Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Research

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 71 - 71
1 Aug 2012
Thiagarajah S MacInnes S Yang L Wilkinson J
Full Access

Introduction

Subtle variations in hip morphology associate with risk of hip osteoarthritis (OA). However, validated accurate methods to quantitate hip morphology using plain radiography are lacking. We have developed a Matlab-based software-tool (SHIPs) that measures 19 OA-associated morphological-parameters of the hip using a PACS pelvic radiograph. In this study we evaluated the accuracy and repeatability of the method.

Methods

Software accuracy was assessed by firstly measuring the linear ratio of 2 fixed distances and several angles against a gold-standard test radiograph, and secondly by repeated measurements on a simulated AP radiograph of the pelvis (reformatted from CT-data) that was digitally rotated about 3-axes to determine the error associated with pelvic mal-positioning. Repeatability was assessed using 30-AP Pelvic radiographs analysed twice (intra-observer), by 2 readers (inter-observer), and finally, using 2 pelvic radiographs taken in 23 subjects (n=46 radiographs) taken same day after re-positioning (short-term clinical-practice variability), and was expressed as coefficient of variation (CV%).