As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach. A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30).Aims
Methods
Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages.Aims
Methods
Aims. The aim of this retrospective cohort study was to assess and investigate the safety and efficacy of using a distal tibial osteotomy compared to proximal osteotomy for limb lengthening in children. Methods. In this study, there were 59 consecutive
The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device. This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.Aims
Methods
To compare changes in gait kinematics and walking speed 24 months after conventional (C-MLS) and minimally invasive (MI-MLS) multilevel surgery for children with diplegic cerebral palsy (CP). A retrospective analysis of 19 children following C-MLS, with mean age at surgery of 12 years five months (seven years ten months to 15 years 11 months), and 36 children following MI-MLS, with mean age at surgery of ten years seven months (seven years one month to 14 years ten months), was performed. The Gait Profile Score (GPS) and walking speed were collected preoperatively and six, 12 and 24 months postoperatively. Type and frequency of procedures as part of MLS, surgical adverse events, and subsequent surgery were recorded.Aims
Methods
Preoperative talar valgus deformity ≥ 15° is considered a contraindication for total ankle arthroplasty (TAA). We compared operative procedures and clinical outcomes of TAA in patients with talar valgus deformity ≥ 15° and < 15°. A matched cohort of patients similar for demographics and components used but differing in preoperative coronal-plane tibiotalar valgus deformity ≥ 15° (valgus, n = 50; 52% male, mean age 65.8 years (SD 10.3), mean body mass index (BMI) 29.4 (SD 5.2)) or < 15° (control, n = 50; 58% male, mean age 65.6 years (SD 9.8), mean BMI 28.7 (SD 4.2)), underwent TAA by one surgeon. Preoperative and postoperative radiographs, Ankle Osteoarthritis Scale (AOS) pain and disability and 36-item Short Form Health Survey (SF-36) version 2 scores were collected prospectively. Ancillary procedures, secondary procedures, and complications were recorded.Aims
Methods
Aims. Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator. Patients and Methods. A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had
The aim of this study was to present a series of patients with
aseptic failure of a total ankle arthroplasty (TAA) who were treated
with fusion of the hindfoot using a nail. A total of 23 TAAs, in 22 patients, were revised for aseptic
loosening and balloon osteolysis to a hindfoot fusion by a single
surgeon (NH) between January 2012 and August 2014. The procedure
was carried out without bone graft using the Phoenix, Biomet Hindfoot
Arthrodesis Nail. Preoperative investigations included full blood
count, CRP and ESR, and radiological investigations including plain
radiographs and CT scans. Postoperative plain radiographs were assessed
for fusion. When there was any doubt, CT scans were performed.Aims
Patients and Methods
The aim of this double-blind prospective randomised controlled
trial was to assess whether low intensity pulsed ultrasound (LIPUS)
accelerated or enhanced the rate of bone healing in adult patients
undergoing distraction osteogenesis. A total of 62 adult patients undergoing limb lengthening or bone
transport by distraction osteogenesis were randomised to treatment
with either an active (n = 32) or a placebo (n = 30) ultrasound
device. A standardised corticotomy was performed in the proximal
tibial metaphysis and a circular Ilizarov frame was used in all
patients. The rate of distraction was also standardised. The primary
outcome measure was the time to removal of the frame after adjusting
for the length of distraction in days/cm for both the per protocol
(PP) and the intention-to-treat (ITT) groups. The assessor was blinded
to the form of treatment. A secondary outcome was to identify covariates affecting
the time to removal of the frame.Aims
Patients and Methods
To present our experience of using a combination of intra-articular
osteotomy and external fixation to treat different deformities of
the knee. A total of six patients with a mean age of 26.5 years (15 to
50) with an abnormal hemi-joint line convergence angle (HJLCA) and
mechanical axis deviation (MAD) were included. Elevation of a tibial
hemiplateau or femoral condylar advancement was performed and limb
lengthening with correction of residual deformity using a circular
or monolateral Ilizarov frame.Aims
Patients and Methods
The last decade has seen a considerable increase
in the use of in total ankle arthroplasty (TAA) to treat patients
with end-stage arthritis of the ankle. However, the longevity of
the implants is still far from that of total knee and hip arthroplasties. The aim of this review is to outline a diagnostic and treatment
algorithm for the painful TAA to be used when considering revision
surgery. Cite this article:
Aims. Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to
Aims. We wished to examine the effectiveness of tibial lengthening
using a two ring Ilizarov frame in skeletally immature patients.
This is a potentially biomechanically unstable construct which risks
the loss of axial control. Patients and Methods. We retrospectively reviewed a consecutive series of 24 boys and
26 girls, with a mean age of 8.6 years (4 to 14), who underwent
52
Two types of fracture, early and late, have been
reported following limb lengthening in patients with achondroplasia (ACH)
and hypochondroplasia (HCH). We reviewed 25 patients with these conditions who underwent 72
segmental limb lengthening procedures involving the femur and/or
tibia, between 2003 and 2011. Gender, age at surgery, lengthened
segment, body mass index, the shape of the callus, the amount and
percentage of lengthening and the healing index were evaluated to determine
predictive factors for the occurrence of early (within three weeks
after removal of the fixation pins) and late fracture (>
three weeks
after removal of the pins). The Mann‑Whitney U test and Pearson’s
chi-squared test for univariate analysis and stepwise regression
model for multivariate analysis were used to identify the predictive factor
for each fracture. Only one patient (two tibiae) was excluded from
the analysis due to excessively slow formation of the regenerate,
which required supplementary measures. A total of 24 patients with
70 limbs were included in the study. There were 11 early fractures in eight patients. The shape of
the callus (lateral or central callus) was the only statistical
variable related to the occurrence of early fracture in univariate
and multivariate analyses. Late fracture was observed in six limbs
and the mean time between removal of the fixation pins and fracture
was 18.3 weeks (3.3 to 38.4). Lengthening of the tibia, larger healing
index, and lateral or central callus were related to the occurrence
of a late fracture in univariate analysis. A multivariate analysis
demonstrated that the shape of the callus was the strongest predictor
for late fracture (odds ratio: 19.3, 95% confidence interval: 2.91
to 128). Lateral or central callus had a significantly larger risk
of fracture than fusiform, cylindrical, or concave callus. Radiological monitoring of the shape of the callus during distraction
is important to prevent early and late fracture of lengthened limbs
in patients with ACH or HCH. In patients with thin callus formation,
some measures to stimulate bone formation should be considered as
early as possible. Cite this article:
Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent
This study evaluated the effect of limb lengthening
on longitudinal growth in patients with achondroplasia. Growth of
the lower extremity was assessed retrospectively by serial radiographs
in 35 skeletally immature patients with achondroplasia who underwent
bilateral limb lengthening (Group 1), and in 12 skeletally immature
patients with achondroplasia who did not (Group 2). In Group 1,
23 patients underwent only
Lengthening of the humerus is now an established
technique. We compared the complications of humeral lengthening
with those of femoral lengthening and investigated whether or not
the callus formation in the humerus proceeds at a higher rate than
that in the femur. A total of 24 humeral and 24 femoral lengthenings
were performed on 12 patients with achondroplasia. We measured the
pixel value ratio (PVR) of the lengthened area on radiographs and
each radiograph was analysed for the shape, type and density of
the callus. The quality of life (QOL) of the patients after humeral
lengthening was compared with that prior to surgery. The complication
rate per segment of humerus and femur was 0.87% and 1.37%, respectively.
In the humerus the PVR was significantly higher than that of the
femur. Lower limbs were associated with an increased incidence of
concave, lateral and central callus shapes. Humeral lengthening
had a lower complication rate than lower-limb lengthening, and QOL
increased significantly after humeral lengthening. Callus formation
in the humerus during the distraction period proceeded at a significantly
higher rate than that in the femur. These findings indicate that humeral lengthening has an important
role in the management of patients with achondroplasia.
Septicaemia resulting from meningococcal infection is a devastating illness affecting children. Those who survive can develop late orthopaedic sequelae from growth plate arrests, with resultant complex deformities. Our aim in this study was to review the case histories of a series of patients with late orthopaedic sequelae, all treated by the senior author (CFB). We also describe a treatment strategy to address the multiple deformities that may occur in these patients. Between 1997 and 2009, ten patients (seven girls and three boys) were treated for late orthopaedic sequelae following meningococcal septicaemia. All had involvement of the lower limbs, and one also had involvement of the upper limbs. Each patient had a median of three operations (one to nine). Methods of treatment included a combination of angular deformity correction, limb lengthening and epiphysiodesis. All patients were skeletally mature at the final follow-up. One patient with bilateral below-knee amputations had satisfactory correction of her right amputation stump deformity, and has complete ablation of both her proximal tibial growth plates. In eight patients length discrepancy in the lower limb was corrected to within 1 cm, with normalisation of the mechanical axis of the lower limb. Meningococcal septicaemia can lead to late orthopaedic sequelae due to growth plate arrests. Central growth plate arrests lead to limb-length discrepancy and the need for lengthening procedures, and peripheral growth plate arrests lead to angular deformities requiring corrective osteotomies and ablation of the damaged physis. In addition, limb amputations may be necessary and there may be altered growth of the stump requiring further surgery. Long-term follow-up of these patients is essential to recognise and treat any recurrence of deformity.
Survivors of infantile meningococcal septicaemia often develop progressive skeletal deformity as a result of physeal damage at many sites, particularly in the lower limb. Distal tibial physeal arrest typically occurs with sparing of the distal fibular physis leading to a rapidly progressive varus deformity. There have been reports of isolated cases of this deformity, but to our knowledge there have been no papers which specifically describe the development of the deformity and the options for treatment. Surgery to correct this deformity is complex because of the patient’s age, previous scarring and the multiplanar nature of the deformity. The surgical goal is to restore leg-length equality and the mechanical axis at the end of growth. Surgery should be planned and staged throughout growth in order to achieve the best functional results. We report our experience in six patients (seven ankles) with this deformity, who were managed by corrective osteotomy using a programmable circular fixator.
We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties. Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful