Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 67 - 67
1 Feb 2020
Yoshida K Fukushima K Sakai R Uchiyama K Takahira N Ujihira M
Full Access

Introduction

Primary stability is achieved by the press fit technique, where an oversized component is inserted into an undersized reamed cavity. The major geometric design of an acetabular shell is hemispherical type. On the other one, there are the hemielliptical type acetabular shells for enhanced peripheral contact.

In the case of developmental dysplasia of the hip (DDH), the aseptic loosening may be induced by instability due to decreased in the contact area between the acetabular shell and host bone.

The aim of this study was to assess the effect of reaming size on the primary stability of two different outer geometry shells in DDH models.

Materials and methods

The authors evaluated hemispherical (Continuum Acetabular Shell, Zimmer Biomet G.K.) and hemielliptical (Trabecular Metal Modular Acetabular Shell, Zimmer Biomet G.K.) acetabular shells. Both shells had a 50 mm outer diameter and same tantalum 3D highly porous surface.

An acetabular bone model was prepared using a solid rigid polyurethane foam block with 20 pcf density (Sawbones, Pacific Research Laboratories Inc.) as a synthetic bone substrate. Press fit conditions were every 1 mm from 4 mm under reaming to 2 mm over reaming. To simulate the acetabular dysplasia the synthetic bone substrate was cut diagonally at 40°. Where, the acetabular inclination and cup-CE angle were assumed to 40° and 10°, respectively.

Acetabular components were installed with 5 kN by a uniaxial universal testing machine (Autograph AGS-X, Shimadzu Corporation).

Primary stability was evaluated by lever-out test. The lever-out test was performed in 4 mm undersized to 2 mm oversized reaming conditions. Lever out moment was calculated from the multiplication of the maximum load and the moment arm for primary stability of the shell. The sample size was 6 for each shell type.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 142 - 142
1 May 2016
Yoneo T Nakao M Sakai R Fukushima K Uchiyama K Takahira N Mabuchi K
Full Access

Introduction

On the basis of a proposal by Noble, the marrow cavity form can be classified into three categories: normal, champagne-fluted and stovepipe. In the present study, three typical finite element femoral models were created using CT data based on Noble's three categories. The purpose was to identify the relationship of stress distribution of the surrounding areas between femoral bone marrow cavity form and hip stems. The results shed light on whether the distribution of the high-stress area reflects the stem design concept. In order to improve the results of THA, researchers need to consider the instability of a stem design based on the stress distributioin and give feedback on future stem selection.

Methods

As analyzing object, we selected SL-PLUS and BiCONTACT stems. To develop finite element models, two parts (cortical bone and stem) were constructed using four-node tetrahedral elements. The model consisted of about 60,000 elements. The material characteristics were defined by the combination of mass density, elastic coefficient, and Poisson's ratio. Concerning the analysis system, HP Z800 Workstation was used as hardware and LS-DYNA Ver. 971 as software. The distal end of the femur was constrained in all directions. On the basis of ISO 7206 Part 4,8 that specifies a method of endurance testing for joint prostheses, the stem was tilted 10°, and a 1500 N resultant force in the area around the hip joint was applied to the head at an angle of 25° with the long axis. Automatic contact with a consideration of slip was used.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 76 - 76
1 May 2016
Nakao M Uchida K Sakai R Naruse K Takaso M Mabuchi K
Full Access

Background

Residual stress remains in bone tissues after press-fit-fixation of a joint prosthesis, recently employed for joint arthroplasty. The response of bone tissues to the residual stress is, however, unknown because it is not physiological. This unnatural stimulus may have adverse effects on bone tissues, including causing thigh pain or bone resorption. In the present study, we designed an experimental method to apply a stationary load from inside an animal femur using a loop spring of titanium alloy with super elasticity. The femoral response was assessed based on the migration of the wire into bone twelve weeks after implantation. As the results, wire migration was noted in 10 of 11 cases.

Methods

We developed a method using a loop spring made of super elastic titanium alloy, which can maintain sufficient stress in a rat femur for a prolonged period. This titanium alloy, which contains 43.94% titanium and 56.06% nickel, was supplied as a wire (WDL1, Actment Co., Ltd., Kasukabe, Japan). In the present study, an experimental method was designed to apply a stationary load from inside a rat femur by inserting a loop spring made of super elastic wire.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 73 - 73
1 May 2016
Tanaka K Sakai R Mabuchi K
Full Access

Introduction

Post cam is useful to realize the intrinsic stability of a posterior-stabilized (PS) knee prosthesis replaced for a case with the severe degeneration. Some retrieval studies reveal the ultrahigh molecular weight polyethylene (UHMWPE) deformation or severe failure of the tibial post of PS knee. Strength of the tibial post of available design is obviously insufficient to prevent the severe deformation. The large size post might, however, shorten the range of knee motion. Therefore, minimally required size of the post should be clarified for polyethylene inserts. In the present study, we performed finite element (FE) analysis assumed the mechanical conditions of a tibial post in a PS knee and aimed to design criterion of a post of polyethylene insert of a knee prosthesis.

Method

The shape of three commercially available knee prostheses, product A, B, and C was referred as PS knee prosthesis. The contour of the metallic femoral component and the UHMWPE insert were digitized by a computed tomography apparatus. Three dimensional finite elements were generated by modeling software (Simpleware, Ltd. UK) as four-node tetrahedral elements. In FE analysis, we used LS-DYNA ver.971 (Livemore Software Technology Corp. USA) as the software and Endeaver Pro-4500 (EPSON Corp. Japan) as the hardware. These bottoms of the tibial insert were fully constrained. The value of 30MPa was defined as yield stress of UHMWPE. 500N posterior load was applied to each femoral component at 10 degree hyperextension. Then, 1000N anterior load at 120 degree flexion, after tibial insert was located 10 degree internal rotation (Fig. 1). These loads were assumed to realize the two types of tibial post impingement under several kinds of knee motions. The distributed values of von Mises stress and plastic strain on the tibial post were shown as the results of the analysis.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 143 - 143
1 Jan 2016
Fukushima K Sakai R Uchiyama K Moriya M Yamamoto T Takahira N Mabuchi K Takaso M
Full Access

Introduction

According to proposal of Noble, the femoral bone marrow cavity form of patients who underwent Total Hip Arthroplasty (THA) can be classified under 3 categories; those are Stovepipe, Normal and Champagne-fluted. We developed typical sodium chloride femoral model was created by 3D prototyping technique. The purpose was to identify the relationship of pressure zone of the surrounding areas between femoral bone marrow cavity form and hip stem.

Materials and Method

As opponent clarified stem design concept Zweymüller type model was used. According to CT data with the patients who underwent THA, the sodium chloride femoral model was custom-made and selected as the representative model based on Noble's 3 categories. Eight models of each category were used to performed mechanical test.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 79 - 79
1 Jan 2016
Nakao M Fukushima K Sakai R Takahira N Uchiyama K Yamamoto T Moriya M Mabuchi K
Full Access

Introduction

On the basis of a proposal by Noble, the marrow cavity form can be classified into three categories: stovepipe, normal, and champagne-fluted. In the present study, three typical finite element femoral models were created using CT data based on Noble's three categories. The purpose was to identify the relationship of stress distribution of the surrounding areas between femoral bone marrow cavity form and hip stem. The results shed light on whether the distribution of the high-stress area reflects the stem design concept. In order to improve the results of THA, researchers need to consider the instability of a stem design based on the pressure zone and give feedback on future stem selection.

Methods

To develop finite element models, two parts (cortical bone and stem) were constructed using four-node tetrahedral elements. The model consisted of about 40,000 elements. The material characteristics were defined by the combination of mass density, elastic coefficient, and Poisson's ratio. Concerning the analysis system, HP Z800 Workstation(HP, Japan) was used as hardware and LS-DYNA Ver. 971 (Livermore Software Technology Corporation, USA) as software. The distal end of the femur was constrained in all directions. On the basis of ISO 7206 Part 4,8 that specifies a method of endurance testing for joint prostheses, the stem was tilted 10°, and a 500 N resultant force in the area around the hip joint was applied to the head at an angle of 25° with the long axis. Automatic contact with a consideration of slip was used. Von Mises stress during a 1.0 s period after loading was analyzed, and stress distribution in the stem and its maximum value were calculated.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 2 - 2
1 Jan 2016
Mabuchi K Uchida K Ito T Shimizu K Nakao M Naruse K Sakai R
Full Access

INTRODUCTION

It is generally accepted that strong hammering is necessary for the press fit fixation of a joint prosthesis. In this regard, large stress must remain within bone tissues for a long period. This residual stress is, however, some different from the feasible mechanical stimuli for bone tissues because that is stationary, continuous and directed from within outward unlike physiological conditions. The response on this residual stress, which may induce the disorder of the fixation of implant, has not been discussed, yet.

In the present study, we designed an experimental method to exert a stationary load from inside of a femur of a rat by inserting a loop spring made from a super elastic wire of titanium alloy. Response of the femur was assessed by bone morphology mainly about the migration of the wire into the bone twelve weeks after the implantation.

MATERIALS AND METHODS

We developed a method using a loop spring made of super elastic wire of titanium alloy, which can maintain sufficient magnitude of stress in a rat femur during the experimental period. The loop spring was fabricated with a wire of 0.4 mm diameter before the quenching process. Eleven Wistar rats of ten weeks old were used for the experiments. The loop spring was inserted the right femur, as shown in Figure 1. The left femur was remained intact. The compressive load was added from within outward of bone marrow when the spring was compressed with the insertion into a bone marrow of a rat femur, as shown in Figure 2. The average contact stress was calculated by dividing the elastic force by the spring and bone contact area. The contact stress was distributed from 62 to 94 MPa, which are sufficiently lower than the yield stress of cortical bone [1]. The assessment of bone morphology around the implanted loop spring was performed by micro-CT imaging after the twelve weeks of cage activity.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 62 - 62
1 Jan 2016
Tanaka K Hasegawa K Sakai R Mabuchi K
Full Access

Introduction

Post cam structure, which is the main structure of posterior-stabilized design (PS), is useful to realize the intrinsic stability of a knee prosthesis replaced for a case with the severe degeneration. A large size post might, however, shorten the range of knee motion. On the other hand, retrieval studies sometimes reveal the ultrahigh molecular weight polyethylene (UHMWPE) deformation or severe failure of the tibial post of PS knee. Strength of a tibial post of available design is obviously insufficient to prevent the severe deformation. Therefore, minimally required size of the post should be clarified for polyethylene inserts. In the present study, we performed finite element (FE) analysis assumed the mechanical conditions of a tibial post in a PS knee and aimed to design criterion of a post of polyethylene insert of a knee prosthesis.

Method

The shape of one commercially available knee prosthesis was referred as a posterior-stabilized knee prosthesis. The contour of the metallic femoral component was traced and digitized by hand. The contour of the UHMWPE insert was digitized by a micro computed tomography apparatus. Three dimensional finite elements were generated by a modeling software (Simpleware, Ltd. UK) as total 83000 four-noded tetrahedral elements. The bottom of the tibial insert was fully constrained. Load on femoral component was assumed to realize the tibial post impingement under several kinds of knee motions. Posterior load 100 N or 500N at the 10 degree hyperextension, anterior load 500N or 1000N during 120 degree flexion were applied (Fig. 1). The software of FE analysis was LS-DYNA ver.971 (Livemore Software Technology Corp. USA). The hardware was Endeaver Pro-4500 (EPSON Corp. Japan). The distributed values of von Mises stress and plastic strain of the tibial post were shown as the results of the analysis.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 247 - 247
1 Mar 2013
Sakai R Takahira N Uchiyama K Yamamoto T Fukushima K Tanaka K Uchijima D Itoman M Mabuchi K
Full Access

Fracture during total hip arthroplasty occurs partly because the acquisition of fixation at the time of stem implantation depends on the operator's experience and sensation due to the absence of definite criteria. Therefore, an objective evaluation method to determine whether the stem has been appropriately implanted is necessary. We clarified the relationship between the hammering sound frequency during stem implantation and internal stress in a femoral model, and evaluated the possible usefulness of hammering sound frequency analysis for preventing intraoperative fracture.

Three types of cementless stem were used. Orthopedists performed stem insertion using a procedure similar to that employed in routine operation. Stress was estimated by finite element analysis using the hammering force calculated from the loading sensor as a loading condition, and frequency analysis of hammering sound data obtained using a microphone was performed (Fig. 1).

Finite element analysis showed a decrease in the hammering sound frequency with an increase in the estimated maximum stress (Fig. 2, 3). When a decrease in frequency was observed, adequate hammering had already been performed to achieve press-fit stability. Therefore, there is a possibility that the continuation of hammering induces intraoperative fractures that become a problem. Based on the relationship between stress and frequency, the evaluation of changes in frequency may be useful for preventing the development of intraoperative fractures.

When a decrease in frequency is observed, the hammering force should be reduced thereafter. Hammering sound frequency analysis may allow the prediction of bone fractures that can be visually confirmed, and may be a useful objective evaluation method for the prevention of intraoperative bone fracture.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 195 - 195
1 Sep 2012
Uchijima D Hiraki Y Katori T Tanaka K Sakai R Mabuchi K
Full Access

Fluid film lubricating ability of a total hip prosthesis depends on the profile accuracies including surface-roughness or the sphericity of a head or a cup. Therefore, surface polishing is important. It was, however, difficult to polish the central portion of a cup or head using the conventional rotating machine. In the present study, we developed a polishing method combining a pendulum machine and a robotic arm. The effect of the accuracy improvement by this method was evaluated by the friction measurements on some test specimens.

Nine balls and a cup of Co-Cr-Mo alloy that were polished by a conventional process using a rotating machine were prepared for the prototype. The average diameter of the balls was 31.9648 mm with the sphericity of 0.0028 μm. The inside diameter of the cup was 31.9850 mm with the sphericity of 0.0044 μm. We combined a robotic arm and a pendulum apparatus to enable the further polishing. The ability of both automatic centering and change in the sliding direction was accomplished by this system. The sliding direction has been changed 180 times every ten degrees. The total distance of polishing was 120 m under vertical load of 100 N in a bath of saline solution containing abrasive grains of silicate of the diameter of 2μm. The surface roughness of the central portion of the cup, which is important area for the fluid film lubrication decreased from Ra 20.2 μm before the polishing to Ra 18.7 μm after the polishing.

A pendulum type friction tester was used for the assessment of the improvement of the lubricating ability by the polishing. The measurement was run over at 10 times under the conditions of the load of 600 N in a bath of saline solution. As the result, the frictional coefficients decreased from 0.1456–0.1720 before polishing to 0.1250–0.1300 after polishing. The polishing effect was, however, observed only at the specimens that radial clearances did not exceed the value of 50 μm.

The present results indicated that the surface polishing of the central portion of hip prostheses must improve the lubrication ability and the radial clearance before the finishing process should be chinked as possible.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 96 - 96
1 Mar 2010
Ogawa N Sakai R Sato K Obata S Itoman M Mabuchi K
Full Access

The primary fixation of cementless hip prostheses is related to the shape of the stem. When there is a complication of loading in several directions, the mechanical fixation of a hip stem is considered to provide good primary fixation. The purpose of this study was to evaluate whether the IMC stem with its characteristic fixation method, which was developed by a group at Kitasato University, contributes to primary fixation by finite element analysis.

Analysis was performed at a friction coefficient of 0.1 with automatic contact, under the restriction of the distal femoral end. The following three loading conditions were applied:

step loading of the joint resultant force in the region around the hip stem,

loading in the rotational direction, simulating torsion, and

loading of the femoral head equivalent to that during walking.

Micromotion of the IMC stem along the x-, y-, and z-axes direction was calculated by simulation, and the stress distributed on the stem and femur was determined.

Micromotion along the z-axis, which is a clinical problem in hip prosthesis stems, was lower in the IMC stem than in other stems reported. Micromotion of the stem along the z-axis was low, indicating a low risk of sinking. The interlocking mechanism, which is a characteristic of the IMC stem, functioned to suppress its micromotion, indicating that the locking method of this stem contributed to the stability. Since no stress concentration was detected, it was considered that there are no risks of breakage of the IMC stem and femur.

It was suggested that effective fixation of the finite element model of the IMC stem can be achieved because the micromotion and stress level are appropriate for primary fixation.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 152 - 152
1 Mar 2009
Matsuyama J Ohnishi I Sakai R Miyasaka K Harada A Bessho M Ohashi S Matsumoto T Nakamura K
Full Access

The most important issue in the assessment of fracture healing is to acquire information on the restoration of mechanical integrity of the bone. To measure bending stiffness at the healing fracture site, we focused on the use of echo tracking (ET) that was a technique measuring minute displacement of bone surface by detecting a wave pattern in a radiofrequency echo signal with an accuracy of 2.6 μ. The purpose of this study was to assure that the ET system could quantitatively assess the progress, retardation or arrest of healing by detecting bending stiffness at the fracture site.

With the ET system, eight tibial fractures in 7 patients with an average age of 37 years (range: 24–69) were measured. Two tibiae in 2 patients were treated conservatively with a cast, and 6 tibiae in 5 patients were treated with internal fixation (intramedullary nailing: 4, plating: 1, screw 1). Patients assumed supine position, and the affected lower leg was held horizontally with the antero-medial aspect faced upwards. The fibula head and the lateral malleolus were supported and held tight by a Vacufix ®. A 7.5 Hz ultrasound probe was placed on each antero-medial aspect of the proximal and distal fragments along its long axis. Each probe was equipped with a multi-ET system with 5 tracking points with each span of 10 mm. A load of 25 N was applied at a rate of 5 N/second using a force gauge parallel to the direction of the probe and these probes detected the bending angle between the proximal and distal fragments. An ET angle was defined as the sum of the inclinations of both fragments. In the patients treated with a cast, the contralateral side was also measured and served as a control. Fracture healing was assessed time sequentially with an interval of 2 or 3 weeks during the treatment.

None of the patients complained of pain, or no other complication related to this measurement occurred. In the patient (patient:M) treated with a cast, the ET angle exponentially decreased as time elapsed (y = 1.4035e-0.1053x, R = 0.9754) and the radiographic appearance showed normal healing. Including this case, in all patients with radiographic normal healing, the ET angle exponentially decreased. However, in patients with retarded healing (patient:N), the decrease of the angle was extremely slow(y = 0.2769e-0.0096x, R = 0.815). In patients with non union (patient:T), the angle stayed at the same level.

With this method, noninvasive assessment of bending stiffness at the healing site was achieved. Bending angle measured by ET diminished over time exponentially in patients with normal healing. On the contrary, in patients with healing arrest, no significant decrease of the bending angle was recognized. It was demonstrated that the echo tracking method could be applicable clinically to evaluate fracture healing as a versatile, quantitative and noninvasive technique.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 404 - 404
1 Apr 2004
Sakai R
Full Access

It is widely accepted that a wide contact area between bone and artificial materials is necessary in the fixation of hip joint prostheses. It is also considered be applied to the proximal femur. However, these two concepts are contradictory in that the contact point cannot be determined in a wide contact area. Therefore, in this study, we revised the available concepts to improve the method of fixation of joint prostheses using finite-element-method (FEM) analysis of the equivalent mathematical models.

The first model was designed based on the intramedullary cruciate fixation stem. In this model, the total contact area was limited to the small area of legs-on-plane. Another model was designed based on the fit-and-fill-type stem, and in this model, the total contact area of the plane-on-plane was equal to the total surface area of the stem. In the plane-on-plane model, there was an unstable vibrating pattern in stress distribution, and we considered that deterministic chaos existed in the stress filled wide


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 401 - 401
1 Apr 2004
Sakai R
Full Access

It is widely accepted that a wide contact area between bone and artificial materials is necessary in the fixation of hip joint prostheses. It is also considered important that the load should be applied to the proximal femur. However, these two concepts are contradictory in that the contact point cannot be determined in a wide contact area. Therefore, in this study, we revised the available concepts to improve the method of fixation of joint prostheses using finite-element-method (FEM) analysis of the equivalent mathematical models. The first model was designed based on the intramedullary cruciate fixation stem. In this model, the total contact area was limited to the small area of legs-on-plane. Another model was designed based on the fit-and-fill-type stem, and in this model, the total contact area of the plane-on-plane was equal to the total surface area of the stem. In the plane-on-plane model, there was an unstable vibrating pattern in stress distribution, and we considered that deterministic cha