The Cementless Oxford Unicompartmental Knee Replacement
(OUKR) was developed to address problems related to cementation,
and has been demonstrated in a randomised study to have similar
clinical outcomes with fewer radiolucencies than observed with the
cemented device. However, before its widespread use it is necessary
to clarify contraindications and assess the complications. This
requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless
OUKRs in 881 patients at a minimum follow-up of one year. All patients
had radiological assessment aligned to the bone–implant interfaces
and clinical scores. Analysis was performed at a mean of 38.2 months
(19 to 88) following surgery. A total of 17 patients died (comprising
19 knees (1.9%)), none as a result of surgery; there were no tibial
or femoral loosenings. A total of 19 knees (1.9%) had significant
implant-related complications or required revision. Implant survival
at six years was 97.2%, and there was a partial radiolucency at
the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies.
There was no significant increase in complication rate compared
with cemented fixation (p = 0.87), and no specific contraindications
to
Aims. Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost. Methods. In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA. Results. Cemented fixation had the lowest implant survival in all age groups, with a total ten-year survivorship of 97.0% (95% confidence interval (CI) 95.8 to 97.8) in the cemented group, 97.6% (95% CI 96.9 to 98.1) in the hybrid group, and 97.9% (95% CI 96.9 to 98.6) in the cementless group. This was not statistically significant (p = 0.092). There was no age group where cemented fixation outperformed hybrid or
To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications? A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group.Aims
Methods
There has been a recent increase in interest
for non-cemented fixation in total knee arthroplasty (TKA), however
the superiority of cement fixation is an ongoing debate. Whereas the results based on Level III and IV evidence show similar
survivorship rates between the two types of fixation, Level I and
II evidence strongly support cemented fixation. United Kingdom,
Australia, Sweden, and New Zealand registry data show lower failure
rates and greater usage of cemented than non-cemented fixation.
Case series studies have also indicated greater functional outcomes
and lower revision rates among cemented TKAs. Non-cemented fixation
involves more patellofemoral complications, including increased
susceptibility to wear due to a thinner polyethylene bearing on
the cementless metal-backed component. The combination of results
from registry data, prospective randomised studies, and meta-analyses
support the current superiority of cemented fixation in TKAs.
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite coating, has improved the outcomes of uncemented
TKA over time.
We reviewed 25 patients in whom a MUTARS megaprosthesis with a conical fluted stem had been implanted. There were three types of stem: a standard stem was used in 17 cases (three in the proximal femur, nine in the distal femur and five proximal tibia), a custom-made proximal femoral stem in four cases and a custom-made distal femoral stem in four cases. The mean age of the patients was 40.1 years (17 to 70) and the mean follow-up was for 2.5 years (0.9 to 7.4). At follow-up two patients had died from their disease: one was alive with disease and 22 were disease-free. One of 23 prostheses had been removed for infection and another revised to a cemented stem. The mean Musculoskeletal Tumor Society score was 24.9 (12 to 30) and the mean Karnofsky index was 82% (60% to 100%). There was no radiological evidence of loosening or subsidence. Stem stress shielding was seen in 11 patients and was marked in five of these. There were five complications, rupture of the extensor mechanism of the knee after extra-articular resection in two patients, deep venous thrombosis in one, septic loosening in one, and dislocation of the hip in one. The survival rate after seven years was 87% (95% confidence interval (CI) 83 to 91) for the patients and 95% (95% CI 91 to 99) for the megaprosthesis. A longer follow-up is needed to confirm these encouraging results.
Our aim was to examine the clinical and radiographic outcomes
in 257 consecutive Oxford unicompartmental knee arthroplasties (OUKAs)
(238 patients), five years post-operatively. A retrospective evaluation was undertaken of patients treated
between April 2008 and October 2010 in a regional centre by two
non-designing surgeons with no previous experience of UKAs. The
Oxford Knee Scores (OKSs) were recorded and fluoroscopically aligned
radiographs were assessed post-operatively at one and five years.Aims
Patients and Methods
A retrospective review was performed of patients
undergoing primary cementless total knee replacement (TKR) using
porous tantalum performed by a group of surgical trainees. Clinical
and radiological follow-up involved 79 females and 26 males encompassing
115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up
was 7 years (2 to 11). Tibial and patellar components were porous
tantalum monoblock implants, and femoral components were posterior
stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological
assessments were made for implant positioning, alignment, radiolucencies,
lysis, and loosening. There was 95.7% survival of implants. There
was no radiological evidence of loosening and no osteolysis found.
No revisions were performed for aseptic loosening. Average tibial
component alignment was 1.4° of varus (4°of valgus to 9° varus),
and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean
femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral
alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar
tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient
satisfaction with improvement in pain was 91%. Cementless TKR incorporating
porous tantalum yielded good clinical and radiological outcomes
at a mean of follow-up of seven-years. Cite this article:
The cementless Oxford unicompartmental knee replacement
has been demonstrated to have superior fixation on radiographs and
a similar early complication rate compared with the cemented version.
However, a small number of cases have come to our attention where,
after an apparently successful procedure, the tibial component subsides into
a valgus position with an increased posterior slope, before becoming
well-fixed. We present the clinical and radiological findings of
these six patients and describe their natural history and the likely
causes. Two underwent revision in the early post-operative period,
and in four the implant stabilised and became well-fixed radiologically with
a good functional outcome. This situation appears to be avoidable by minor modifications
to the operative technique, and it appears that it can be treated
conservatively in most patients. Cite this article:
Aims. Cementless total knee arthroplasty (TKA) offers the potential for strong biological fixation compared with cemented TKA where fixation is achieved by the mechanical integration of the cement. Few mid-term results are available for newer cementless TKA designs, which have used additive manufacturing (3D printing). The aim of this study was to present mid-term clinical outcomes and implant survivorship of the cementless Stryker Triathlon Tritanium TKA. Methods. This was a single institution registry review of prospectively gathered data from 341 cementless Triathlon Tritanium TKAs at four to 6.8 years follow-up. Outcomes were determined by comparing pre- and postoperative Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) scores, and pre- and postoperative 12-item Veterans RAND/Short Form Health Survey (VR/SF-12) scores. Aseptic loosening and revision for any reason were the endpoints which were used to determine survivorship at five years. Results. At mid-term follow-up, the mean KOOS JR score improved significantly from 33.14 (0 t0 85, standard deviation (SD) 21.88) preoperatively to 84.12 (15.94 to 100, SD 20.51) postoperatively (p < 0.001), the mean VR/SF-12 scores improved significantly from physical health (PH), 31.21 (SD 5.32; 23.99 to 56.77) preoperatively to 42.62 (SD 10.72; 19.38 to 56.82) postoperatively (p < 0.001) and the mental health (MH), 38.15 (SD 8.17; 19.06 to 60.75) preoperatively to 55.09 (SD 9.64; 19.06 to 66.98) postoperatively (p < 0.001). A total of 11 revisions were undertaken, with an overall revision rate of 2.94%, including five for periprosthetic joint infection (1.34%), three for loosening (0.80%), two for instability (0.53%), and one for pain (0.27%). The overall survivorship was 97.06% and survivorship for aseptic loosening as the endpoint was 98.40%, with a 99.5% survivorship of the 3D-printed tibial component. Conclusion. This 3D-printed cementless total knee system shows excellent survivorship at mid-term follow-up. This design and the ability to obtain
Aims. Cementless unicompartmental knee arthroplasty (UKA) has advantages over cemented UKA, including improved fixation, but has a higher risk of tibial plateau fracture, particularly in Japanese patients. The aim of this multicentre study was to determine when cementless tibial components could safely be used in Japanese patients based on the size and shape of the tibia. Methods. The study involved 212 cementless Oxford UKAs which were undertaken in 174 patients in six hospitals. The medial eminence line (MEL), which is a line parallel to the tibial axis passing through the tip of medial intercondylar eminence, was drawn on preoperative radiographs. Knees were classified as having a very overhanging medial tibial condyle if this line passed medial to the medial tibial cortex. They were also classified as very small if a size A/AA tibial component was used. Results. The overall rate of fracture was 8% (17 out of 212 knees). The rate was higher in knees with very overhanging condyles (Odds ratio (OR) 13; p < 0.001) and with very small components (OR 7; p < 0.001). The OR was 21 (p < 0.001) in those with both very overhanging condyles and very small components. In all, 69% of knees (147) had neither very overhanging nor very small components, and the fracture rate in these patients was 1.4% (2 out of 147 knees). Males had a significantly reduced risk of fracture (OR 0.13; p = 0.002), probably because no males required very small components and females were more likely to have very overhanging condyles (OR 3; p = 0.013). 31% of knees (66) were in males and in these the rate of fracture was 1.5% (1 out of 66 knees). Conclusion. The rate of tibial plateau fracture in Japanese patients undergoing cementless UKA is high. We recommend that cemented tibial fixation should be used in Japanese patients who require very small components or have very overhanging condyles, as identified from preoperative radiographs. In the remaining 69% of knees
Aims. Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA),
Aims. Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for
Aims. Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or mechanical failure of the implant. Methods. Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) aged 50 years and younger (mean 42 years (20 to 50)). There were 87 patients (96 hips) that were followed for a minimum of 15 years (mean 17.3 years (15 to 21)) for analysis. Posterior approach was used with
Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising coating for metallic surfaces, and several types of silver-coated arthroplasty implants are in clinical use today. However, silver can also exert toxic effects on eukaryotic cells both in the immediate vicinity of the coated implants and systemically. In most clinically-used implants, silver coatings are applied on bulk components that are not in direct contact with bone, such as in partial or total long bone arthroplasties used in tumour or complex revision surgery. These implants differ considerably in the coating method, total silver content, and silver release rates. Safety issues, such as the occurrence of argyria, have been a cause for concern, and the efficacy of silver coatings in terms of preventing PJI is also controversial. The application of silver coatings is uncommon on parts of implants intended for
Early implants for total knee replacement were fixed to bone with cement. No firm scientific reason has been given for the introduction of cementless knee replacement and the long-term survivorship of such implants has not shown any advantage over cemented forms. In a randomised, prospective study we have compared cemented and uncemented total knee replacement and report the results of 139 prostheses at five years. Outcome was assessed both clinically by independent examination using the Nottingham knee score and radiologically using the Knee Society scoring system. Independent statistical analysis of the data showed no significant difference between cemented and
The purpose of this study was to assess the long-term results (more than ten years) of two types of cemented ulnar component with type-5 Kudo total elbow arthroplasty in a consecutive series of 56 patients (60 elbows) with rheumatoid arthritis, and to compare the results in elbows above and below a Larsen grade IV. There was no radiolucency around the humeral component. Patients in whom a metal-backed ulnar component and a porous-coated stem were used had better clinical results and significantly less progression of radiolucent line formation around the ulnar component. They also had a significantly better long-term survival than patients with an all-polyethylene ulnar component. The clinical results of arthroplasty using all-polyethylene ulnar components were inferior, regardless of the degree of joint destruction. We conclude that the type-5 Kudo total elbow arthroplasty with
The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity. Cite this article:
In recent years, the use of a collared cementless femoral prosthesis has risen in popularity. The design intention of collared components is to transfer some load to the resected femoral calcar and prevent implant subsidence within the cancellous bone of the metaphysis. Conversely, the load transfer for a cemented femoral prosthesis depends on the cement-component and cement-bone interface interaction. The aim of our study was to compare the three most commonly used collared cementless components and the three most commonly used tapered polished cemented components in patients aged ≥ 75 years who have undergone a primary total hip arthroplasty (THA) for osteoarthritis (OA). Data from the Australian Orthopaedic Association National Joint Replacement Registry from 1 September 1999 to 31 December 2022 were analyzed. Collared cementless femoral components and cemented components were identified, and the three most commonly used components in each group were analyzed. We identified a total of 11,278 collared cementless components and 47,835 cemented components. Hazard ratios (HRs) from Cox proportional hazards models, adjusting for age and sex, were obtained to compare the revision rates between the groups.Aims
Methods
Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.Aims
Methods
The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs. A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.Aims
Methods
The aim of this study was to describe the demographic details of patients who sustain a femoral periprosthetic fracture (PPF), the epidemiology of PPFs, PPF characteristics, and the predictors of PPF types in the UK population. This is a multicentre retrospective cohort study including adult patients presenting to hospital with a new PPF between 1 January 2018 and 31 December 2018. Data collected included: patient characteristics, comorbidities, anticoagulant use, social circumstances, level of mobility, fracture characteristics, Unified Classification System (UCS) type, and details of the original implant. Descriptive analysis by fracture location was performed, and predictors of PPF type were assessed using mixed-effects logistic regression models.Aims
Methods
The aim of this study was to investigate the rate of revision for distal femoral arthroplasty (DFA) performed as a primary procedure for native knee fractures using data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR). Data from the AOANJRR were obtained for DFA performed as primary procedures for native knee fractures from 1 September 1999 to 31 December 2020. Pathological fractures and revision for failed internal fixation were excluded. The five prostheses identified were the Global Modular Arthroplasty System, the Modular Arthroplasty System, the Modular Universal Tumour And Revision System, the Orthopaedic Salvage System, and the Segmental System. Patient demographic data (age, sex, and American Society of Anesthesiologists grade) were obtained, where available. Kaplan-Meier estimates of survival were used to determine the rate of revision, and the reasons for revision and mortality data were examined.Aims
Methods
To analyze the short-term outcome of two types of total wrist arthroplasty (TWA) in terms of wrist function, migration, and periprosthetic bone behaviour. A total of 40 patients suffering from non-rheumatoid wrist arthritis were enrolled in a randomized controlled trial comparing the ReMotion and Motec TWAs. Patient-rated and functional outcomes, radiological changes, blood metal ion levels, migration measured by model-based radiostereometric analysis (RSA), bone mineral density (BMD) measured by dual-energy X-ray absorptiometry (DXA), complications, loosening, and revision rates at two years were compared.Aims
Methods
Higher osteoblastic bone activity is expected in aseptic loosening and painful unicompartmental knee arthroplasty (UKA). However, insights into normal bone activity patterns after medial UKAs are lacking. The aim of this study was to identify the evolution in bone activity pattern in well-functioning medial mobile-bearing UKAs. In total, 34 patients (13 female, 21 male; mean age 62 years (41 to 79); BMI 29.7 kg/m2 (23.6 to 42.1)) with 38 medial Oxford partial UKAs (20 left, 18 right; 19 cementless, 14 cemented, and five hybrid) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively, and at one and two years postoperatively. Changes in mean osteoblastic activity were investigated using a tracer localization scheme with volumes of interest (VOIs), reported by normalized mean tracer values. A SPECT/CT registration platform additionally explored cortical tracer evolution in zones of interest identified by previous experimental research.Aims
Methods
Revision total hip arthroplasty (THA) is projected
to increase by 137% from the years 2005 to 2030. Reconstruction of
the femur with massive bone loss can be a formidable undertaking.
The goals of revision surgery are to create a stable construct,
preserve bone and soft tissues, augment deficient host bone, improve
function, provide a foundation for future surgery, and create a
biomechanically restored hip. Options for treatment of the compromised femur
include: resection arthroplasty, allograft prosthetic composite
(APC), proximal femoral replacement,
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce
The aim of this study was to evaluate the performance of first-generation annealed highly cross-linked polyethylene (HXLPE) in cementless total hip arthroplasty (THA). We retrospectively evaluated 29 patients (35 hips) who underwent THA between December 2000 and February 2002. The survival rate was estimated using the Kaplan-Meier method. Hip joint function was evaluated using the Japanese Orthopaedic Association (JOA) score. Two-dimensional polyethylene wear was estimated using Martell’s Hip Analysis Suite. We calculated the wear rates between years 1 and 5, 5 and 10, 10 and 15, and 15 and final follow-up.Aims
Methods
We have carried out a long-term survival analysis of a prospective, randomised trail comparing cemented with
We aimed to compare the implant survival, complications, readmissions, and mortality of Vancouver B2 periprosthetic femoral fractures (PFFs) treated with internal fixation with that of B1 PFFs treated with internal fixation and B2 fractures treated with revision arthroplasty. We retrospectively reviewed the data of 112 PFFs, of which 47 (42%) B1 and 27 (24%) B2 PFFs were treated with internal fixation, whereas 38 (34%) B2 fractures underwent revision arthroplasty. Decision to perform internal fixation for B2 PFFs was based on specific radiological (polished femoral components, intact bone-cement interface) and clinical criteria (low-demand patient). Median follow-up was 36.4 months (24 to 60). Implant survival and mortality over time were estimated with the Kaplan-Meier method. Adverse events (measured with a modified Dindo-Clavien classification) and 90-day readmissions were additionally compared between groups.Aims
Methods
The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up. A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted.Aims
Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
The aims of this study were to validate the outcome of total elbow arthroplasty (TEA) in patients with rheumatoid arthritis (RA), and to identify factors that affect the outcome. We searched PubMed, MEDLINE, Cochrane Reviews, and Embase from between January 2003 and March 2019. The primary aim was to determine the implant failure rate, the mode of failure, and risk factors predisposing to failure. A secondary aim was to identify the overall complication rate, associated risk factors, and clinical performance. A meta-regression analysis was completed to identify the association between each parameter with the outcome.Aims
Methods
We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up. Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips).Aims
Patients and Methods
The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.Aims
Materials and Methods
We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis. The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR.Aims
Patients and Methods
Our intention was to investigate if the highly porous biological fixation surfaces of a new 3D-printed total knee arthroplasty (TKA) achieved adequate fixation of the tibial and patellar components to the underlying bone. A total of 29 patients undergoing primary TKA consented to participate in this prospective cohort study. All patients received a highly porous tibial baseplate and metal-backed patella. Patient-reported outcomes measures were recorded and implant migration was assessed using radiostereometric analysis.Aims
Patients and Methods
Shoulder arthroplasty using short humeral components is becoming increasingly popular. Some such components have been associated with relatively high rates of adverse radiological findings. The aim of this retrospective review was to evaluate the radiological humeral bone changes and mechanical failure rates with implantation of a short cementless humeral component in anatomical (TSA) and reverse shoulder arthroplasty (RSA). A total of 100 shoulder arthroplasties (35 TSA and 65 RSA) were evaluated at a mean of 3.8 years (3 to 8.3). The mean age at the time of surgery was 68 years (31 to 90). The mean body mass index was 32.7 kg/m2 (17.3 to 66.4).Aims
Patients and Methods
We have investigated those factors which influence the range of movement after total knee arthroplasty, including sex, age, preoperative diagnosis and preoperative flexion deformity and flexion range. We also compared cemented and
The aim of this meta-analysis was to compare the outcome of total elbow arthroplasty (TEA) undertaken for rheumatoid arthritis (RA) with TEA performed for post-traumatic conditions with regard to implant failure, functional outcome, and perioperative complications. We completed a comprehensive literature search on PubMed, Web of Science, Embase, and the Cochrane Library and conducted a systematic review and meta-analysis. Nine cohort studies investigated the outcome of TEA between RA and post-traumatic conditions. The preferred reporting items for systematic reviews and meta-analysis (Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA)) guidelines and Newcastle-Ottawa scale were applied to assess the quality of the included studies. We assessed three major outcome domains: implant failures (including aseptic loosening, septic loosening, bushing wear, axle failure, component disassembly, or component fracture); functional outcomes (including arc of range of movement, Mayo Elbow Performance Score (MEPS), and the Disabilities of the Arm, Shoulder and Hand (DASH) questionnaire), and perioperative complications (including deep infection, intraoperative fracture, postoperative fracture, and ulnar neuropathy).Aims
Materials and Methods
The aim of this study was to report the mid-term clinical outcome
of cemented unlinked J-alumina ceramic elbow (JACE) arthroplasties
when used in patients with rheumatoid arthritis (RA). We retrospectively reviewed 87 elbows, in 75 patients with RA,
which was replaced using a cemented JACE total elbow arthroplasty
(TEA) between August 2003 and December 2012, with a follow-up of
96%. There were 72 women and three men, with a mean age of 62 years
(35 to 79). The mean follow-up was nine years (2 to 14). The clinical condition
of each elbow before and after surgery was assessed using the Mayo
Elbow Performance Index (MEPI, 0 to 100 points). Radiographic loosening
was defined as a progressive radiolucent line of >1 mm that was
completely circumferential around the prosthesis.Aims
Patients and Methods
The treatment of patients with allergies to metal in total joint arthroplasty is an ongoing debate. Possibilities include the use of hypoallergenic prostheses, as well as the use of standard cobalt-chromium (CoCr) alloy. This non-designer study was performed to evaluate the clinical outcome and survival rates of unicondylar knee arthroplasty (UKA) using a standard CoCr alloy in patients reporting signs of a hypersensitivity to metal. A consecutive series of patients suitable for UKA were screened for symptoms of metal hypersensitivity by use of a questionnaire. A total of 82 patients out of 1737 patients suitable for medial UKA reporting cutaneous metal hypersensitivity to cobalt, chromium, or nickel were included into this study and prospectively evaluated to determine the functional outcome, possible signs of hypersensitivity, and short-term survivorship at a minimum follow-up of 1.5 years.Aims
Patients and Methods
In the initial development of total shoulder arthroplasty (TSA),
the humeral component was usually fixed with cement. Cementless
components were subsequently introduced. The aim of this study was
to compare the long-term outcome of cemented and cementless humeral
components in arthroplasty of the shoulder. All patients who underwent primary arthroplasty of the shoulder
at our institution between 1970 and 2012 were included in the study.
There were 4636 patients with 1167 cemented humeral components and
3469 cementless components. Patients with the two types of fixation
were matched for nine different covariates using a propensity score
analysis. A total of 551 well-balanced pairs of patients with cemented
and cementless components were available after matching for comparison
of the outcomes. The clinical outcomes which were analysed included loosening
of the humeral component determined at revision surgery, periprosthetic
fractures, post-operative infection and operating time.Aims
Patients and Methods
The aim of this study was to report the implant survival and patient-reported outcome measures (PROMs) in a consecutive series of patients aged less than 50 years at the time of arthroplasty using the Birmingham Hip Resurfacing system (BHR), with a minimum follow-up of ten years. A total of 226 patients with osteoarthritis of the hip, who underwent BHR and presented to a single surgeon, were included in the study. Survival of the implant was confirmed by cross-checking with the Australian Orthopaedic Association National Joint Replacement Registry. Kaplan–Meier survival curves with 95% confidence intervals (CIs) were constructed. Pre- and postoperative PROMs were compared with Aims
Patients and Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
The primary purpose of this study of metal-on-metal
(MoM) hip resurfacing was to compare the effect of using a cementless
or cemented femoral component on the subsequent bone mineral density
(BMD) of the femoral neck. This was a single-centre, prospective, double-blinded control
trial which randomised 120 patients (105 men and 15 women) with
a mean age of 49.4 years (21 to 68) to receive either a cemented
or cementless femoral component. Follow-up was to two years. Outcome
measures included total and six-point region-of-interest BMD of
the femoral neck, radiological measurements of acetabular inclination,
neck-shaft and stem-shaft angles, and functional outcome scores
including the Harris hip score, the Western Ontario and McMaster
Universities Osteoarthritis Index and the University of California
at Los Angeles activity scale. In total, 17 patients were lost to follow-up leaving 103 patients
at two years. There were no revisions in the cementless group and
three revisions (5%) in the cemented group (two because of hip pain
and one for pseudotumour). The total BMD was significantly higher in the cementless group
at six months (p <
0.001) and one year (p = 0.01) than in the
cemented group, although there was a loss of statistical significance
in the difference at two years (p = 0.155). All patient outcomes improved significantly: there were no significant
differences between the two groups. The results show better preservation of femoral neck BMD with
a cementless femoral component after two years of follow-up. Further
investigation is needed to establish whether this translates into
improved survivorship. Cite this article:
Periprosthetic femoral fractures (PFF) following total hip arthroplasty
(THA) are devastating complications that are associated with functional
limitations and increased overall mortality. Although cementless
implants have been associated with an increased risk of PFF, the
precise contribution of implant geometry and design on the risk
of both intra-operative and post-operative PFF remains poorly investigated.
A systematic review was performed to aggregate all of the PFF literature
with specific attention to the femoral implant used. A systematic search strategy of several journal databases and
recent proceedings from the American Academy of Orthopaedic Surgeons
was performed. Clinical articles were included for analysis if sufficient
implant description was provided. All articles were reviewed by
two reviewers. A review of fundamental investigations of implant
load-to-failure was performed, with the intent of identifying similar
conclusions from the clinical and fundamental literature.Aims
Patients and Methods
The aim of this study was to identify patient- and surgery-related
risk factors for sustaining an early periprosthetic fracture following
primary total hip arthroplasty (THA) performed using a double-tapered
cementless femoral component (Bi-Metric femoral stem; Biomet Inc.,
Warsaw, Indiana). A total of 1598 consecutive hips, in 1441 patients receiving
primary THA between January 2010 and June 2015, were retrospectively
identified. Level of pre-operative osteoarthritis, femoral Dorr
type and cortical index were recorded. Varus/valgus placement of
the stem and canal fill ratio were recorded post-operatively. Periprosthetic
fractures were identified and classified according to the Vancouver
classification. Regression analysis was performed to identify risk
factors for early periprosthetic fracture.Aims
Patients and Methods
Our aim was to prepare a systematic review and meta-analysis
to compare the outcomes of cemented and cementless hemiarthroplasty
of the hip, in elderly patients with a fracture of the femoral neck,
to investigate the mortality, complications, length of stay in hospital,
blood loss, operating time and functional results. A systematic review and meta-analysis was conducted following
the Preferred Reporting Items for Systematic Reviews and Meta-Analyses
guidelines on randomised controlled trials (RCTs), studying current
generation designs of stem only. The synthesis of results was done
of pooled data, with a fixed effects or random effects model, based on
heterogeneity.Aims
Materials and Methods
The optimal method of tibial component fixation remains uncertain
in total knee arthroplasty (TKA). Hydroxyapatite coatings have been
applied to improve bone ingrowth in uncemented designs, but may
only coat the directly accessible surface. As peri-apatite (PA)
is solution deposited, this may increase the coverage of the implant
surface and thereby fixation. We assessed the tibial component fixation
of uncemented PA-coated TKAs Patients were randomised to PA-coated or cemented TKAs. In 60
patients (30 in each group), radiostereometric analysis of tibial
component migration was evaluated as the primary outcome at baseline,
three months post-operatively and at one, two and five years. A
linear mixed-effects model was used to analyse the repeated measurements.Aims
Patients and Methods
The most common reasons for revision of unicompartmental
knee arthroplasty (UKA) are loosening and pain. Cementless components
may reduce the revision rate. The aim of this study was to compare
the fixation and clinical outcome of cementless and cemented Oxford
UKAs. A total of 43 patients were randomised to receive either a cemented
or a cementless Oxford UKA and were followed for two years with
radiostereometric analysis (RSA), radiographs aligned with the bone–implant
interfaces and clinical scores. The femoral components migrated significantly during the first
year (mean 0.2 mm) but not during the second. There was no significant
difference in the extent of migration between cemented and cementless
femoral components in either the first or the second year. In the
first year the cementless tibial components subsided significantly
more than the cemented components (mean 0.28 mm ( As second-year migration is predictive of subsequent loosening,
and as radiolucency is suggestive of reduced implant–bone contact,
these data suggest that fixation of the cementless components is
at least as good as, if not better than, that of cemented devices. Cite this article:
Bactericidal levels of antibiotics are difficult
to achieve in infected total joint arthroplasty when intravenous antibiotics
or antibiotic-loaded cement spacers are used, but intra-articular
(IA) delivery of antibiotics has been effective in several studies.
This paper describes a protocol for IA delivery of antibiotics in
infected knee arthroplasty, and summarises the results of a pharmacokinetic
study and two clinical follow-up studies of especially difficult
groups: methicillin-resistant Cite this article:
To determine the effect of a change in design of a cementless
ceramic acetabular component in fixation and clinical outcome after
total hip arthroplasty We compared 342 hips (302 patients) operated between 1999 and
2005 with a relatively smooth hydroxyapatite coated acetabular component
(group 1), and 337 hips (310 patients) operated between 2006 and
2011 using a similar acetabular component with a macrotexture on
the entire outer surface of the component (group 2). The mean age of
the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70)
in group 2. The mean follow-up was 12.7 years (10 to 17) for group
1 and 7.2 years (4 to 10) for group 2.Aims
Patients and Methods
It has been suggested that cemented fixation of total hip arthroplasty
(THA) is associated with an increased peri-operative mortality compared
with cementless THA. Our aim was to investigate this through a nationwide
matched cohort study adjusting for age, comorbidity, and socioeconomic
background. A total of 178 784 patients with osteoarthritis who underwent
either cemented or cementless THA from the Swedish Hip Arthroplasty
Register were matched with 862 294 controls from the general population.
Information about the causes of death, comorbidities, and socioeconomic
background was obtained. Mortality within the first 90 days after
the operation was the primary outcome measure.Aims
Patients and Methods
Until now, there has been no consensus as to whether stemmed
acetabular components are appropriate for use in patients undergoing
revision total hip arthroplasty (THA) who have major acetabular
defects or pelvic discontinuity. We wished to address this deficiency
in the literature. We carried out a retrospective study of 35 patients (six men
and 29 women) with a mean age of 68 years (37 to 87), with major
acetabular defects who underwent revision THA between 2000 and 2012.Aim
Patients and Methods
The purpose of this prospective, randomised study
was to evaluate the clinical and radiological results comparing the
identical cemented or cementless NexGen total knee prostheses implanted
bilaterally in the same patient. Sequential simultaneous bilateral
total knee replacements were performed in 50 patients (100 knees).
There were 39 women and 11 men with a mean age of 58.4 years (51
to 67) who received a cemented prosthesis in one knee and a cementless
prosthesis in the other. The mean follow-up was 13.6 years (13 to
14). At final review, the mean Knee Society scores (96.2 (82 to
100)
We report the long-term survival of a prospective randomised consecutive series of 501 primary knee replacements using the press-fit condylar posterior cruciate ligament-retaining prosthesis. Patients received either cemented (219 patients, 277 implants) or cementless (177 patients, 224 implants) fixation. Altogether, 44 of 501 knees (8.8%) underwent revision surgery (24 cemented This single-surgeon series, with no loss to follow-up, provides reliable data of the revision rates of one of the most commonly-used total knee replacements. The survival of the press-fit condylar total knee replacement remained good at 15 years, irrespective of the method of fixation.
Our aim was to determine the success rate of repeated debridement and two-stage cementless revision arthroplasty according to the type of infected total hip replacement (THR). We enrolled 294 patients (294 hips) with an infected THR in the study. There were 222 men and 72 women with a mean age of 55.1 years (24.0 to 78.0). The rate of control of infection after the initial treatment and after repeated debridement and two-stage revisions was determined. The clinical (Harris hip score) and radiological results were evaluated. The mean follow-up was 10.4 years (5.0 to 14.0). The eventual rate of control of infection was 100.0% for early superficial post-operative infection, 98.4% for early deep post-operative infection, 98.5% for late chronic infection and 91.0% for acute haematogenous infection. Overall, 288 patients (98%) maintained a functioning THR at the latest follow-up. All the allografts appeared to be united and there were no failures. These techniques effectively controlled infection and maintained a functional THR with firm fixation in most patients. Repeated debridement and two-stage or repeated two-stage revisions further improved the rate of control of infection after the initial treatment and increased the likelihood of maintaining a functional THR.
This was a randomised controlled trial studying
the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised
that it would have similar early clinical results and micromovement
to those of a standard-length tapered Synergy metaphyseal fixation
stem. Using radiostereometric analysis (RSA) we compared the two
stems in 43 patients. A short metaphyseal fixation stem was used
in 22 patients and a Synergy stem in 21 patients. No difference
was found in the clinical outcomes pre- or post-operatively between
groups. RSA showed no significant differences two years post-operatively
in mean micromovement between the two stems (except for varus/valgus tilt
at p = 0.05) (subsidence 0.94 mm ( Cite this article:
We have reviewed the rate of revision of fully cemented, hybrid and uncemented primary total hip replacements (THRs) registered in the New Zealand Joint Registry between 1999 and December 2006 to determine whether there was any statistically significant difference in the early survival and reason for revision in these different types of fixation. The percentage rate of revision was calculated per 100 component years and compared with the reason for revision, the type of fixation and the age of the patients. Of the 42 665 primary THRs registered, 920 (2.16%) underwent revision requiring change of at least one component. Fully-cemented THRs had a lower rate of revision when considering all causes for failure (p <
0.001), but below the age of 65 years uncemented THRs had a lower rate (p <
0.01). The rate of revision of the acetabular component for aseptic loosening was less in the uncemented and hybrid groups compared with that in the fully cemented group (p <
0.001), and the rate of revision of cemented and uncemented femoral components was similar, except in patients over 75 years of age in whom revision of cemented femoral components was significantly less frequent (p <
0.02). Revision for infection was more common in patients aged below 65 years and in cemented and hybrid THRs compared with cementless THRs (p <
0.001). Dislocation was the most common cause of revision for all types of fixation and was more frequent in both uncemented acetabular groups (p <
0.001). The experience of the surgeon did not affect the findings. Although cemented THR had the lowest rate of revision for all causes in the short term (90 days), uncemented THR had the lowest rate of aseptic loosening in patients under 65 years of age and had rates comparable with international rates of aseptic loosening in those over 65 years.
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,
and therefore, improve patient satisfaction post-operatively. Cite this article:
We compared peri-prosthetic bone mineral density between identical cemented and cementless LCS rotating platform total knee arthroplasties. Two matched cohorts had dual energy x-ray absorptiometry scans two years post-operatively using a modified validated densitometric analysis protocol, to assess peri-prosthetic bone mineral density. The knee that was not operated on was also scanned to enable the calculation of a relative bone mineral density difference. Oxford Knee and American Knee Society scores were comparable in the two cohorts. Statistical analysis revealed no significant difference in absolute, or relative peri-prosthetic bone mineral density with respect to the method of fixation. However, the femoral peri-prosthetic bone mineral density and relative bone mineral density difference were significantly decreased, irrespective of the method of fixation, particularly in the anterior distal portion of the femur, with a mean reduction in relative bone mineral density difference of 27%. There was no difference in clinical outcome between the cemented and cementless LCS total knee arthroplasty. However, both produce stress-shielding around the femoral implants. This leads us to question the use of more expensive cementless total knee components.
The aims of this retrospective study were to
compare the mid-term outcomes following revision total knee replacement
(TKR) in 76 patients (81 knees) <
55 years of age with those
of a matched group of primary TKRs based on age, BMI, gender and
comorbid conditions. We report the activity levels, functional scores,
rates of revision and complications. Compared with patients undergoing
primary TKR, those undergoing revision TKR had less improvement
in the mean Knee Society function scores (8.14 (–55 to +60) Young patients undergoing revision TKR should be counselled that
they can expect somewhat less improvement and a higher risk of complications
than occur after primary TKR. Cite this article:
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
We assessed the orientation of the acetabular
component in 1070 primary total hip arthroplasties with hard-on-soft, small
diameter bearings, aiming to determine the size and site of the
target zone that optimises outcome. Outcome measures included complications,
dislocations, revisions and ΔOHS (the difference between the Oxford
Hip Scores pre-operatively and five years post-operatively). A wide
scatter of orientation was observed (2 This study demonstrated that with traditional technology surgeons
can only reliably achieve a target zone of ±15°. As the optimal
zone to diminish the risk of dislocation is also ±15°, surgeons
should be able to achieve this. This is the first study to demonstrate
that optimal orientation of the acetabular component improves the
functional outcome. However, the target zone is small (± 5°) and
cannot, with current technology, be consistently achieved. Cite this article:
Femoral revision after cemented total hip replacement
(THR) might include technical difficulties, following essential cement
removal, which might lead to further loss of bone and consequently
inadequate fixation of the subsequent revision stem. Femoral impaction allografting has been widely used in revision
surgery for the acetabulum, and subsequently for the femur. In combination
with a primary cemented stem, impaction grafting allows for femoral
bone restoration through incorporation and remodelling of the impacted
morsellized bone graft by the host skeleton. Cavitary bone defects
affecting meta-physis and diaphysis leading to a wide femoral shaft,
are ideal indications for this technique. Cancellous allograft bone
chips of 1 mm to 2 mm size are used, and tapered into the canal
with rods of increasing diameters. To impact the bone chips into
the femoral canal a prosthesis dummy of the same dimensions of the definitive
cemented stem is driven into the femur to ensure that the chips
are very firmly impacted. Finally, a standard stem is cemented into
the neo-medullary canal using bone cement. To date several studies have shown favourable results with this
technique, with some excellent long-term results reported in independent
clinical centres worldwide. Cite this article:
Cementless femoral stems are currently preferred
for total hip replacement (THR) in the United States. Improvements
in stem design, instrumentation and surgical technique have made
this technology highly successful, reproducible, and applicable
to the vast majority of patients requiring a THR. However, there
are ongoing developments in some aspects of stem design that influence
clinical results, the incidence of complications and their inherent adaptability
in accommodating the needs of individual patients. Here we examine
some of these design features. Cite this article:
Tapered, fluted, modular, titanium stems are
increasingly popular in the operative management of Vancouver B2 and
selected B3 peri-prosthetic femoral fractures. We have reviewed
the results at our institution looking at stem survival and clinical
outcomes and compared this with reported outcomes in the literature.
Stem survival at a mean of 54 months was 96% in our series and 97%
for combined published cases. Review of radiology showed maintenance
or improvement of bone stock in 89% of cases with high rates of
femoral union. Favourable clinical outcome scores have reported
by several authors. No difference in survival or clinical scores
was observed between B2 and B3 fractures. Tapered stems are a useful
option in revision for femoral fracture across the spectrum of femoral
bone deficiency. Cite this article:
Many different lengths of stem are available
for use in primary total hip replacement, and the morphology of
the proximal femur varies greatly. The more recently developed shortened
stems provide a distribution of stress which closely mimics that
of the native femur. Shortening the femoral component potentially
comes at the cost of decreased initial stability. Clinical studies
on the performance of shortened cemented and cementless stems are promising,
although long-term follow-up studies are lacking. We provide an
overview of the current literature on the anatomical features of
the proximal femur and the biomechanical aspects and clinical outcomes
associated with the length of the femoral component in primary hip
replacement, and suggest a classification system for the length
of femoral stems. Cite this article:
We randomised 62 knees to receive either cemented or cementless versions of the Oxford unicompartmental knee replacement. The implants used in both arms of the study were similar, except that the cementless components were coated with porous titanium and hydroxyapatite. The tibial interfaces were studied with fluoroscopically-aligned radiographs. At one year there was no difference in clinical outcome between the two groups. Narrow radiolucent lines were seen at the bone-implant interfaces in 75% of cemented tibial components. These were partial in 43%, and complete in 32%. In the cementless implants, partial radiolucencies were seen in 7% and complete radiolucencies in none. These differences are statistically significant (p <
0.0001) and imply satisfactory bone ingrowth into the cementless implants.
The optimum cementing technique for the tibial
component in cemented primary total knee replacement (TKR) remains
controversial. The technique of cementing, the volume of cement
and the penetration are largely dependent on the operator, and hence
large variations can occur. Clinical, experimental and computational
studies have been performed, with conflicting results. Early implant
migration is an indication of loosening. Aseptic loosening is the
most common cause of failure in primary TKR and is the product of
several factors. Sufficient penetration of cement has been shown
to increase implant stability. This review discusses the relevant literature regarding all aspects
of the cementing of the tibial component at primary TKR. Cite this article:
The aim of this prospective randomised study
was to compare the clinical and radiological results of a cemented
all-polyethylene Ultima acetabular component with those of a cementless
porous-coated acetabular component (PFC) following total hip replacement
(THR). A total of 287 patients received either a polyethylene acetabular
component (group A) or a cobalt–chromium porous-coated component
(group B) with an identical cemented femoral component and 28 mm
cobalt-chromium head, thus making it the largest study of its type.
Patients were evaluated radiologically and clinically using the
Harris hip score (HHS). Group A comprised 183 patients (73 male,
110 female) with a mean age of
71.3 years (55 to 89). Group B comprised 104 patients (48 male,
56 female) with a mean age of 69.8 years (56 to 89). A total of
16 patients (13 in Group A, three in Group B) did not have post-operative
data for analysis. The mean follow-up in group A was 7.52 years
(0.4 to 15.0) and in Group B 7.87 years (0.5 to 14.0). At final follow-up the mean HHS was similar between groups A
and B (74.5 (25 to 100) and 78.0 (37 to 100), respectively; p =
0.068). The total number of revisions for any cause was 28, 17 of
which were in group A and 11 in group B. The ten-year survivorship
was 86.8% (95% confidence interval (CI) 78.4 to 92.1) and 89.2%
(95% CI 78.3 to 94.8) for groups A and B, respectively (log-rank
p-value = 0.938). A total of 20 cemented and two cementless acetabular
components had evidence of acetabular radiolucencies or acetabular
component migration at last follow-up (p = 0.001). These results indicate that patients with a cemented all-polyethylene
and cementless porous-coated polyethylene lined acetabular component
have similar long-term clinical outcomes.
Degenerative problems of the hip in patients
with childhood and adult onset neuromuscular disorders can be challenging
to treat. Many orthopaedic surgeons are reluctant to recommend total
hip replacement (THR) for patients with underlying neuromuscular
disorders due to the perceived increased risks of dislocation, implant loosening,
and lack of information about the functional outcomes and potential
benefits of these procedures in these patients. Modular femoral
components and alternative bearings which facilitate the use of
large femoral heads, constrained acetabular components and perhaps
more importantly, a better understanding about the complications
and outcomes of THR in the patient with neuromuscular disorders,
make this option viable. This paper will review the current literature
and our experience with THR in the more frequently encountered neuromuscular
disorders. Cite this article:
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in England and Wales. This retrospective
cohort study records survival time to revision following primary
cementless THR with the most common combination (accounting for
almost a third of all cementless THRs), and explores risk factors independently
associated with failure, using data from the National Joint Registry
for England and Wales. Patients with osteoarthritis who had a DePuy
Corail/Pinnacle THR implanted between the establishment of the registry
in 2003 and 31 December 2010 were included within analyses. There
were 35 386 procedures. Cox proportional hazard models were used
to analyse the extent to which the risk of revision was related
to patient, surgeon and implant covariates. The overall rate of
revision at five years was 2.4% (99% confidence interval 2.02 to
2.79). In the final adjusted model, we found that the risk of revision
was significantly higher in patients receiving metal-on-metal (MoM:
hazard ratio (HR) 1.93, p <
0.001) and ceramic-on-ceramic bearings
(CoC: HR 1.55, p = 0.003) compared with the best performing bearing
(metal-on-polyethylene). The risk of revision was also greater for
smaller femoral stems (sizes 8 to 10: HR 1.82, p <
0.001) compared
with mid-range sizes. In a secondary analysis of only patients where body
mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m2 significantly
increased the risk of revision (HR 1.55, p = 0.002). The influence
of the bearing on the risk of revision remained significant (MoM:
HR 2.19, p <
0.001; CoC: HR 2.09,
p = 0.001). The risk of revision was independent of age, gender,
head size and offset, shell, liner and stem type, and surgeon characteristics. We found significant differences in failure between bearing surfaces
and femoral stem size after adjustment for a range of covariates
in a large cohort of single-brand cementless THRs. In this study
of procedures performed since 2003, hard bearings had significantly
higher rates of revision, but we found no evidence that head size
had an effect. Patient characteristics, such as BMI and American
Society of Anesthesiologists grade, also influence the survival
of cementless components. Cite this article:
We present a retrospective series of 170 cemented titanium straight-stem femoral components combined with two types of femoral head: cobalt-chromium (CoCr) alloy (114 heads) and alumina ceramic (50 heads). Of the study group, 55 patients (55 stems) had died and six (six stems) were lost to follow-up. At a mean of 13.1 years (3 to 15.3) 26 stems had been revised for aseptic loosening. The mean follow-up time for stable stems was 15.1 years (12.1 to 16.6). Survival of the stem at 15 years was 75.4% (95% confidence interval (CI) 67.3 to 83.5) with aseptic failure (including radiological failure) as the end-point, irrespective of the nature of the head and the quality of the cement mantle. Survival of the stem at 15 years was 79.1% (95% CI 69.8 to 88.4) and 67.1% (95% CI 51.3 to 82.9) with the CoCr alloy and ceramic heads, respectively. The quality of the cement mantle was graded as a function of stem coverage: stems with complete tip coverage (type 1) had an 84.9% (95% CI 77.6 to 92.2) survival at 15 years, compared with those with a poor tip coverage (type 2) which had a survival of only 22.4% (95% CI 2.4 to 42.4). The poor quality of the cement mantle and the implantation of an alumina head substantially lowered the survival of the stem. In our opinion, further use of the cemented titanium alloy straight-stem femoral components used in our series is undesirable.
We report the use of an allograft prosthetic composite for reconstruction of the skeletal defect in complex revision total hip replacement for severe proximal femoral bone loss. Between 1986 and 1999, 72 patients (20 men, 52 women) with a mean age of 59.9 years (38 to 78) underwent reconstruction using this technique. At a mean follow-up of 12 years (8 to 20) 57 patients were alive, 14 had died and one was lost to follow-up. Further revision was performed in 19 hips at a mean of 44.5 months (11 to 153) post-operatively. Causes of failure were aseptic loosening in four, allograft resorption in three, allograft nonunion in two, allograft fracture in four, fracture of the stem in one, and deep infection in five. The survivorship of the allograft-prosthesis composite at ten years was 69.0% (95% confidence interval 67.7 to 70.3) with 26 patients remaining at risk. Survivorship was statistically significantly affected by the severity of the pre-operative bone loss (Paprosky type IV; p = 0.019), the number of previous hip revisions exceeding two (p = 0.047), and the length of the allograft used (p = 0.005).
We carried out a prospective study of 118 hydroxyapatite-coated, cementless total knee replacements in patients who were ≤ 55 years of age and who had primary (92; 78%) or post-traumatic (26; 22%) osteoarthritis. The mean period of follow-up was 7.9 years (5 to 12.5). The Knee Society clinical scores improved from a pre-operative mean of 98 (0 to 137) to a mean of 185 (135 to 200) at five years, and 173 (137 to 200) at ten years. There were two revisions of the tibial component because of aseptic loosening, and one case of polyethylene wear requiring further surgery. There was no osteolysis or progressive radiological loosening of any other component. At 12 years, the overall rate of implant survival was 97.5% (excluding exchange of spacer) and 92.1% (including exchange of spacer). Cementless total knee replacement can achieve excellent long-term results in young, active patients with osteoarthritis. In contrast to total hip replacement, polyethylene wear, osteolysis and loosening of the prosthesis were not major problems for these patients, although it is possible that this observation could change with longer periods of follow-up.
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
When performing total hip replacement (THR) in high dislocated hips, the presence of soft-tissue contractures means that most surgeons prefer to use a femoral shortening osteotomy in order to avoid the risk of neurovascular damage. However, this technique will sacrifice femoral length and reduce the extent of any leg-length equalisation. We report our experience of 74 THRs performed between 2000 and 2008 in 65 patients with a high dislocated hip without a femoral shortening osteotomy. The mean age of the patients was 55 years (46 to 72) and the mean follow-up was 42 months (12 to 78). All implants were cementless except for one resurfacing hip implant. We attempted to place the acetabular component in the anatomical position in each hip. The mean Harris hip score improved from 53 points (34 to 74) pre-operatively to 86 points (78 to 95) at final follow-up. The mean radiologically determined leg lengthening was 42 mm (30 to 66), and the mean leg-length discrepancy decreased from 36 mm (5 to 56) pre-operatively to 8.5 mm (0 to 18) postoperatively. Although there were four (5%) post-operative femoral nerve palsies, three had fully resolved by six months after the operation. No loosening of the implant was observed, and no dislocations or infections were encountered. Total hip replacement without a femoral shortening osteotomy proved to be a safe and effective surgical treatment for high dislocated hips.
Pathological fractures of the humerus are associated with pain, morbidity, loss of function and a diminished quality of life. We report our experience of stabilising these fractures using polymethylmethacrylate and non-locking plates. We undertook a retrospective review over 20 years of patients treated at a tertiary musculoskeletal oncology centre. Those who had undergone surgery for an impending or completed pathological humeral fracture with a diagnosis of metastatic disease or myeloma were identified from our database. There were 63 patients (43 men, 20 women) in the series with a mean age of 63 years (39 to 87). All had undergone intralesional curettage of the tumour followed by fixation with intramedullary polymethylmethacrylate and plating. Complications occurred in 14 patients (22.2%) and seven (11.1%) required re-operation. At the latest follow-up, 47 patients (74.6%) were deceased and 16 (25.4%) were living with a mean follow-up of 75 months (1 to 184). A total of 54 (86%) patients had no or mild pain and 50 (80%) required no or minimal assistance with activities of daily living. Of the 16 living patients none had pain and all could perform activities of daily living without assistance. Intralesional resection of the tumour, filling of the cavity with cement, and plate stabilisation of the pathological fracture gives immediate rigidity and allows an early return of function without the need for bony union. The patient’s local disease burden is reduced, which may alleviate tumour-related pain and slow the progression of the disease. The cemented-plate technique provides a reliable option for the treatment of pathological fractures of the humerus.
Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR. A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by >
3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47). We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.
This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
Peri-prosthetic femoral fracture after total hip replacement (THR) is associated with a poor outcome and high mortality. However, little is known about its long-term incidence after uncemented THR. We retrospectively reviewed a consecutive series of 326 patients (354 hips) who had received a CLS Spotorno replacement with an uncemented, straight, collarless tapered titanium stem between January 1985 and December 1989. The mean follow-up was 17 years (15 to 20). The occurrence of peri-prosthetic femoral fracture during follow-up was noted. Kaplan-Meier survival analysis was used to estimate the cumulative incidence of fracture. At the last follow-up, 86 patients (89 hips) had died and eight patients (eight hips) had been lost to follow-up. A total of 14 fractures in 14 patients had occurred. In ten hips, the femoral component had to be revised and in four the fracture was treated by open reduction and internal fixation. The cumulative incidence of peri-prosthetic femoral fracture was 1.6% (95% confidence interval 0.7 to 3.8) at ten years and 4.5% (95% confidence interval 2.6 to 8.0) at 17 years after the primary THR. There was no association between the occurrence of fracture and gender or age at the time of the primary replacement. Our findings indicate that peri-prosthetic femoral fracture is a significant mode of failure in the long term after the insertion of an uncemented CLS Spotorno stem. Revision rates for this fracture rise in the second decade. Further research is required to investigate the risk factors involved in the occurrence of late peri-prosthetic femoral fracture after the implantation of any uncemented stem, and to assess possible methods of prevention.
Revision after failed femoral components may
be technically demanding due to loss of peri-prosthetic bone. This retrospective
study evaluated the long-term results of femoral revision using
the cementless Wagner Self-Locking stem. Between 1992 and 1998,
68 consecutive hips in 66 patients underwent femoral revision using
this implant. A total of 25 patients died from unrelated causes
without further revision; the remaining 41 hips in 41 patients (12 men
and 29 women) with a mean age of 61 years (29 to 80) were reviewed
at a mean follow-up of 13.9 years (10.4 to 15.8). A transfemoral
approach was used in 32 hips. A total of five stems required further
revision because of infection in two, progressive subsidence in
two and recurrent dislocation in one. Four hips had dislocated and
eight stems had subsided ≥ 10 mm. The mean Harris hip score improved
from 33 points pre-operatively to 75 points at final follow-up (p
<
0.001). In all, 33 stems (91.7%) showed radiological signs
of stable bone fixation. The cumulative survival rates at 15.8 years
with femoral revision for any reason and for stem failure as the
endpoints were 92.0% (95% confidence interval (CI) 86.0% to 98.4%)
and 96.6% (95% CI 92.2% to 100%), respectively. The survivorship
with revision and ≥ 10 mm migration of the stem as the endpoint
was 83.6% (95% CI 76.6% to 91.4%). This study shows quite good survival and moderate clinical outcome
when using a monoblock tapered titanium stem for supporting the
regeneration of bone in complex revision hip surgery.
We have evaluated the difference in the migration patterns over two years of two cementless stems in a randomised, controlled trial using radiostereophotogrammetric analysis (RSA). The implants studied were the Furlong HAC stem, which has good long-term results and the Furlong Active stem, which is a modified version of the former designed to minimise stress concentrations between the implant and bone, and thus to improve fixation. A total of 23 Furlong HAC and 20 Furlong Active stems were implanted in 43 patients. RSA examinations were carried out immediately post-operatively and at six, 12 and 24 months post-operatively. The subsidence during the first year in the Furlong HAC stem, was approximately one-third that of the Furlong Active stem, the measured mean subsidence of the femoral head at six months being 0.27 mm (95% confidence interval (CI) 0.03 to 0.51) and 0.99 mm (95% CI 0.38 to 1.60), respectively (p = 0.03). One Active stem continued to subside during the second year. All hips, regardless of the type of stem were clinically successful as judged by the Oxford hip score and a derived pain score without any distinction between the two types of stem. The initial stability of the Furlong Active stem was not as good as the established stem which might compromise osseo-integration to the detriment of long-term success. The changes in the geometry of the stem, to minimise stress have affected the attainment of initial stability.
Endoprosthetic reconstruction following resection of 31 tumours of the proximal femur in 30 patients was performed using a Wagner SL femoral revision stem. The mean follow-up was 25.6 months (0.6 to 130.0). Of the 28 patients with a metastasis, 27 died within a mean follow-up period of 18.1 months (0.6 to 56.3) after the operation, and the remaining patient was excluded from the study 44.4 months post-operatively when the stem was removed. The two patients with primary bone tumours were still alive at the latest follow-up of 81.0 and 130.0 months, respectively. One stem only was removed for suspected low-grade infection 44.4 months post-operatively. The worst-case survival rate with removal of the stem for any cause and/or loss to follow-up was 80.0% (95% confidence interval 44.9 to 100) at 130.0 months. The mean Karnofsky index increased from 44.2% (20% to 70%) pre-operatively to 59.7% (0% to 100%) post-operatively, and the mean Merle d’Aubigné score improved from 4.5 (0 to 15) to 12.0 (0 to 18). The mean post-operative Musculoskeletal Tumour Society score was 62.4% (3.3% to 100%). The Wagner SL femoral revision stem offers an alternative to special tumour prostheses for the treatment of primary and secondary tumours of the proximal femur. The mid-term results are very promising, but long-term experience is necessary.
The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite coating in a series of 52 men (six bilateral) and 40 women (two bilateral), has been described previously. None required revision for aseptic loosening. We have extended the follow-up to 20 years with a minimum of 17 years. The mean age of the patients at total hip replacement was 58.9 years (19 to 84). Six patients were lost to follow-up, but were included up to their last clinical review. A total of 22 patients (22 hips) had died, all from causes unrelated to their surgery. There have been 43 re-operations for failure of the acetabular component. However, in 38 of these the stem was not revised since it remained stable and there was no associated osteolysis. Two of the revisions were for damage to the trunnion after fracture of a modular ceramic head, and in another two, removal of the femoral component was because of the preference of the surgeon. In all cases the femoral component was well fixed, but could be extracted at the time of acetabular revision. In one case both components were revised for deep infection. There has been one case of aseptic loosening of the stem which occurred at 14 years. This stem had migrated distally by 7.6 mm in ten years and 8.4 mm at the time of revision at which stage it was found to be rotationally loose. With hindsight this component had been undersized at implantation. The survivorship for the stem at 17 years with aseptic loosening as the endpoint was 98.6% (95% confidence interval 95.9 to 100) when 62 hips were at risk. All remaining stems had a satisfactory clinical and radiological outcome. The Freeman proximally hydroxyapatite-coated femoral component is therefore a dependable implant and its continued use can be recommended.
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.
Second-generation metal-on-metal bearings were introduced as a response to the considerable incidence of wear-induced failures associated with conventional replacements, especially in young patients. We present the results at ten years of a consecutive series of patients treated using a metal-on-metal hip resurfacing. A distinct feature of the bearings used in our series was that they had been subjected to double-heat treatments during the post-casting phase of their manufacture. In the past these bearings had not been subjected to thermal treatments, making this a unique metal-on-metal bearing which had not been used before in clinical practice. We report the outcome of 184 consecutive hips (160 patients) treated using a hybrid-fixed metal-on-metal hip resurfacing during 1996. Patients were invited for a clinicoradiological follow-up at a minimum of ten years. The Oxford hip score and anteroposterior and lateral radiographs were obtained. The mean age at operation was 54 years (21 to 75). A series of 107 consecutive hips (99 patients) who received the same prosthesis, but subjected to a single thermal treatment after being cast, between March 1994 and December 1995, were used as a control group for comparison. In the 1994 to 1995 group seven patients (seven hips) died from unrelated causes and there were four revisions (4%) for osteolysis and aseptic loosening. In the 1996 group nine patients died at a mean of 6.9 years after operation because of unrelated causes. There were 30 revisions (16%) at a mean of 7.3 years (1.2 to 10.9), one for infection at 1.2 years and 29 for osteolysis and aseptic loosening. Furthermore, in the latter group there were radiological signs of failure in 27 (24%) of the 111 surviving hips. The magnitude of the problem of osteolysis and aseptic loosening in the 1996 cohort did not become obvious until five years after the operation. Our results indicate that double-heat treatments of metal-on-metal bearings can lead to an increased incidence of wear-induced osteolysis.
Using meta-analysis we compared the survival and clinical outcomes of cemented and uncemented techniques in primary total knee replacement. We reviewed randomised controlled trials and observational studies comparing cemented and uncemented fixation. Our primary outcome was survival of the implant free of aseptic loosening. Our secondary outcome was joint function as measured by the Knee Society score. We identified 15 studies that met our final eligibility criteria. The combined odds ratio for failure of the implant due to aseptic loosening for the uncemented group was 4.2 (95% confidence interval (CI) 2.7 to 6.5) (p <
0.0001). Subgroup analysis of data only from randomised controlled trials showed no differences between the groups for odds of aseptic loosening (odds ratio 1.9, 95% CI 0.55 to 6.40, p = 0.314). The weighted mean difference for the Knee Society score was 0.005 (95% CI −0.26 to 0.26) (p = 0.972). There was improved survival of the cemented compared to uncemented implants, with no statistically significant difference in the mean Knee Society score between groups for all pooled data.
This prospective study presents the ten-year (5 to 16) clinical and radiological results of 55 primary total hip replacements (THR) using a cementless modular femoral component (S-ROM). All patients had a significant anatomical abnormality which rendered the primary THR difficult. The mean Harris hip score was 36 (12 to 72) pre-operatively, 83 (44 to 100) at five years, and 85 (45 to 99) at ten years. The Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and short-form (SF)-12 scores were recorded from the year 2000. The mean SF-12 score at five years after surgery was 45.24 (22.74 to 56.58) for the physical component and 54.14 (29.20 to 66.61) for the mental component. By ten years the SF-12 scores were 42.86 (21.59 to 58.95) and 51.03 (33.78 to 61.40), respectively. The mean WOMAC score at five years post-operatively was 25 (0 to 59), and at ten years was 27 (2 to 70). No femoral components were radiologically loose, although five had osteolysis in Gruen zone 1, three had osteolysis in zone 7, and two showed osteolysis in both zones 1 and 7. No osteolysis was observed around or distal to the prosthetic sleeve. No femoral components were revised, although three hips underwent an acetabular revision and two required a liner exchange. At a mean of ten years’ follow-up the S-ROM femoral component implanted for an anatomically difficult primary THR has excellent clinical and radiological results.
We report the long-term outcome of a modified second-generation cementing technique for fixation of the acetabular component of total hip replacement. An earlier report has shown the superiority of this technique assessed by improved survival compared with first-generation cementing. The acetabular preparation involved reaming only to the subchondral plate, followed by impaction of the bone in the anchorage holes. Between 1978 and 1993, 287 total hip replacements were undertaken in 244 patients with a mean age of 65.3 years (21 to 90) using a hemispherical Weber acetabular component with this modified technique for cementing and a cemented femoral component. The survival with acetabular revision for aseptic loosening as the endpoint was 99.1% (95% confidence interval 97.9 to 100 after ten years and 85.5% (95% confidence interval 74.7 to 96.2) at 20 years. Apart from contributing to a long-lasting fixation of the component, this technique also preserved bone, facilitating revision surgery when necessary.
A clinical and radiological study was conducted on 97 total hip replacements performed for congenital hip dislocation in 79 patients between 1989 and 1998 using a three-dimensional custom-made cementless stem. The mean age at operation was 48 years (17 to 72) and the mean follow-up was for 123 months (83 to 182). According to the Crowe classification, there were 37 class I, 28 class II, 13 class III and 19 class IV hips. The mean leg lengthening was 25 mm (5 to 58), the mean pre-operative femoral anteversion was 38.6° (2° to 86°) and the mean correction in the prosthetic neck was −23.6° (−71° to 13°). The mean Harris hip score improved from 58 (15 to 84) to 93 (40 to 100) points. A revision was required in six hips (6.2%). The overall survival rate was 89.5% (95% confidence interval 89.2 to 89.8) at 13 years when two hips were at risk. This custom-made cementless femoral component, which can be accommodated in the abnormal proximal femur and will correct the anteversion and frontal offset, provided good results without recourse to proximal femoral corrective osteotomy.
We retrospectively reviewed, ten years after surgery, 100 consecutive total hip replacements in which the Duraloc 300 cup had been used. Post-operative radiographs were analysed for placement of the cup and interface gaps and follow-up radiographs for lucent lines, osteolysis, wear and migration. All the components were found to be stable with no evidence of loosening. The mean rate of wear was 0.12 mm/year. Three hips developed acetabular osteolysis at the level of the apex hole. Two have successfully undergone bone grafting without removal of the implants and one patient is awaiting surgery. The Duraloc 300 cup has a survival of 100% at ten years with no aseptic loosening and a low incidence of pelvic osteolysis.