Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 87 - 87
1 Jan 2016
Clarke I Sufficool D Bowsher JG Savisaar C Burgett-Moreno M Donaldson T
Full Access

Introduction

Hip simulators proved to be valuable, pre-clinical tests for assessing wear. Preferred implant positioning has been with cup mounted above head, i.e. ‘Anatomical’ (Figs. 1a-c) 1,2 while the ‘Inverted’ test (cup below head) was typically preferred in debris studies (Figs. 1d-f).3,4 In an Anatomical study, wear patterns on cups and heads averaged 442 and 1668 mm² area, respectively, representing 8% and 30% of available hemi-surface (Table 1), i.e. the head pattern was ×3.8 times larger than cup. This concept of wear patterns is illustrated well in the ‘pin-on-disk’ test (Fig. 1) in which the oscillating pin has the ‘contained’ wear area (CWP) and the large wear track on the disk is the ‘distributed’ pattern (DWP). Hip simulators also create CWP and DWP patterns, site dependant on whether Anatomical (Fig. 1a-c) or ‘Inverted’ (Fig. 1d-f) test. However there is scant foundation as to clinical merits of either test mode. Retrieval studies of MOM bearings have indicated that cups have the larger wear patterns, i.e. contrary to simulator tests running Anatomical cups (Table 1).5 Therefore we compared Anatomical and Inverted cup modes using 38mm and 40mm MOM in two 5-million cycle simulator studies.

Methods

38mm and 40mm MOM bearings were run in Anatomical mode (study-1) and Inverted (study-2) mode, respectively, in a hip simulator. Lubricant was bovine serum diluted to provide protein concentration 17 mg/ml. Wear was measured gravimetrically and wear-rates calculated by linear regression. Wear patterns were assessed by stereomicroscopy and compared to algorithms using standard spherical equations.