Recent publications have drawn attention to the fact that some brands of joint replacement may contain variants which perform significantly worse (or better) than their ‘siblings’. As a result, the National Joint Registry has performed much more detailed analysis on the larger families of knee arthroplasties in order to identify exactly where these differences may be present and may hitherto have remained hidden. The analysis of the Nexgen knee arthroplasty brand identified that some posterior-stabilized combinations have particularly high revision rates for aseptic loosening of the tibia, and consequently a medical device recall has been issued for the Nexgen ‘option’ tibial component which was implicated. More elaborate signal detection is required in order to identify such variation in results in a routine fashion if patients are to be protected from such variation in outcomes between closely related implant types. Cite this article:
Access to joint replacement is being restricted for patients with comorbidities in a number of high-income countries. However, there is little evidence on the impact of comorbidities on outcomes. The purpose of this study was to determine the safety and effectiveness of hip and knee arthroplasty in patients with and without comorbidities. In total, 312,079 hip arthroplasty and 328,753 knee arthroplasty patients were included. A total of 11 common comorbidities were identified in administrative hospital records. Safety risks were measured by assessing length of hospital stay (LOS) and 30-day emergency readmissions and mortality. Effectiveness outcomes were changes in Oxford Hip or Knee Scores (OHS/OKS) (scale from 0 (worst) to 48 (best)) and in health-related quality of life (EQ-5D) (scale from 0 (death) to 1 (full health)) from immediately before, to six months after, surgery. Regression analysis was used to estimate adjusted mean differences (LOS, change in OHS/OKS/EQ-5D) and risk differences (readmissions and mortality).Aims
Methods
The aim of this study was to determine whether patients with
metal-on-metal (MoM) arthroplasties of the hip have an increased
risk of cardiac failure compared with those with alternative types
of arthroplasties (non-MoM). A linkage study between the National Joint Registry, Hospital
Episodes Statistics and records of the Office for National Statistics
on deaths was undertaken. Patients who underwent elective total
hip arthroplasty between January 2003 and December 2014 with no
past history of cardiac failure were included and stratified as
having either a MoM (n = 53 529) or a non-MoM (n = 482 247) arthroplasty.
The primary outcome measure was the time to an admission to hospital
for cardiac failure or death. Analysis was carried out using data
from all patients and from those matched by propensity score.Aims
Patients and Methods
Tuberculosis (TB) infection of bones and joints accounts for
6.7% of TB cases in England, and is associated with significant
morbidity and disability. Public Health England reports that patients
with TB experience delays in diagnosis and treatment. Our aims were
to determine the demographics, presentation and investigation of
patients with a TB infection of bones and joints, to help doctors
assessing potential cases and to identify avoidable delays. This was a retrospective observational study of all adults with
positive TB cultures on specimens taken at a tertiary orthopaedic
centre between June 2012 and May 2014. A laboratory information
system search identified the patients. The demographics, clinical presentation,
radiology, histopathology and key clinical dates were obtained from
medical records.Aims
Patients and Methods
The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single stem design. This retrieval study involved 48 total hip arthroplasties (THAs) with CoCr V40 trunnions paired with either CoCr (n = 21) or ceramic (n = 27) heads. The taper junction of all hips was evaluated for fretting/corrosion damage and volumetric material loss using a roundness-measuring machine. We used linear regression analysis to investigate taper damage differences after adjusting for potential confounding variables.Objectives
Methods
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
The National Joint Registry for England, Wales and Northern Ireland
(NJR) has extended its scope to report on hospital, surgeon and
implant performance. Data linkage of the NJR to the London Implant
Retrieval Centre (LIRC) has previously evaluated data quality for
hip primary procedures, but did not assess revision records. We analysed metal-on-metal hip revision procedures performed
between 2003 and 2013. A total of 69 929 revision procedures from
the NJR and 929 revised pairs of components from the LIRC were included.Aims
Methods
Arthroplasty registries are important for the
surveillance of joint replacements and the evaluation of outcome. Independent
validation of registry data ensures high quality. The ability for
orthopaedic implant retrieval centres to validate registry data
is not known. We analysed data from the National Joint Registry
for England, Wales and Northern Ireland (NJR) for primary metal-on-metal
hip arthroplasties performed between 2003 and 2013. Records were
linked to the London Implant Retrieval Centre (RC) for validation.
A total of 67 045 procedures on the NJR and 782 revised pairs of
components from the RC were included. We were able to link 476 procedures
(60.9%) recorded with the RC to the NJR successfully. However, 306
procedures (39.1%) could not be linked. The outcome recorded by the
NJR (as either revised, unrevised or death) for a primary procedure
was incorrect in 79 linked cases (16.6%). The rate of registry-retrieval
linkage and correct assignment of outcome code improved over time.
The rates of error for component reference numbers on the NJR were
as follows: femoral head category number 14/229 (5.0%); femoral head
batch number 13/232 (5.3%); acetabular component category number
2/293 (0.7%) and acetabular component batch number 24/347 (6.5%). Registry-retrieval linkage provided a novel means for the validation
of data, particularly for component fields. This study suggests
that NJR reports may underestimate rates of revision for many types
of metal-on-metal hip replacement. This is topical given the increasing
scope for NJR data. We recommend a system for continuous independent
evaluation of the quality and validity of NJR data. Cite this article:
In patients with a tumour affecting the distal
ulna it is difficult to preserve the function of the wrist following extensive
local resection. We report the outcome of 12 patients (nine female,
three male) who underwent excision of the distal ulna without local
soft-tissue reconstruction. In six patients, an aggressive benign
tumour was present and six had a malignant tumour. At a mean follow-up
of 64 months (15 to 132) the mean Musculoskeletal Tumour score was
64% (40% to 93%) and the mean DASH score was 35 (10 to 80). The
radiological appearances were satisfactory in most patients. Local
recurrence occurred in one patient with benign disease and two with
malignant disease. The functional outcome was thus satisfactory
at a mean follow-up in excess of five years, with a relatively low
rate of complications. The authors conclude that complex reconstructive
soft-tissue procedures may not be needed in these patients. Cite this article:
The pre-operative differentiation between enchondroma,
low-grade chondrosarcoma and high-grade chondrosarcoma remains a
diagnostic challenge. We reviewed the accuracy and safety of the
radiological grading of cartilaginous tumours through the assessment
of, first, pre-operative radiological and post-operative histological agreement,
and second the rate of recurrence in lesions confirmed as high-grade
on histology. We performed a retrospective review of major long
bone cartilaginous tumours managed by curettage as low grade between
2001 and 2012. A total of 53 patients with a mean age of 47.6 years
(8 to 71) were included. There were 23 men and 30 women. The tumours
involved the femur (n = 20), humerus (n = 18), tibia (n = 9), fibula
(n = 3), radius (n = 2) and ulna (n = 1). Pre-operative diagnoses
resulted from multidisciplinary consensus following radiological
review alone for 35 tumours, or with the addition of pre-operative
image guided needle biopsy for 18. The histologically confirmed diagnosis
was enchondroma for two (3.7%), low-grade chondrosarcoma for 49
(92.6%) and high-grade chondrosarcoma for two (3.7%). Three patients
with a low-grade tumour developed a local recurrence at a mean of 15
months (12 to 17) post-operatively. A single high-grade recurrence
(grade II) was treated with tibial diaphyseal replacement. The overall
recurrence rate was 7.5% at a mean follow-up of 4.7 years (1.2 to
12.3). Cartilaginous tumours identified as low-grade on pre-operative
imaging with or without additional image-guided needle biopsy can
safely be managed as low-grade without pre-operative histological
diagnosis. A few tumours may demonstrate high-grade features histologically,
but the rates of recurrence are not affected. Cite this article:
We reviewed the outcome of 69 uncemented, custom-made,
distal femoral endoprosthetic replacements performed in 69 patients
between 1994 and 2006. There were 31 women and 38 men with a mean
age at implantation of 16.5 years (5 to 37). All procedures were
performed for primary malignant bone tumours of the distal femur.
At a mean follow-up of 124.2 months (4 to 212), 53 patients were
alive, with one patient lost to follow-up. All nine implants (13.0%)
were revised due to aseptic loosening at a mean of 52 months (8
to 91); three implants (4.3%) were revised due to fracture of the
shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic
fracture. Bone remodelling associated with periosteal cortical thinning
adjacent to the uncemented intramedullary stem was seen in 24 patients
but this did not predispose to failure. All aseptically loose implants
in this series were diagnosed to be loose within the first five
years. The results from this study suggest that custom-made uncemented
distal femur replacements have a higher rate of aseptic loosening
compared to published results for this design when used with cemented
fixation. Loosening of uncemented replacements occurs early indicating
that initial fixation of the implant is crucial. Cite this article:
The management of failed autologous chondrocyte
implantation (ACI) and matrix-assisted autologous chondrocyte implantation
(MACI) for the treatment of symptomatic osteochondral defects in
the knee represents a major challenge. Patients are young, active
and usually unsuitable for prosthetic replacement. This study reports
the results in patients who underwent revision cartilage transplantation
of their original ACI/MACI graft for clinical or graft-related failure.
We assessed 22 patients (12 men and 10 women) with a mean age of
37.4 years (18 to 48) at a mean of 5.4 years (1.3 to 10.9). The
mean period between primary and revision grafting was 46.1 months
(7 to 89). The mean defect size was 446.6 mm2 (150 to
875) and they were located on 11 medial and two lateral femoral condyles,
eight patellae and one trochlea. The mean modified Cincinnati knee score improved from 40.5 (16
to 77) pre-operatively to 64.9 (8 to 94) at their most recent review
(p <
0.001). The visual analogue pain score improved from 6.1
(3 to 9) to 4.7 (0 to 10) (p = 0.042). A total of 14 patients (63%)
reported an ‘excellent’ (n = 6) or ‘good’ (n = 8) clinical outcome,
5 ‘fair’ and one ‘poor’ outcome. Two patients underwent patellofemoral
joint replacement. This study demonstrates that revision cartilage
transplantation after primary ACI and MACI can yield acceptable
functional results and continue to preserve the joint. Cite this article:
Metal-on-metal (MOM) total hip arthroplasty using large diameter femoral heads offer clinical advantages however the failure rates of these hips is unacceptably high. Retrieved hips have a wide range of wear rates of their bearing and taper surfaces and there is no agreement regarding the cause of failure. Detailed visual inspection is the first step in the forensic examination of failed hip components and may help explain the mechanisms of failure. The aim of this study was to determine if there was a correlation between the results of detailed inspections and the volumetric wear of the bearing and taper surfaces of retrieved hips. Detailed, non-destructive macroscopic and stereomicroscopic examinations of 89 retrieved MOM hip components were performed by a single experienced examiner using quantitative assessment to document the severity of 10 established damage features: Light scratches, Moderate scratches, Heavy scratches, Embedded particles, Discolouration, Haziness, Pitting, Visible wear zone, Corrosion, Fretting Each surface was considered in terms of zones comprising of quadrants (cup, head, and taper) and subquadrants (cup and head), Figure 1. Each zone was scored on a scale of 0 to 3 by determining the percentage of the surface area of the zone that exhibited the feature in question: a score of 0=0%, 1<25%, 25%<2<75%, 3>75%. The sum of the scores of each zone was used for the assessment of each damage feature. The volume of wear at the surfaces of each hip was measured with a Zeiss Prismo coordinate measuring machine (cup and head) and a Talyrond 365 roundness measurement instrument (taper), using previously reported methods1, 2. Simple linear regression models were used to asses the univariable associations between the inspection scores and wear volumes. Multiple linear regression models were subsequently used to asses the simultaneous contribution of the inspection scores, found significant in univariable analyses, on the wear outcome variables. All statistical analysis was performed using Stata/IC version 12.1 (StataCorp, USA) and throughout a p value < 0.05 was considered statistically significant.Introduction
Method
There has been widespread concern regarding the adverse tissue reactions after metal-on-metal (MoM) total hip replacements (THR). Concerns have also been expressed with mechanical wear from micromotion and fretting corrosion at the head/stem taper junction in total hip replacements. In order to understand the interface mechanism a study was undertaken in order to investigate the effect of surface finish and contact area associated with modular tapers in total hip replacements with a single combination of materials of modular tapers. An inverted hip replacement setup was used (ASTM F1875-98). 28 mm Cobalt Chrome (CoCr) femoral heads were coupled with either full length (standard) or reduced length (mini) 12/14 Titanium (Ti) stem tapers. These Ti stem tapers had either a rough or smooth surface finish whilst all the head tapers had a smooth surface finish. Wear and corrosion of taper surfaces were compared after samples were sinusoidally loaded between 0.1 kN and 3.1 kN for 10 million cycles at 4 Hz. In test 1 rough mini stem tapers were compared with rough standard stem tapers whilst in test 2 rough mini stem tapers were compared with smooth mini stem tapers. Surface parameters and profiles were measured before and after testing. Electrochemical static and dynamic corrosion tests were performed between rough mini stem tapers and smooth mini stem tapers under loaded and non-loaded conditions.Introduction:
Methods:
It has been suggested that corrosion and fretting at the taper junctions of stemmed metal-on-metal hip replacements may contribute to their high failure rates. A peer-reviewed semi-quantitative scoring system [Goldberg et al., 2002] has been used to visually assess the severity of corrosion and fretting of the taper junction but has not been validated using multiple examiners. The aim of this study was to assess the inter-observer variability of this method. Macroscopic and stereomicroscopic examinations of the femoral head and stem tapers of 100 retrieved large diameter metal on metal (MOM) hip components were performed by two independent observers using the methods defined by Goldberg et al. [2002] to quantify corrosion and fretting. Scores ranging from 1 (none) to 4 (severe) were assigned to the medial, lateral, posterior and anterior quadrants of the neck taper and the distal and proximal regions of the head taper. An overall score was then assigned to each surface as a whole. Cohen's weighted Kappa statistic (κ) was used to measure the inter-observer agreement. A quadratic weighting scheme, that allocated weights to the importance of disagreements that are proportional to the square of the number of categories apart, was used to take account of scaled disagreement. Kappa values were assessed using previously established criteria where κ ≤ 0 = poor, 0.01 to 0.20 = slight, 0.21 to 0.40 = fair, 0.41 to 0.60 = moderate, 0.61 to 0.80 = substantial, 0.81 to 1 = almost perfect. A sample size of 100 was used in order to detect a coefficient of 0.60 to within 0.25 with 95% confidence with two experienced observers. Statistical analysis was performed using Stata/IC version 12.1 (StataCorp, College Station, TC, USA) and a p value < 0.05 was considered statistically significant.Introduction
Method