Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 25 - 25
1 May 2016
Baba Y Maruyama D Yamamoto K Nakagawa S Nakashima Y Nagayama N
Full Access

Introduction

Total knee arthroplasty (TKA) has achieved excellent clinical outcomes and functional performances. However, there is a need for greater implant longevity and higher flexion by younger and Asian patients. We determined the relationship between mobility and stability of TKA product because they are essential for much further functional upgrading. This research evaluated the geometry characteristics of femorotibial surfaces quantitatively by measuring their force of constraint by computer simulation and mechanical test.

Methods

We measured the force of constraint of femorotibial surfaces in order to evaluate the property of femorotibial surfaces. A total knee system was used for this evaluation, and has an asymmetrical joint surface, which restores the anatomical jointline in both sagittal and coronal planes, and is expected to permit normal kinematics, with cruciate-retaining fixed type.

We performed computer simulation using finite element analyses (FEA) and mechanical tests using knee simulator to measure the force of constraint regarding anterior-posterior (AP) and internal-external (IE) rotational direction in extension position, 90-degree flexion and a maximum flexion of 140-degree. In the FEA, Young's modulus and Poisson's ratio were set to 213 GPa and 0.3 for Co-Cr-Mo alloy as the femoral component, and 1 GPa and 0.3 for UHMWPe as the tibial insert, respectively. The force load to AP direction of tibial tray was measured when the femoral component moved plus or minus 10 millimeters. The moment load to IE rotational direction of tibial tray was measured when the femoral component moved plus or minus 20 degrees. The vertical load of 710 N was loaded on the femoral component during these measurements.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 36 - 36
1 May 2016
Shiomi T Yamamura M Takahashi S Suzuka T Nakagawa S
Full Access

The purpose of this study was to evaluate in vivo fit and fill analysis of tapered wedge-type stem in total hip arthroplasty (THA) with computed tomography (CT)-based navigation system. 100 THAs were all performed through the posterolateral approach, with patients in the lateral decubitus position. Each cohort of 50 consecutive primary cementless THAs with was compared with and without CT-based navigation system. The post-operative antero-posterior (AP) hip radiographs were obtained two weeks after the operation. All radiographic fit and fill measurements in the proximal and distal areas were analyzed by two of the authors who were both blinded to the use of CT-based navigation system. The type of the fit in the cementless stem was divided into three types. The fit of the stem was classified as Type I, if there was both proximal and distal engagement (maximum proximal to distal engagement difference of 2 mm or greater), Type II when there was proximal engagement only, and Type III when there was distal engagement only. The fill parameters such as mean stem-to-canal ratios and mean minimum and maximum gaps between the stems to the cortical bone in proximal and distal sections were compared. There was a significantly better overall canal fit obtained by THA with CT-based navigation system compared to without the navigation system (p<0.01). With CT-based navigation system, 42 of 50 stems (84%) were categorized as Type I fit compared to 31 of 50 stems (62%) without the navigation system. As to Type II fit, There are significantly more stems without the navigation system (26%) compared to with it (12%). There were better canal fills of the stems obtained by THA with CT-based navigation system both in proximal (94%) and distal sections (88%) of the femur compared to without the navigation system (proximal 88%/distal 82%) (p<0.05). Excellent radiographic fit and fill has been previously reported to potentially correlate with improved clinical outcomes. The stems obtained by THA with CT-based navigation system had a significantly better canal fit demonstrated by higher proportion of Type I and lower proportion of Type II fits, compared to without the navigation system. The stems with the navigation system had also significantly better proximal and distal canal fill.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 232 - 232
1 Dec 2013
Baba Y Yamamoto K Maruyama D Sugimoto T Nakagawa S Nakashima Y
Full Access

Background:

Total knee arthroplasty (TKA) has achieved excellent clinical outcomes and functional performances. However, younger and Asian patients require even greater implant longevity and higher flexion. It is necessary for much further functional upgrading to design TKA with mobility and stability. Therefore, we determined the relationship between mobility and stability of TKA.

Methods:

We evaluated the force of constraint of femorotibial surfaces in two types of designs in order to measure the property of femorotibial surfaces. The anatomical geometry knee (AGK) has an asymmetrical design, which restores the anatomical jointline in both sagittal and coronal planes, and is expected to permit normal kinematics, with cruciate-retaining fixed type. The functional designed knee (FDK) has a symmetrical design, and enhances concave femorotibial surfaces with cruciate-retaining mobile type.

We performed mechanical tests to measure the force of constraint regarding anterior-posterior (AP) and internal-external (IE) rotational direction in extension position, 90-degree flexion and a maximum flexion of 140-degree. The force load to AP direction of tibial tray was measured when the femoral component moved plus or minus 10 millimeters. The moment load to IE rotational direction of tibial tray was measured when the femoral component moved plus or minus 20 degrees. The vertical load of 710N has been loaded on the femoral component during this test.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_VIII | Pages 29 - 29
1 Mar 2012
Ichiseki T Kaneuji A Sugimori T Fukui K Kitamura K Mikami T Nakagawa S Matsumoto T
Full Access

Introduction

Recently, oxidative stress has been implicated in the development of osteonecrosis. Here we focused on vitamins with marked antioxidant potency to see whether their use might prevent the development of osteonecrosis associated with corticosteroid administration.

Methods

Fifteen male Japanese white rabbits weighing about 3.5 kg were injected once into the right gluteal muscle with methylprednisolone (MPSL) 40 mg/kg (S Group). Ten other rabbits, in addition, received consecutive daily intravenous injections of vitamin E 50 mg/kg starting from the day of MPSL administration (E Group), and 10 other animals similarly received consecutive daily intravenous injections of vitamin C 30 mg/kg (C Group). All animals were euthanized 2 weeks after MPSL administration, and femurs were extracted, and stained with hematoxylin-eosin. Blood levels of glutathione (GSH) were also measured.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 8 | Pages 1133 - 1136
1 Nov 2004
Tokuhara Y Kadoya Y Nakagawa S Kobayashi A Takaoka K

Varus and valgus joint laxity of the normal living knee in flexion was assessed using MRI. Twenty knees were flexed to 90° and were imaged in neutral and under a varus-valgus stress in an open MRI system. The configuration of the tibiofemoral joint gap was studied in slices which crossed the epicondyles of the femur.

When a varus stress was applied, the lateral joint gap opened by 6.7 ± 1.9 mm (mean ± sd; 2.1 to 9.2) whereas the medial joint gap opened by only by a mean of 2.1 ± 1.1 mm (0.2 to 4.2). These discrepancies indicate that the tibiofemoral flexion gap in the normal knee is not rectangular and that the lateral joint gap is significantly lax. These results may be useful for adequate soft-tissue balancing and bone resection in total knee arthroplasty and reconstruction surgery on ligaments.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 6 | Pages 925 - 931
1 Aug 2004
Pinskerova V Johal P Nakagawa S Sosna A Williams A Gedroyc W Freeman MAR

MRI studies of the knee were performed at intervals between full extension and 120° of flexion in six cadavers and also non-weight-bearing and weight-bearing in five volunteers. At each interval sagittal images were obtained through both compartments on which the position of the femoral condyle, identified by the centre of its posterior circular surface which is termed the flexion facet centre (FFC), and the point of closest approximation between the femoral and tibial subchondral plates, the contact point (CP), were identified relative to the posterior tibial cortex.

The movements of the CP and FFC were essentially the same in the three groups but in all three the medial differed from the lateral compartment and the movement of the FFC differed from that of the CP. Medially from 30° to 120° the FFC and CP coincided and did not move anteroposteriorly. From 30° to 0° the anteroposterior position of the FFC remained unchanged but the CP moved forwards by about 15 mm. Laterally, the FFC and the CP moved backwards together by about 15 mm from 20° to 120°. From 20° to full extension both the FFC and CP moved forwards, but the latter moved more than the former. The differences between the movements of the FFC and the CP could be explained by the sagittal shapes of the bones, especially anteriorly.

The term ‘roll-back’ can be applied to solid bodies, e.g. the condyles, but not to areas. The lateral femoral condyle does roll-back with flexion but the medial does not, i.e. the femur rotates externally around a medial centre. By contrast, both the medial and lateral contact points move back, roughly in parallel, from 0° to 120° but they cannot ‘roll’.

Femoral roll-back with flexion, usually imagined as backward rolling of both condyles, does not occur.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 450 - 456
1 Apr 2004
Nakagawa S Johal P Pinskerova V Komatsu T Sosna A Williams A Freeman MAR

The posterior cruciate ligament (PCL) was imaged by MRI throughout flexion in neutral tibial rotation in six cadaver knees, which were also dissected, and in 20 unloaded and 13 loaded living (squatting) knees. The appearance of the ligament was the same in all three groups. In extension the ligament is curved concave-forwards. It is straight, fully out-to-length and approaching vertical from 60° to 120°, and curves convex-forwards over the roof of the intercondylar notch in full flexion. Throughout flexion the length of the ligament does not change, but the separations of its attachments do.

We conclude that the PCL is not loaded in the unloaded cadaver knee and therefore, since its appearance in all three groups is the same, that it is also unloaded in the living knee during flexion. The posterior fibres may be an exception in hyperextension, probably being loaded either because of posterior femoral lift-off or because of the forward curvature of the PCL. These conclusions relate only to everyday life: none may be drawn with regard to more strenuous activities such as sport or in trauma.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 8 | Pages 1199 - 1200
1 Nov 2000
Nakagawa S Kadoya Y Todo S Kobayashi A Sakamoto H Freeman MAR Yamano Y

We studied active flexion from 90° to 133° and passive flexion to 162° using MRI in 20 unloaded knees in Japanese subjects. Flexion over this arc is accompanied by backward movement of the medial femoral condyle of 4.0 mm and by backward movement laterally of 15 mm, i.e., by internal rotation of the tibia. At 162° the lateral femoral condyle lies posterior to the tibia.