Aims. Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants –
Aims. Revision total hip arthroplasty in patients with Vancouver type B3 fractures with Paprosky type IIIA, IIIB, and IV femoral defects are difficult to treat. One option for Paprovsky type IIIB and IV defects involves
Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the
The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes. We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).Aims
Methods
Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants. A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed.Aims
Methods
We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions. Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour.Aims
Methods
Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.Aims
Methods
We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed. All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation.Aims
Methods
Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results. We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (Aims
Patients and Methods
Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR. Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.Aims
Patients and Methods
Recurrent infection following two-stage revision total hip arthroplasty
(THA) for prosthetic joint infection (PJI) is a devastating complication.
The purpose of this study was to report the survival of repeat two-stage
revision hip arthroplasty, describe complications, and identify
risk factors for failure. We retrospectively identified 19 hips (19 patients) that had
undergone repeat two-stage revision THA for infection between 2000
to 2013. There were seven female patients (37%) and the mean age
was 60 years (30 to 85). Survival free from revision was assessed
via Kaplan–Meier analysis. The patients were classified according
to the Musculoskeletal Infection Society (MSIS) system, and risk
factors for failure were identified. Mean follow-up was four years
(2 to 11).Aims
Patients and Methods
This study aimed to compare the change in health-related quality
of life of patients receiving a traditional cemented monoblock Thompson
hemiarthroplasty compared with a modern cemented modular polished-taper
stemmed hemiarthroplasty for displaced intracapsular hip fractures. This was a pragmatic, multicentre, multisurgeon, two-arm, parallel
group, randomized standard-of-care controlled trial. It was embedded
within the WHiTE Comprehensive Cohort Study. The sample size was
964 patients. The setting was five National Health Service Trauma
Hospitals in England. A total of 964 patients over 60 years of age who
required hemiarthroplasty of the hip between February 2015 and March
2016 were included. A standardized measure of health outcome, the
EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and
at four months following the operation.Aims
Patients and Methods
To evaluate the hypothesis that failed osteosynthesis of periprosthetic
Vancouver type B1 fractures can be treated successfully with stem
revision using a transfemoral approach and a cementless, modular,
tapered revision stem with reproducible rates of fracture healing,
stability of the revision stem, and clinically good results. A total of 14 patients (11 women, three men) with a mean age
of 72.4 years (65 to 90) undergoing revision hip arthroplasty after
failed osteosynthesis of periprosthetic fractures of Vancouver type
B1 were treated using a transfemoral approach to remove the well-fixed
stem before insertion of a modular, fluted titanium stem which obtained
distal fixation. These patients were clinically and radiologically
followed up for a mean 52.2 months (24 to 144).Aims
Patients and Methods
Loss or absence of proximal femoral bone in revision total hip
arthroplasty (THA) remains a significant challenge. While the main
indication for the use of proximal femoral replacements (PFRs) is
in the treatment of malignant disease, they have a valuable role
in revision THA for loosening, fracture and infection in patients
with bone loss. Our aim was to determine the clinical outcomes,
implant survivorship, and complications of PFRs used in revision
THA for indications other than malignancy. A retrospective review of 44 patients who underwent revision
THA using a PFR between 2000 and 2013 was undertaken. Their mean
age was 79 years (53 to 97); 31 (70%) were women. The bone loss
was classified as Paprosky IIIB or IV in all patients. The mean
follow-up was six years (2 to 12), at which time 22 patients had
died and five were lost to follow-up.Aims
Patients and Methods
There is increasing global awareness of adverse
reactions to metal debris and elevated serum metal ion concentrations
following the use of second generation metal-on-metal total hip
arthroplasties. The high incidence of these complications can be
largely attributed to corrosion at the head-neck interface. Severe
corrosion of the taper is identified most commonly in association
with larger diameter femoral heads. However, there is emerging evidence
of varying levels of corrosion observed in retrieved components
with smaller diameter femoral heads. This same mechanism of galvanic
and mechanically-assisted crevice corrosion has been observed in
metal-on-polyethylene and ceramic components, suggesting an inherent
biomechanical problem with current designs of the head-neck interface. We provide a review of the fundamental questions and answers
clinicians and researchers must understand regarding corrosion of
the taper, and its relevance to current orthopaedic practice. Cite this article:
The goals of this study were to define the risk factors, characteristics,
and chronology of fractures in 5417 revision total hip arthroplasties
(THAs). From our hospital’s prospectively collected database we identified
all patients who had undergone a revision THA between 1969 and 2011
which involved the femoral stem. The patients’ medical records and
radiographs were examined and the relevant data extracted. Post-operative
periprosthetic fractures were classified using the Vancouver system.
A total of 5417 revision THAs were identified.Aims
Patients and Methods