header advert
Results 1 - 8 of 8
Results per page:
Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 2 - 2
1 Feb 2021
Pizzamiglio C Fattori A Rovere F Poon P Pressacco M
Full Access

Background

Stemless prostheses are recognized to be an effective solution for anatomic total shoulder arthroplasty (TSA) while providing bone preservation and shortest operating time. Reverse shoulder arthroplasty (RSA) with stemless has not showed the same effectiveness, as clinical and biomechanical performances strongly depend on the design. The main concern is related to stability and bone response due to the changed biomechanical conditions; few studies have analyzed these effects in anatomic designs through Finite Element Analysis (FEA), however there is currently no study analyzing the reverse configuration. Additionally, most of the studies do not consider the effect of changing the neck-shaft angle (NSA) resection of the humerus nor the proper assignment of spatial bone properties to the bone models used in the simulations. The aim of this FEA study is to analyze bone response and primary stability of the SMR Stemless prosthesis in reverse with two different NSA cuts and two different reverse angled liners, in bone models with properties assigned using a quantitative computed tomography (QCT) methodology.

Methods

Sixteen fresh-frozen cadaveric humeri were modelled using the QCT-based finite element methodology. The humeri were CT-scanned with a hydroxyapatite phantom to allow spatial bone properties assignment [Fig. 1]. Two implanted SMR stemless reverse configurations were considered for each humerus: a 150°-NSA cut with a 0° liner and a 135°-NSA cut with a 7° sloped liner [Fig. 2]. A 105° abduction loading condition was simulated on both the implanted reverse models and the intact (anatomic) humerus; load components were derived from previous dynamic biomechanical simulations on RSA implants for the implanted stemless models and from the OrthoLoad database for the intact humeri. The postoperative bone volume expected to resorb or remodel [Fig. 3a] in the implanted humeri were compared with their intact models in sixteen metaphyseal regions of interest (four 5-mm thick layers parallel to the resection and four anatomical quadrants) by means of a three-way repeated measures ANOVA followed by post hoc tests with Bonferroni correction. In order to evaluate primary stability, micromotions at the bone-Trabecular Titanium interface [Fig. 3b] were compared between the two configurations using a Wilcoxon matched-pairs signed-rank test. The significance level α was set to 0.05.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 17 - 17
1 Feb 2020
Fattori A Negro ND Gunsallus K Lipman J Hotchkiss R Figgie M Wright T Pressacco M
Full Access

Introduction

Total Elbow Arthroplasty (TEA) is recognized as an effective treatment solution for patients with rheumatoid arthritis or for traumatic conditions. Current total elbow devices can be divided into linked or unlinked design. The first design usually presents a linking element (i.e. an axle) to link together the ulnar and humeral components to stabilize the joint; the second one does not present any linkage and the stability is provided by both intrinsic design constraints and the soft tissues. Convertible modular solutions allow for an intraoperative decision to link or unlink the prosthesis; the modular connections introduce however additional risks in terms of both mechanical strength and potential fatigue and fretting phenomena that may arise not only due to low demand activities loads, but also high demand (HD) ones that could be even more detrimental. The aim of this study was to assess the strength of the modular connection between the axle and the ulnar component in a novel convertible elbow prosthesis design under simulated HD and activities of daily living (ADLs) loading.

Methods

A novel convertible total elbow prosthesis (LimaCorporate, IT) comprising both ulnar and humeral components that can be linked together by means of an axle, was used. Both typical ADLs and HD torques to be applied to the axle were determined based on finite element analysis (FEA); the boundary load conditions for the FEA were determined based on kinematics analysis on real patients in previous studies. The FEA resultant moment acting on the axle junction during typical ADLs (i.e. feeding with 7.2lbs weight in hand) was 3.2Nm while for HD loads (i.e. sit to stand) was 5.7 Nm. In the experimental setup, 5 axle specimens coupled with 5 ulnar bodies through a tapered connection (5 Nm assembly torque) were fixed to a torque actuator (MTS Bionix) and submerged in a saline solution (9g/l). A moment of 3.2 Nm was applied to the axle for 5M cycles through a fixture to test it under ADLs loading. After 5M cycles, the axles were analyzed with regards to fretting behavior and then re-assembled to test them against HD loading by applying 5.7 Nm for 200K cycles (corresponding to 20 years function).


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 120 - 120
1 Feb 2020
Gonzalez FQ Fattori A Lipman J Negro ND Brial C Figgie M Hotchkiss R Pressacco M Wright T
Full Access

Introduction

The interaction between the mobile components of total elbow replacements (TER) provides additional constraint to the elbow motion. Semi-constrained TER depend on a mechanical linkage to avoid dislocation and have greater constraint than unconstrained TER that rely primarily in soft tissue for joint stability. Greater constraint increases the load transfer to the implant interfaces and the stresses in the polyethylene components. Both of these phenomena are detrimental to the longevity of TER, as they may result in implant loosening and increased damage to the polyethylene components, respectively[1]. The objective of this work was to compare the constraint profile in varus-valgus and internal-external rotation and the polyethylene stresses under loads from a common daily activity between two semi-constrained TER, Coonrad/Morrey (Zimmer-Biomet) and Discovery® (DJO), and an unconstrained TER, TEMA (LimaCorporate).

Methods

We developed finite element (FE) models of the three TER mechanisms. To reduce computational cost, we did not include the humeral and ulnar stems. Materials were linear-elastic for the metallic components (ETi6Al4V=114.3 GPa, ECoCr=210 GPa, v=0.33) and linear elastic-plastic for the polyethylene components (E=618 MPa, v=0.46; SY=22 MPa; SU=230.6 MPa; εU=1.5 mm/mm). The models were meshed with linear tetrahedral elements of sizes 0.4–0.6 mm. We assumed a friction coefficient of 0.02 between metal and polyethylene. In all simulations, the ulnar component was fixed and the humeral component loaded. We computed the constraint profiles in full extension by simulating each mechanism from 8° varus to 8° valgus and from 8° internal to 8° external rotation. All other degrees-of-freedom except for flexion extension were unconstrained. Then, we identified the instant during feeding that generated the highest moments at the elbow[2], and we applied the joint forces and moments to each TER to evaluate the stresses in the polyethylene. To validate the FE results, we experimentally evaluated the constraint of the design with highest polyethylene stresses in pure internal-external rotation and compared the results against those from a FE model that reproduced the experimental setup (Fig.1-a).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 78 - 78
1 Apr 2019
Fattori A Kontaxis A Chen X Vidoni G Castagna A Pressacco M
Full Access

Introduction

Reverse Shoulder Arthroplasty (RSA) is recognized to be an effective solution for rotator cuff deficient arthritic shoulders, but there are still concerns about impingement and range of motion (ROM). Several RSA biomechanical studies have shown that humeral lateralization can increase ROM in planar motions (e.g. abduction). However, there is still a debate whether humeral lateralization should be achieved with a larger sphere diameter or by lateralizing the center of rotation (COR). The latter has shown to decrease the deltoid moment arm and increase shear forces, where the former may pose challenges in implanting the device in small patients. The aim of this study was to evaluate how humeral lateralization achieved by varying COR lateral offset and glenosphere diameter in a reverse implant can affect impingement during activities of daily living (ADLs).

Methods

Nine shoulder CT scans were obtained from healthy subjects. A reverse SMR implant (LimaCorporate, IT) was virtually implanted on the glenoid and humerus (neck-shaft angle 150°) as per surgical technique using Mimics software (Materialise NV). Implant positioning was assessed and approved by a senior surgeon. The 3D models were imported into a validated shoulder computational model (Newcastle Shoulder Model) to study the effects of humeral lateralization. The main design parameters considered were glenosphere diameter (concentric Ø36mm, Ø40mm, Ø44mm) and COR offset (standard, +2mm, +5mm), for a total of 9 combinations for each subject; −10°, 0° and 10° humeral components versions were analyzed. The model calculated the percentage of impingement (intra-articular, contact of cup with scapula neck and glenoid border; extra-articular, contact of humerus with acromion and coracoid) during 5 ADLs (hand to opposite shoulder, hand to back of head, hand to mouth, drink from mug and place object to head height).


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 72 - 72
1 Apr 2018
Riva F Boccalon M Buttazzoni E Pressacco M
Full Access

Introduction

Advantages of ceramic materials for hip joint prosthesis are recognized to be high hardness, scratch resistance, improved wettability, lower friction and lower wear than CoCr surfaces. Recent studies suggest the use of ceramic femoral head reduce fretting corrosion at stem taper junction compared to metal-on-metal taper junction[1]. Continuous improvement of ceramic materials for orthopedic lead to the development of a resurfacing ceramic-on-ceramic hip joint prosthesis. The main differences of resurfacing heads respect to standard heads are their anatomical dimension and the empty shape suitable to cover the femoral bone and to connect with the resurfacing stem. Ceramic is essentially a brittle material and its strength is influenced by the minimum thickness in the stressed area. Ceramic resurfacing head minimum thickness is comparable with ceramic revision head already on the market. The aim of this study is to develop a mechanical pre-clinical analysis verification process for the newly developed system.

Materials and methods

The empty shape of the ceramic resurfacing head may influence its strength in a crush loading scenario. Although this is not a physiological condition this test represents the most severe loading for a resurfacing head. Also comparative analysis can be done considering the yield point of conventional metal resurfacing heads reported by the FDA Center for Devices and Radiological Health. For this reason a static unsupported head strength test is performed by applying a compressive load perpendicular to the head axis along the equatorial plane[2](Fig.1). Resurfacing ceramic head made in ZTA is suitable both for a resurfacing stem and an adaptor to be coupled with a standard stem. Mechanical test was performed on worst case resurfacing head size both with resurfacing stem and standard stem based and on FE non linear analysis performed in ANSYS 17.2 according the following material properties: ZTA ceramic (modulus of elasticity E, Poisson ratio ν and density ρ of 348GPa, 0.23 and 4.25g/cm3 respectively), and Ti6Al4V (E=114GPa, ν=0.33 and ρ=4.43g/cm3). For comparison purposes unsupported test was performed on standard head Ø28#S both in Biolox®Delta and Biolox®Forte ceramic. At least three components were used for each test and the average values was compared with predicates[2]. Static compressive load was applied with MTS hydraulic actuators with load cell of 100kN.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 113 - 113
1 Mar 2017
Riva F De Bona F Benasciutti D Pressacco M
Full Access

Introduction

Modularity allows surgeons to use femoral heads of various materials, diameters and offsets to achieve the best possible outcome, nevertheless the fretting corrosion behaviour of modular junctions can be significantly affected. The aim of this study was to assess physiological friction moment and lubrication ratio in order to compare various tribological materials against different bearing sizes. This data is important as lubrication will affect the friction, wear and torque generated which may lead directly to the production of debris or to enhanced corrosion at modular junctions.

Materials and methods

Hip joints were tested in lubricant condition on a hip simulator following the ISO14242-3 configuration. Three samples for each combinations were examined: 1) 36mm metal-on-metal made in CoCrMo 2) 36mm ceramic-on-ceramic made in ZTA 3) 58mm resurfacing metal-on-metal made in CoCrMo 4) 57mm resurfacing ceramic-on-ceramic made in ZTA. Preconditioning and dynamic loading steps were spaced out by rest periods (Fig. 1) and the entire series was repeated three times for each combination. Strains were measured on the Ti6Al4V neck's femoral stem with three couples of biaxial strain gauges and were converted into friction moments by means of analytical formulas. Mean maximum moment M and lubrication ratio λ were calculated. MSTART-UP and MTURN-OFF were respectively the first three and last three peak moment sampled for each consecutive step.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 61 - 61
1 Feb 2017
Khan H Riva F Pressacco M Meswania J Panagiotidou A Coathup M Blunn G
Full Access

Background

Complications of metal-on-metal hip resurfacing, leading to implant failure, include femoral notching, neck fracture, and avascular necrosis. Revision arthroplasty options include femoral-only revision with a head, however mis-matching radial clearance could accelerate metal ion release. Alternatively, revision of a well-fixed acetabular component could lead to further bone loss, complicating revision surgery. We have developed a ceramic hip resurfacing system with a titanium-ceramic taper junction; taking advantage of the low frictional torque and wear rates that ceramic affords. Taking a revision scenario into account, the ceramic head has a deep female taper for the resurfacing stem, but also a superficial tapered rim. Should revision to this resurfacing be required, any femoral stem with a 12/14 taper can be implanted, onto which a dual taper adaptor is attached. The outer diameter of the taper adaptor then becomes the male taper for the superficial taper of the ceramic head; ultimately allowing retention of the acetabular component. In an in-vitro model, we have compared the fretting corrosion of this taper adaptor to existing revision taper options: a titanium-cobalt chrome (Ti-CoCr) taper junction, and a titanium-titanium sleeve-ceramic (Ti-Ti-Cer) taper junction.

Methods

To simulate gait, sinusoidal cyclical loads between 300N-2300N, at a frequency of 3Hz was applied to different neck offsets generating different bending moments and torques. Bending moment and frictional torque were tested separately. An electrochemical assessment using potentiostatic tests at an applied potential of 200mV, was used to measure the fretting current (μA) and current amplitude (μA). In a short term 1000 cycle test with bending moment, four neck lengths (short to x-long) were applied. For frictional torque, four increments of increasing torque (2-4-6-8Nm) were applied. In a long-term test using the taper adaptor, the combination of worst-case scenario of bending and torque were applied, and fretting currents measured every million cycles, up to 10 million cycles.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 60 - 60
1 Feb 2017
Khan H Meswania J Riva F Pressacco M Panagiotidou A Coathup M Blunn G
Full Access

Background

Hip resurfacing has advantages for the young active patient with arthritis; maintaining a large range of motion, preserving bone stock, and reduced dislocation risk. However high serum metal ion levels with metal-on-metal resurfacing, and their clinical implications, has led to a decline in the use of hip resurfacing. Ceramic bearing surfaces display the lowest frictional torque and excellent wear rates. Recent developments have enabled large, strong ceramic materials to be used as resurfacing components. Any wear debris that is generated from these articulations is inert. However an all-ceramic hip resurfacing could be at risk of fracture at the head-stem junction. A new ceramic hip resurfacing system with a titanium-ceramic modular taper junction has been developed. The introduction of a taper introduces the potential for fretting corrosion; we sought to determine the extent of this in an in-vitro model, and compared this prosthesis to the conventional 12/14 titanium-cobalt chrome (Ti6Al4V-CoCr) taper junction.

Methods

To simulate the gait cycle, sinusoidal cyclical loads between 300N-2300N, at a frequency of 3Hz, were applied to different head-neck offsets generating different bending moments and torques. The effect of increasing the bending moment and frictional torque were tested separately. Furthermore, the resurfacing head was mounted in a fixture held with just the stem, thus representing complete bone resorption under the head. An electrochemical assessment using potentiostatic tests at an applied potential of 200mV, was used to measure the fretting current (μA) and current amplitude (μA). In a short-term 1000 cycle test, six neck lengths (short to xxx-long) of the Ti6Al4V-CoCr taper were compared to the standard neutral (concentric), and 3mm A/P offset stem options for the resurfacing design. To represent frictional torque, four increments of increasing torque (2-4-6-8Nm) were applied to both tapers. In a long term test with the resurfacing stem, the worst-case scenario of the eccentric offset option and 8Nm of torque were applied, and potentiostatic measurements were taken every million cycles, up to 10 million cycles.