Langton D Jameson S Van Oursouw M De Smet K Nargol A
Full Access

Background: Definitive cup position for the reduction of blood metal ion levels has yet to be established.

Methods: Samples for serum metal ion analysis are taken routinely from patients under the care of the two senior authors of this paper. Both are high volume experienced hip resurfacing surgeons, one based in England, the other in Belgium. Metal ion results from two centres from patients with unilateral joints were correlated to size and orientation of femoral and acetabular components, UCLA activity score, age, time post surgery and post operative femoral head/neck ratios. EBRA software was used to assess cup inclination and anteversion on standing radiographs.

Cup orientation in vivo was compared to explant analysis of 60 retrieved resurfacing components using a coordinate measuring machine.

Results: Three resurfacing devices were studied. There were 620 results in total. Only femoral size and cup inclination/anteversion were found to have any effect on ion levels. In all devices, metal ion levels were inversely related to femoral size (p< 0.05). The device providing the smallest acetabular coverage arc was associated with the highest metal ion levels. Consistent throughout the implants, lowest ion levels were associated with cups with radiological inclination of 40–50° and anteversion 10–20°. Cup inclination angles lower than 40° were associated with posterior edge loading and likely sub-luxation of the femoral component.

Conclusion: The greater the coverage angle provided by the acetabular component, the greater the tolerance to suboptimal position. Lowest ion levels were found in well positioned lower clearance devices. Cup inclination/ante-version angles of 45/15+/−5° were associated with low ion levels in all three devices. We do not recommend cups to be placed with inclination angles below 40°.