The aim of this study was to determine whether the rates of revision
for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle
components varied according to the year of the initial operation,
and compare these with the rates of revision for other designs of
MoM THA. Data from the National Joint Registry for England and Wales included
36 mm MoM THAs with Pinnacle acetabular components which were undertaken
between 2003 and 2012 with follow-up for at least five years (n
= 10 776) and a control group of other MoM THAs (n = 13 817). The
effect of the year of the primary operation on all-cause rates of revision
was assessed using Cox regression and interrupted time-series analysis.Aims
Patients and Methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
Advocates of debridement, antibiotics and implant retention (DAIR)
in hip periprosthetic joint infection (PJI) argue that a procedure
not disturbing a sound prosthesis-bone interface is likely to lead
to better survival and functional outcome compared with revision.
This case-control study aims were to compare outcome of DAIRs for
infected primary total hip arthroplasty (THA) with outcomes following
primary THA and two-stage revision of infected primary THAs. We retrospectively reviewed all DAIRs, performed for confirmed
infected primary hip arthropasty (n = 82) at out institution, between
1997 and 2013. Data recorded included full patient information and
type of surgery. Outcome measures included complications, mortality,
implant survivorship and functional outcome. Outcome was compared with
two control groups matched for gender and age; a cohort of primary
THAs (n = 120) and a cohort of two-stage revisions for infection
(n = 66).Aims
Patients and Methods
Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of England, Wales and Northern Ireland to determine
if a cemented femoral stem brand was associated with the risk of
having revision for a PF after primary THA. All cemented femoral
stem brands with more than 10 000 primary operations recorded in
the NJR were identified. The four most commonly used cemented femoral
stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem
(n = 15 113) and Charnley (n = 20 182). We compared the revision risk ratios due to PF amongst the stems
using a Poisson regression model adjusting for patient factors.
Compared with the Exeter V40, the age, gender and ASA grade adjusted
revision rate ratio was 3.89 for the cemented CPT stem (95% confidence
interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to
1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70).Aims
Patients and Methods
The aims of this study were to compare the diagnostic test characteristics
of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI)
alone, and ultrasound combined with MARS-MRI for identifying intra-operative
pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients
undergoing revision surgery. This retrospective diagnostic accuracy study involved 39 patients
(40 MoMHRs). The time between imaging modalities was a mean of 14.6
days (0 to 90), with imaging performed at a mean of 5.3 months (0.06
to 12) before revision. The prevalence of intra-operative pseudotumours
was 82.5% (n = 33).Aims
Methods
We assessed the orientation of the acetabular
component in 1070 primary total hip arthroplasties with hard-on-soft, small
diameter bearings, aiming to determine the size and site of the
target zone that optimises outcome. Outcome measures included complications,
dislocations, revisions and ΔOHS (the difference between the Oxford
Hip Scores pre-operatively and five years post-operatively). A wide
scatter of orientation was observed (2 This study demonstrated that with traditional technology surgeons
can only reliably achieve a target zone of ±15°. As the optimal
zone to diminish the risk of dislocation is also ±15°, surgeons
should be able to achieve this. This is the first study to demonstrate
that optimal orientation of the acetabular component improves the
functional outcome. However, the target zone is small (± 5°) and
cannot, with current technology, be consistently achieved. Cite this article:
There is great variability in acetabular component
orientation following hip replacement. The aims of this study were
to compare the component orientation at impaction with the orientation
measured on post-operative radiographs and identify factors that
influence the difference between the two. A total of 67 hip replacements
(52 total hip replacements and 15 hip resurfacings) were prospectively
studied. Intra-operatively, the orientation of the acetabular component
after impaction relative to the operating table was measured using
a validated stereo-photogrammetry protocol. Post-operatively, the
radiographic orientation was measured; the mean inclination/anteversion
was 43° ( This study demonstrated that in order to achieve a specific radiographic
orientation target, surgeons should implant the acetabular component
5° less inclined and 8° more anteverted than their target. Great
variability (2 Cite this article:
The orientation of the acetabular component is
influenced not only by the orientation at which the surgeon implants
the component, but also the orientation of the pelvis at the time
of implantation. Hence, the orientation of the pelvis at set-up
and its movement during the operation, are important. During 67
hip replacements, using a validated photogrammetric technique, we
measured how three surgeons orientated the patient’s pelvis, how
much the pelvis moved during surgery, and what effect these had
on the final orientation of the acetabular component. Pelvic orientation
at set-up, varied widely (mean (± 2, standard deviation ( Cite this article:
There is a North Atlantic divide, with cementless
femoral stems being used more frequently in the USA and cemented
stems being used more frequently in many countries in Europe. This
is primarily because different cemented stems have been used on
different sides of the Atlantic and the results of the cemented
stems in the US have often been poor, whereas the results of the
stems used in Europe have been good. In the National registers in
Europe, cemented stems have tended to achieve better results than
cementless. Cite this article:
The Exeter femoral stem is a double-tapered highly
polished collarless cemented implant with good long-term clinical
results. In order to determine why the stem functions well we have
undertaken a long-term radiostereometric analysis (RSA) study. A total of 20 patients undergoing primary Exeter total hip replacement
for osteoarthritis using the Hardinge approach were recruited and
followed with RSA for ten years. The stems progressively subsided
and internally rotated with posterior head migration. The mean subsidence
was 0.7 mm (95% confidence interval (CI) 0.5 to 0.9) at two years
and 1.3 mm (95% CI 1.0 to 1.6) at ten years. The mean posterior
migration of the head was 0.7 mm (95% CI 0.5 to 0.9) at two years
and 1.2 mm (95% CI 1.0 to 1.4) at ten years. There was no significant
cement restrictor migration. The Exeter stem continues to subside slowly into the cement mantle
in the long term. This appears to compress the cement and the cement
bone interface, contributing to secure fixation in the long term. Cite this article:
Recent events have highlighted the importance
of implant design for survival and wear-related complications following
metal-on-metal hip resurfacing arthroplasty. The mid-term survival
of the most widely used implant, the Birmingham Hip Resurfacing
(BHR), has been described by its designers. The aim of this study
was to report the ten-year survival and patient-reported functional
outcome of the BHR from an independent centre. In this cohort of 554 patients (646 BHRs) with a mean age of
51.9 years (16.5 to 81.5) followed for a mean of eight years (1
to 12), the survival and patient-reported functional outcome depended
on gender and the size of the implant. In female hips (n = 267)
the ten-year survival was 74% (95% confidence interval (CI) 83 to
91), the ten-year revision rate for pseudotumour was 7%, the mean
Oxford hip score (OHS) was 43 ( This study supports the ongoing use of resurfacing in young active
men, who are a subgroup of patients who tend to have problems with
conventional THR. In contrast, the results in women have been poor
and we do not recommend metal-on-metal resurfacing in women. Continuous
follow-up is recommended because of the increasing incidence of
pseudotumour with the passage of time.
Pseudotumours (abnormal peri-prosthetic soft-tissue reactions)
following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have
been associated with elevated metal ion levels, suggesting that
excessive wear may occur due to edge-loading of these MoM implants.
This study aimed to quantify The duration and magnitude of edge-loading Objectives
Methods
The aim of this retrospective cohort study was
to identify any difference in femoral offset as measured on pre-operative
anteroposterior (AP) radiographs of the pelvis, AP radiographs of
the hip and corresponding CT scans in a consecutive series of 100
patients with primary end-stage osteoarthritis of the hip (43 men
and 57 women with a mean age of 61 years (45 to 74) and a mean body
mass index of 28 kg/m2 (20 to 45)). Patients were positioned according to a standardised protocol
to achieve reproducible projection and all images were calibrated.
Inter- and intra-observer reliability was evaluated and agreement
between methods was assessed using Bland-Altman plots. In the entire cohort, the mean femoral offset was 39.0 mm (95%
confidence interval (CI) 37.4 to 40.6) on radiographs of the pelvis,
44.0 mm (95% CI 42.4 to 45.6) on radiographs of the hip and 44.7
mm (95% CI 43.5 to 45.9) on CT scans. AP radiographs of the pelvis
underestimated femoral offset by 13% when compared with CT (p <
0.001).
No difference in mean femoral offset was seen between AP radiographs
of the hip and CT (p = 0.191). Our results suggest that femoral offset is significantly underestimated
on AP radiographs of the pelvis but can be reliably and accurately
assessed on AP radiographs of the hip in patients with primary end-stage
hip osteoarthritis. We, therefore, recommend that additional AP radiographs of the
hip are obtained routinely for the pre-operative assessment of femoral
offset when templating before total hip replacement.
Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.
Pseudotumour is a rare but important complication of metal-on-metal hip resurfacing that occurs much more commonly in women than in men. We examined the relationship between head-neck ratio (HNR) and pseudotumour formation in 18 resurfaced hips (18 patients) revised for pseudotumour and 42 asymptomatic control resurfaced hips (42 patients). Patients in whom pseudotumour formation had occurred had higher pre-operative HNR than the control patients (mean 1.37 ( We suggest that reducing the size of the femoral head, made possible by a high pre-operative HNR, increases the risk of impingement and edge loading, and may contribute to high wear and pseudotumour formation. As the incidence of pseudotumour is low in men, it appears safe to perform resurfacing in men. However, this study suggests that it is also reasonable to resurface in women with a pre-operative HNR ≤ 1.3.
We have evaluated the difference in the migration patterns over two years of two cementless stems in a randomised, controlled trial using radiostereophotogrammetric analysis (RSA). The implants studied were the Furlong HAC stem, which has good long-term results and the Furlong Active stem, which is a modified version of the former designed to minimise stress concentrations between the implant and bone, and thus to improve fixation. A total of 23 Furlong HAC and 20 Furlong Active stems were implanted in 43 patients. RSA examinations were carried out immediately post-operatively and at six, 12 and 24 months post-operatively. The subsidence during the first year in the Furlong HAC stem, was approximately one-third that of the Furlong Active stem, the measured mean subsidence of the femoral head at six months being 0.27 mm (95% confidence interval (CI) 0.03 to 0.51) and 0.99 mm (95% CI 0.38 to 1.60), respectively (p = 0.03). One Active stem continued to subside during the second year. All hips, regardless of the type of stem were clinically successful as judged by the Oxford hip score and a derived pain score without any distinction between the two types of stem. The initial stability of the Furlong Active stem was not as good as the established stem which might compromise osseo-integration to the detriment of long-term success. The changes in the geometry of the stem, to minimise stress have affected the attainment of initial stability.
Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by edge loading due to poor orientation of the acetabular component. Our aim was to determine the optimal acetabular orientation to minimise the risk of pseudotumour formation. We matched 31 hip resurfacings revised for pseudotumour formation with 58 controls who had a satisfactory outcome from this procedure. The radiographic inclination and anteversion angles of the acetabular component were measured on anteroposterior radiographs of the pelvis using Einzel-Bild-Roentgen-Analyse software. The mean inclination angle (47°, 10° to 81°) and anteversion angle (14°, 4° to 34°) of the pseudotumour cases were the same (p = 0.8, p = 0.2) as the controls, 46° (29° to 60°) and 16° (4° to 30°) respectively, but the variation was greater. Assuming an accuracy of implantation of ± 10° about a target position, the optimal radiographic position was found to be approximately 45° of inclination and 20° of anteversion. The incidence of pseudotumours inside the zone was four times lower (p = 0.007) than outside the zone. In order to minimise the risk of pseudotumour formation we recommend that surgeons implant the acetabular component at an inclination of 45° (± 10) and anteversion of 20° (± 10) on post-operative radiographs. Because of differences between the radiographic and the operative angles, this may be best achieved by aiming for an inclination of 40° and an anteversion of 25°.
The cause of fracture of the femoral neck after hip resurfacing is poorly understood. In order to evaluate the role of avascular necrosis we compared 19 femoral heads retrieved at revision for fracture of the femoral neck and 13 retrieved for other reasons. We developed a new technique of assessing avascular necrosis in the femoral head by determining the percentage of empty osteocyte lacunae present. Femoral heads retrieved as controls at total hip replacement for osteoarthritis and avascular necrosis had 9% ( In the fracture group the percentage of empty lacunae was 71% ( We conclude that fracture after resurfacing of the hip is associated with a significantly greater percentage of empty osteocyte lacunae within the trabecular bone. This indicates established avascular necrosis and suggests that damage to the blood supply at the time of surgery is a potent risk factor for fracture of the femoral neck after hip resurfacing.
The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure. We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 μm/year (2.75 to 25.4) than the 1.79 μm/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 μm/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 μm/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred. Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour.