Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 8 - 8
1 Jun 2023
Harris PC Lacey S Perdomo A Ramsay G
Full Access

Introduction

The vast majority of orthopaedic surgeons use C-arm fluoroscopy in the operating theatre when building a circular external fixator. In the absence of previous research in this area, we hypothesised that the surgeon who builds a circular external fixator is exposed to a greater amount of radiation purely as a result of the presence of the metallic fixator in the x-ray beam. The aim of our study therefore was to investigate how the presence of a circular external fixator affects the radiation dose to the surgeon and the surgical assistant.

Materials & Methods

A simulated environment was created using a radiolucent operating table, an acrylic lower limb phantom (below knee segment), various configurations of metalic circular external fixation, and a standard size C-arm image intensifier.

The variables investigated were 1. the amount of metal in the beam 2. the orientation of the beam (PA vertical vs lateral) 3. the horizonal distance of the person from the beam (surgeon vs assistant) and 4. the vertical distance of the various body parts from the beam (e.g. thyroid, groin).

In terms of radiation dose, we recorded two things : 1. the dose produced by the image intensifier 2. the dose rate at standardised positions in the operating theatre. The latter was done using a solid-state survey sensor. These positions represented both where the surgeon and surgical assistant typically stand plus the heights of their various body regions relative to the operating table..