Advertisement for orthosearch.org.uk
Results 1 - 20 of 26
Results per page:
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 63 - 63
19 Aug 2024
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A
Full Access

Vitamin E-doped cross-linked polyethylene (VEPE) has encouraged the use of larger heads in thinner liners in total hip arthroplasty (THA). However, there are concerns about wear and mechanical failure of the thin liner, especially when metal heads are used. The aim of this randomized controlled trial was to investigate if the use of a large metal head in a thin VEPE liner would increase polyethylene wear compared with a standard 32-mm metal head and to compare periacetabular radiolucencies and patient-reported outcomes in THA.

96 candidates for uncemented THA were randomly allocated to either the largest possible metal head (36–44 mm) that could be fitted in the thinnest available VEPE liner (intervention group) or a standard 32-mm metal head (control group). The primary outcome was proximal head penetration (PHP) measured with model-based radiostereometric analysis (RSA). Secondary outcomes were periacetabular radiolucencies and patient-reported outcomes. The mid-term results of the trial at 5 years are presented.

Median total PHP (interquartile range) was -0.04 mm (−0.12 to 0.02) in the intervention group and -0.03 mm (=0.14 to 0.05) in the control group (p=0.691). The rates of periacetabular radiolucencies were 1/44 and 4/42 (p=0.197), respectively. Patient-reported hip function and health-related quality of life did not differ between the groups, but participants in the intervention group reported a higher level of activity (median UCLA rank 7 vs 6, p=0.020). There were 5 revisions caused by dislocations (2), periprosthetic fracture (1), stem subsidence (1), or iliopsoas impingement (1).

Large metal heads in thin VEPE liners did not increase liner wear and were not associated with liner failure 5 years after THA.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 3 - 9
1 Mar 2024
Halken CH Bredgaard Jensen C Henkel C Gromov K Troelsen A

Aims

This study aimed to investigate patients’ attitudes towards day-case hip and knee arthroplasty and to describe patient characteristics associated with different attitudes, with the purpose of providing an insight into the information requirements for patients that surgeons should address when informing patients about day-case surgery.

Methods

A total of 5,322 patients scheduled for hip or knee arthroplasty between 2016 and 2022 were included in the study. Preoperatively, patients were asked if they were interested in day-case surgery (‘Yes’, ‘Do not know’, ‘No’). Patient demographics including age, BMI, sex, and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension three-level questionnaire (EQ-5D-3L) were examined within each attitude group. Additionally, changes in attitude were assessed among patients who had completed the questionnaire in association with prior hip or knee arthroplasty.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 31 - 37
1 Mar 2024
Bunyoz KI Tsikandylakis G Mortensen K Gromov K Mohaddes M Malchau H Troelsen A

Aims

In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased taper corrosion, with the release of metal ions and adverse reactions. The aim of this study was to investigate the relationship between the size of the femoral head and the levels of metal ions in the blood in these patients.

Methods

A total of 96 patients were enrolled at two centres and randomized to undergo MoP THA using either a 32 mm metal head or a femoral head of between 36 mm and 44 mm in size, being the largest possible to fit the thinnest available polyethylene insert. The levels of metal ions and patient-reported outcome measures (Oxford Hip Score, University of California, Los Angeles Activity Scale) were recorded at two and five years postoperatively.


Bone & Joint Open
Vol. 4, Issue 12 | Pages 923 - 931
4 Dec 2023
Mikkelsen M Rasmussen LE Price A Pedersen AB Gromov K Troelsen A

Aims

The aim of this study was to describe the pattern of revision indications for unicompartmental knee arthroplasty (UKA) and total knee arthroplasty (TKA) and any change to this pattern for UKA patients over the last 20 years, and to investigate potential associations to changes in surgical practice over time.

Methods

All primary knee arthroplasty surgeries performed due to primary osteoarthritis and their revisions reported to the Danish Knee Arthroplasty Register from 1997 to 2017 were included. Complex surgeries were excluded. The data was linked to the National Patient Register and the Civil Registration System for comorbidity, mortality, and emigration status. TKAs were propensity score matched 4:1 to UKAs. Revision risks were compared using competing risk Cox proportional hazard regression with a shared γ frailty component.


Bone & Joint Open
Vol. 4, Issue 6 | Pages 457 - 462
26 Jun 2023
Bredgaard Jensen C Gromov K Petersen PB Jørgensen CC Kehlet H Troelsen A

Aims

Medial unicompartmental knee arthroplasty (mUKA) is an advised treatment for anteromedial knee osteoarthritis. While long-term survival after mUKA is well described, reported incidences of short-term surgical complications vary and the effect of surgical usage on complications is less established. We aimed to describe the overall occurrence and treatment of surgical complications within 90 days of mUKA, as well as occurrence in high-usage centres compared to low-usage centres.

Methods

mUKAs performed in eight fast-track centres from February 2010 to June 2018 were included from the Lundbeck Foundation Centre for Fast-track Hip and Knee Replacement Database. All readmissions within 90 days of surgery underwent chart review and readmissions related to the surgical wound or the prosthesis were recorded. Centres were categorized as high-usage centres when using mUKA in ≥ 20% of annual knee arthroplasties. The occurrence of complications between high- and low-usage centres were compared using Fisher’s exact test.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 34 - 34
23 Jun 2023
Bunyoz KI Tsikandylakis G Mortensen K Gromov K Mohaddes M Malchau H Troelsen A
Full Access

In metal-on-polyethylene (MoP) THA large femoral metal heads are designed to increase stability and to reduce dislocation risk. The increased head size could lead to increased taper corrosion with the release of metal ions and adverse reactions. Using blood ion measurements, we aimed to investigate the association between femoral head size and metal-ion release after MoP THA.

96 patients were enrolled at two centers and randomized to receive either a 32-mm metal head or a 36–44 mm metal head (the largest possible fitting the thinnest available polyethylene insert). Blood metal ions and PROMs (OHS, UCLA) were measured at two- and five-year follow-ups.

Both 2- and 5-year median chrome, cobalt, and titanium levels were below taper corrosion indicative ion levels. At 5 years, median chrome, cobalt, and titanium levels were 0.5 μg/L (0.50–0.62), 0.24 μg/L (0.18–0.30), and 1.16 μg/L (1.0–1.68) for the 32-mm group, and 0.5 μg/L (0.5–0.54), 0.23 μg/L (0.17–0.39), and 1.30 μg/L (1–2.05) for the 36–44 mm group, with no difference between groups (p=0.825, p=1.000, p=0.558). At 2 years, 7 (32-mm) versus 4 (36–44-mm) patients had elevated ions. At 5 years, 6 (32-mm) versus 7 (36–44-mm) patients had elevated ions. There was no difference in either OHS (p=0.665) or UCLA (p=0.831) between patients with or without elevated blood metal ions.

5 years after the insertion of MoP THAs, we found no differences in the blood metal ion levels between 32 mm heads and 36–44 mm heads and no corrosion-related revisions. As taper corrosion can debut after 5 years, there is still a need for long-term follow-up studies on the association between head size and corrosion in MoP THA.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 80 - 80
23 Jun 2023
Halken CH Jensen CB Henkel C Gromov K Troelsen A
Full Access

The interest in day-case hip arthroplasty is increasing; however, there are conflicting results regarding readmission risk, and little is known about patients’ attitude towards day-case surgery. We aimed to investigate differences in 30-day readmission rates between day-case patients and single-overnight-stay patients following total hip arthroplasty (THA) and explore patients’ attitude towards day-case surgery.

From the Danish National Patient Register we identified 29,486 THAs (1353 day-case THAs and 28,133 single-overnight-stay THAs) performed between 2010 and 2020. Day-case surgery was defined as discharge on the day-of-surgery. Overnight admissions within 30 days of surgery were considered readmissions. Differences in readmission rates between day-case and single-overnight-stay patients were investigated using logistic regression adjusted for patient characteristics and year of surgery. In a single-center descriptive study, 2395 hip arthroplasty patients from 2016 to start-2023 answered a questionnaire on patient characteristics, attitude towards day-case surgery and patient-reported outcome measures (PROM).

The overall 30-day readmission rate was 4.4% (CI: 4.2–4.6%) with no difference between day-case (4.4%) and single-overnight-stay THAs (4.4%) (odds-ratio: 1.2 [CI: 0.91–1.6]). In the descriptive study answers to the question whether patients were interested in being discharged on the day of surgery, were: “Yes” = 41%, “Do not know” = 20%, “No” = 39%. Patients responding “No” had lower preoperative PROM-scores, were older (“No” = 70.2 y vs. “Yes” = 65.3 y), and more often female (“No” = 72% vs. “Yes” = 52% female).

Based on nationwide data from 2010–2020, day-case THA patients were not more likely to be readmitted compared to single-overnight-stay patients. However, most patients were not positive towards day-case surgery. While surgeons may consider day-case surgery to be safe, patients are not intuitively positive, indicating an attitude mismatch. To achieve high success rate and patient satisfaction in day-case surgery, shared-decision making initiatives within day-case surgery are needed.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 441 - 447
23 May 2022
Mikkelsen M Wilson HA Gromov K Price AJ Troelsen A

Aims

Treatment of end-stage anteromedial osteoarthritis (AMOA) of the knee is commonly approached using one of two surgical strategies: medial unicompartmental knee arthroplasty (UKA) or total knee arthroplasty (TKA). In this study we aim to investigate if there is any difference in outcome for patients undergoing UKA or TKA, when treated by high-volume surgeons, in high-volume centres, using two different clinical guidelines. The two strategies are ‘UKA whenever possible’ vs TKA for all patients with AMOA.

Methods

A total of 501 consecutive AMOA patients (301 UKA) operated on between 2013 to 2016 in two high-volume centres were included. Centre One employed clinical guidelines for the treatment of AMOA allowing either UKA or TKA, but encouraged UKA wherever possible. Centre Two used clinical guidelines that treated all patients with a TKA, regardless of wear pattern. TKA patients were included if they had isolated AMOA on preoperative radiographs. Data were collected from both centres’ local databases. The primary outcome measure was change in Oxford Knee Score (OKS), and the proportion of patients achieving the patient-acceptable symptom state (PASS) at one-year follow-up. The data were 1:1 propensity score matched before regression models were used to investigate potential differences.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1206 - 1214
1 Jul 2021
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A

Aims

We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty.

Methods

In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 762 - 768
1 Apr 2021
Ban I Kristensen MT Barfod KW Eschen J Kallemose T Troelsen A

Aims

To compare the functionality of adults with displaced mid-shaft clavicular fractures treated either operatively or nonoperatively and to compare the relative risk of nonunion and reoperation between the two groups.

Methods

Based on specific eligibility criteria, 120 adults (median age 37.5 years (interquartile range (18 to 61)) and 84% males (n = 101)) diagnosed with an acute displaced mid-shaft fracture were recruited, and randomized to either the operative (n = 60) or nonoperative (n = 60) treatment group. This randomized controlled, partially blinded trial followed patients for 12 months following initial treatment. Functionality was assessed by the Constant score (CS) (assessor blinded to treatment) and Disability of the Arm, Shoulder and Hand (DASH) score. Clinical and radiological evaluation, and review of patient files for complications and reoperations, were added as secondary outcomes.


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1167 - 1175
14 Sep 2020
Gromov K Petersen PB Jørgensen CC Troelsen A Kehlet H

Aims

The aim of this prospective multicentre study was to describe trends in length of stay and early complications and readmissions following unicompartmental knee arthroplasty (UKA) performed at eight different centres in Denmark using a fast-track protocol and to compare the length of stay between centres with high and low utilization of UKA.

Methods

We included data from eight dedicated fast-track centres, all reporting UKAs to the same database, between 2010 and 2018. Complete ( > 99%) data on length of stay, 90-day readmission, and mortality were obtained during the study period. Specific reasons for a length of stay of > two days, length of stay > four days, and 30- and 90-day readmission were recorded. The use of UKA in the different centres was dichotomized into ≥ 20% versus < 20% of arthroplasties which were undertaken being UKAs, and ≥ 52 UKAs versus < 52 UKAs being undertaken annually.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1177 - 1178
1 Oct 2019
Troelsen A Haddad FS


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 960 - 969
1 Aug 2019
Odgaard A Laursen MB Gromov K Troelsen A Kristensen PW Schrøder H Madsen F Overgaard S

Aims

The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.

Materials and Methods

Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 768 - 778
1 Jul 2019
Galea VP Rojanasopondist P Ingelsrud LH Rubash HE Bragdon C Huddleston III JI Malchau H Troelsen A

Aims

The primary aim of this study was to quantify the improvement in patient-reported outcome measures (PROMs) following total hip arthroplasty (THA), as well as the extent of any deterioration through the seven-year follow-up. The secondary aim was to identify predictors of PROM improvement and deterioration.

Patients and Methods

A total of 976 patients were enrolled into a prospective, international, multicentre study. Patients completed a battery of PROMs prior to THA, at three months post-THA, and at one, three, five, and seven-years post-THA. The Harris Hip Score (HHS), the 36-Item Short-Form Health Survey (SF-36) Physical Component Summary (PCS), the SF-36 Mental Component Summary (MCS), and the EuroQol five-dimension three-level (EQ-5D) index were the primary outcomes. Longitudinal changes in each PROM were investigated by piece-wise linear mixed effects models. Clinically significant deterioration was defined for each patient as a decrease of one half of a standard deviation (group baseline).


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 20 - 20
1 Oct 2018
Galea V Connelly J Matuszak S Botros MA Rojanasopondist P Nielsen C Huddleston J Bragdon C Malchau H Troelsen A
Full Access

Introduction

The aim of this study was to evaluate the effects of posterior tibial slope (PTS) and posterior condylar offset (PCO) on patient-reported pain and function one year after TKA.

Methods

A total of 500 patients from 11 clinics in 6 countries were enrolled into a prospective, multicenter study. All patients were indicated for primary TKA for OA and received components from a single manufacturer. All liners were made from vitamin-E stabilized, highly crosslinked (95 kGy) polyethylene; 54.7% were posterior stabilized (PS) and the remaining were cruciate-retaining.

The Knee Injury and Osteoarthritis Outcome Score (KOOS) was administered at the one-year follow-up visit. The KOOS pain and activities in daily life (ADL) sub-scores were dichotomized and served as the primary outcomes. Dichotomization was done with the patient acceptable symptom state (PASS), defined by previous studies as the value of the PROM above which patients deem their state as acceptable (84.5 points for KOOS pain and 83.0 points for KOOS ADL).

Plain lateral radiographs were taken and assessed for PTS (Figure 1) and PCO (Figure 2). PTS was categorized as above (excessive flexion), within (ideal), or below (extension) the safe zone of 0° − 7° of flexion. PCO increases or decreases of greater than 3mm were compared against no change (≤ 3mm). Each of the two sagittal positioning metrics was tested against the KOOS pain and ADL PASS at one year.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 88 - 88
1 Jan 2018
Søballe K Troelsen A Andreasen C
Full Access

The goal of periacetabular osteotomy (PAO) is to delay or prevent osteoarthritic development in dysplastic hips. However, it is unclear whether the surgical goals are achieved and if so in which patients. This information is essential to select appropriate patients for a durable PAO that achieves its goals.

We therefore (1) determined hip survival rates; (2) determined how many preserved hips were functionally successful after PAO; and (3) identified demographic, clinical, and radiographic factors predicting failure after PAO.

We reviewed 316 patients (401 hips) who had PAO. We evaluated radiographic parameters and obtained WOMAC scores. Through inquiry to the National Registry of Patients, we identified conversions to THA. Risk factors for conversion to THA were assessed. Mean followup was 8 years (range, 4–12 years).

We conclude that PAO can be performed with a good outcome at medium to long term follow-up, suggesting that it may be applied by experienced surgeons with satisfactory results. To further improve the outcome, focus should be on the potential negative influence of parameters that are easily assessed, such as the patients age (>45) preoperative grade of osteoarthritis, incongruency, and severe acetabular dysplasia. Perioperatively, the surgeon should attempt to achieve hip congruence and a CE angle between 30° to 40° to improve the durability of PAO.

Level II, prognostic study.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_22 | Pages 89 - 89
1 Dec 2017
Lange J Troelsen A Solgaard S Ørsnes T Otte K Nielsen PT Lamm M Krarup N Mikkelsen S Zawadski A Søballe K
Full Access

Aim

Our aim was to evaluate cementless one-stage revision in chronic periprosthetic hip joint infection.

Method and patients

The study was performed as a multicentre, proof-of-concept, observational study with prospective data collection. Inclusion of patients with a chronic periprosthetic hip joint infection (PJI) were followed by protocolled surgical treatment (cementless one-stage revision - the CORIHA protocol) at one of 8 participating departments of orthopaedic surgery between 2009 – 2014, and the patients enrolled in a 2-year follow-up program. A PJI were diagnosed based on adopted criteria from McPherson and Zimmerli. At the time of initiation of the study in 2009, the collaborating departments performed approximately one-fourth of all nationwide primary HJR and more than one-third of all revisions.

In total 56 PJI patients with a median age of 72 years and a median pre-operative ASA score of 2 met the established eligibility criteria and accepted to participate; 31 (55%) were males.

The cohort had a mean follow-up time of 4.0 years, with all patients followed for minimum 2 years.

The primary outcome were relapse described as re-revision due to infection (regardless of considered as a relapse or new infection). This was evaluated by competing risk analysis (competing risks: aseptic revisions and death).

Secondary, all-cause mortality was evaluated by survival analysis.

The study was approved by the local Committees on Biomedical Research Ethics.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 577 - 584
1 May 2017
Nebergall AK Greene ME Laursen MB Nielsen PT Malchau H Troelsen A

Aims

The objective of this five-year prospective, blinded, randomised controlled trial (RCT) was to compare femoral head penetration into a vitamin E diffused highly cross-linked polyethylene (HXLPE) liner with penetration into a medium cross-linked polyethylene control liner using radiostereometric analysis.

Patients and Methods

Patients scheduled for total hip arthroplasty (THA) were randomised to receive either the study E1 (32 patients) or the control ArComXL polyethylene (35 patients). The median age (range) of the overall cohort was 66 years (40 to 76).


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 451 - 457
1 Apr 2017
Gromov K Bersang A Nielsen CS Kallemose T Husted H Troelsen A

Aims

The aim of this study was to identify patient- and surgery-related risk factors for sustaining an early periprosthetic fracture following primary total hip arthroplasty (THA) performed using a double-tapered cementless femoral component (Bi-Metric femoral stem; Biomet Inc., Warsaw, Indiana).

Patients and Methods

A total of 1598 consecutive hips, in 1441 patients receiving primary THA between January 2010 and June 2015, were retrospectively identified. Level of pre-operative osteoarthritis, femoral Dorr type and cortical index were recorded. Varus/valgus placement of the stem and canal fill ratio were recorded post-operatively. Periprosthetic fractures were identified and classified according to the Vancouver classification. Regression analysis was performed to identify risk factors for early periprosthetic fracture.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 128 - 128
1 Feb 2017
Bragdon C Galea V Donahue G Lindgren V Troelsen A Marega L Muratoglu O Malchau H
Full Access

Introduction

Studies of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA) have reported high complication and failure rates due to elevated metal ion levels. These rates were shown to be especially high for the Articular Surface Replacement (ASR) HRA, possibly due to its unique design. Associations between metal ion concentrations and various biological and mechanical factors have been reported. Component positioning as measured by acetabular inclination has been shown to be of especially strong consequence in metal ion production in ASR HRA patients, but few studies have evaluated acetabular anteversion as an independent variable.

The primary objective of this study was to evaluate the associations between component orientation, quantified by acetabular inclination and anteversion, and blood metal ions. Secondly, we sought to report whether conventional safe zones apply to MoM HRA implants or if these implants require their own positioning standards.

Methods

We conducted a multi-center, prospective study of 512 unilateral ASR HRA patients enrolled from September 2012 to June 2015. At time of enrollment our patients were a mean of 7 (3–11.5) years from surgery. The mean age at surgery was 56 years and 24% were female. All subjects had complete demographic and surgical information and blood metal ions. In addition, each patient had valid AP pelvis and shoot-though lateral radiographs read by 5 validated readers measuring acetabular abduction and anteversion, and femoral offset. A multivariate logistic regression was used with high cobalt or chromium (greater than or equal to 7ppb) as the dependent variable. The independent variables were: female gender, UCLA activity score, age at surgery, femoral head size, time from surgery, femoral offset, acetabular abduction, and acetabular anteversion.