Results 1 - 9 of 9
Results per page:

Aims

The efficacy of saline irrigation for treatment of implant-associated infections is limited in the presence of porous metallic implants. This study evaluated the therapeutic efficacy of antibiotic doped bioceramic (vancomycin/tobramycin-doped polyvinyl alcohol composite (PVA-VAN/TOB-P)) after saline wash in a mouse infection model implanted with titanium cylinders.

Methods

Air pouches created in female BalBc mice by subcutaneous injection of air. In the first of two independent studies, pouches were implanted with titanium cylinders (400, 700, and 100 µm pore sizes) and inoculated with Staphylococcus aureus (1 × 103 or 1 × 106 colony-forming units (CFU)/pouch) to establish infection and biofilm formation. Mice were killed after one week for microbiological analysis. In the second study, pouches were implanted with 400 µm titanium cylinders and inoculated with S. aureus (1 × 103 or 1 × 106 CFU/pouch). Four groups were tested: 1) no bacteria; 2) bacteria without saline wash; 3) saline wash only; and 4) saline wash plus PVA-VAN/TOB-P. After seven days, the pouches were opened and washed with saline alone, or had an additional injection of PVA-VAN/TOB-P. Mice were killed 14 days after pouch wash.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 31 - 31
11 Apr 2023
Powell D Wu B Dietz P Bou-Akl T Ren W Markel D
Full Access

Failure of osseointegration and periprosthetic joint infection (PJI) are the two main reasons of implant failure after total joint replacement (TJR). Nanofiber (NF) implant surface coating represents an alternative local drug eluting device that improves osseointegration and decreases the risk of PJI. The purpose of this study was to investigate the therapeutic efficacies of erythromycin (EM)-loaded coaxial PLGA/PCL-PVA NF coating in a rat S. aureus-infected tibia model.

NF coatings with 100mg and 1000mg EM were prepared. NF without EM was included as positive control. 56 Sprague Dawley rats were divided into 4 groups. A titanium pin (1.0-mm x 8 mm) was placed into the tibia through the intercondylar notch. S. aureus (SA) was introduced by both direct injection of 10 μl broth (1 × 104 CFU) into the medullary cavity and single dip of Ti pins into a similar solution prior to insertion. Rats were sacrificed at 8 and 16 weeks after surgery. The outcome measurements include μCT based quantitative osteolysis evaluation and hard tissue histology.

Results: EM-NF coating (EM100 and EM1000) reduced osteolysis at 8 and 16 weeks, compared to EM0 and negative control. The effective infection control by EM-NFs was further confirmed by hard tissue section analysis. The Bone implant contact (BIC) and bone area fraction Occupancy (BAFO) within 200 µm of the surface of the pins were used to evaluate the osseointegration and new bone formation around the implants. At 16 weeks, the bone implant contact (BIC) of EM 100 (35.08%) was higher than that of negative control (3.43%) and EM0 (0%). The bone area fraction occupancy within 200 µm (BAFO) of EM100 (0.63 mm2) was higher than that of negative control (0.390 mm2) and EM0 (0.0 mm2). The BAFO of EM100 was also higher than that of EM1000 (0.3mm2).

There was much less osteolysis observed with EM100 and EM1000 NF coatings at 16 weeks, as compared to EM0 positive control, p=0.08 and p=0.1, respectively. Osseointegration and periprosthetic bone formation was enhanced by EM-NFs, especially EM100. Data from this pilot study is promising for improving implant surface fabrication strategies.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 138 - 138
4 Apr 2023
Markel D Dietz P Wu B Bou-Akl T Ren W
Full Access

The efficacy of saline irrigation for the treatment of periprosthetic infection (PJI) is limited in the presence of infected implants. This study evaluated the efficacy of vancomycin/tobramycin-doped polyvinyl alcohol (PVA)/ceramic composites (PVA-VAN/TOB-P) after saline irrigation in a mouse pouch infection model.

3D printed porous titanium (Ti) cylinders (400, 700 and 100 µm in pore size) were implanted into mice pouches, then inoculated with S. aureus at the amounts of 1X103 CFU and 1X106 CFU per pouch, respectively. Mice were randomized into 4 groups (n=6 for each group): (1) no bacteria; (2) bacteria without saline wash; 3) saline wash only, and (4) saline wash+PVA-VAN/TOB-P. After seven days, pouches were washed out alone or with additional injection of 0.2 ml of PVA-VAN/TOB-P. Mice were sacrificed 14 days after pouch wash. Bacteria cultures of collected Ti cylinders and washout fluid and histology of pouch tissues were performed.

The low-grade infection (1X103 CFU) was more significant in 400 µm Ti cylinders than that in Ti cylinders with larger pore sizes (700 and 1000 µm (p<0.05). A similar pattern of high-grade infection (1X106 CFU) was observed (p<0.05). For the end wash, the bacteria burden (0.49±0.02) in saline wash group was completely eradicated by the addition of PVA-VAN/TOB-P (0.005±0.001, p<0.05).

We noticed that 400 µm Ti cylinders have the highest risk of implant infection. Our data supported that the effect of saline irrigation was very limited in the presence of contaminated porous Ti cylinders. PVA-VAN/TOB-P was biodegradable, biocompatible, and was effective in eradicating bacteria retention after saline irrigation in a mouse model of low grade and high-grade infection. We believe that PVA-VAN/TOB-P represents an alternative to reduce the risk of PJI by providing a sustained local delivery of antibiotics.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 144 - 144
4 Apr 2023
Markel D Chatterji R Bou-Akl T Wu B Dietz P Ren W
Full Access

Irrigation with antiseptic agents, antibiotics, and surfactants are used for treatment and prevention of infections. Despite desirable microbicidal actions, studies have demonstrated cytotoxic effects on host tissue that may impair healing. This study investigated the extent of tissue damage caused by commonly used irrigation solutions in the presence or absence of infection.

Air pouches created in 60 balb/c mice were divided into two groups (n=30): infected with Staphylococcus aureus and control. One week later the infected group was subdivided into 5 subgroups (n=6) based on irrigation solutions and by day 0 (immediately) and day7 after irrigation (n=3). Solutions included Saline, Bacitracin, Clorpactin, Irrisept and Bactisure. In infected group wash fluid was collected for quantitative analysis of bacterial growth. At the specified times mice were sacrificed, pouch tissue sent for histology, and sections analyzed for inflammation, necrosis, and edema.

Inflammation decreased in infected vs sterile pouches for all solutions except Bacitracin day 0 and for all solutions day 7 with significance in all except Bacitracin (p<0.05). On day 0, necrosis increased in infected vs sterile pouches in Bacitracin (p=0.006), Irrisept (p=0.18), or Bactisure (p=0.07); however, on day 7, necrosis significantly decreased in infected pouches for all solutions (p<0.05) except for Clorpactin (p=0.18). Edema decreased in infected vs sterile pouches on day 0 for all solutions with significance in saline, Irrisept, and Bacitracin (p<0.05). On day 7, infected pouches had decreased edema in saline, Bacitracin, and Bactisure (p<0.05) and increased in Irrisept (p<0.05) and Clorpactin (p=0.069) compared to sterile pouches. Bacterial culture of washouts demonstrated that Clorpactin, Irrisept and Bactisure controlled the infection, whereas saline and Bacitracin showed bacterial multiplication 3.9 × 10^7 CFU/ml and 6.7 × 10^7 CFU/ml respectively. Bacitracin wash showed significantly more bacteria growth compared to Clorpactin (p=0.024), Irrisept (p=0.025) and Bactisure (p=0.025).

Tissue damage varied with irrigation solutions and the presence or absence of infection. Presence of bacteria appeared to lead to less tissue inflammation and edema. Tissue necrosis varied over time with different solutions. Surgeons must weigh risks and benefits when selecting solutions and determining when to irrigate.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 140 - 140
4 Apr 2023
Fry M Ren W Bou-Akl T Wu B Cizmic Z Markel D
Full Access

Extensor mechanism and abductor reconstructions in total joint arthroplasty are problematic. Growing tendon into a metallic implant would have great reconstructive advantages. With the introduction of porous metal implants, it was hoped that tendons could be directly attached to implants. However, the effects of the porous metal structure on tissue growth and pore penetration is unknown. In this rat model, we investigated the effect of pore size on tendon repair fixation using printed titanium implants with differing pore sizes.

There were four groups of six Sprague Dawley rats (n = 28) plus control (n=4). Implants had pore sizes of 400µm (n=8), 700µm (n=8), and 1000µm (n=8). An Achilles tendon defect was created, and the implant positioned and sutured between the cut ends. Harvest occurred at 12-weeks. Half the specimens underwent tensile load to failure testing, the other half fixed and processed for hard tissue analysis.

Average load to failure was 72.6N for controls (SD 10.04), 29.95N for 400µm (SD 17.95), 55.08N for 700µm (SD 13.47), and 63.08N for 1000µm (SD 1.87). The load to failure was generally better in the larger pore sizes. Histological evaluation showed that there was fibrous tendon tissue within and around the implant material, with collagen fibers organized in bundles. This increases as the pore diameter increases.

Printing titanium implants allows for precise determination of pore size and structure. Our results showed that tendon repair utilizing implants with 700µm and 1000µm pores exhibited similar load to failure as controls. Using a defined pore structure at the attachment points of tendons to implants may allow predictable tendon to implant reconstruction at the time of revision arthroplasty.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1035 - 1041
1 Sep 2019
Markel DC Bou-Akl T Rossi MD Pizzimenti N Wu B Ren W

Aims

The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components.

Patients and Methods

A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 3 - 3
1 Jun 2017
Song W Chen L Bergum C Zhou Z Shen M Ren W Yu X Markel D
Full Access

Bone ingrowth is desired with uncemented hip implants. Infection is clearly undesirable. We have worked on developing a nanofiber coating for implants that would enhance bone formation while inhibiting infection. Few studies have focused on developing an implant surface nanofiber (NF) coating to prevent infection and enhance osseointegration by local drug release. In this study, coaxial doxycycline (Doxy)-doped polycaprolactone/polyvinyl alcohol (PCL/PVA) Nanofibers were directly deposited on the titanium (Ti) implant surface during electrospinning.

The interaction of loaded Doxy with both PVA and PCL NFs was characterized by Raman spectroscopy. The bonding strength of Doxy-doped NF coating on Ti implants was confirmed by a stand single-pass scratch test. The improved implant osseointegration by PCL/PVA NF coatings in vivo was confirmed by scanning electron microscopy, histomorphometry and micro computed tomography at 2, 4 and 8 weeks after implantation. The bone contact surface (%) changes of NF coating group (80%) is significantly higher than that of no NF group (< 5%, p<0.05). Finally, we demonstrated that Doxy-doped NF coating effectively inhibited bacterial infection and enhanced osseointegration in an infected (Staphylococcus aureus) tibia implantation rat model. Doxy released from NF coating inhibited bacterial growth up to 8 weeks in vivo. The maximal push-in force of Doxy-NF coating (38 N) is much higher than that of NF coating group (6.5 N) 8 weeks after implantation (p<0.05), which was further confirmed by quantitative histological analysis and micro computed tomography.

These findings indicate that coaxial PCL/PVA NF coating doped with Doxy and/or other drugs have great potential in enhancing implant osseointegration and preventing infection.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 136 - 136
1 Feb 2017
Ren W Markel D
Full Access

Few studies have been reported focusing on developing implant surface nanofiber (NF) coating to prevent infection and enhance osseointegration by local drug release. In this study, coaxial doxycycline (Doxy)-doped polycaprolactone/polyvinyl alcohol (PCL/PVA) NFs were directly deposited on the titanium (Ti) implant surface during electrospinning. The bonding strength of Doxy-doped NF coating on Ti implants was confirmed by a stand single-pass scratch test. The improved implant osseointegration by PCL/PVA NF coatings in vivo was confirmed by scanning electron microscopy, histomorphometry and micro computed tomography at 2, 4 and 8 weeks after implantation. The bone contact surface (%) changes of NF coating group (80%) is significantly higher than that of no NF group (< 5%, p<0.05). Finally, we demonstrated that Doxy-doped NF coating effectively inhibited bacterial infection and enhanced osseointegration in an infected (Staphylococcus aureus) tibia implantation rat model. Doxy released from NF coating inhibited bacterial growth up to 8 weeks in vivo. The maximal push-in force of Doxy-NF coating (38 N) is much higher than that of NF coating group (6.5 N) 8 weeks after implantation (p<0.05), which was further confirmed by quantitative histological analysis and micro computed tomography. These findings indicate that coaxial PCL/PVA NF coating doped with Doxy and/or other drugs have great potential in enhancing implant osseointegration and preventing infection.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 445 - 445
1 Nov 2011
Ren W Zhang R Hawkins M Markel D
Full Access

We have demonstrated that erythromycin (EM) inhibits wear debris-induced macrophage activation and osteo-clastogenesis (both in vitro and in vivo) through targeting NF-κB signalling. Our clinical trial further verified that oral EM can be efficiently delivered to periprosthetic tissue and improve local inflammation. The purpose of this study was to assess the efficacy of periprosthetic EM delivery in a rat osteolysis model.

The PA coated titanium (Ti) pin (Stryker) was loaded with EM (8 μl = 2.8 mg/pin). Drug release assay showed around 25% of loaded EM was remained in the PA layer 24 hours after loading. Rats were divided into three groups:

saline control (n=5);

UHMWPE particle injection (n=7), and

UHMWPE particle injection with EM treatment (n=7).

Uncoated Ti pins were pressfit inserted into right tibia following the injection 200 μl of either UHMWPE particles (5 mg/ml) or saline (control). The revision surgeries were performed 6 weeks after the first surgery. The previous implanted pins were replaced with new Ti pins either with or without EM coating. Rats were then sacrificed one month after “revision surgery”, and the knee joint samples were collected for μCT and histology analysis.

μCT analysis showed that the value of bone volume (bv/tv) in the group treated with EM (0.26 ± 0.07) was significantly higher than the group untreated (0.14 ± 0.04), while there was no significant difference between EM treated group and the saline control group (0.15 ± 0.11). The parameters of cancellous bone structure all pointed a trend of better structure in EM treated group than other two groups. However, this difference did not reach statistical significance. Histology analysis (H& E staining) demonstrated that in the saline control the tibia retained a smooth endocortical surface with a prominent periprosthetic membrane. In the EM-treated group, endocortical erosion was reduced and the peri-prosthetic tissue appeared thinner than uncoated pins.

The overall cellularity of periprosthetic membranes from the EM-treated group was decreased compared to the untreated group. Analysis of membrane thickness revealed a significantly thinner membrane in EM-treated group compared with untreated group and saline control (p< 0.05).

The results of this study seem to indicate that an EM coated Ti pin provided a sufficient drug source to effectively treat wear debris-induced periprosthetic inflammation and osteolysis.