Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 504 - 511
1 Apr 2016
Ajami S Blunn GW Lambert S Alexander S Foxall Smith M Coathup MJ

Aims

To assess the extent of osteointegration in two designs of shoulder resurfacing implants. Bony integration to the Copeland cylindrical central stem design and the Epoca RH conical-crown design were compared.

Patients and Methods

Implants retrieved from six patients in each group were pair-matched. Mean time to revision surgery of Copeland implants was 37 months (standard deviation (sd) 23; 14 to 72) and Epoca RH 38 months (sd 28; 12 to 84). The mean age of patients investigated was 66 years (sd 4; 59 to 71) and 58 years (sd 17; 31 to 73) in the Copeland and Epoca RH groups respectively. None of these implants were revised for loosening.


The Bone & Joint Journal
Vol. 96-B, Issue 8 | Pages 1077 - 1081
1 Aug 2014
Nuttall D Birch A Haines JF Trail IA

Resurfacing of the humeral head is commonly used within the UK to treat osteoarthritis (OA) of the shoulder. We present the results of a small prospective randomised study of this procedure using the Global CAP prosthesis with two different coatings, Porocoat and DuoFix hydroxyapatite (HA). We followed two groups of ten patients with OA of the shoulder for two years after insertion of the prosthesis with tantalum marker beads, recording pain, Constant–Murley and American Shoulder and Elbow Surgeons (ASES) outcome scores, and using radiostereometric analysis to assess migration. The outcomes were similar to those of other series, with significant reductions in pain (p = 0.003) and an improvement in the Constant (p = 0.001) and ASES scores (p = 0.006). The mean migration of the prosthesis three months post-operatively was 0.78 mm (0.51 to 1.69) and 0.72 mm (0.33 to 1.45) for the Porocoat and DuoFix groups, respectively. Analysis of variance indicated that the rate of migration reached a plateau after three months post-operatively in both groups. At follow-up of two years the mean migration was 1 mm (sd 1 (0.25 to 3.32)); in the Porocoat group and 0.8 mm (sd 0.4 (0.27 to 1.45)) in the DuoFix HA group. Significant migration of the prosthesis was seen in one patient who had received an anterior humeral bone graft. This prosthesis was later revised after 2.7 years.

The addition of a coating of HA to the sintered surface does not improve fixation of this prosthesis.

Cite this article: Bone Joint J 2014;96-B:1077–81.