Advertisement for orthosearch.org.uk
Results 1 - 15 of 15
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 696 - 703
1 Apr 2021
Clough TM Ring J

Aims

We report the medium-term outcomes of a consecutive series of 118 Zenith total ankle arthroplasties (TAAs) from a single, non-designer centre.

Methods

Between December 2010 and May 2016, 118 consecutive Zenith prostheses were implanted in 114 patients. Demographic, clinical, and patient-reported outcome measures (PROMs) data were collected. The endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan-Meier survival curves were generated with 95% confidence intervals (CIs) and the rate of failure calculated for each year.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 423 - 429
1 Mar 2021
Diez-Escudero A Hailer NP

Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising coating for metallic surfaces, and several types of silver-coated arthroplasty implants are in clinical use today. However, silver can also exert toxic effects on eukaryotic cells both in the immediate vicinity of the coated implants and systemically. In most clinically-used implants, silver coatings are applied on bulk components that are not in direct contact with bone, such as in partial or total long bone arthroplasties used in tumour or complex revision surgery. These implants differ considerably in the coating method, total silver content, and silver release rates. Safety issues, such as the occurrence of argyria, have been a cause for concern, and the efficacy of silver coatings in terms of preventing PJI is also controversial. The application of silver coatings is uncommon on parts of implants intended for cementless fixation in host bone, but this option might be highly desirable since the modification of implant surfaces in order to improve osteoconductivity can also increase bacterial adhesion. Therefore, an optimal silver content that inhibits bacterial colonization while maintaining osteoconductivity is crucial if silver were to be applied as a coating on parts intended for bone contact. This review summarizes the different methods used to apply silver coatings to arthroplasty components, with a focus on the amount and duration of silver release from the different coatings; the available experience with silver-coated implants that are in clinical use today; and future strategies to balance the effects of silver on bacteria and eukaryotic cells, and to develop silver-coated titanium components suitable for bone ingrowth.

Cite this article: Bone Joint J 2021;103-B(3):423–429.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 39 - 45
1 Jan 2021
Fang X Cai Y Mei J Huang Z Zhang C Yang B Li W Zhang W

Aims

Metagenomic next-generation sequencing (mNGS) is useful in the diagnosis of infectious disease. However, while it is highly sensitive at identifying bacteria, it does not provide information on the sensitivity of the organisms to antibiotics. The purpose of this study was to determine whether the results of mNGS can be used to guide optimization of culture methods to improve the sensitivity of culture from intraoperative samples.

Methods

Between July 2014 and October 2019, patients with suspected joint infection (JI) from whom synovial fluid (SF) was obtained preoperatively were enrolled. Preoperative aspirated SF was analyzed by conventional microbial culture and mNGS. In addition to samples taken for conventional microbial culture, some samples were taken for intraoperative culture to optimize the culture method according to the preoperative mNGS results. The demographic characteristics, medical history, laboratory examination, mNGS, and culture results of the patients were recorded, and the possibility of the optimized culture methods improving diagnostic efficiency was evaluated.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 2 - 8
1 Jun 2019
Aggarwal VK Weintraub S Klock J Stachel A Phillips M Schwarzkopf R Iorio R Bosco J Zuckerman JD Vigdorchik JM Long WJ

Aims

We studied the impact of direct anterior (DA) versus non-anterior (NA) surgical approaches on prosthetic joint infection (PJI), and examined the impact of new perioperative protocols on PJI rates following all surgical approaches at a single institution.

Patients and Methods

A total of 6086 consecutive patients undergoing primary total hip arthroplasty (THA) at a single institution between 2013 and 2016 were retrospectively evaluated. Data obtained from electronic patient medical records included age, sex, body mass index (BMI), medical comorbidities, surgical approach, and presence of deep PJI. There were 3053 male patients (50.1%) and 3033 female patients (49.9%). The mean age and BMI of the entire cohort was 62.7 years (18 to 102, sd 12.3) and 28.8 kg/m2 (13.3 to 57.6, sd 6.1), respectively. Infection rates were calculated yearly for the DA and NA approach groups. Covariates were assessed and used in multivariate analysis to calculate adjusted odds ratios (ORs) for risk of development of PJI with DA compared with NA approaches. In order to determine the effect of adopting a set of infection prevention protocols on PJI, we calculated ORs for PJI comparing patients undergoing THA for two distinct time periods: 2013 to 2014 and 2015 to 2016. These periods corresponded to before and after we implemented a set of perioperative infection protocols.


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1262 - 1269
1 Sep 2016
Pinder EM Bottle A Aylin P Loeffler MD

Aims

To determine whether there is any difference in infection rate at 90 days between trauma operations performed in laminar flow and plenum ventilation, and whether infection risk is altered following the installation of laminar flow (LF).

Patients and Methods

We assessed the impact of plenum ventilation (PV) and LF on the rate of infection for patients undergoing orthopaedic trauma operations. All NHS hospitals in England with a trauma theatre(s) were contacted to identify the ventilation system which was used between April 2008 and March 2013 in the following categories: always LF, never LF, installed LF during study period (subdivided: before, during and after installation) and unknown. For each operation, age, gender, comorbidity, socio-economic deprivation, number of previous trauma operations and surgical site infection within 90 days (SSI90) were extracted from England’s national hospital administrative Hospital Episode Statistics database. Crude and adjusted odds ratios (OR) were used to compare ventilation groups using hierarchical logistic regression. Subanalysis was performed for hip hemiarthroplasties.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 461 - 467
1 Apr 2016
Abdel MP Watts CD Houdek MT Lewallen DG Berry DJ

Aim and Methods

The goals of this study were to define the risk factors, nature, chronology, and treatment strategies adopted for periprosthetic femoral fractures in 32 644 primary total hip arthroplasties (THAs).

Results

There were 564 intra-operative fractures (1.7%); 529 during uncemented stem placement (3.0%) and 35 during cemented stem placement (0.23%). Intra-operative fractures were more common in females and patients over 65 years (p < 0.001). The majority occurred during placement of the femoral component (60%), and involved the calcar (69%). There were 557 post-operative fractures (20-year probability: 3.5%; 95% confidence interval (CI) 3.2 to 3.9); 335 fractures after placement of an uncemented stem (20-year probability: 7.7%; 95% CI 6.2 to 9.1) and 222 after placement of a cemented stem (20-year probability: 2.1%; 95% CI 1.8 to 2.5). The probability of a post-operative fracture within 30 days after an uncemented stem was ten times higher than a cemented stem. The most common post-operative fracture type was a Vancouver AG (32%; n = 135), with 67% occurring after a fall. In all, 36% (n = 152) were treated with revision arthroplasty.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 442 - 451
1 Apr 2016
Lacny S Bohm E Hawker G Powell J Marshall DA

Aims

We aimed to assess the comparability of data in joint replacement registries and identify ways of improving the comparisons between registries and the overall monitoring of joint replacement surgery.

Materials and Methods

We conducted a review of registries that are full members of the International Society of Arthroplasty Registries with publicly available annual reports in English. Of the six registries which were included, we compared the reporting of: mean age, definitions for revision and re-operation, reasons for revision, the approach to analysing revisions, and patient-reported outcome measures (PROMs) for primary and revision total hip arthroplasty (THA) and hip resurfacing arthroplasty (HRA).


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 486 - 491
1 Apr 2014
Jämsen E Puolakka T Peltola M Eskelinen A Lehto MUK

We evaluated the duration of hospitalisation, occurrence of infections, hip dislocations, revisions, and mortality following primary hip and knee replacement in 857 patients with Parkinson’s disease and compared them with 2571 matched control patients. The data were collected from comprehensive nationwide Finnish health registers. The mean follow-up was six years (1 to 13). The patients with Parkinson’s disease had a longer mean length of stay (21 days [1 to 365] vs 13 [1 to 365] days) and an increased risk for hip dislocation during the first post-operative year (hazard ratio (HR) 2.33, 95% confidence intervals (CI) 1.02 to 5.32). There was no difference in infection and revision rates, and one-year mortality. In longer follow-up, patients with Parkinson’s disease had higher mortality (HR 1.94, 95% CI 1.68 to 2.25) and only 274 (34.7%) were surviving ten years after surgery. In patients with Parkinson’s disease, cardiovascular and psychiatric comorbidity were associated with prolonged hospitalisation and cardiovascular diseases also with increased mortality.

Cite this article: Bone Joint J 2014;96-B:486–91.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 1001 - 1006
1 Jul 2013
Esteban J Alvarez-Alvarez B Blanco A Fernández-Roblas R Gadea I Garcia-Cañete J Sandoval E Valdazo M

We have designed a prospective study to evaluate the usefulness of prolonged incubation of cultures from sonicated orthopaedic implants. During the study period 124 implants from 113 patients were processed (22 osteosynthetic implants, 46 hip prostheses, 54 knee prostheses, and two shoulder prostheses). Of these, 70 patients had clinical infection; 32 had received antibiotics at least seven days before removal of the implant. A total of 54 patients had sonicated samples that produced positive cultures (including four patients without infection). All of them were positive in the first seven days of incubation. No differences were found regarding previous antibiotic treatment when analysing colony counts or days of incubation in the case of a positive result. In our experience, extending incubation of the samples to 14 days does not add more positive results for sonicated orthopaedic implants (hip and knee prosthesis and osteosynthesis implants) compared with a conventional seven-day incubation period.

Cite this article: Bone Joint J 2013;95-B:1001–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 998 - 998
1 Jul 2011


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1407 - 1413
1 Nov 2008
Kontakis G Koutras C Tosounidis T Giannoudis P

We performed a comprehensive systematic review of the literature to examine the role of hemiarthroplasty in the early management of fractures of the proximal humerus. In all, 16 studies dealing with 810 hemiarthroplasties in 808 patients with a mean age of 67.7 years (22 to 91) and a mean follow-up of 3.7 years (0.66 to 14) met the inclusion criteria. Most of the fractures were four-part fractures or fracture-dislocations.

Several types of prosthesis were used. Early passive movement on the day after surgery and active movement after union of the tuberosities at about six weeks was described in most cases. The mean active anterior elevation was to 105.7° (10° to 180°) and the mean abduction to 92.4° (15° to 170°). The incidence of superficial and deep infection was 1.55% and 0.64%, respectively. Complications related to the fixation and healing of the tuberosities were observed in 86 of 771 cases (11.15%). The estimated incidence of heterotopic ossification was 8.8% and that of proximal migration of the humeral head 6.8%. The mean Constant score was 56.63 (11 to 98). At the final follow-up, no pain or only mild pain was experienced by most patients, but marked limitation of function persisted.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 194 - 199
1 Feb 2008
Sosna A Pokorny D Hromádka R Jahoda D Barták V Pinskerová V

The results of proximal humeral replacement following trauma are substantially worse than for osteoarthritis or rheumatoid arthritis. The stable reattachment of the lesser and greater tuberosity fragments to the rotator cuff and the restoration of shoulder biomechanics are difficult. In 1992 we developed a prosthesis designed to improve fixation of the tuberosity fragments in comminuted fractures of the proximal humerus. The implant enables fixation of the fragments to the shaft of the prosthesis and the diaphyseal fragment using screws, washers and a special toothed plate. Between 1992 and 2003 we used this technique in 50 of 76 patients referred to our institution for shoulder reconstruction after trauma. In the remaining 26, reconstruction with a prosthesis and nonabsorbable sutures was performed, as the tuberosity fragments were too small and too severely damaged to allow the use of screws and the toothed plate.

The Constant score two years post-operatively was a mean of 12 points better in the acute trauma group and 11 points better in the late post-traumatic group than in the classical suture group.

We recommend this technique in patients where the tuberosity fragments are large enough to allow fixation with screws, washers and a toothed plate.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 887 - 892
1 Jul 2007
Cohen AT Skinner JA Warwick D Brenkel I

Little is known about the efficacy of graduated compression stockings in preventing venous thromboembolism after hip surgery. We conducted a prospective, randomised single-blind study to determine whether the addition of compression stockings to fondaparinux conferred any additional benefit.

The study included 874 patients, of whom 795 could be evaluated (400 in the fondaparinux group and 395 in the fondaparinux plus compression stocking group). Fondaparinux was given post-operatively for five to nine days, either alone or combined with wearing stockings, which were worn for a mean 42 days (35 to 49).

The study outcomes were venous thromboembolism, or sudden death before day 42. Duplex ultrasonography was scheduled within a week of day 42. Safety outcomes were bleeding and death from venous thromboembolism.

The prevalence of deep-vein thrombosis was similar in the two groups 5.5% (22 of 400) in the fondaparinux group and 4.8 (19 of 395) in the fondaparinux plus stocking group (odds ratio 0.88, 95% confidence interval 0.46 to 1.65, p = 0.69). Major bleeding occurred in only one patient.

The addition of graduated compression stockings to fondaparinux appears to offer no additional benefit over the use of fondaparinux alone.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 716 - 720
1 Jun 2006
Hardoon SL Lewsey JD Gregg PJ Reeves BC van der Meulen JHP

New brands of joint prosthesis are released for general implantation with limited evidence of their long-term performance in patients. The CUSUM continuous monitoring method is a statistical testing procedure which could be used to provide prospective evaluation of brands as soon as implantation in patients begins and give early warning of poor performance. We describe the CUSUM and illustrate the potential value of this monitoring tool by applying it retrospectively to the 3M Capital Hip experience. The results show that if the clinical data and methodology had been available, the CUSUM would have given an alert to the underperformance of this prosthesis almost four years before the issue of a Hazard Notice by the Medical Devices Agency. This indicates that the CUSUM can be a valuable tool in monitoring joint prostheses, subject to timely and complete collection of data. Regional or national joint registries provide an opportunity for future centralised, continuous monitoring of all hip and knee prostheses using these techniques


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 950 - 954
1 Jul 2005
Khatri M Stirrat AN

We present the outcome of 47 Souter-Strathclyde replacements of the elbow with a mean follow-up of 82 months (12 to 129). The clinical results were assessed using a condition-specific outcome measure. The mean total score (maximum 100) before the operation was 47.21 and improved to 79.92 (p < 0.001). The mean pain score (maximum 50) improved from 21.41 to 46.70 (p < 0.001) and the mean functional component of the score (maximum 30) from 11.19 to 18.65 (p < 0.001). There was negligible change in the score for the range of movement although a significant improvement in mean flexion from 124° to 136° was noted (p < 0.001).

Revision surgery was required in four patients, for dislocation, wound dehiscence and early infection in one, late infection in two and aseptic loosening in one. The cumulative survival was 75% at nine years for all causes of failure and 97% at ten years for aseptic loosening alone. Our study demonstrates the value of the Souter-Strathclyde total elbow arthroplasty in providing relief from pain and functional improvement in rheumatoid patients.