This study compared the pullout forces of the initial implantation and the “cement-in-cement” revision technique for short and standard-length (125 mm vs. 150 mm) Exeter® V40 femoral stems used in total hip arthroplasty (THA). The idea that the pullout force for a double taper slip stem is relative to the force applied to the femur and that “cement-in-cement” revision provides the same reproduction of force. A total sample size of 15 femoral stems were tested (Short, n = 6 and Standard, n = 9). 3D printed fixtures for repeatable sample preparation were used to minimize variance during testing. To promote stem subsidence and to simulate an in vivo environment, the samples were placed in an incubator at 37°C at 100% humidity and experienced a constant compressive loading of 1335 N for 14 days. The samples underwent a displacement-controlled pullout test. After the initial pullout test, “cement-in-cement” revision will be performed and tested similar to the initial implantation to observe the efficacy of the revision technique. To compare the pullout forces between the two groups, a Kruskal-Wallis test using a significance level of 0.05 was conducted. The mean maximum pullout force for the short and standard-length femoral stems were 3939 ± 1178 N and 5078 ± 1168 N, respectively. The Kruskal-Wallis test determined no statistically significant difference between the two groups for the initial implantation (p = 0.13). The “cement-in-cement” revision pullout force will be conducted in future testing. This study demonstrated the potential use of short stem designs for THA as it provides similar levels of fixation as the standard-length femoral stem. The potential benefits for using a short stem design would be providing similar load transfer to the proximal femur, preserving proximal metaphyseal femoral bone in primary replacement, and reducing the invasiveness during revision.