Stem dissociation in modular revision knee replacement due to failure of the frictional lock of the Morse taper has been reported in the literature. However, the medium and long-term implications of stem dissociation are unknown, as clinical outcomes have not been reported. We report a series of 10 cases in which there was intra-operative dissociation of the tibial stem. Between 1994 and 1999, 98 patients underwent revision total knee replacement for aseptic loosening at our institution. Ten of these patients were noted to have tibial stem dissociation, apparent on the immediate post-operative radiographs. The senior author (RWP) performed all procedures and used a standardized operative technique. The Co-ordinate modular knee revision system was used in all cases. The quality of the bone was noted in all the cases intra-operatively; and was graded as 1) sound bone, 2) soft but intact, 3) soft and fractured cortex. Our study demonstrates that the tibial stem dissociation did not cause any significant detriment to the clinical outcome on minimum follow-up of six years in nine cases where the tibial metaphyseal cortical rim was intact. In one case, where the medial tibial plateau had a cortical defect, the prosthesis drifted into varus mal-alignment and the patient required a further revision for aseptic loosening. We therefore question whether long canal filling tibial stems are necessary in all revision total knee replacements particularly when the cortical rim is intact and a non-constrained poly-ethylene insert is used.
The aim of our study was to assess the impact that HV has on patients’ quality of life and to correlate each of the above angles to SF-36 sub-scales.