Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1043 - 1049
1 Aug 2016
Huijbregts HJTAM Khan RJK Fick DP Hall MJ Punwar SA Sorensen E Reid MJ Vedove SD Haebich S

Aims

We conducted a randomised controlled trial to assess the accuracy of positioning and alignment of the components in total knee arthroplasty (TKA), comparing those undertaken using standard intramedullary cutting jigs and those with patient-specific instruments (PSI).

Patients and Methods

There were 64 TKAs in the standard group and 69 in the PSI group.

The post-operative hip-knee-ankle (HKA) angle and positioning was investigated using CT scans. Deviation of > 3° from the planned position was regarded as an outlier. The operating time, Oxford Knee Scores (OKS) and Short Form-12 (SF-12) scores were recorded.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_7 | Pages 12 - 12
1 May 2015
Huijbregts H Punwar S McMurray D Sorensen E Fick D Khan R
Full Access

Eligible patients were randomly allocated to PMI or standard intramedullary jigs. Smith and Nephew's patient specific cutting blocks (Visionaire) were used for PMI. Postoperative component positioning was investigated using the ‘Perth CT protocol’. Deviation of more than 3° from the recommended position was regarded as an outlier. Exact Mann-Whitney U test was used to compare component positioning and difference in proportion of outliers was calculated using Chi Squared analysis.

Fifty-five knees were enrolled in the standard instrumentation group and fifty-two knees in the PMI group.

Coronal femoral alignment was 0.7 ± 1.9° (standard) vs 0.5 ± 1.6° (PMI) (P=0.33). Outliers 9.4% vs 7.4% (P=0.71). Coronal tibial alignment was 0.4 ± 1.5° (standard) vs 0.6 ± 1.4° (PMI) (P=0.56). Outliers 1.9% vs 1.9% (P=0.99). Sagittal femoral alignment was 0.6 ± 1.5° (standard) vs 1.3 ± 1.9° (PMI) (P=0.07). Outliers 3.8% vs 13.2% (P=0.09). Tibial slope was 1.7 ± 1.9 ° (standard) vs 1.8 ± 2.7° (PMI) (P=0.88). Outliers 13.2% vs 24.1% (P=0.15). External rotation of femoral component was 0.6 ± 1.4° (standard) vs 0.2 ± 1.8° (PMI) (P=0.14). Outliers: 3.8% vs 5.6% (P=0.66).

Compared to standard intramedullary jigs, patient matched instrumentation does not improve component positioning or reduce alignment outliers.