The mechanisms controlling and limiting movement and serving to transmit load between the femur and the tibia are discussed. Having accounted for the transmission of all components of force and couple across the joint and noted the load-bearing role of the menisci, some principles which might guide the design of knee prostheses are deduced. It is shown that current designs transgress some of these principles. An experimental prosthesis is then described, which incorporates analogues of the natural menisci. The possible practical application of this novel principle has been studied in cadaveric human joints and in living patients.
The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad).Aims
Methods
The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.Aims
Materials and Methods
The aim of this study was to compare the survivorship and radiographic
outcomes at ten-year follow-up of three prospective consecutive
series of patients each of which received a different design of
cementless femoral components for total hip arthroplasty (THA). In Cohort 1, 91 consecutive patients (100 hips) underwent THA
with a cementless porous-coated anatomic femoral stem (PCA) between
October 1983 and January 1986. In Cohort 2, 86 consecutive patients
(100 hips) underwent THA with an extensively porous-coated cementless
femoral stem (Prodigy) between June 1994 and October 1997. In Cohort
3, 88 consecutive patients (100 hips) underwent THA with a proximally
porous-coated triple-tapered cementless stem (Summit) between April
2002 and October 2003. All three groups underwent prospective clinical
and radiographic evaluation.Aims
Patients and Methods
Progressive retroversion of a cemented stem is
predictive of early loosening and failure. We assessed the relationship
between direct post-operative stem anteversion, measured with CT,
and the resulting rotational stability, measured with repeated radiostereometric
analysis over ten years. The study comprised 60 cemented total hip
replacements using one of two types of matt collared stem with a
rounded cross-section. The patients were divided into three groups
depending on their measured post-operative anteversion (<
10°,
10° to 25°, >
25°). There was a strong correlation between direct
post-operative anteversion and later posterior rotation. At one
year the <
10° group showed significantly more progressive retroversion
together with distal migration, and this persisted to the ten-year
follow-up. In the <
10° group four of ten stems (40%) had been
revised at ten years, and an additional two stems (20%) were radiologically
loose. In the ‘normal’ (10° to 25°) anteversion group there was
one revised (3%) and one loose stem (3%) of a total of 30 stems,
and in the >
25° group one stem (5%) was revised and another loose (5%)
out of 20 stems. This poor outcome is partly dependent on the design
of this prosthesis, but the results strongly suggest that the initial
rotational position of cemented stems during surgery affects the
subsequent progressive retroversion, subsidence and eventual loosening.
The degree of retroversion may be sensitive to prosthetic design
and stem size, but <
10° of anteversion appears deleterious to
the long-term outcome for cemented hip prosthetic stems. Cite this article:
Aims. Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years. Materials and Methods. We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single
Literature surrounding artificial intelligence (AI)-related applications for hip and knee arthroplasty has proliferated. However, meaningful advances that fundamentally transform the practice and delivery of joint arthroplasty are yet to be realized, despite the broad range of applications as we continue to search for meaningful and appropriate use of AI. AI literature in hip and knee arthroplasty between 2018 and 2021 regarding image-based analyses, value-based care, remote patient monitoring, and augmented reality was reviewed. Concerns surrounding meaningful use and appropriate methodological approaches of AI in joint arthroplasty research are summarized. Of the 233 AI-related orthopaedics articles published, 178 (76%) constituted original research, while the rest consisted of editorials or reviews. A total of 52% of original AI-related research concerns hip and knee arthroplasty (n = 92), and a narrative review is described. Three studies were externally validated. Pitfalls surrounding present-day research include conflating vernacular (“AI/machine learning”), repackaging limited registry data, prematurely releasing internally validated prediction models, appraising model architecture instead of inputted data, withholding code, and evaluating studies using antiquated regression-based guidelines. While AI has been applied to a variety of hip and knee arthroplasty applications with limited clinical impact, the future remains promising if the question is meaningful, the methodology is rigorous and transparent, the data are rich, and the model is externally validated. Simple checkpoints for meaningful AI adoption include ensuring applications focus on: administrative support over clinical evaluation and management; necessity of the advanced model; and the novelty of the question being answered. Cite this article:
This study compared patient-reported outcomes of three total knee arthroplasty (TKA) designs from one manufacturer: one cruciate-retaining (CR) design, and two cruciate-sacrificing designs, anterior-stabilized (AS) and posterior-stabilized (PS). Patients scheduled for primary TKA were included in a single-centre, prospective, three-armed, blinded randomized trial (n = 216; 72 per group). After intraoperative confirmation of posterior cruciate ligament (PCL) integrity, patients were randomly allocated to receive a CR, AS, or PS design from the same TKA system. Insertion of an AS or PS design required PCL resection. The primary outcome was the mean score of all five subscales of the Knee injury and Osteoarthritis Outcome Score (KOOS) at two-year follow-up. Secondary outcomes included all KOOS subscales, Oxford Knee Score, EuroQol five-dimension health questionnaire, EuroQol visual analogue scale, range of motion (ROM), and willingness to undergo the operation again. Patient satisfaction was also assessed.Aims
Methods
The aim of this study was to estimate the clinical and economic burden of dislocation following primary total hip arthroplasty (THA) in England. This retrospective evaluation used data from the UK Clinical Practice Research Datalink database. Patients were eligible if they underwent a primary THA (index date) and had medical records available 90 days pre-index and 180 days post-index. Bilateral THAs were excluded. Healthcare costs and resource use were evaluated over two years. Changes (pre- vs post-THA) in generic quality of life (QoL) and joint-specific disability were evaluated. Propensity score matching controlled for baseline differences between patients with and without THA dislocation.Aims
Methods
The aim of this study was to investigate the outcome of periprosthetic fractures of the humerus and to assess the uniformity of the classifications used for these fractures (including those around elbow and/or shoulder arthroplasties) by performing a systematic review of the literature. A systematic search was conducted using the National Institute for Health and Care Excellence Healthcare Databases Advance Search. For inclusion, studies had to report clinical outcomes following the management of periprosthetic fractures of the humerus. The protocol was registered on the PROSPERO database.Aims
Methods
The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection. From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.Aims
Methods
We present a patient who underwent delayed sub-periosteal hemipelvectomy for control of infection and to enable soft-tissue cover after trauma. At four months after amputation, clinical examination and radiographs demonstrated almost complete re-ossification of the hemipelvis. This has allowed the patient to regain sitting balance and to use a walking
The use and variety of stemless humeral components in anatomical total shoulder arthroplasty (TSA) have proliferated since their advent in 2004. Early outcomes are reassuring but independent mid-term results are scarce. This independent study reports a consecutive series of 143 Eclipse stemless shoulder prostheses with a minimum five-year (5 to 10) follow-up. Outcomes of 143 procedures undertaken for all indications in 131 patients were reviewed, with subset analysis of those for osteoarthritis (OA) (n = 99). The primary outcome was the Oxford Shoulder Score (OSS) at a minimum of five years. Secondary outcomes were ranges of motion and radiological analysis of humeral radiolucency, rotator cuff failure, and glenoid loosening.Aims
Methods
Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up.Aims
Methods
This systematic review asked which patterns of complications are associated with the three reverse total shoulder arthroplasty (RTSA) prosthetic designs, as classified by Routman et al, in patients undergoing RTSA for the management of cuff tear arthropathy, massive cuff tear, osteoarthritis, and rheumatoid arthritis. The three implant design philosophies investigated were medial glenoid/medial humerus (MGMH), medial glenoid/lateral humerus (MGLH), and lateral glenoid/medial humerus (LGMH). A systematic review of the literature was performed via a search of MEDLINE and Embase. Two reviewers extracted data on complication occurrence and patient-reported outcome measures (PROMs). Meta-analysis was conducted on the reported proportion of complications, weighted by sample size, and PROMs were pooled using the reported standardized mean difference (SMD). Quality of methodology was assessed using Wylde’s non-summative four-point system. The study was registered with PROSPERO (CRD42020193041).Aims
Methods
The results of proximal humeral replacement following trauma are substantially worse than for osteoarthritis or rheumatoid arthritis. The stable reattachment of the lesser and greater tuberosity fragments to the rotator cuff and the restoration of shoulder biomechanics are difficult. In 1992 we developed a
Modern total knee arthroplasty (TKA) prostheses are designed to restore near normal kinematics including high flexion. Kneeling is a high flexion, kinematically demanding activity after TKA. The debate about design choice has not yet been informed by six-degrees-of-freedom in vivo kinematics. This prospective randomized clinical trial compared kneeling kinematics in three TKA designs. In total, 68 patients were randomized to either a posterior stabilized (PS-FB), cruciate-retaining (CR-FB), or rotating platform (CR-RP) design. Of these patients, 64 completed a minimum one year follow-up. Patients completed full-flexion kneeling while being imaged using single-plane fluoroscopy. Kinematics were calculated by registering the 3D implant models onto 2D-dynamic fluoroscopic images and exported for analysis.Aims
Methods
The combination of an irreparable tear of the rotator cuff and destructive arthritis of the shoulder joint may cause severe pain, disability and loss of independence in the aged. Standard anatomical shoulder replacements depend on a functioning rotator cuff, and hence may fail in the presence of tears in the cuff. Many designs of non-anatomical constrained or semi-constrained prostheses have been developed for cuff tear arthropathy, but have proved unsatisfactory and were abandoned. The DePuy Delta III reverse
The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS). A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively.Aims
Methods
Many knee replacement prostheses, embodying various principles of design, are now available and there is need for a method by which valid comparisons of results can be made. An important criterion of success is durability, so the length of time the prostheses have been in situ must be taken into account. Such a method is proposed here and is applied to the results of 673 knee replacements, of nine different types, implanted at the same hospital between 1970 and 1983. A prosthesis was considered to have failed if it had been removed or persistently caused severe pain. Two types of prosthesis were found to be significantly less successful than the other seven, between which none consistently showed significant superiority. Results for the seven types were similar despite the facts that they had been used for knees with different degrees of damage, some as secondary implants, and that they were of different design and at different stages of technical development. The more recently introduced types of