Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 396 - 396
1 Jul 2008
Jaiswal P Gupta A Strokes O Pollock R Cannon S Briggs T Blunn G
Full Access

Aim: To study the remodelling of cortical bone around the passive growing component of an expandable endoprosthesis.

Introduction: When inserting the passive component of an expandable prosthesis a polyethylene sleeve is commonly used. The sleeve migrates towards the lateral cortex and causes a cortical reaction and hence the use of the sleeve has been discarded recently. This study quantifies the amount of cortical reaction and degree of cortical drift in patients that had sleeves and those that did not.

Material And Methods: We reviewed X-rays and case notes of all patients that had an expandable endoprosthesis in a 20 year period. The thickness of medial and lateral cortices of the tibial diaphysis was measured at 6 months and on the last follow up radiograph. The distance from the edge of the sleeve (or prosthesis) from the cortical edge was also compared. Retrieved components also had their histology reviewed.

Results: The sleeve shifted laterally on average by 2mm (range 0.5 – 3mm) and touched the cortex. This was associated with an increase in lateral cortical thickness by 2.27mm (range 1 – 3mm). When the sleeve was used the prosthesis was inserted in the mid-line. When the sleeve was not used the tibial component tended to be inserted in valgus.

Conclusions: The presence of a sleeve is associated with a cortical reaction and the sleeve tends to migrate laterally. The clinical implications of this and the evolution of the design will be discussed in the meeting.