header advert
Results 1 - 20 of 53
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 49 - 49
23 Jun 2023
McCalden R Pomeroy E Naudie D Vasarhelyi E Lanting B MacDonald S Howard J
Full Access

Subsidence remains a concern when utilizing modern tapered fluted titanium (TFT) femoral stems and may lead to leg length discrepancy, impingement, instability and failure to obtain stem osseointegration. This study aims to compare stem subsidence across three modern TFT stems. Our secondary aim was to investigate the influence of bicortical contact or ‘scratch fit’ on subsidence, as well as the role of intraoperative imaging in maximizing this bicortical contact and preventing stem subsidence.

A retrospective review of 271 hip arthroplasties utilizing modern TFT stems in a single institution was performed. Three stem designs were included in the analysis: one monoblock TFT stem (n=91) and two modular TFT stems (Modular A [n=90]; Modular B [n=90]). Patient demographics, Paprosky femoral bone loss classification, bi-cortical contact, utilization of intra-operative imaging and stem subsidence (comparison of initial post-operative radiograph to the latest follow up radiograph - minimum three months) were recorded.

There was no statistically significant difference in the amount of subsidence between the three stems (Monoblock: 2.33mm, Modular A: 3.43mm, Modular B: 3.02mm; p=0.191). There was no statistical difference in subsidence >5mm between stems (Monoblock: 9.9%, Modular A: 22.2%, Modular B: 16.7%). Subgroup analysis based on femoral bone loss grading showed no difference in subsidence between stems. Increased bicortical contact was strongly associated with reduced subsidence (p=0.004). Intra-operative imaging was used in 46.5% (126/271) of cases; this was not correlated with bicortical contact (p=0.673) or subsidence (p=0.521). Across all groups, only two stems were revised for subsidence (0.7%).

All three modern TFT stems were highly successful and associated with low rates of subsidence, regardless of modular or monoblock design. Surgeons should select the stem that they feel is most clinically appropriate.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 16 - 16
1 Nov 2021
McCalden R Salipas A Teeter M Somerville L Naudie D MacDonald S
Full Access

The purpose of this study was to precisely measure the 10-year polyethylene wear rate of primary total hips using Radiostereometric analysis (RSA) comparing Oxidized Zirconium (OxZi) to cobalt chrome (CoCr) femoral heads articulating with highly cross-linked polyethylene (XLPE).

RSA was performed on 46 patients who underwent total hip arthroplasty − 23 who received OxZi femoral heads and 23 who received CoCr heads in combination with XLPE at a minimum of 10 years follow-up. All patients had identical THR systems implanted except for the femoral head utilized. The Centre Index method was utilised to assess total wear rates (from index surgery until final evaluation) using a dedicated RSA Software program (UmRSA Digital Measure v.2.2.1). In addition, the Martell technique was used to subtract head penetration occurring in the first 1–2 years (i.e. bedding-in phase) to allow calculation of the ‘steady state’ wear rates.

There were no significant differences in demographics (i.e age, BMI, gender) between the groups. The average time from surgery of the RSA examinations was 11.7 and 12.6 years for the CoCr and OxZi groups respectively. Using the Centre Index Method to calculate total head penetration, wear rates were slightly higher in the OxZi group (0.048 +/− 0.021mm/year) compared to the CoCr group (0.035 +/− 0.017mm/year) with no statistical difference between the groups (p= 0.02). After correction for the bedding-in period, there was no statistically significant difference in mean ‘steady-state’ wear rate between OxZi (0.031 +/− 0.021mm/year) and CoCr (0.024 +/− 0.019mm/year) at 10 year follow up (p= 0.24). A comparison of preoperative and postoperative SF12, HHS, and WOMAC scores showed no statistical difference between the groups.

RSA demonstrated the 10 year in-vivo wear rates of both bearing combinations to be well below the threshold for osteolysis. There was no significant difference between either the total or ‘steady-state’ wear rates of the OxZi and CoCr groups at 10 years.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_7 | Pages 78 - 78
1 Jul 2020
Somerville L Clout A MacDonald S Naudie D McCalden RW Lanting B
Full Access

While Oxidized Zirconium (OxZr) femoral heads matched with highly cross-linked polyethylene (XLPE) have demonstrated the lowest rate of revision compared to other bearing couples in the Australian National Joint Registry, it has been postulated that these results may, in part, be due to the fact that a single company offers this bearing option with a limited combination of femoral and acetabular prostheses. The purpose of this study was to assess clinical and radiographic outcomes in a matched cohort of total hip replacements (THR) utilizing an identical cementless femoral stem and acetabular component with either an Oxidized Zirconium (OxZr) or Cobalt-Chrome (CoCr) femoral heads at a minimum of 10 years follow-up.

We reviewed our institutional database to identify all patients whom underwent a THR with a single cementless femoral stem, acetabular component, XLPE liner and OxZr femoral head with a minimum of 10 years of follow-up. These were then matched to patients who underwent a THR with identical prosthesis combinations with CoCr femoral head by gender, age and BMI. All patients were prospectively evaluated with WOMAC, SF-12 and Harris Hip Score (HHS) preoperatively and postoperatively at 6 weeks, 3 months, 1 and 2 years and every 2 years thereafter. Charts and radiographs were reviewed to determine the revision rates and survivorship (both all cause and aseptic) at 10 years for both cohorts. Paired analysis was performed to determine if differences exist in patient reported outcomes.

There were 208 OxZr THRs identified which were matched with 208 CoCr THRs. There was no difference in average age (OxZr, 54.58 years, CoCr, 54.75 years), gender (OxZr 47.6% female, CoCr 47.6% female), and average body max index (OxZr, 31.36 kg/m2, CoCr, 31.12 kg/m2) between the two cohorts. There were no significant differences preoperatively in any of the outcome scores between the two groups (WOMAC (p=0.449), SF-12 (p=0.379), HHS(p=0.3718)). Both the SF12 (p=0.446) and the WOMAC (p=0.278) were similar between the two groups, however the OxZr THR cohort had slightly better HHS compared to the CoCr THR cohort (92.6 vs. 89.7, p=0.039). With revision for any reason as the end point, there was no significant difference in 10 years survivorship between groups (OxZr 98.5%, CoCr 96.6%, p=0.08). Similarly, aseptic revisions demonstrated comparable survivorship rates at 10 year between the OxZr (99.5%) and CoCr groups (97.6%)(p=0.15).

Both THR cohorts demonstrated outstanding survivorship and improvement in patient reported outcomes. The only difference was a slightly better HHS score for the OxZr cohort which may represent selection bias, where OxZr implants were perhaps implanted in more active patients. Implant survivorship was excellent and not dissimilar for both the OxZr and CoCr groups at 10 years. Therefore, with respect to implant longevity at the end of the first decade, there appears to be no clear advantage of OxZr heads compared to CoCr heads when paired with XLPE for patients with similar demographics. Further follow-up into the second and third decade may be required to demonstrate if a difference does exist.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 56 - 56
1 Feb 2020
Broberg J Howard J Lanting B Vasarhelyi E Yuan X Naudie D Teeter M
Full Access

Introduction

Surgeons performing a total knee replacement (TKR) have two available techniques available to help them achieve the proper bone resections and ligament tension – gap balancing (GB) and measured resection (MR). GB relies on balancing ligaments prior to bony resections whereas bony resections are made based on anatomical landmarks in MR. Many studies have been done to compare the joint kinematics between the two techniques, however the results have been varied. These studies were not done with anatomically designed prostheses. The Journey II (Smith & Nephew, Memphis, TN) is one such design which attempts to mimic the normal knee joint structure to return more natural kinematics to the joint, with emphasis on eliminating both paradoxical anterior motion and reduced posterior femoral rollback. Given the design differences between anatomical and non-anatomical prostheses, it is important to investigate whether one technique provides superior kinematics when an anatomical design is used. We hypothesize that there will be no difference between the two techniques.

Methods

A total of 56 individuals were recruited to receive a Journey II prosthesis and randomized evenly to groups where the GB technique or MR technique is used. For all patients in the study, a series of radiostereometric analysis (RSA) images were acquired at 3-months post-operatively at different knee flexion angles, ranging in 20° increments from 0° to 120°. Model-based RSA software (RSACore, Leiden, Netherlands) was used to obtain the 3D positions and orientations of the femoral and tibial implant components, which were in turn used to obtain kinematic measures (contact locations and magnitude of excursion) for each condyle.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 55 - 55
1 Feb 2020
Broberg J Howard J Lanting B Vasarhelyi E Yuan X McCalden R Naudie D Teeter M
Full Access

Introduction

Despite improvements in the survivorship of total knee replacements (TKR) over the years, patient satisfaction following TKR has not improved, with approximately 20% of patients recording dissatisfaction with their new knee joint. It is unclear why many patients feel this way, but it may relate in part to implant designs that do not provide a “natural” feeling knee. Implant manufacturers continue to introduce new concepts for implant design, which are essential for reaching the goal of a “normal” knee after TKR surgery. The Journey II TKR (Smith & Nephew) was developed with this goal in mind. Its anatomical design attempts to mimic the normal knee joint structure to return more natural kinematics to the joint, with emphasis on eliminating both paradoxical anterior motion and reduced posterior femoral rollback. Our objective is to examine patients receiving the Journey II TKR to measure the knee joint contact kinematics of the Journey II TKR compared to a non-anatomically designed implant by the same manufacturer. We hypothesize that the Journey II TKR will have more natural contact kinematics that differ from the non-anatomically designed implant.

Methods

A total of 28 individuals were recruited to receive a Journey II TKR, matching an existing prior cohort with a non-anatomical design from the same manufacturer (Legion TKR, Smith & Nephew). For both groups, a series of radiostereometric analysis (RSA) images were acquired at different knee flexion angles, ranging in 20° increments from 0° to 120°. Model-based RSA software (RSACore, Leiden, Netherlands) was used to obtain the 3D positions and orientations of the femoral and tibial implant components, which were in turn used to obtain kinematic measures (contact locations and magnitude of excursion) for each condyle. Results from the Journey II TKR group at 3 months post-operation were compared to the 2-year post-operative measurements from the Legion TKR group.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 35 - 35
1 Oct 2018
MacDonald SJ Garach M Lanting B McCalden RW Vasarhelyi E Naudie D Howard J
Full Access

Introduction

The infection rate after total joint arthroplasty (TJA) has been shown to be 1–2% in multiple series and registry data. Irrigation, debridement, and polyethylene exchange (IDPE) is a common first line treatment in many cases of acute prosthetic joint infection (PJI). The reinfection rate in open IDPE procedures is variable with studies showing reinfection rates of 10–70% depending on various patient and microbial factors. Our pilot study aimed to determine if the bacterial load in infected total joints was sufficiently reduced by IDPE to allow for the use of post-debridement cultures as an independent marker of procedural success.

Methods

46 prosthetic joint infections underwent irrigation and debridement using 6L of normal saline and 3L of a normal saline and bacitracin mixture prior to the insertion of a new polyethylene liner. This protocol utilized a single equipment setup with all surgical members donning new gloves prior to polyethylene exchange. Between 3 and 5 intraoperative cultures were obtained both prior to and after debridement as per the surgeon's standard protocol. A two-tailed student's t-test was used to evaluate for any differences in the rate of positive culture between these two groups.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 30 - 30
1 Aug 2018
MacDonald S Garach M Lanting B McCalden R Vasarhelyi E Naudie D Howard J
Full Access

The infection rate after total joint arthroplasty (TJA) has been shown to be 1–2% in multiple series and registry data. Irrigation, debridement, and polyethylene exchange (IDPE) is a common first line treatment in many cases of acute prosthetic joint infection (PJI). The reinfection rate in open IDPE procedures is variable with studies showing reinfection rates of 10–70% depending on various patient and microbial factors. Our pilot study aimed to determine if the bacterial load in infected total joints was sufficiently reduced by IDPE to allow for the use of post-debridement cultures as an independent marker of procedural success.

46 prosthetic joint infections underwent irrigation and debridement using 6L of normal saline and 3L of a normal saline and bacitracin mixture prior to the insertion of a new polyethylene liner. This protocol utilized a single equipment setup with all surgical members donning new gloves prior to polyethylene exchange. Between 3 and 5 intraoperative cultures were obtained both prior to and after debridement as per the surgeon's standard protocol. A two-tailed student's t-test was used to evaluate for any differences in the rate of positive culture between these two groups.

Of all pre- and post-debridement cultures sampled 66.5% and 60.7% of cultures were positive respectively. No significant difference in the rate of positive intraoperative culture was found between pre-debridement and post-debridement groups (p = 0.52). In 32 of 46 (69%) cases there was no difference in the total number of positive cultures despite a thorough debridement.

Our data shows that open debridement of PJI does not provide a sterile environment, and post-debridement cultures should not be used as an independent marker of procedural success. The role of an irrigation and debridement to reduce the bacterial burden and potentiate the clearance of an infection is established but its efficacy is unclear, and the inability to create a post-debridement sterile environment is a concern.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 16 - 16
1 Aug 2018
McCalden R Ponnusamy K Marsh J Somerville L MacDonald S Naudie D Lanting B Howard J Vasarhelyi E
Full Access

The purpose of this study is to compare 90-day costs and outcomes for primary total hip arthroplasty (THA) patients between a non-obese (BMI 18.5–24.9) versus overweight (25–29.9), obese (30–34.9), severely-obese (35–39.9), morbidly-obese (40–44.9), and super-obese (45+) cohorts.

We conducted a retrospective review of a prospective institutional database of primary THA patients from 2006–2013, including patients with a minimum of three-year follow-up. Thirty-three super-obese patients were identified, and the other five cohorts were randomly selected in a 2:1 ratio (total n = 363). Demographics, 90-day outcomes (costs, reoperations, and readmissions), and outcomes after three years (revisions and change scores for SF12, HHS, and WOMAC) were collected. Costs were determined using unit costs from our institutional administrative data for all in-hospital resource utilization. Comparisons between the non-obese and other groups were made with Kruskal-Wallis tests for non-normal data and chi-square and Fisher's exact test for categorical data.

The 90-day costs in the morbidly-obese ($13,134 ± 7,250 mean ± standard deviation, p <0.01) and super-obese ($15,604 ± 6,783, p <0.01) cohorts were statistically significantly greater than the non-obese cohorts ($10,315 ± 1,848). Only the super-obese cohort had statistically greater 90-day reoperation and readmission rates than the non-obese cohort (18.2% vs 0%, p <0.01 and 21.2% vs 4.5%, p=0.02, respectively). In addition, reoperations and septic revisions after 3 years were greater in the super-obese cohort compared to the non-obese cohort 21.2% vs 3.0% (p = 0.01), and 18.2% vs 1.5% (p= 0.01), respectively. There were no other statistical differences between the other cohorts with the non-obese cohort at 90-days or after 3 years. Improvements in SF12, HHS, and WOMAC were comparable in all cohorts.

The 90-day costs of a primary total hip arthroplasty for morbidly-obese (BMI 40–44.9) and super-obese (BMI>45) are significantly greater than for non-obese patients, yet these patients have comparable improvements in outcome scores. Health care policies, when based purely on the economic impact of health care delivery, may place morbidly-obese and super-obese patients at risk of losing arthroplasty care, thereby denying them access to the comparable quality of life improvements.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_9 | Pages 30 - 30
1 May 2018
Spiegelberg B Lanting B Howard J Teeter M Naudie D
Full Access

Background

There has been a trend in the evolution of total hip arthroplasty towards increased modularity, with this increase in modularity come some potentially harmful consequences. Modularity at the neck shaft junction has been linked to corrosion, adverse reaction to metal debris and pseudotumor formation.

The aim of this retrieval study is to assess whether the surface integrity of the polyethylene (PE) liner is affected by metal wear debris in a single implant design series of THA revised for trunnionosis.

Method

A retrieval analysis of thirty dual-taper modular neck hip prostheses was performed, the mean time from implantation to revision was 2.7 years (1.02–6.2). The PE liners were analysed using a scanning electron microscope with an energy dispersive spectrometer to assess for metal particles embedded on the liner surface. Serum metal ion levels and inflammatory markers were also analysed.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 26 - 26
1 Jan 2018
MacDonald S Howard J Goyal P Yuan X Lanting B Teeter M Naudie D McCalden R
Full Access

Lewinnek's safe zone recommendation to minimise dislocations was a target of 5–25° for anteversion angle and 30–50° for inclination angle. Subsequently, it was demonstrated that mal-positioning of the acetabular cup can also lead to edge loading, liner fracture, and greater conventional polyethylene wear. The purpose of this study was to measure the effect of acetabular cup position on highly crosslinked polyethylene wear in total hip arthroplasty (THA) at long-term follow-up.

We identified all patients that underwent primary THA with a minimum of 10 years follow-up using an institutional database in London, Ontario, Canada. Patients with a single implant design consisting of a 28 mm cobalt chromium head and highly crosslinked polyethylene liner (ram extruded, GUR 1050, 100 kGy gamma irradiated, remelted, ethylene oxide sterilised) were selected for inclusion. In total, 85 hips from 79 recruited patients were analysed. Patients underwent a supine radiostereometric analysis (RSA) exam in which the x-ray sources and detectors were positioned to obtain an anterior-posterior and cross-table lateral radiograph. Acetabular cup anteversion angle, inclination angle, and 3D penetration rate (including wear and creep) were measured from the stereo radiograph pairs.

At a mean follow-up of 13 years (range, 10–17 years) the mean penetration rate was 0.059 mm/year (95% CI: 0.045 to 0.073 mm/year). Mean anteversion angle was 18.2° (range, −14 to 40°) and mean inclination angle was 43.6° (range, 27 to 61°). With respect to the Lewinnek safe zone, 67% hips met the target for anteversion angle, 77% met the target for inclination angle, and 51% met the target for both. There was no correlation between anteversion angle and penetration rate (r = −0.14, p = 0.72) or between inclination angle and penetration rate (r = 0.11, p = 0.35). There was also no difference (p = 0.07) in penetration rate between hips located within the Lewinnek safe zone for both anteversion angle and inclination angle (mean 0.057 mm/year, 95% CI: 0.036 to 0.079 mm/year) and those outside the safe zone (mean 0.062 mm/year, 95% CI: 0.042 to 0.083 mm/year).

Acetabular cup position had no effect on the wear rate of highly crosslinked polyethylene at long-term follow-up. Although care should still be taken to correctly position the acetabular cup for stability, highly crosslinked polyethylene is a forgiving bearing material that can withstand a wide range of cup positions without negatively impacting longevity due to wear.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 46 - 46
1 Mar 2017
Teeter M Howard J Vasarhelyi E Yuan X McCalden R Naudie D
Full Access

Background

Patient specific instrumentation (PSI) for total knee replacement (TKR) has demonstrated mixed success in simplifying the operation, reducing its costs, and improving limb alignment. Evaluation of PSI with tools such as radiostereometric analysis (RSA) has been limited, especially for cut-through style guides providing mechanical alignment. The primary goal of the present study was to compare implant migration following TKR using conventional and PSI surgical techniques, with secondary goals to examine whether the use of PSI reduces operative time, instrumentation, and surgical waste.

Methods

The study was designed as a prospective, randomized controlled trial of 50 patients, with 25 patients each in the PSI and conventional groups, powered for the RSA analysis. Patients in the PSI group received an MRI and standing 3-foot x-rays to construct patient-specific cut-through surgical guides for the femur and tibia with a mechanical alignment. All patients received the same posterior-stabilized implant, with marker beads inserted in the bone around the implants to enable RSA imaging. Intraoperative variables such as time, number of instrumentation trays used, and mass of surgical waste were recorded. Patients underwent supine RSA exams at multiple time points (2&6 weeks, 3&6 months and yearly) with 6 months data currently available. Migration of the tibial and femoral components was calculated using model-based RSA software. WOMAC, SF-12, EQ5D, and UCLA outcome measures were recorded pre-operatively and post-operatively.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 66 - 66
1 Mar 2017
MacLean C Lanting B Vasarhelyi E Naudie D McAuley J Howard J McCalden R MacDonald S
Full Access

Background

The advent of highly cross-linked polyethylene has resulted in improved wear rates and reduced osteolysis with at least intermediate follow-up when compared to conventional polyethylene. However, the role of alternative femoral head bearing materials in decreasing wear is less clear. The purpose of this study was to determine in-vivo polyethylene wear rates across ceramic, Oxinium, and cobalt chrome femoral head articulations.

Methods

A review of our institutional database was performed to identify patients who underwent a total hip arthroplasty using either ceramic or oxidized zirconium (Oxinium) femoral head components on highly cross-linked polyethylene between 2008 and 2011. These patients were then matched on implant type, age, sex and BMI with patients who had a cobalt chrome bearing implant during the same time period. RSA analysis was performed using the center index method to measure femoral head penetration (polyethylene wear). Secondary quality of life outcomes were collected using WOMAC and HHS Scores. Paired analyses were performed to detect differences in wear rate (mm/year) between the cobalt chrome cohorts and their matched ceramic and Oxinium cohorts. Additional independent group comparisons were performed by analysis of variance with the control groups collapsed to determine wear rate differences between all three cohorts.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 55 - 55
1 Dec 2016
Lanting B Thoren J Yuan X McCalden R McAuley J MacDonald S Vasarhelyi E Howard J Naudie D Teeter M
Full Access

Adequate fixation of implant components is an important goal for all arthroplasty procedures. Aseptic loosening is one of the leading causes of revision surgery in total knee arthroplasty. Radiostereometric analysis (RSA) is an imaging technique to measure implant migration, with established migration thresholds for well-fixed, at risk, and unacceptably migrating components. The purpose of the present study was to examine the long-term fixation of a cemented titanium fixed bearing polished tibial baseplate.

Patients enrolled in a previous two-year prospective trial were recalled at ten years. All patients received a cemented, posterior-stabilised total knee replacement of the same design implanted by one of three surgeons. Of the original 35 patients, 16 were available for long-term follow-up, with one patient lost to follow-up, nine patients deceased, and a further nine patients unwilling to return to the clinic. Each patient underwent RSA imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes as well as maximum total point motion (MTPM) was compared between all time points (baseline, six weeks, three months, six months, one year, two years) up to the ten year follow-up visits. Outcome scores including the Knee Society Score (KSS), WOMAC, SF-12, and UCLA Activity Score were recorded.

At ten years, the mean migrations of the tibial component were less than 0.1 mm and 0.1 degree in all planes relative to the post-operative RSA exam. There was no significant difference in tibial component migration between time points. However, MTPM increased significantly over time (p = 0.002), from 0.23 ± 0.18 mm at six weeks to 0.42 ± 0.20 mm at ten years. At one year, 13 patients had an acceptable MTPM level, three patients had an ‘at risk’ level, and no patient had an ‘unacceptable’ level. No patients were revised at ten years. WOMAC and KSS were significantly improved (p < 0.0001) at the latest follow-up compared to pre-operatively, but there was no difference in SF-12. The median UCLA Activity Score at latest follow-up was six (range, two to eight).

The tibial baseplate demonstrated solid fixation at ten years. No patients had an unacceptable MTPM level at one year and no patients were revised at ten years, supporting the use of RSA to predict long-term loosening risk. The low level of tibial baseplate migration found in the present study correlates to the low rate of revision for this implant as reported in individual studies and in joint replacement registries.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 84 - 84
1 Dec 2016
Nyland M Lanting B Somerville L Vasarhelyi E Naudie D McAuley J McCalden R MacDonald S Howard J
Full Access

Infection following total hip arthroplasty (THA) represents a devastating complication and is one of the main causes for revision surgery. This complication may be treated by irrigation and debridement with head and polyethylene exchange (IDHPE) or a two-stage revision (2SR). Previous studies have reported on the eradication success rates but few have reported patient outcome scores. The purpose of this study was to report patient outcome scores for both IDHPE and 2SR and compare these to a non-infected matched cohort. We hypothesised that both cohorts would have worse outcomes than the control group, and that those who failed an initial IDHPE and required a 2SR would have a worse outcome than those treated initially with a 2SR.

A retrospective review identified 137 patients from our institutional arthroplasty database who had an infected primary THA between 1986–2013. We excluded patients with less than one-year follow-up. Mean follow-up was 60 months (12–187 months). A control cohort was identified and matched according to age and Charlton Comorbidity Index (CCI). Harris Hip Scores, Short Form 12 and WOMAC scores were compared between our control group and our infected cohort.

Sixty-eight patients were treated with a 2SR and 69 patients were treated with an IDHPE. There was a 59% success rate in eradicating the infection with an IDHPE. All of the 28 patients who failed an IDHPE later went on to a 2SR. Outcome scores for the 2SR cohort were significantly worse than the non-infected controls (p0.05). There was no difference in outcome scores when comparing our 2SR cohort to our failed IDHPE (p>0.05).

Previous studies have focused on eradication rates. However, it is important to consider patient outcome scores when deciding the best treatment. Infected patients treated with a successful IDHPE had similar outcomes to non-infected patients. Patients that failed IDHPE and went onto 2SR had similar outcomes to those that had a 2SR alone. IDHPE should still be considered in the treatment algorithm of infected THA.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 59 - 59
1 Dec 2016
Sisko Z Teeter M Lanting B Howard J McCalden R Naudie D MacDonald S Vasarhelyi E
Full Access

Previous retrieval studies demonstrate increased tibial baseplate roughness leads to higher polyethylene backside wear in total knee arthroplasty (TKA). Micromotion between the polyethylene backside and tibial baseplate is affected by the locking mechanism design and can further increase backside wear. The purpose of this study was to examine modern locking mechanisms, in the setting of both roughened and polished tibial baseplates, on backside tibial polyethylene wear.

Five TKA models were selected, all with different tibial baseplate and/or locking mechanism designs. Six retrieval tibial polyethylenes from each TKA model were matched based on time in vivo (TIV), age at TKA revision, BMI, gender, number of times revised, and revision reason. Two observers scored each polyethylene backside according to a visual damage score and individual damage modes. Primary outcomes were mean damage score and individual damage modes. Demographics were compared by one-way ANOVA. Damage scores and modes were analysed by the Kruskal-Wallis test and Dunn's multiple comparisons test.

There were no differences among the groups based on TIV (p=0.962), age (p=0.651), BMI (p=0.951), gender, revision number, or reason for revision. There was a significant difference across groups for mean total damage score (p=0.029). The polished tibial design with a partial peripheral capture locking mechanism and anterior constraint demonstrated a significantly lower score compared to one of the roughened tibial designs with a complete peripheral-rim locking mechanism (13.0 vs. 22.1, p=0.018). Otherwise, mean total damage scores were not significant between groups. As far as modes of wear, there were identifiable differences among the groups based on abrasions (p=0.005). The polished design with a tongue-in-groove locking mechanism demonstrated a significantly higher score compared to both groups with roughened tibial baseplates (5.83 vs. 0.83, p=0.024 and 5.83 vs. 0.92, p=0.033). Only the two designs with roughened tibial baseplates demonstrated dimpling (5.67 and 8.67) which was significant when compared against all other groups (p0.99). No other significant differences were identified when examining burnishing, cold flow, scratching, or pitting. No polyethylene components exhibited embedded debris or delamination.

Total damage scores were similar between all groups except when comparing one of the polished TKA design to one of the roughened designs. The other TKA model with a roughened tibial baseplate had similar damage scores to the polished designs, likely due to its updated locking mechanism. Dimpling wear patterns were specific for roughened tibial baseplates while abrasive wear patterns were identified in the design with a tongue-in-groove locking mechanism. Our study showed even in the setting of a roughened tibial baseplate, modern locking mechanisms decrease backside wear similar to that of other current generation TKA designs.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 59 - 59
1 Nov 2016
Goyal P Yuan X Teeter M McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Lanting B Howard J
Full Access

Studies that have previously examined the relationship between inclination angle and polyethylene wear have shown increased wear of conventional polyethylene with high inclination angles. To date, there have been no long term in vivo studies examining the correlation between cup position and polyethylene wear with highly crosslinked polyethylene.

An institutional arthroplasty database was used to identify patients who had metal-on-highly crosslinked polyethylene primary total hip arthroplasty (THA) using the same component design with a minimum follow up of 10 years ago. A modified RSA examination setup was utilised, recreating standard anteroposterior (AP) and cross-table lateral exams in a single biplane RSA acquisition. Three dimensional head penetration was measured using the centre index method. The same radiographs were used to measure inclination angle and anteversion. Spearman correlation was used to show an association between the parameters of acetabular position and wear rate.

A total of 43 hips were included for analysis in this study. Average follow-up was 12.3 ± 1.2 years. The average linear wear rate was calculated to be 0.066 ± 0.066 mm/year. Inclination angle was not correlated with polyethylene wear rate (p=0.82). Anteversion was also not correlated with polyethylene wear rate (p=0.11). There was no statistical difference between wear rates of hips within Lewinnek's “safe zone” and those outside this “safe zone” (p=0.11). Males had a higher wear rate of 0.094 ± 0.089 mm/year compared to females with a wear rate of 0.046 ± 0.032 mm/year (p=0.045).

At long term follow up of greater than 10 years, highly cross linked polyethylene has very low wear rates. This excellent tribology is independent of acetabular position, but gender did impact wear rates. Due to the low wear rates, follow-up of even longer term is suggested to examine variables affecting wear.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 56 - 56
1 Nov 2016
Nadeau R Howard J Ralley F Somerville L Naudie D
Full Access

Tranexamic acid (TEA), an antifibrinolytic agent, is routinely used for reduction of blood loss in total hip arthroplasty (THA). However, use of intravenous (IV) TEA has been questioned due to safety concerns and a lack of biochemical data in the arthroplasty literature. Tranexamic acid given topically as a periarticular solution is a promising alternative route of administration. The purpose of this study is to identify differences in systemic absorption for intravenous and topical TEA administered during primary THA.

In a blinded randomised controlled trial of patients undergoing primary cementless total hip arthroplasty, 29 participants received a weight-based bolus infusion of intravenous TEA (20 mg/kg) 10 minutes prior to skin incision. Conversely, 15 participants received a 1.5 g bolus dose of TEA administered topically into the periarticular region of the operative hip at the time of arthrotomy closure. A blood sample was drawn one hour post-administration for measurement of serum TEA concentration (µg/mL) by tandem mass spectrometry. In addition to comparing mean concentration levels for both treatment arms, each sample concentration was referenced to a pre-determined TEA concentration threshold of 10 µg/mL, a value known to represent 80% tissue plasminogen activator (tPA) inhibition in vivo.

Those participants receiving topical TEA had four-fold lower TEA levels at one hour postoperatively (mean 12.44 ± 17.59 versus 52.54 ± 23.94 µg/mL, p<0.05).

These results demonstrate significantly lower circulating TEA at one hour after topical administration. Intravenous TEA must travel through the intravascular compartment in order to reach the operative hip. Topical administration of TEA targets bleeding tissues within the surgical field without necessitating parenteral administration. This results in less inhibition of tPA away from the operative site, potentially decreasing the risk of developing a pro-thrombotic state postoperatively. Correlating these results with outcomes from clinical efficacy trials comparing intravenous and topical TEA use in THA will further clarify optimal dosing strategies.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 87 - 87
1 Nov 2016
Matz J Morden D Teeter M McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Howard J Lanting B
Full Access

Complications involving the patellofemoral joint are a source of anterior knee pain, instability, and dysfunction following total knee arthroplasty. “Overstuffing” the patello-femoral joint refers to an increase in the thickness of the patellofemoral joint after a total knee replacement compared to the preoperative thickness. While biomechanical studies have indicated that overstuffing the patellofemoral joint may lead to adverse clinical outcomes, limited clinical evidence exists to support this notion. The purpose of this study is to evaluate the effect of changing the thickness of the patellafemoral joint on functional outcomes following total knee arthroplasty.

Our institutional arthroplasty database was used to identify 1347 patients who underwent a primary total knee arthroplasty between 2006 and 2012 with the same component design. Standard preoperative and postoperative anteroposterior, lateral, and skyline radiographs were collected and measured for patello-femoral overstuffing. These measurements included anterior patellar displacement, anterior femoral offset, and anteroposterior femoral size. These measurements were correlated with patient outcome data using WOMAC, KSS scores, and postoperative range of motion. Multiple linear regression analysis was used to assess the association between stuffing and functional outcomes.

A total of 1031 patients who underwent total knee arthroplasty were included. Increased anterior patellar displacement, a measure of patellofemoral joint thickness, was associated with decreased WOMAC scores (p=0.02). Anterior femoral offset (p=0.210) and anteroposterior femoral size (p=0.091) were not significantly associated with patient functional outcomes. Postoperative range of motion (ROM) was not associated with patellofemoral stuffing (p=0.190).

The current study demonstrated that functional outcomes are adversely affected by patellofemoral overstuffing. Based on these results, caution is encouraged against increasing the thickness of the patellofemoral joint, particularly on the patellar side of the joint.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 64 - 64
1 Nov 2016
Teeter M Naudie D McCalden R Yuan X MacDonald S
Full Access

The philosophy of cemented total hip arthroplasty (THA) femoral components has become polarised. At one extreme are polished, collarless, tapered devices that are expected to subside; at the other extreme are roughened, non-tapered implants with a collar designed not to subside. Radiostereometric analysis (RSA) allows the accurate measurement of implant movement and has been extensively used for measurement of the in vivo migration of implants. The degree of migration as measured by RSA during the first years after surgery has been shown to correlate with the long-term performance of cemented femoral implants. The purpose of this study was to review the two-year RSA results of two different designs of primary cemented THA stems.

Data from two previous prospective RSA trials with two-year follow-up were pooled. The first group included 36 patients who received a Spectron (Smith & Nephew, Memphis, USA) cemented stem. The second group included 13 patients who received an Exeter (Stryker, Mahwah, USA) cemented stem, and 15 patients who received a CPCS (Smith & Nephew, Memphis, USA) cemented stem. All patients underwent RSA examinations shortly post-operation, at 6 weeks, 3 months, 6 months, 1 years, and 2 years. Migration and rotation of the femoral stems was measured at each time point relative to the post-operative exam, and compared between the two groups.

There was no difference in age at surgery (Spectron 78 ± 6 years, Exeter/CPCS 77 ± 5 years, p = 0.43), BMI (Spectron and Exeter/CPCS 28 ± 5 kg/m2, p = 0.92), or percentage of male patients (Spectron 23% male, Exeter/CPCS 21% male) between the implant groups. Subsidence was significantly greater (p < 0.0001) at all time points from three months to two years for the Exeter and CPCS stems (0.94 ± 0.39 mm at two years) compared to the Spectron stem (0.05 ± 0.16 mm at two years). There was no significant difference between the stem types for medial-lateral translation (p = 0.07) or anterior-posterior translation (p = 0.49), or for anterior-posterior tilt (p = 0.15), internal-external rotation (p = 0.89), or varus-valgus rotation (p = 0.05).

Implant material, design, and surface finish are all factors in the long-term performance of cemented femoral hip implants. In this study, both femoral stem designs had a magnitude of sub-sidence that was within the limits of what is considered to be safe with respect to long-term performance. The continuous subsidence of the Exeter and CPCS stems is consistent with previous reports in the literature.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 57 - 57
1 Nov 2016
Lanting B Tan S Lau A Teeter M Del Balso C McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Howard J
Full Access

Trunnionosis in modular hip arthroplasty has recently been recognised to be clinically important. Gaining an understanding of how the material interface at the head-trunnion affects the tribology at the modular junctions has current clinical implications as well as an implication on future implant selection and material choice. This matched-cohort study aims to compare tribocorrosion between ceramic and cobalt-chromium trunnions and to investigate other factors that contribute to the difference in tribocorrosion if present.

All hip prostheses retrieved between 1999 and 2015 at one centre were reviewed. Fifty two ceramic heads were retrieved, and these were matched to a cobalt-chromium cohort according to taper design, head size, neck length and implantation time in that order. The two cohorts were similar in male:female ratio (p=0.32) and body mass index (p=0.15) though the ceramic group was younger than the cobalt-chromium group (56.6 (+/−)13.5 years for ceramic group vs 66.3 (+/−14.4) years for cobalt-chromium group; p=0.001). There was no significant difference in the reasons for revision between the two groups (p=0.42). The femoral head trunnions were examined by two independent observers using a previously published 4-point scoring technique. The trunnions were divided into three zones: apex, middle and base. The observers were blinded to clinical and manufacturing data where possible.

Ceramic head trunnions demonstrated a lower median fretting and corrosion score at the base zone (p<0.001), middle zone (p<0.001) and in the combined score (p<0.001). In a subgroup analysis by head size, ceramic heads had a lower fretting and corrosion score at 28mm head diameter (p<0.001). Within the ceramic group, taper design had a significant effect on fretting and corrosion in the apex zone (p=0.04). Taper design also had a similar effect in the cobalt-chromium group in the apex zone (p=0.03). For the ceramic trunnions, the largest effect was contributed by the difference between the 11/13 taper and the 12/14 taper. For the cobalt-chromium trunnions, the largest effect was contributed by the difference between the 5 degree 38′ 37″ taper and type 1 taper.

Ceramic head trunnions showed a significantly lower fretting and corrosion score as compared to cobalt-chromium trunnions. Ceramic heads had a lower score than cobalt-chromium heads at 28mm head diameter. Taper design had an effect on fretting and corrosion within each cohort.