Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 71 - 71
23 Jun 2023
Sedel L
Full Access

Since 1977 we did implant ceramic on ceramic THR in younger and active population. In 1991 we published (JBJS B N°4) data's in a consecutive group of patients under 50 years of age.

What about the same patients for more than 30 years? Eighty-six hips in 75 patients, 41 males 34 females, mean age 43 (18 to 50), mean weight 68 kg (36 to 100), Charnley class: 38 A, 28 (38 hips) B, 9 patients (10 hips) C. Sixty-six primary procedures, 20 revisions (18 failed arthroplasties: 6 THR, 5 resurfacing, four single cup, two hemiarthroplasty, one bipolar), one after acetabular fractures. Four hips previously infected.

Eight Patients deceased (8 hips) prosthesis still in place, ten lost to follow-up before 2 years, eight hips in 8 patients were revised before the review, partially followed: from 2 to 20 years: 35, completely followed: 25 hips in 23 patients resuming in: No pain in 20, slight pain in 2, severe disability in 3 not related to the hip, no radiolucent lines in 22, radiolucent lines in 3, no osteolysis in 25.

Revision for: early sepsis in one, socket loosening in 8 (3 revisions cases), femoral head fractures in 2: one extra small head (22mm) for Crowe 3 DDH, one fractured at 24 years.

Inertness, stability related to fibrous tissue generation, no noise.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 144 - 144
1 Nov 2021
García-Rey E Saldaña-Quero L Sedel L
Full Access

Introduction and Objective

Despite pure alumina have shown excellent long-term results in patients undergoing total hip arthroplasty (THA), alumina matrix composites (AMCs) composed of alumina and zirconium oxide are more commonly used. There are no comparative studies between these two different ceramics. We performed a retrospective case-control study to compare results and associated complications between AMC from two manufacturers and those with pure alumina from another manufacturer.

Materials and Methods

480 uncemented THAs with ceramic on ceramic (CoC) bearing surfaces (288 men and 192 women; mean age of 54.1 ± 12.4 years), were implanted from 2010 to 2015. Group 1: 281 THAs with pure alumina; Group 2A: 142 with AMC bearing in a trabecular titanium cup. Group 2B: 57 hips with AMC bearing with a porous-coated cup.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 18 - 18
1 Nov 2021
Sedel L Bizot P Garcia-Cimbrelo E
Full Access

Fracture risks are the most common argument against the use of Ceramic on ceramic (CoC) hip implants. Question: is ceramic material at risk in case of severe local trauma?

Over a long period, we tried to identify patients with a CoC prosthesis (Ceraver Osteal°)who did sustain a trauma. This was conducted in three different institutions.

Eleven patients were found: 9 males and 2 females aged 17 to 70 years at time of index surgery. Accident occurred 6 months to 15 years after index: one car accident, five motorcycle accident, five significant trauma after a fall, including one ski board accident. Consequences of these trauma were: six fractures of the acetabulum with socket loosening in 4 that needed revision, two femoral shaft fracture, one orifed and one stem exchanged, one traumatic hip dislocation associated to loosening of the socket revised at 10 years, and one traumatic loosening of the socket. Ten had no consequence on ceramic integrity. One experienced a fracture of the patella from a dashboard trauma, a liner shipping was discovered during socket revision 2 years later. This is the only case of possible relation between trauma and ceramic fracture. In a more recent longitudinal study on 1856 CoC prosthesis performed from 2010 to 2021, 29 severe traumas were identified with no consequence on Ceramic material.

From this limited case study, it can be assumed that Pure Alumina Ceramic well designed and manufactured, will not break after a significant trauma.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 4 - 4
1 Apr 2018
Pitto R Sedel L
Full Access

The objective of this study was to determine whether the bearing surface is a risk factor for revision after late dislocation in total hip arthroplasty (THA).

Data from primary THAs were extracted from the New Zealand Joint Registry over a 13-year period. The mean age of patients was 68.9 years; 53.2% were female. The surgical approach used was posterior in 66% of THAs, lateral in 29% and anterior in 5%. There were 53,331 (65.1%) metal-on-polyethylene THAs, 14,093 (17.2%) ceramic-on-polyethylene, 8,177 (10.0%) ceramic-on-ceramic, 461 (0.5%) ceramic-on-metal, 5910, and (7.2%) metal-on-metal.

The primary endpoint was late revision for dislocation, with ‘late’ defined as greater than one year post-operatively. 73,386 hips were available for analysis. The overall revision rate was 4.3% (3,130 THAs), 1.1% (836) were revised for dislocation. Only 0.65% (470) hips were revised for dislocation after the first post-operative year.

The unadjusted hazard ratios (HR) showed significantly higher rates of revision for dislocation in ceramic-on-polyethylene (HR 2.48; p=0.001) and metal-on-polyethylene (HR 2.00; 95% p =0.007) compared to ceramic-on-ceramic. However, when adjusted for head size, age and surgical approach, only ceramic-on-polyethylene (HR 2.10; p=0.021) maintained a significantly higher rate of revision, whereas metal-on-polyethylene approached significance (HR 1.76; 95% p = 0.075).

In New Zealand, dislocation is the most common reason for revision, ahead of aseptic loosening of the acetabular component. The relationships between bearing materials and risk of revision for late dislocation is controversial. However, in this study ceramic-on-ceramic shows lower risk rates for revision than other bearing surface combinations. Low wear and less debris, limited peri-articular inflammatory reaction and an healthy fibrotic pseudo-capsule are potential factors determining long-term stability of the hip joint.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 58 - 58
1 Apr 2018
Garcia-Rey E Garcia-Cimbrelo E Sedel L
Full Access

Introduction

Acetabular fractures management is controversial since, despite a good anatomical reduction, clinical outcome is not satisfactory very often and the probability of a total hip arthroplasty (THA) is high. Surgical treatment include long operating times, large approach, blood loss, neural and muscle damage, and a high risk of failure and secondary osteoarthritis related to bone necrosis, cartilage damage, and bone loss.

We hypothesized that the acetabular fracture management affected the clinical and radiological outcome of THA after posttraumatic arthritis.

Materials and Methods

We compared 49 patients (49 hips) initially treated conservatively followed some months later by THA in conjunction with acetabular reconstruction (group 1); and 29 patients (29 hips) who had undergone THA after a failed osteosynthesis (group 2). There were more associated fractures according to Letournel in group 2. The mean age was 59.3±15.8 years for group 1 and 52.9±15.2 years for group 2. The mean delay between fracture and THA was 75.4±5 months for group 1 and 59.4±5 for group 2. The mean follow-up was 11.7 in group 1 and 10.2 in group 2. Preoperative bone defect was similar. We used bone autograft in 13 hips (26.5%) in group 1 and four (13.6%) in group 2. We used acetabular reconstruction plates in 2 hips with a pelvic discontinuity in group 1. Complications, clinical outcome according to Harris Hip Score, and radiological reconstruction were compared. Two-way ANOVA with repeated measures were used for comparison.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 49 - 49
1 Jan 2018
Garcia-Cimbrelo E Garcia-Rey E Sedel L
Full Access

Different pathologies, deformities, bone defects, previous surgeries and polyethylene wear limit the survival of total hip arthroplasty (THA) in young patients. We compare preoperative status and outcome in 171 young and very young patients (207 hips) who underwent a ceramic-on-ceramic THA. Sixty-three (77 hips) were less than 30 years old (group 1) and 108 (130) were between 30 and 40 years old. Mean follow-up was 11.3 years. Two-way ANOVA with repeated measures were used to analyse clinical and radiological changes.

Juvenile rheumatoid arthritis (JRA) was the most frequent diagnosis in group 1 and avascular necrosis in group 2. Charnley class type C, low activivity, previous surgery and osteoporotic bone were more frequent in group 1 patients. There were 2 cup revisions for aseptic loosening in group 1 and 4 in group 2. Survivorship analysis at 15 years was 96.7% (95% IC 92.2 a 100) for group 1 and 96.1% (95% IC 92.2 to 100) for group 2 (p=0.749). Despite the worse preoperative status in group 1 patients, clinical outcome was similar in both groups. Severe dysplasia had the worst clinical result and avascular necrosis the best. Patients with Charnley class C and JRA showed the most improvement. Radiographic reconstruction of the abduction angle was worse in group 1 (p=0.02). No osteolysis or complications derived from ceramic use were found.

We conclude that despite the worse preoperative status in group 1 patients, clinical and radiographic results were good in both groups of patients who received a ceramic-on-ceramic THA.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 64 - 64
1 Jan 2018
Sedel L Odri G Sanchez J Sverzut J Bizot P Laredo J
Full Access

Recent studies about hip stability after total hip arthroplasties (THA) concerned differences regarding bearings: ceramic on ceramic (CoC) presenting less dislocations on the long term compared to metal or ceramic on polyethylene. The hypothesis is a difference in the healing process of periarticular tissues, with a stronger fibrous tissue for the first one, and more foreign body reaction, joint effusion with the others.

NMR Imaging of the pelvis showing both hips using novel MR MAVRIC program for metal artefacts suppression, were performed in 10 patients, 15 THA and 2 non-pathological contralateral hips. Eight hips had CoC bearings, 3 of which were impacted cementless bulky ceramic implant, and 5 had a metal back. 7 hips had CoP bearings, 4 of which were cemented.

Native capsules showed a mean thickness of 6.6mm. For CoC bearings, capsule thickness ranged from 7mm to 9.6 mm with a mean thickness of 8mm. For CoP bearings, capsule thickness ranged from 3mm to 8.4mm, with a mean thickness of 6.1mm. Neocapsule appeared clearly in all COC bearings observed, while for CoP, sometimes it was less dense with fatty aspect, 3 hips out of 7 having a very thin capsule under 4mm.

It is possible to observe and quantify new capsule after THR and measure differences although not significant regarding bearings on limited number of samples. More patients might be included, but the tendencies observed here might explain better long term stability in vivo observed with Coc.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 143 - 143
1 Mar 2017
Sedel L
Full Access

Introduction

In the year 1977 we started to use ceramic on ceramic total hip (Ceraver*). The prosthesis was cemented on both sides. Ceramic was medical grade with relative large grains and high porosity. The stem was made of titanium alloy smooth and oxidized.

In 1990 we published the results of 86 hips in 75 patients who were less than 50 years of age at time of operation (1). Recently we tried to reach the same patients, looking specifically to those who could have more than 20 or 30 years follow-up.

Material and methods

This study design included all patients operated between 1977 and 1986 and having less than 50 years of age at time of surgery. Eighty six hips in 75 patients, 34 females and 41 men. Mean age was 43 (from 18 to 50) nine had a BMI in excess. 66 hips were performed primaries while 20 consisted in revision procedure including 6 total hip revisions, 5 resurfacing, 4 single cup, 3 hemiarthroplasty, and 2 acetabula fractures. Four of these had a past history of infection. Preoperative diagnosis were secondary osteoarthritis in 41, AVN in 26, primary OA in 3, Rheumatoid arthritis in 12, tuberculosis in 2 and hemophilic in two.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 322 - 322
1 Mar 2013
Sedel L
Full Access

Starting in 1977 a new cemented stem made of titanium alloy (with vanadium) was designed regarding some principle: rectangular shape, smooth surface covered with thin layer of titanium oxide, filling the medullar cavity. As a consequence: a thin layer of cement. It was designed with a collar. Initial Cementing technique used dough cement, vent tube and finger packing; then we applied cement retractor low viscosity cement and sometimes Harris Syringe. At the moment we went back to initial technique plus a cement retractor made of polyethylene. Many papers looked at long term follow up results depicting about 98 to 100 percent survivors at 10 years and 95 to 98% at 20 years (Hernigou, Hamadouche, Nizard, El Kaim).

Clinical as well as radiological results are available.

Radiological results depicted some radiolucent lines that appeared at the very long term. They could be related to friction between the stem and the cement. As advocated by Robin Ling, he called “French paradox” the fact that if a cemented prosthesis is smooth and fills the medullary cavity, long term excellent results could be expected.

This was the case with stainless steel Kerboull shape, the Ling design (Exeter)and the Ceraver design.

The majority of these stems were implanted with an all alumina bearing system. And in some occasion, when revision had to be performed, the stem was left in place (108 cases over 132 revisions)

Our experience over more than 5000 stems implanted is outstanding (see figure 1: aspect after 30 years).

Discussion other experience with cemented titanium stem were bad (Sarmiento, Fare). We suspect that this was related either to the small size of this flexible material, or to the roughness of its surface.

If one uses titanium cemented stem it must be large enough and extra smooth.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 532 - 533
1 Nov 2011
Nich C Marchadier A Sedel L Petite H Hamadouche M
Full Access

Purpose of the study: Oestrogen depletion leads to osteoclastic hyperactivity and subsequent postmenopausal osteoporosis. Little is known about interactions with bone absorption induced by wear particles from joint bearings. The purpose of this study was to evaluate bone response to polyethylene (PE) particles in a mouse model of oestrogen deficiency.

Material and methods: Particles of PE were implanted in the calvaria of seven non-ovariectomised mice and in seven ovariectomised mice (OVX). Fourteen mice were operated on without implantation of the particles (7 non-OVX and 7 OVX, control groups). The mice were sacrificed at two weeks. The crania were studied under a microscanner and histologically without decalcification.

Results: The microscanner showed that particles of PE induced a significant decrease in bone thickness in non-OVX mice (p=0.04), while the thickness remained unchanged in OVX mice who had received the particles (p=0.40). After implantation of the PE particles, the number of osteoclasts per mm of bone perimeter was 2.84±1.6 in the non-OVX mice and 1.74±1.3 in OVX mice (p=0.004). Compared with controls, the mean loss of bone was 12±10% in the non-OVX mice versus 4.7±0.9%in the OVX mice (p=0.004).

Discussion: The volume of osteolysis induced by PE particles was smaller in OVX mice compared with non-OVX mice.

Conclusion: These results suggest that a deficit in oestrogens has a protective effect against bone adsorption induced by PE particles.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 431 - 432
1 Nov 2011
Nizard R Cogan A Hannouche D Raould A Sedel L
Full Access

Hard-on-hard bearing surface have been accepted as a valuable alternative for young and active patients needing a hip replacement because these combinations are resistant to wear. Initial development of alumina-on-alumina bearings faced complications such as fractures, and socket loosening. But, with the increasing number of prostheses implanted, noise occurrence appeared as a new complication. The primary aim of the present survey was to quantify the prevalence of having noise in a population receiving alumina-on-alumina hip arthroplasty.

Two hundred and eighty-four ceramic-on-ceramic hips were performed in 238 patients (126 males and 112 females) from January 2003 to December 2004. The average age at the index operation was 52.4 ± 13.4 years (range, 13 to 74 years). We used the same type of prosthesis for all patients manufactured in all cases by Ceraver-Osteal®. Clearance between femoral and insert was between 20 and 50 microns in order to achieve minimal wear. The survey was conducted by an independent surgeon who did not participated in patients care during the last 6 months of 2007. He interviewed the patients by phone with a standardized questionnaire (appendix) that aimed to assess if noise was present and the characteristics of this noise if present. No suggestion was done on how they could describe the noise and they felt free to use the word that they considered to be the most adapted. Satisfaction was evaluated asking if the patient was very satisfied, satisfied or dissatisfied with its prosthesis.

When the noise was present, the X-ray was independently evaluated to assess if sign of component fracture was present.

Four patients (six hips) died of unrelated cause during the follow-up. Three patients (three hips) lived outside France and could not be followed (1.3%). Nine patients (ten hips) could not be traced and are considered lost to follow-up (3.8%). Two hundred and twenty-two patients with 265 hips were therefore surveyed. Among these 265 hips, 28 experienced noise generation (10.6%). It was defined as a snap for 6 patients, as a cracking sound by 6, as rustling by 6 patients, as a squeaking by 7 patients (2.6%), a tinkling by 2 patients, one patient was unable to define the sound she felt. No factor related to the patient influenced the occurrence of noise. Twelve patients were dissatisfied with the result of the hip prosthesis, 5 of them experienced noise (41.7%); 210 were satisfied or very satisfied 23 of them experienced noise (11%); this difference was significant (p=0.002). No patients required revision for noise.

The origins of noise occurrence are unknown but several hypotheses can be suggested.

Squeaking may be due to absence of sufficient lubrication. Other types of noise can be due to microseparation, occult dislocation, impingement between the femoral neck and the acetabular rim but demonstration remain an issue.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 445 - 446
1 Nov 2011
Sedel L Delambre J Nizard R Hannouche D
Full Access

Introduction: While an increasing number of authors have reported on the long-term results of primary alumina total hip arthroplasty (THA) [1], strategies for revising a ceramic-on-ceramic THA are debated in the literature. According to some authors [2], the reimplantation of a ceramic head on a well-fixed femoral stem is inadvisable, as it may lead to a fracture of the newly implanted head. The aim of the present study was to evaluate the incidence of this specific issue, and to report on the clinical and radiological results of the revised hips.

Methods: Between January 1977 and December 2005, 138 consecutive alumina-alumina revision hip arthroplasties were performed in 127 patients. There were 79 women (62.2%) and 48 men (37.8%), with an average age of 67 years (range, 32–91 years). Among these, an isolated acetabular revision was performed in 108 cases. The reason for revision was aseptic loosening of the acetabular component in 98 hips, pain in 7, fracture of an alumina liner in 2, and recurrent dislocation in 1. The revised socket was a cemented alumina in 56 hips, a threaded screw-in titanium with an alumina core in 34, a pressfit titanium with an alumina core in 11, and bulk alumina in 7. Acetabular bone stock losses were classified according to the AAOS system. Most of the hips had a contained type II defect (86%). In all cases, the femoral stem was left in place and the acetabular component alone was revised. At the time of revision surgery, an aluminaalumina combination was implanted in 27 hips, an aluminapolyethylene combination in 56, a metal-poly-ethylene combination in 15, and a zirconia-polyethylene in 10. Overall, a ceramic head was reimplanted on a used femoral taper in 59 cases. Acetabular reconstruction with allografts supported with the Kerboull acetabular reinforcement device was performed in 31% of the hips.

Results: The mean follow-up period was 78 ± 37 months. Thirteen patients (15 hips) died a mean 37 months after surgery. Sixteen patients were lost to follow-up. Postoperatively, five hips had a recurrent dislocation, 2 a deep infection, 9 a trochanteric nonunion (21,3%), 6 a transient nerve palsy. 18 hips required a re-revision surgery, 12 of which for aseptic loosening of the acetabular component. Among the 59 ceramic heads implanted on a well-fixed stem, no fracture of the head occurred at a mean 81 months follow-up. Of the original 108 hips, 77 were available for clinical evaluation and 75 for radiological evaluation at least 2 years after surgery. The mean Merle d’Aubigné score increased from 10.1 ± 2.1 to 16.7 ± 1.1 at the latest follow-up (p< 0.001). Forty-five hips were graded excellent or very good (60%), 26 good (34,6%), 3 fair (4%), and 1 poor (1.3%). When revision for aseptic loosening was considered as a failure, the overall survival rate at 8 years was 96.3 ± 1.8%.

Discussion & Conclusions: In the present study, aseptic loosening of the acetabular component was the main reason for revision surgery. Osteolysis around ceramic implants was moderate and was related to the migration of the socket. Among the ceramic heads implanted on a used titanium trunnion, no fracture was observed. This approach is possible, in so far as careful inspection does not show any major imperfection of the morse taper [3]. As for other bearing surfaces, the management of aseptic loosening of al-al prostheses is based on the amount of osteolysis around the loosened socket.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 89 - 89
1 Mar 2010
Sedel L
Full Access

Femoral revision is frequent, due to femoral loosening, thigh pain, recurrent dislocation, osteolysis or sepsis. Whatever the reason, with the exception of some difficult septic cases, our strategic approach is similar. Some of our expertise concern femoral stem retrieval. Our reconstruction strategy is different if we are revising total hip in active and young patient or if it is an old and inactive one.

First step is always an large “en bloc” tissue excision. For old and inactive, it is sometimes possible to retain the stem if not loosed and perform a “in cement” cementation; In this group we select usually metal or alumina on polyethylene couple and cemented implants; In young and active, we select alumina on alumina combination which resumed in cementless acetabular fixation, and cementless or cemented stem.

Stem retrieval of a well fixed cementless stem is performed via a large transtrochanteric approach associated with a transfemoral one. Repair is performed using cerclage and long cemented stem.

Cement retrieval is performed since 9 years using Ultra sound (Oscar*) material, which in our hand is very successful specially for cement retractor retrieval. Then medullary canal is reamed in order to get a bloody healthy bone receive either a cemented or sometimes a cementless stem, depending on the bone quality.

To compensate femoral bone destruction and enhance cemented stem fixation, we used a modified Ling technique replacing allogenic morcellised bone by hydroxyapatite granules. Granules of 5 mm in diameter are made of 70% HA and 30% of β TCP. Mechanical resistance is excellent and biological activity is high. Thus stem stability can be obtained easily. This can be done either with a cemented or a cementless stem (about 60 cases).

In case of very severe bone loss and osteolysis, we performed massive allogenic bone transplant associated with long cemented stem and distal HA granules with cement.(17 cases).

As we usually performed one stage revision for septic cases, strategy is not different; It is only in selected cases with many sepsis recurrence and specially aggressive bacteria that we performed a two stage procedure.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 253 - 253
1 Jul 2008
FERREIRA E POTIER E MIR L PETITE H SEDEL L
Full Access

Purpose of the study: Several studies have demonstrated the usefulness of mesenchymatous stem cells (MSC) for cell therapy aimed at favoring bone tissue healing. Bone morphogenesis proteins (BMP) orient MSC towards osteoblastic differentiation. Since they are rapidly degraded in the organism, these proteins require a continuous release system to potentialize their biological activity in a controlled localized manner. We evaluated the usefulness of using the electroporation technique to insert a BMP transgene into the MSC of rats to enable sufficient transient expression of BMP genes to enable satisfactory bone healing. We first developed electroporation conditions for rat MSC and checked cell viability after the electric shock. Secondly, in order to obtain quantitative and/or temporal BMP expression, we tested the influence of different promoters on transcription actvity.

Material and methods: To determine the electroporation parameters, MCS were transfected with the pCMV-LacZ plasmid using two electric impulsions: a series of eight 100 impulsions/μs at high voltage (900-170V/cm) followed or not by a series of eight 12.5 ms low-voltage impulsions (60 V/cm). After determining the electroporation conditions, six plasmids carrying different promoters were electroporated.

Results: The best transfection rate in rat MSC was obtained with a series of 8 impulsions at 1500 V/cm. Before the electrical shock, the suspended rat MSC had to be incubated at ambient temperature to favor cell survival. Proliferation of electroporated cells was comparable to that of non electroporated cells. Surprisingly, addition of low-voltage pulses significantly decreased the efficacy of transfection. In addition, MSC transfected with the promoters GAPDH and beta-actin presented a beta-galactoside activity (at 48 h) superior to that obtained with the pCMV promoter.

Discussion: After optimization of these parameters, we demonstrated that MCS can be effectively transfected by electroporation. The following steps will be to check for long-term expression of beta-galactoside by electroporated MSC, transfection of MSC with plasmids or the BMP-2 gene controlled by these same promoters and monitoring promoter activity as a function of the stage of MSC differentiation.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 254 - 254
1 Jul 2008
VIATEAU V GUILLEMIN G CALANDO Y OUDINA K SEDEL L HANNOUCHE D PETITE H
Full Access

Purpose of the study: The objective of this study was to establish an experimental sheep model for a surgical procedure which has been clinically successful for repairing major loss of bone stock: the Masquelet technique.

Material and methods: A 25 mm bone defect was created in a metatarsal bone then filled with a cement filler. After six weeks, the cement was removed after opening the neoformed pseudosynovial membrane. The cavity was left empty in group 1 (n=6) or filled with a morcelized cancellous autograft harvested from the iliac crests in group 1 (n=6).

Results: The surgery was well tolerated in all animals which were able to used the injured limb the day after the operation. Radiographic images and histological findings 24 weeks after surgery demonstrated that healing had not been achieved in all of the animals in group 1. Inversely, healing was achieved in all animals in group 2 at 24 weeks. Immunohistochemistry of the neoformed pseudosynovial demonstrated :

an abundant vascular network,

presence of cells expressing transcription factor CBFA1,

very few inflammatory CD14+ cells (macrophages),

an extracellular matrix positive for type I collagen.

Conclusion: The sheep metatarsal model is a model of critical size with low morbidity. This model could be used to:

evaluate new therapeutic strategies for bone regeneration in conditions close to clinical situations,

study the role of the membrane in bone repair.

The presence of a pseudosynovial membrane might:

be a barrier against the diffusion of bone morphogenetic proteins outside the lesion and

potentially be a reservoir of stem and vascular cells which could be useful for new technologies.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 280 - 280
1 Jul 2008
HAMOU C HANNOUCHE D RAOULD A NIZARD R SEDEL L
Full Access

Purpose of the study: Complex fracture-dislocation of the elbow, and subsequent surgical treatment, is often the source of a stiff joint. The purpose of this study was to assess the efficacy of a therapeutic protocol combining systematic insertion of a dynamic external fixator allowing early mobilization of the elbow with restitution of the radial height and the coronoid process.

Material and methods: This consecutive series of ten patient, six men and four women, mean age 49 years, age range 27–67 years, underwent surgery from 2002 to 2004. Three patients presented a posterior Monteggia fracture (two type IIA, one type IId associated with comminutive fracture of the trochlea), four patients presented a dislocation associated with a Masson 4 fracture of the radial head and two presented inveterated dislocations diagnosed three weeks after the traumatic event. In all, seven patients presented a fracture of the radial head and six a fracture of the coronoid process. For all patients, the operation consisted in stabilization with a dynamic external fixator of the elbow associated or not with restoration of the radial height with a radial head prosthesis (n=4) and reconstitution of the coronoid process (n=6). The lateral ligaments had to be reinserted in four elbows. The comminutive fracture of the rochlea was treated with an iliac crest graft.

Results: One patient died early. One patient presented pin tract infection and four developed heterotopic ossifications. At mean follow-up of twelve months, outcome was excellent in our patients, good in four, fair in one (Mayo clinic classification). All patients had a stable elbow. The mean range of motion was 89° flexion-extension and 145° pronationsupination.

Conclusion: In this series, systematic use of external fixation for complex fracture-dislocation of the elbow joint yielded satisfactory results when the element stabilizing the joint were appropriately restored and when rehabilitation was undertaken early.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 253 - 254
1 Jul 2008
POTIER E FERREIRA E BETTON O MEUNIER A SEDEL L PETITE H
Full Access

Purpose of the study: Cell therapy proposes to fill gaps left by bone stock loss using osteocompetent cells (mesenchymatous stem cells, MSC). Preclinical results have been promising but still require improvement particularly concerning stress to the MSC during in vivo implantation. Stress results from sudden transfer i) from oxygen medium (21% O2) to a hypoxic medium (0–5% O2 because O2 diffusion is limited to 200 mm from a blood vessel), ii) a cell support to an osteoconductor support, et iii) a rich medium (fecal calf serum, FCS) to a medium with a limited supply of nutrients, hormones and growth factors diffusing from the environing biological fluids. The purpose of this study was to evaluate in vitro the impact of these different factors on MSC survival.

Material and methods: Human MSC(hMSC) harvested from bone marrow (n= 5 donors) and sheep MSC (sMSC) obtained with a preclinical model (n = 5 animal donors) were exposed for 48 h(hMSC) or 72h (sMSC) to the following transfers: i) rich medium (10% FCS) to poor medium (1% FCS), ii) plastic support to osteo-conductor supports (alumina, calcium carbonate), and iii) oxygen medium (21% O2) to hypoxic medium (6% O2). sMSC were also exposed to prolonged hypoxia (48–120h). Cell death was determined using image analysis after live/dead cell staining.

Results: The results demonstrated that MSC are: i) sensitive to a decrease from 10% to 0% FCS; 14% death of hMSC and 17% death of sMSC), ii) sensitive to transfer onto osteoconductor supports (sMSC on calcium carbonate: 23%), iii) very sensitive to prolonged hypoxia (120h) when combined with decreased FCS (sMSC: 23%; hMSC: 98%). A complementary study on the influence of hypoxia on differentiation properties of surviving sMSC is under way.

Conclusion: If the in vivo results concord with the in vitro results, i.e. if massive cell death is observed 4 days after implantation due to hypoxia, the current transplantation conditions will have to be revisited. Acceleration of neovascularization of in vivo implants which would shorten the period of hypoxia should allow better survival of implanted sMSC.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 253 - 253
1 Jul 2008
DEGAT M DUBREUCQ G MEUNIER A DAHRI-CORREIA L PETITE H SEDEL L LOGEART-AVRAMOGLOU D
Full Access

Purpose of the study: Bone morphogenetic proteins (BMPs) are osteoinducing proteins which play a primordial role in bone repair. To obtain optimal mineralization in vivo, high doses of heparin binding growth factor must be used. Studies have demonstrated that functionalized dextranes (FD) present affinity for heparin binding growth factor. We studied the capacity of dextrane derivatives to interact with BMP-2 and potentialize its biological activity in vitro.

Material and methods: Different soluble FD were obtained by random substitution of carbosymethyl (CM), benzylamide (B) and sulfate (Su) groups on native dextrane chains. Gel electrophoresis was used to study the affinity of the anionic FDs for BMP-2. The effect of polymers on osteoinduction activity of BMP-2 was evaluated by histochemistry. ALP (an early marker) synthesized by mypoblasts C2C12 were dosed seven days after injection in presence of BMP-2 associated or not with polymers. IN addition, expression of osteocalcin (late marker) was quantified by RT-PCR.

Results: Electrophoresis demonstrated that DMCB and DMCBSu interacted with BMP-2. These interactions appeared to increase with B level but decreased with Su level. We worked with FD1, a DMCB with a high affinity for BMP-2. The ALP activity was clearly potentialized when BMP-2 was associated with heparin and even better with FD1. Expression of osteocalcin was also amplified with the FD1-BMP-2 association. The influence on the biological activity of BMP-2 of FD, presenting different degrees of substitution, was also tested. Only FDs containing a high concentration of B expressed affinity for BMP-2, potentializing the biological activity of the protein.

Discussion: Dextanes functionalized with a high rate of benzylamide substitution interact with BMP-2 while sulfate substitution limits such interaction. Only FDS which interact with BMP-2 can potentialize the protein’s biological activity in vitro. Two hypotheses can be put forward: i) FD presents BMP-2 to its receptor cell, ii) FD protects BMP-2 from proteolytic degradation or capture by antagonists. The capacity of FD1 to potentialize the biological activity of BMP-2 could be a way of reducing the quantity of growth factor needed for optimal bone repair.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 285 - 285
1 Jul 2008
NIZARD R LACHERE A RADMANESH O HANNOUCHE D SEDEL L
Full Access

Purpose of the study: Computer assisted surgery for total knee arthroplasty is widely used in Europe. The reliability of these systems appears to be very good with bone cuts within 3° of the planned mechanical axis. Nevertheless, the relationship between intraoperative measurements provided by the navigation system and the postoperative gonometry can be used to assess the quality of realignment. The purpose of this work was to determine this relationship.

Material and methods: Thirty-three knees operated on with the Navitrack® system were assessed. Two senior operaters performed all procedures. An Omnia® cemented prosthesis with an ultracongruent fixed plateau was used in all cases. Intraoperative measurements wer made with the definitive prosthesis after cement solidification without stress on the knee. Postoperative gonometry was undertaken when the intraoperative flexion had disappeared six weeks to six months after the operation. Gonometric measurements were made by an independent operator using a computerized system operating on digitalized x-rays. The gonometric protocol had been standardized previously and only the gonometric measurements in compliance with this protocol were retained for analysis. The difference between measurements was analyzed with the t test for paired variables. Search for correlations was also performed.

Results: On average the intraoperative deviation was 0.8±0.8° (3° valgus to 2.4° varus). The postoperative gonometry showed 1.7±1.1° (3.4° valgus to 4.3° varus). The mean difference between the intraoperative axis and the measured postoperative axis was significant (p< 0.0001). There was no significant correlation between intraoperative and postoperative measurements.

Discussion: Computer-assisted navigation systems have their limitations which should be measured. The present findings would demonstrate a significant difference of minimal amplitude between the intraoperative measurement and the postoperative gonometry. Although the clinical pertinence of this difference remains to be demonstrated, it must be kept in mind for safe use of these navigation systems.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 41 - 41
1 Mar 2006
Sedel L Jin Z Rieker C Grigoris P Roberts P
Full Access

Since 1977 we did implant all alumina (Al2O3) bearings total hip prostheses. A lot of data were documented concerning tissue reaction, in vivo wear behaviour, fractures, and clinical outcome. Ceramic materials retrieved at revision were analysed. In some cases, wear was as low as a few microns for a 15-year period in use. This is two thousand times less than a regular metal on polyethylene sliding couple. and 100 times less than a metal on metal prosthesis. Fracture mechanism is related to crack propagation into the material. During the first period, the fracture rate was in the range of 2%; it then dropped to less than 0.1 %. Few fractures could not be explained by technical or design mistakes. Clinical outcome: More than 4000 total hips in selected young and /or active patients were implanted. In a recently reported series of consecutive patients operated by P.Boutin during the year 1980, 118 hips in 106 patients were included. Mean age was 62 years. At the twenty-year follow-up evaluation, forty-five patients (fifty-one hips) were still alive and had not been revised, twenty-five patients (twenty-five hips) had undergone revision of either or both components, twenty-seven patients (thirty hips) had died from unrelated causes, and nine patients (twelve hips) were lost to follow-up. The mean Merle d’Aubigné hip score was 16.2 ± 1.8 at the latest follow-up. Survival of the cup at twenty years with revision for any reason as the end-point was 85.6 percent for cementless cups versus 61.2 % for cemented cups, respectively. Survival of the stem at twenty years with revision for any reason as the end-point was 84.9 % for cementless stems versus 87.3% for cemented stems. Wear of the prosthetic components was undetectable on plain radiographs. No fracture of the alumina socket or head was recorded. Another study concerned a more recent design of the socket which consisted in a metal back titanium alloy shell covered with a pure titanium mesh with an alumina liner. The nine year survival rate was 98.4% with revision for aseptic loosening as the end point. Conclusion This alumina on alumina bearing provides interesting results without any physical limitation specially in young and active patients.