Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 85 - 85
1 May 2016
Asada S Ouyang Y Jones H Ismaily S Noble P
Full Access

Introduction

Restoration of knee function after total knee arthroplasty (TKA) often entails a balance between normal kinematics and normal knee stability, especially in performing demanding physical activities. The ultra-congruent (UC) knee design prioritizes stability over kinematics through close conformity between the femoral component and the tibial insert in extension. This configuration is intended to provide AP stability in the absence of the posterior cruciate ligament during activities that would otherwise cause anterior femoral subluxation. In this study we examine the kinematics of an ultra-congruent knee design in comparison with the intact knee and with conventional articulations used in PCL-retaining (CR) and PCL-substituting (PS) TKR designs.

Materials and Methods

The 3D tibio-femoral kinematics of 6 fresh frozen cadaveric human knees were tested during loaded simulation of squatting in a computer-controlled knee testing rig. Muscle forces were simulated by loading rectus femoris and vastus intermedius (150N), vastus lateralis (100N), vastus medialis (75N), and the hamstring muscles (60N) (total: 385N). Testing was performed on the intact knee, and after implanting a standard design of total knee prosthesis with the posterior cruciate ligament intact (CR-TKA), resected (PCL-substituting insert; PS-TKA), and a UC insert (UC-TKA group). The 3D positions of the tibia and femur were tracked with a high resolution 12 camera motion analysis system (Motion Analysis Inc.) and used to position 3D CT reconstructions of each bone. The translation and rotation of the femur with respect to the tibia were calculated by projecting the femoral transcondylar axis onto a plane normal to the longitudinal anatomical axis of the tibia coincident with the transverse axis of the tibial plateau.