Hughes LD Chamberlain KA Sloan A Choudry Q Robinson H
Full Access

MHRA guidance for patients with metal on metal hip replacements (MOM) was provided in 2012 and updated in 2017 to assist in the early detection of soft tissue reactions due to metal wear debris. A large number of metal on metal hip replacements were undertaken at our trust. A program of recall for all patients with metal on metal hip replacements was undertaken and MHRA guidelines implemented.

Identification and recall of all patients from 2001 with MOM hip replacements using theatre logs, patient records and consultant log books. Two consultant review of X-rays and patient records. Postal questionnaires and GP requests for cobalt & chromium blood tests. Two consultant led MOM review clinics undertaken with metal artefact reduction scans (MARS) performed following consultation in 2017.

674 patients, 297 available for review. 59 refused follow up. 87 moved out of area, 36 untraceable, 26 not MOM, 147 RIP, 22 already revised. From 297 patients 126 female, 171 male, age range 39 – 95 yrs. 126 resurfacing, 171 MOM THR. 26 patients with elevated metal ions, MARS performed of which 17 positive, 9 negative. Of 17 positive scans 10 patients asymptomatic, 7 waiting revision.

A time consuming effort and additional resource was needed and supported by the trust. From 297 hips 17 positive MARs were identified (5.7%). A new database registry has been developed to track MOM patients, clinics set up for ongoing follow up with radiological protocols for imaging. An arthroplasty advanced nurse practitioner (ANP) is now trained reviewing patients independently.