femoral component was inserted. The extension gap was measured with and without the femoral component in place in 80 patients with varus osteoarthritis undergoing posterior-stabilised TKR. The effect of a post-operative increase in the size of the femoral posterior condyles was also evaluated. The results showed that placement of the femoral component significantly reduced the medial and lateral extension gaps by means of 1.0 mm and 0.9 mm, respectively (p < 0.0001). The extension gap was reduced when a larger femoral component was selected relative to the thickness of the resected posterior condyle. When the post-operative posterior lateral condyle was larger than that pre-operatively, 17 of 41 knees (41%) showed a decrease in the extension gap of > 2.0 mm. When a specially made femoral trial component with a posterior condyle enlarged by 4 mm was tested, the medial and lateral extension gaps decreased further by means of 2.1 mm and 2.8 mm, respectively.

If the thickness of the posterior condyle is expected to be larger than that pre-operatively, it should be recognised that the extension gap is likely to be altered. This should be taken into consideration when preparing the extension gap.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 794 - 798
1 Jun 2010
Minoda Y Ikebuchi M Kobayashi A Iwaki H Inori F Nakamura H

Bone mineral density (BMD) around the femoral component has been reported to decrease after total knee replacement (TKR) because of stress shielding. Our aim was to determine whether a cemented mobile-bearing component reduced the post-operative loss of BMD. In our study 28 knees receiving a cemented fixed-bearing TKR were matched with 28 receiving a cemented mobile-bearing TKR. They underwent dual-energy x-ray absorptiometry, pre-operatively and at three weeks and at three, six, 12, 18 and 24 months post-operatively. The patients were not taking medication to improve the BMD.

The pre-operative differences in the BMD of the femoral neck, wrist, lumbar spine and knee in the two groups were not significant. The BMD of the femur decreased postoperatively in the fixed-bearing group, but not the mobile-bearing group. The difference in the post-operative change in the BMD in the two groups was statistically significant (p < 0.05) at 18 and 24 months.

Our findings show that a cemented mobile-bearing TKR has a favourable effect on the BMD of the distal femur after TKR in the short term. Further study is required to determine the long-term effects.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 475 - 480
1 Apr 2009
Matsumoto T Kuroda R Kubo S Muratsu H Mizuno K Kurosaka M

We have developed a new tensor for total knee replacements which is designed to assist with soft-tissue balancing throughout the full range of movement with a reduced patellofemoral joint. Using this tensor in 40 patients with osteoarthritis we compared the intra-operative joint gap in cruciate-retaining and posterior-stabilised total knee replacements at 0°, 10°, 45°, 90° and 135° of flexion, with the patella both everted and reduced.

While the measurement of the joint gap with a reduced patella in posterior-stabilised knees increased from extension to flexion, it remained constant for cruciate-retaining joints throughout a full range of movement. The joint gaps at deep knee flexion were significantly smaller for both types of prosthetic knee when the patellofemoral joint was reduced (p < 0.05).


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 915 - 918
1 Jul 2007
Hanratty BM Thompson NW Wilson RK Beverland DE

We have studied the concept of posterior condylar offset and the importance of its restoration on the maximum range of knee flexion after posterior-cruciate-ligament-retaining total knee replacement (TKR). We measured the difference in the posterior condylar offset before and one year after operation in 69 patients who had undergone a primary cruciate-sacrificing mobile bearing TKR by one surgeon using the same implant and a standardised operating technique. In all the patients true pre- and post-operative lateral radiographs had been taken.

The mean pre- and post-operative posterior condylar offset was 25.9 mm (21 to 35) and 26.9 mm (21 to 34), respectively. The mean difference in posterior condylar offset was + 1 mm (−6 to +5). The mean pre-operative knee flexion was 111° (62° to 146°) and at one year postoperatively, it was 107° (51° to 137°).

There was no statistical correlation between the change in knee flexion and the difference in the posterior condylar offset after TKR (Pearson correlation coefficient r = −0.06, p = 0.69).


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 467 - 470
1 Apr 2007
Kim Y Kim J Kim D

We performed a prospective, randomised study to compare the results and rates of complications of primary total knee replacement performed using a quadriceps-sparing technique or a standard arthrotomy in 120 patients who had bilateral total knee replacements carried out under the same anaesthetic. The clinical results, pain scales, surgical and hospital data, post-operative complications and radiological results were compared.

No significant differences were found between the two groups with respect to the blood loss, knee score, function score, pain scale, range of movement or radiological findings. In contrast, the operating time (p = 0.0001) and the tourniquet time (p < 0.0001) were significantly longer in the quadriceps-sparing group, as was the rate of complications (p = 0.0468).

We therefore recommend the use of a standard arthrotomy with the shortest possible skin incision for total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 316 - 322
1 Mar 2007
Pearse EO Caldwell BF Lockwood RJ Hollard J

We carried out an audit on the result of achieving early walking in total knee replacement after instituting a new rehabilitation protocol, and assessed its influence on the development of deep-vein thrombosis as determined by Doppler ultrasound scanning on the fifth post-operative day. Early mobilisation was defined as beginning to walk less than 24 hours after knee replacement.

Between April 1997 and July 2002, 98 patients underwent a total of 125 total knee replacements. They began walking on the second post-operative day unless there was a medical contraindication. They formed a retrospective control group. A protocol which allowed patients to start walking at less than 24 hours after surgery was instituted in August 2002. Between August 2002 and November 2004, 97 patients underwent a total of 122 total knee replacements. They formed the early mobilisation group, in which data were prospectively gathered. The two groups were of similar age, gender and had similar medical comorbidities. The surgical technique and tourniquet times were similar and the same instrumentation was used in nearly all cases. All the patients received low-molecular-weight heparin thromboprophylaxis and wore compression stockings post-operatively.

In the early mobilisation group 90 patients (92.8%) began walking successfully within 24 hours of their operation. The incidence of deep-vein thrombosis fell from 27.6% in the control group to 1.0% in the early mobilisation group (chi-squared test, p < 0.001). There was a difference in the incidence of risk factors for deep-vein thrombosis between the two groups. However, multiple logistic regression analysis showed that the institution of an early mobilisation protocol resulted in a 30-fold reduction in the risk of post-operative deep-vein thrombosis when we adjusted for other risk factors.