Advertisement for orthosearch.org.uk
Results 1 - 9 of 9
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_16 | Pages 40 - 40
17 Nov 2023
Kuder I Jones G Rock M van Arkel R
Full Access

Abstract

Objectives

Ultrasound speckle tracking is a safe and non-invasive diagnostic tool to measure soft tissue deformation and strain. In orthopaedics, it could have broad application to measure how injury or surgery affects muscle, tendon or ligament biomechanics. However, its application requires custom tuning of the speckle-tracking algorithm then validation against gold-standard reference data. Implementing an experiment to acquire these data takes months and is expensive, and therefore prohibits use for new applications. Here, we present an alternative optimisation approach that automatically finds suitable machine and algorithmic settings without requiring gold-standard reference data.

Methods

The optimisation routine consisted of two steps. First, convergence of the displacement field was tested to exclude the settings that would not track the underlying tissue motion (e.g. frame rates that were too low). Second, repeatability was maximised through a surrogate optimisation scheme. All settings that could influence the strain calculation were included, ranging from acquisition settings to post-processing smoothing and filtering settings, totalling >1,000,000 combinations of settings. The optimisation criterion minimised the normalised standard deviation between strain maps of repeat measures. The optimisation approach was validated for the medial collateral ligament (MCL) with quasi-static testing on porcine joints (n=3), and dynamic testing on a cadaveric human knee (n=1, female, aged 49). Porcine joints were fully dissected except for the MCL and loaded in a material-testing machine (0 to 3% strain at 0.2 Hz), which was captured using both ultrasound (>14 repeats per specimen) and optical digital image correlation (DIC). For the human cadaveric knee (undissected), 3 repeat ultrasound acquisitions were taken at 18 different anterior/posterior positions over the MCL while the knee was extended/flexed between 0° and 90° in a knee extension rig. Simultaneous optical tracking recorded the position of the ultrasound transducer, knee kinematics and the MCL attachments (which were digitised under direct visualisation post testing). Half of the data collected was used for optimisation of the speckle tracking algorithms for the porcine and human MCLs separately, with the remaining unseen data used as a validation test set.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 74 - 74
17 Apr 2023
Theodoridis K Hall T Munford M Van Arkel R
Full Access

The success of cementless orthopaedic implants relies on bony ingrowth and active bone remodelling. Much research effort is invested to develop implants with controllable surface roughness and internal porous architectures that encourage these biological processes. Evaluation of these implants requires long-term and costly animal studies, which do not always yield the desired outcome requiring iteration. The aim of our study is to develop a cost-effective method to prescreen design parameters prior to animal trials to streamline implant development and reduce live animal testing burden.

Ex vivo porcine cancellous bone cylinders (n=6, Ø20×12mm) were extracted from porcine knee joints with a computer-numerically-controlled milling machine under sterile conditions within 4 hours of animal sacrifice. The bone discs were implanted with Ø6×12mm additive manufactured porous titanium implants and were then cultured for 21days. Half underwent static culture in medium (DMEM, 10% FBS, 1% antibiotics) at 37°C and 5% CO2. The rest were cultured in novel high-throughput stacked configuration in a bioreactor that simulated physiological conditions after surgery: the fluid flow and cyclic compression force were set at 10ml/min and 10–150 N (1Hz,5000 cycles/day) respectively. Stains were administered at days 7 and 14. Samples were evaluated with widefield microscopy, scanning electron microscopy (SEM) and with histology.

More bone remodelling was observed on the samples cultured within the bioreactor: widefield imaging showed more remodelling at the boundaries between the implant-bone interface, while SEM revealed immature bone tissue integration within the pores of the implant. Histological analysis confirmed these results, with many more trabecular struts with new osteoid formation on the samples cultured dynamically compared to static ones.

Ex vivo bone can be used to analyse new implant technologies with lower cost and ethical impact than animal trial. Physiological conditions (load and fluid flow) promoted bone ingrowth and remodelling.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 75 - 75
1 Dec 2021
Stoddart J Garner A Tuncer M Cobb J van Arkel R
Full Access

Abstract

Objectives

There is renewed interest in bi-unicondylar arthroplasty (Bi-UKA) for patients with medial and lateral tibiofemoral osteoarthritis, but a spared patellofemoral compartment and functional cruciate ligaments. The bone island between the two tibial components may be at risk of tibial eminence avulsion fracture, compromising function. This finite element analysis compared intraoperative tibial strains for Bi-UKA to isolated medial unicompartmental arthroplasty (UKA-M) to assess the risk of avulsion.

Methods

A validated model of a large, high bone-quality tibia was prepared for both UKA-M and Bi-UKA. Load totalling 450N was distributed between the two ACL bundles, implant components and collateral ligaments based on experimental and intraoperative measurements with the knee extended and appropriately sized bearings used. 95th percentile maximum principal elastic strain was predicted in the proximal tibia. The effect of overcuts/positioning for the medial implant were studied; the magnitude of these variations was double the standard deviation associated with conventional technique.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 52 - 52
1 Dec 2021
Wang J Hall T Musbahi O Jones G van Arkel R
Full Access

Abstract

Objectives

Knee alignment affects both the development and surgical treatment of knee osteoarthritis. Automating femorotibial angle (FTA) and hip-knee-ankle angle (HKA) measurement from radiographs could improve reliability and save time. Further, if the gold-standard HKA from full-limb radiographs could be accurately predicted from knee-only radiographs then the need for more expensive equipment and radiation exposure could be reduced. The aim of this research is to assess if deep learning methods can predict FTA and HKA angle from posteroanterior (PA) knee radiographs.

Methods

Convolutional neural networks with densely connected final layers were trained to analyse PA knee radiographs from the Osteoarthritis Initiative (OAI) database with corresponding angle measurements. The FTA dataset with 6149 radiographs and HKA dataset with 2351 radiographs were split into training, validation and test datasets in a 70:15:15 ratio. Separate models were learnt for the prediction of FTA and HKA, which were trained using mean squared error as a loss function. Heat maps were used to identify the anatomical features within each image that most contributed to the predicted angles.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 55 - 55
1 Mar 2021
Dandridge O Garner A van Arkel R Amis A Cobb J
Full Access

Abstract

Objectives

The need for gender specific knee arthroplasty is debated. This research aimed to establish whether gender differences in patellar tendon moment arm (PTMA), a composite measure that characterises function of both the patellofemoral and tibiofemoral joints, are a consequence of knee size or other variation.

Methods

PTMA about the instantaneous helical axis was calculated from positional data acquired using optical tracking. First, data post-processing was optimised, comparing four smoothing techniques (raw, Butterworth filtered, generalised cross-validation cubic spline interpolated and combined filtered/interpolated) using a fabricated knee. Then PTMA was measured during open-chain extension for N=24 (11 female) fresh-frozen cadaveric knees, with physiologically based loading and extension rates (420°/s) applied. Gender differences in PTMA were assessed before and after accounting for knee size with epicondylar width.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 102 - 102
1 Mar 2021
Kohli N De Eguilior Caballero JR Ghouse S Van Arkel R
Full Access

Abstract

Introduction

The long-term biological success of cementless orthopaedic prostheses is highly dependent on osteointegration. Pre-clinical testing of new cementless implant technology however, requires live animal testing, which has anatomical, loading, ethical and cost challenges. This proof-of-concept study aimed to develop an in vitro model to examine implant osteointegration under known loading/micromotion conditions.

Methods

Fresh cancellous bone cylinders (n=8) were harvested from porcine femur and implanted with additive manufactured porous titanium implants (Ø4 × 15 mm). To simulate physiological conditions, n=3 bone cylinders were tested in a bioreactor system with a cyclic 30 µm displacement at 1Hz for 300 cycles every day for 15 days in a total of 21 days culture. The chamber was also perfused with culture medium using a peristaltic pump. Control bone cylinders were cultured under static conditions (n=5). Samples were calcein stained at day 7. Post-testing, bone cylinders were formalin fixed and bony ingrowth was measured via microscopy.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 27 - 27
1 Mar 2021
Dandridge O Garner A van Arkel R Amis A Cobb J
Full Access

Abstract

Objectives

Unicompartmental (UKA) and bicompartmental (BCA) knee arthroplasty are associated with improved functional outcomes compared to Total Knee Arthroplasty (TKA) in suitable patients, although the reason is poorly understood. The aim of this study was to measure how the different arthroplasties affect knee extensor function.

Methods

Extensor function was measured for sixteen cadaveric knees and then re-tested following the different arthroplasties. Eight knees underwent medial UKA then BCA, then posterior-cruciate retaining TKA, and eight underwent the lateral equivalents then TKA. Extensor efficiency was calculated for ranges of knee flexion associated with common activities of daily living. Data were analyzed with repeated measures analysis of variance (α=0.05).


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 40 - 40
1 Mar 2021
Karunaseelan KJ van Arkel R Jeffers J
Full Access

Abstract

Objectives

Hip joint laxity after total hip arthroplasty (THA) has been considered to cause microseparation and lead to complications, including wear and dislocation. In the native hip, the hip capsular ligaments may tighten at the limits of range of hip motion and provide a passive stabilising force preventing edge loading and reduce the risk of dislocation. Previous attempts to characterise mechanical properties of hip capsular ligaments have been largely variable and there are no cadaveric studies quantifying the force contributions of each ligament in different hip positions. In this study we quantify the passive force contribution of the hip capsular ligaments throughout a complete range of motion (ROM).

Methods

Nine human cadaveric hip specimens (6 males and 3 females) with mean age of (76.4 ± 9.0 years) were skeletonised, preserving the capsular ligaments. Prepared specimens were tested in a 6 degree of freedom system to assess ROM with 5 Nm torque applied in external and internal rotation throughout hip flexion and extension. Capsular ligaments were resected in a stepwise fashion to assess internal force contributions of the iliofemoral (superior and inferior), pubofemoral, and ischiofemoral ligaments during ROM.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_2 | Pages 1 - 1
1 Jan 2019
Logishetty K Van Arkel R Muirhead-Allwood S Ng G Cobb J Jeffers J
Full Access

The hip's capsular ligaments (CL) passively restrain extreme range of motion (ROM) by wrapping around the native femoral head/neck, and protect against impingement and instability. We compared how CL function was affected by device (hip resurfacing arthroplasty, HRA; dual mobility total hip arthroplasty, DM-THA; and conventional THA, C-THA), and surgical approach (anterior and posterior), with and without CL surgical-repair. We hypothesized that CL function would only be preserved when native head-size (HRA/DM-THA) was restored.

CL function was quantified on sixteen cadaveric hips, by measuring ROM by internally (IR) and externally rotating (ER) the hip in six functional positions, ranging from full extension with abduction to full flexion with adduction (squatting). Native ROM was compared to ROM after posterior capsulotomy (right hips) or anterior capsulotomy (left hips), and HRA, and C-THA and DM-THA, before and after CL repair.

Independent of approach, ROM increased most following C-THA (max 62°), then DM-THA (max 40°), then HRA (max 19°), indicating later CL engagement and reduced biomechanical function with smaller head-size. Dislocations also occurred in squatting after C-THA and DM-THA. CL-repair following HRA restored ROM to the native hip (max 8°). CL-repair following DM-THA reduced ROM hypermobility in flexed positions only and prevented dislocation (max 36°). CL-repair following C-THA did not reduce ROM or prevent dislocation.

For HRA and repair, native anatomy was preserved and ligament function was restored. For DM-THA with repair, ligament function depended on the movement of the mobile-bearing, with increased ROM in positions when ligaments could not wrap around head/neck. For C-THA, the reduced head-size resulted in inferior capsular mechanics in all positions as the ligaments remained slack, irrespective of repair.

Choosing devices with anatomic head-sizes (HRA/DM-THA) with capsular repair may have greater effect than surgical approach to protect against instability in the early postoperative period.