header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 54 - 54
1 Jan 2016
Idei J Sekiguchi M Kubota A Ohikata Y Yamamoto K Tsuchiya K Murase T
Full Access

Introduction

Recently, computer-aided orthopaedic surgery has enabled three dimensional (3D) preoperative planning, navigation systems and patient matched instrument, and they provide good clinical results in total knee arthroplasty. However, the preoperative planning methods and the criteria in total elbow arthroplasty (TEA) still have not sufficiently established due to the uncertainty of 3D anatomical geometry of the elbow joints. In order to clarify the 3D anatomical geometry, this study measured 3D bone models of the normal elbow joints. Additionally this study attempted to apply the 3D preoperative planning to ordinary surgery. Then the postoperative position of implant has evaluated as compared with the position in 3D preoperative planning.

Methods

Three dimensional bone measurements on 4 normal cases were performed. Three dimensional bone models were constructed with CT image using Bone Viewer®(ORTHREE Co., Ltd.). TEA was performed with FINE® Total Elbow System (Nakashima Medical Co., Ltd.) for 3 rheumatoid arthritis (RA) cases (Fig. 1). Three dimensional preoperative planning was based on this bone measurement, and postoperative position of implant were evaluated. The postoperative assessments were evaluated by superimposing preoperative planning image on postoperative CT image using Bone Simulator® (ORTHREE Co., Ltd.). This study only covers humeral part.