Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 43 - 43
10 Feb 2023
Fary C Tripuraneni K Klar B Ren A Abshagen S Verheul R
Full Access

We sought to evaluate the early post-operative active range-of-motion (AROM) between robotic-assisted total knee arthroplasty (raTKA) and conventional TKA (cTKA). A secondary data analysis on a global prospective cohort study was performed. A propensity score method was used to select a matched control of cTKA from the same database using 1:1 ratio, based on age, sex, BMI, and comorbidity index. This resulted in 216 raTKA and cTKA matched cases. Multivariable longitudinal regression was used to evaluate difference in ROM over time and values are reported as least squares means (95% confidence interval). The longitudinal model tested the treatment effect (raTKA vs cTKA), time effect, and their interaction with control on covariance of patients ‘s age, sex, BMI, comorbidity and pre-operative flexion. Logistic regression was used to analyze the active flexion level at one month (cut by 90°) and three months (cut by 110°).

At one-month postoperative the raTKA cases had more AROM for flexion by an average of 5.54 degrees (p<0.001). There was no difference at three months (p=0.228). The raTKA group had a greater improvement from pre-operative values at both one-month, with an average 7.07° (3.6°, 10.5°, p<0.001) more improvement, and at three-months with an average improvement of 4° more (1.61°, 7.24°, p=0.0115). AROM for extension was lower overall in the raTKA group by an average of 0.44° (p=0.029). The raTKA patients had higher odds of achieving ≥90° of flexion at one-month (OR 2.15, 95% CI 1.16, 3.99).

raTKA resulted in greater AROM flexion gains in the early postoperative period than cTKA. Additional research is needed to understand if these earlier gains in AROM are associated with improved patient satisfaction and continued improvements with time.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_22 | Pages 61 - 61
1 Dec 2017
Davies G Bradford N Oliver R Verheul R Bruce W Walsh W
Full Access

Aim

The prevention of surgical-site infection (SSI) is of great importance. Airborne particulate correlates with microbial load and SSI. There are many potential sources of airborne particulates in theatre and from an experimental point of view impossible to control. We evaluated the effectiveness of a novel air decontamination-recirculation system (ADRS) in reducing airborne particles in a laboratory environment and controlled the introduction of particulate using diathermy.

Methods

Airborne particles were measured with and without activation of the ADRS in PC2 laboratory to provide a baseline. Particles were generated in a controlled manner utilising electrocautery ablation of porcine skin tissue. Ablation was performed at 50W power (Cut) for 60 seconds at a constant rate with and without the ADRS operating in the PC2 laboratory. Particles were measured continuously in 30s intervals at two sites 0.5m and 3m from the site of diathermy. Adequate time was allowed for return to baseline between each repetition. Each experiment was repeated 10 times.