The objective of this study was to compare simulated range of motion (ROM) for reverse total shoulder arthroplasty (rTSA) with and without adjustment for scapulothoracic orientation in a global reference system. We hypothesized that values for simulated ROM in preoperative planning software with and without adjustment for scapulothoracic orientation would be significantly different. A statistical shape model of the entire humerus and scapula was fitted into ten shoulder CT scans randomly selected from 162 patients who underwent rTSA. Six shoulder surgeons independently planned a rTSA in each model using prototype development software with the ability to adjust for scapulothoracic orientation, the starting position of the humerus, as well as kinematic planes in a global reference system simulating previously described posture types A, B, and C. ROM with and without posture adjustment was calculated and compared in all movement planes.Aims
Methods
Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey. NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.Aims
Methods
This study aimed to assess the impact of using the metal-augmented glenoid baseplate (AGB) on improving clinical and radiological outcomes, as well as reducing complications, in patients with superior glenoid wear undergoing reverse shoulder arthroplasty (RSA). From January 2016 to June 2021, out of 235 patients who underwent primary RSA, 24 received a superior-AGB after off-axis reaming (Group A). Subsequently, we conducted propensity score matching in a 1:3 ratio, considering sex, age, follow-up duration, and glenoid wear (superior-inclination and retroversion), and selected 72 well-balanced matched patients who received a standard glenoid baseplate (STB) after eccentric reaming (Group B). Superior-inclination, retroversion, and lateral humeral offset (LHO) were measured to assess preoperative glenoid wear and postoperative correction, as well as to identify any complications. Clinical outcomes were measured at each outpatient visit before and after surgery.Aims
Methods
The current evidence comparing the two most common approaches for reverse total shoulder arthroplasty (rTSA), the deltopectoral and anterosuperior approach, is limited. This study aims to compare the rate of loosening, instability, and implant survival between the two approaches for rTSA using data from the Dutch National Arthroplasty Registry with a minimum follow-up of five years. All patients in the registry who underwent a primary rTSA between January 2014 and December 2016 using an anterosuperior or deltopectoral approach were included, with a minimum follow-up of five years. Cox and logistic regression models were used to assess the association between the approach and the implant survival, instability, and glenoid loosening, independent of confounders.Aims
Methods
Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.Aims
Methods
The aim of this study was to longitudinally compare the clinical and radiological outcomes of anatomical total shoulder arthroplasty (aTSA) up to long-term follow-up, when using cemented keel, cemented peg, and hybrid cage peg glenoid components and the same humeral system. We retrospectively analyzed a multicentre, international clinical database of a single platform shoulder system to compare the short-, mid-, and long-term clinical outcomes associated with three designs of aTSA glenoid components: 294 cemented keel, 527 cemented peg, and 981 hybrid cage glenoids. Outcomes were evaluated at 4,746 postoperative timepoints for 1,802 primary aTSA, with a mean follow-up of 65 months (24 to 217).Aims
Methods
The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection. From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.Aims
Methods
Aims. The aim of this study was to report the incidence of implant-related complications, further operations, and their influence on the outcome in a series of patients who underwent primary reverse total shoulder arthroplasty (RTSA). Methods. The prospectively collected clinical and radiological data of 797 patients who underwent 854 primary RTSAs between January 2005 and August 2018 were analyzed. The hypothesis was that the presence of complications would adversely affect the outcome. Further procedures were defined as all necessary operations, including reoperations without change of components, and partial or total revisions. The clinical outcome was evaluated using the absolute and relative Constant Scores (aCS, rCS), the Subjective Shoulder Value (SSV) scores, range of motion, and pain. Results. The overall surgical site complication rate was 22% (188 complications) in 152 patients (156 RTSAs; 18%) at a mean follow-up of 46 months (0 to 169). The most common complications were acromial fracture (in 44 patients, 45 RTSAs; 5.3%), glenoid loosening (in 37 patients, 37 RTSAs; 4.3%), instability (in 23 patients, 23 RTSAs; 2.7%), humeral fracture or loosening of the
Aims. The use and variety of stemless
Aims. Stemless humeral implants have been developed to overcome stem-related complications in total shoulder arthroplasty (TSA). However, stemless implant designs may hypothetically result in less stable initial fixation, potentially affecting long-term survival. The aim of this study is to investigate early fixation and migration patterns of the stemless
Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up.Aims
Methods
Aims. The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up. Methods. A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted. Results. The ROM (p < 0.001), CS (p < 0.001), and DASH score (p < 0.001) showed significant improvements after shoulder arthroplasty for both implants. There were no differences between the groups treated with stemmed or stemless shoulder prosthesis with respect to the mean CS (79.2 (35 to 118) vs 74.4 (31 to 99); p = 0.519) and DASH scores (11.4 (8 to 29) vs 13.2 (7 to 23); p = 0.210). The ten-year unadjusted cumulative survival rate was 95.3% for the stemmed anatomical shoulder prosthesis and 91.5% for the stemless shoulder prosthesis and did not differ between the treatment groups (p = 0.251). The radiological evaluation of the
Aims. The purpose of this study was to report bone adaptive changes after anatomical total shoulder arthroplasty (TSA) using a standard-length hydroxyapatite (HA)-coated
This systematic review asked which patterns of complications are associated with the three reverse total shoulder arthroplasty (RTSA) prosthetic designs, as classified by Routman et al, in patients undergoing RTSA for the management of cuff tear arthropathy, massive cuff tear, osteoarthritis, and rheumatoid arthritis. The three implant design philosophies investigated were medial glenoid/medial humerus (MGMH), medial glenoid/lateral humerus (MGLH), and lateral glenoid/medial humerus (LGMH). A systematic review of the literature was performed via a search of MEDLINE and Embase. Two reviewers extracted data on complication occurrence and patient-reported outcome measures (PROMs). Meta-analysis was conducted on the reported proportion of complications, weighted by sample size, and PROMs were pooled using the reported standardized mean difference (SMD). Quality of methodology was assessed using Wylde’s non-summative four-point system. The study was registered with PROSPERO (CRD42020193041).Aims
Methods
Existing literature indicates that inferiorly inclined glenoid baseplates following reverse total shoulder arthroplasty (RSA) produce better outcomes compared to superiorly inclined baseplates. We aim to compare clinical outcomes for RSAs with superiorly and neutrally/inferiorly inclined lateralized glenospheres. We retrospectively reviewed 154 consecutive patients undergoing RSA between July 2015 and July 2017 by one single-fellowship trained surgeon (AJ). Two raters (KAM and MVS) independently measured glenoid inclination in preoperative and minimum two year follow-up radiographs (anteroposterior/Grashey) using the RSA angle. Inclination was then compared to patient-reported outcomes, range of motion (ROM), and independently assessed degree of scapular notching and staging of heterotopic ossification at two year follow-up.Aims
Methods
The purpose of this study was to determine whether there were long-term differences in outcomes of reverse shoulder arthroplasty (RSA) undertaken for acute proximal humeral fracture versus rotator cuff deficiency with a minimum follow-up of five years. This was a prospective cohort study comparing 67 patients with acute complex proximal humeral fracture and 64 patients with irreparable rotator cuff deficiency who underwent primary RSA. In the fracture group, there were 52 (77.6%) females and 15 (22.4%) males, with a mean age of 73.5 years (51 to 85), while in the arthropathy group, there were 43 (67.1%) females and 21 (32.9%) males, with a mean age of 70.6 years (50 to 84). Patients were assessed by the Constant score, University of California Los Angeles shoulder score (UCLA), short version of the Disability of the Arm Shoulder and Hand score (QuickDASH), and visual analogue scales (VAS) for pain and satisfaction. Radiological evaluation was also performed.Aims
Methods
Scapular notching is thought to have an adverse effect on the outcome of reverse total shoulder arthroplasty (RTSA). However, the matter is still controversial. The aim of this study was to determine the clinical impact of scapular notching on outcomes after RTSA. Three electronic databases (PubMed, Cochrane Database, and EMBASE) were searched for studies which evaluated the influence of scapular notching on clinical outcome after RTSA. The quality of each study was assessed. Functional outcome scores (the Constant-Murley scores (CMS), and the American Shoulder and Elbow Surgeons (ASES) scores), and postoperative range of movement (forward flexion (FF), abduction, and external rotation (ER)) were extracted and subjected to meta-analysis. Effect sizes were expressed as weighted mean differences (WMD).Aims
Methods
Reverse shoulder arthroplasty (RSA) reliably improves shoulder pain and function for a variety of indications. However, the safety and efficacy of RSA in elderly patients is largely unknown. The purpose of this study was to report the mortality, morbidity, complications, reoperations, and outcomes of primary RSA in patients aged > 80 years. Between 2004 and 2013, 242 consecutive primary RSAs were performed in patients aged > 80 years (mean 83.3 years (Aims
Patients and Methods
The aim of this study was to analyze the results of two radiocapitellar prostheses in a large case series followed prospectively, with medium-term follow-up. A total of 31 patients with a mean age of 54 years (27 to 73) were analyzed; nine had primary osteoarthritis (OA) and 17 had post-traumatic OA, three had capitellar osteonecrosis, and two had a fracture. Overall, 17 Lateral Resurfacing Elbow (LRE) and 14 Uni-Elbow Radio-Capitellum Implant (UNI-E) arthroplasties were performed. Pre- and postoperative assessment involved the Mayo Elbow Performance Score (MEPS), the Quick Disabilities of the Arm, Shoulder and Hand (Q-DASH) score, and the modified American Shoulder Elbow Surgeons (m-ASES) score.Aims
Patients and Methods
Aims. A number of methods have been described to remove a well-fixed humeral implant as part of revision shoulder arthroplasty. These include the use of cortical windows and humeral osteotomies. The router bit extraction technique uses a high-speed router bit to disrupt the bone-implant interface. The implant is then struck in a retrograde fashion with a square-tip impactor and mallet. The purpose of this study was to determine the characteristics and frequency of the different techniques needed for the removal of a well-fixed humeral stem in revision shoulder arthroplasty. Patients and Methods. Between 2010 and 2018, 288 revision shoulder arthroplasty procedures requiring removal of a well-fixed