The Exeter Stem can be used with metal femoral head that are made of either cobalt chrome, or stainless steel. The aim of this study was to compare the rates of revision of these two metal femoral head types when used with this femoral component. Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) Data from September 1999 until December 2015 for all primary THRs using an Exeter or an Exeter v40 stem with the diagnosis of osteoarthritis were analysed. Only bearing couples that used a metal head with polyethylene were included. The cumulative percent revision (CPR) calculated using Kaplan-Meier estimates were compared for the two metal head types. CPR were further analysed by age, polyethylene type and head size. Reasons for revision and types of revision were assessed.Introduction/Aims
Method
Femoral stems with exchangeable (modular) necks were introduced
to offer surgeons an increased choice when determining the version,
offset and length of the femoral neck during total hip arthroplasty
(THA). It was hoped that this would improve outcomes and reduce
complications, particularly dislocation. In 2010, the Australian Orthopaedic
Association National Joint Replacement Registry (AOANJRR) first
reported an increased rate of revision after primary THA using femoral
stems with an exchangeable neck. The aim of this study was to provide
a more comprehensive up-to-date analysis of primary THA using femoral
stems with exchangeable and fixed necks. The data included all primary THA procedures performed for osteoarthritis
(OA), reported to the AOANJRR between 01 September 1999 and 31 December
2014. There were 9289 femoral stems with an exchangeable neck and
253 165 femoral stems with a fixed neck. The characteristics of
the patients and prostheses including the bearing surface and stem/neck
metal combinations were examined using Cox proportional hazard ratios
(HRs) and Kaplan-Meier estimates of survivorship. Aims
Materials and Methods